Introduction
The chest radiograph (CXR) is the most common imaging performed with almost all patients
attending the hospital having one. CXR is performed “round the clock” due to its clinical
utility and easy availability. Clinicians are expected to review these radiographs,
make a preliminary diagnosis, and treat patients accordingly. It is very important
to make an accurate assessment of the radiographs as it will not only impact the clinical
management but may also streamline the need for other investigations. Most of the
clinicians and radiologists are comfortable in reviewing a CXR; however, it can be
a daunting task to some juniors. A logical process of interpretation of CXR involves
a review of iatrogenic foreign bodies (IFB) first (lines and drains), followed by
a review of the mediastinum, the lungs, and finally the bony architecture.
With advances in the field of cardiothoracic surgery and increased usage of percutaneous
procedures for placement of devices, more and more IFB are seen on CXRs. Every doctor
interpreting these images needs to be aware of these procedures, normal appearances
of the prosthesis, and should be able to recognize potential complications relating
to these devices/IFB.
In this review, we present a selection of findings on CXR that includes postoperative
appearances, cardiothoracic devices, and their complications. We will review these
devices according to their location within the body, that is, cardiac, vascular, thoracic,
musculoskeletal, and miscellaneous. Some of the well-known devices/lines (central
venous line/chest drains) are not covered in this review.
Cardiac Devices
Defibrillator and Pacemaker
Cardiovascular implantable electronic device is nowadays commonly seen on CXR and
includes permanent pacemaker and implantable cardiac defibrillator (ICD). These devices
are implanted in patients with cardiac rhythm disturbances, to improve cardiac function
and/or prophylactically in patients susceptible to life-threatening arrhythmias. These
are important devices and can have varied appearances on a CXR. These consist of a
pulse generator and some transvenous pacing leads. The pulse generator is often implanted
in a subcutaneous pocket within the pectoralis muscle or in the abdomen ([Fig. 1]). Abdominal placement is preferred in children to accommodate their growth. Pacemaker
leads (generally 1–3) are metallic wires with silicone or polyurethane insulation.
Cephalic or subclavian veins are commonly used access points with jugular and femoral
veins being other alternatives. The right ventricular (RV) lead courses along the
lateral wall of the right atrium, passes through the tricuspid valve with its tip
at the ventricular apex directed anteriorly and inferiorly, just to the left of the
spine.[1] On the lateral film, the lead should be directed anteriorly and inferiorly, to differentiate
it from coronary sinus or left ventricular lead. The right atrial (preferably within
the atrial appendage) lead on the frontal CXR should terminate over the right upper
heart border, with a J-shaped curve and tip-directed upward.[2] On a lateral radiograph, the lead should follow a smooth anterior curve in the midportion
of the heart with its tip angled cephalad into the right atrial appendage[3] ([Figs. 2 ]and [3]). In case of a biventricular pacemaker or cardiac resynchronization therapy (CRT)
device, the left ventricular lead is placed through the coronary sinus into a cardiac
vein, usually along the lateral or posterior free wall of the left ventricle ([Fig. 4]). The left ventricular lead will course inferiorly and laterally on a frontal radiograph
(cannot distinguish from a lead in RV apex) and posteriorly on the lateral view (RV
lead is anterior in location on this view).[4] ICDs are most often composed of a single lead with two shock coils (located in the
region of the brachiocephalic vein–superior vena cava junction and in the RV).[4] The shock coils appear as cylindrical thickening in the lead and can be used to
differentiate ICD from a pacemaker ([Fig. 5]). CRT may also have a defibrillator component (CRT-D) ([Fig. 6]). Pacemaker leads may also be placed along the epicardial surface of the heart rather
than the transvenous position ([Figs. 7 ]and [8]). Important complications of pacemaker insertion include malpositioned/retained
or fractured wires/leads, pneumothorax, Twiddler syndrome, and cardiac perforation
([Figs. 9 ]and [10]). Aberrant vascular anatomy may also simulate malpositioned leads ([Fig. 11]). In patients who may have thrombosis of the upper limb veins, congenital heart
anomalies, or in children who have pacing generator in the abdomen, a transhepatic
route for the wires may be undertaken. In this, venous access is achieved via the
hepatic vein via transcutaneous puncture. The leads are then passed into the inferior
vena cava and then the cardiac chambers.
Fig. 1 Frontal chest radiograph showing a single lead pacemaker device with the pacing lead
within the right ventricle (white arrow). The pulse generator is projected in the
left upper part of the chest (black arrow).
Fig. 2 Frontal (A) and lateral (B) chest radiograph in a patient with dual-chamber pacemaker. The pacing lead in the
right atrium (black arrow) shows the typical J-shaped curve with the tip directed
upwards. The lead in the right ventricle projects anteriorly and inferiorly on the
lateral film.
Fig. 3 Frontal (A) and lateral (B) chest radiograph in a patient with biventricular pacing (cardiac resynchronization
therapy). The right atrial (white dotted arrow) and right ventricle leads (white arrow)
are normal in position. A third lead (black arrow) is noted to pass posteriorly in
the lateral view and in between the other two leads in the frontal view in keeping
with its location in the coronary sinus. Prior sternotomy (white star) is also noted
with intact sternal wires in the midline.
Fig. 4 Frontal chest radiograph in a patient with cardiac resynchronization therapy device
wherein the coronary sinus lead (black arrow) is positioned along the lateral wall
of the left ventricle. The right atrial (dotted arrow) and ventricular leads (white
arrow) are also normal in location.
Fig. 5 Frontal chest radiograph showing single lead defibrillator (implantable cardiac defibrillator)
device. Two shock coils (white arrows) are clearly seen as areas of dense cylindrical
thickening in the leads. Also noted is sternotomy wires (star) and surgical clips
(black arrow) relating to prior coronary artery bypass surgery.
Fig. 6 Frontal chest radiograph showing a cardiac resynchronization therapy-defibrillator
device with two shock coils (white arrow) along the right ventricle lead and a coronary
sinus lead (black arrow). Right atrial lead (black dotted arrow) is also seen. In
the background, a central venous catheter (white dotted arrow) is noted along the
right superior vena cava.
Fig. 7 Frontal chest radiograph in a patient with a hybrid pacing/defibrillator device.
There are three leads along the intravenous route into the heart (right atrium [white
dotted arrow], right ventricle [white arrow], and coronary sinus [not seen well on]),
and a further fourth lead from the pulse generator implanted directly on the epicardial
surface of the heart. Endotracheal tube is also noted (black dotted arrow).
Fig. 8 Flowchart depicting an approach to analyze any cardiac implanted external device.
CIED, cardiovascular implantable electronic device.
Fig. 9 Frontal chest radiograph in a patient postimplantation of dual lead pacing device
showing a left apical pneumothorax (black arrow showing lung edge) secondary to the
procedure.
Fig. 10 Frontal chest radiograph in two different patients with complications relating to
pacing leads. (A) Retained pacemaker lead (white arrow) is noted from a prior pacing device. Also,
note a poorly defined opacity (star) in the left upper lobe partially masked by the
generator box. This turned out to be a malignancy. (B) There is a break in the right ventricular lead (black arrow) for which another lead
was inserted.
Fig. 11 Frontal chest radiograph in a patient with a dual lead pacing device and a variant
anatomy. The leads are seen to pass through a persistent left-sided superior vena
cava (white dotted arrow) into the right ventricle via the coronary sinus. The right
atrial (white arrow) and ventricular (black arrow) leads are seen.
Epicardial Pacing
Epicardial pacing can be either permanent or temporary. Permanent epicardial pacing
involves surgical placement of the epicardial leads directly on the epicardium. The
leads are then tunneled through the chest wall and connected to the pacing box usually
in the subcutaneous pocket of the upper abdomen[5] ([Fig. 12]). Currently, they are used only in unusual circumstances such as patients with recurrent
dislodgement of transvenous systems, patients with prosthetic tricuspid valves or
congenital anomalies such as tricuspid atresia, and in patients underlying CRT in
whom the coronary sinus lead placement was unsuccessful.[5]
Fig. 12 Frontal chest radiograph in two patients with permanent epicardial pacing devices.
(A) Retained pacing leads from an older device (white arrows) and a newer lead (dotted
arrow) is seen with the generator box (black arrow) in the lower abdomen. (B) A generator box in the upper abdomen and a single epicardial pacing lead. Note the
tip of the lead is different from a pacing wire and is embedded into the epicardium
(black arrow).
Temporary epicardial pacing is commonly used following cardiothoracic surgery in which
the pacemaker leads are attached to the epicardium of the atrium and/or ventricle
and then connected to the external pacer.[6] These leads are generally used in temporary treatment of arrhythmias in the perioperative
period, and usually removed by 4 to 5 days after cardiac surgery by transcutaneous
retraction.[7] They are seen on radiographs as coiled up thin wires over the cardiac silhouette
and connected to two small cylindrical external metallic connectors on the surface
([Fig. 13]). Retained surgical swab can sometimes mimic epicardial-pacing wires. At discharge,
most of these leads are removed but some portion of the wires may be retained within
the deeper tissues. Such wires are often left behind without any major impact or complication.
Fig. 13 Frontal chest radiograph (A) and abdominal radiograph (B) in two patients postvalve replacement surgery. The temporary epicardial pacing leads
(white arrows) are seen connected to an external pacing device connector (black arrow).
Thin cylindrical small rods are also seen at the end of the wires (B); these are outside the skin and used to connect to the pacing device. Sternotomy
wires (dotted arrow) are seen in both images with a prosthetic aortic valve (star).
Mediastinal and pleural drains (black dotted arrows) are seen with central venous
line (black star) and endotracheal tube (black arrow head) in part A.
Temporary Transvenous Ventricular Pacing
Patients with acute hemodynamic compromise caused by bradycardia and/or episodes of
asystole should be considered for temporary cardiac pacing.[8] This often occurs in the setting of acute myocardial infarction. Transvenous pacing
is superior to the transcutaneous method because of patient comfort and stability
of pacing.[6] Central venous access is gained through femoral vein (in some cases via the jugular
system) and the pacing lead is advanced into the right atrium and then across the
tricuspid valve with the tip in the RV. On CXR, the lead can be seen coming along
the right paravertebral aspect and then angling along the diaphragm with the tip toward
the cardiac apex ([Fig. 14]). The lead may be displaced with the tip either along the RV outflow tract or in
the atrium.
Fig. 14 Frontal chest radiograph demonstrates temporary transvenous pacing leads (black arrow)
in the right ventricle. The access route is via femoral vein in part A and jugular vein in part B.
Loop Recorder
Loop recorder or reveal device is an insertable cardiac monitor used to record electrocardiogram
(ECG). These devices are used when long-term monitoring of cardiac rhythm is required.
These are patient-activated or automatically activated monitoring systems wherein
the recorded data are analyzed later for the evaluation of suspected cardiac arrhythmias.[7] Reveal devices are subcutaneously placed on the anterior chest wall (usually left
parasternal area extending to the midclavicular line between the first intercostal
space and the fourth rib or in the lower chest—V3 area, between the fourth and fifth
rib) ([Fig. 15]). These devices are usually magnetic resonance imaging compatible and can be confused
as a USB device in patients clothing. There are often no complications related to
these devices. Newer ECG monitoring/recording devices are also being tried in various
institutes as a mode of remote monitoring of patients in postoperative period ([Fig. 16])
Fig. 15 Frontal chest radiograph depicting loop recorder (REVEAL) device (black arrow) in
two patients (A and B). Also note prior sternotomy (star) and a hiatus hernia (dotted arrow) in part B.
Fig. 16 Frontal chest radiograph in a patient postaortic valve replacement (black arrow)
showing remote monitoring device (star), which records the electrocardiogram of the
patients though surface leads and beams the information via Bluetooth to a central
console. Also note right central venous line (white arrow).
Atrial Septal Defect (ASD) Closure Device
Atrial septal defects (ASD) can be closed by open surgical techniques or by percutaneous
catheter-guided means avoiding thoracotomy/sternotomy. The decision on closing of
an ASD depends on the patient symptoms and the amount of shunt from left to the right.
Smaller defects with insignificant shunts are often medically managed. Various types
of percutaneous ASD closure devices are in use, which are placed in position under
fluoroscopic, transesophageal echocardiographic (TOE), or intracardiac echocardiographic
imaging guidance. These are generally self-centering devices constructed from thin
nitinol wires tightly woven onto two flat discs with a connecting waist (sizes 4–40
mm). Polyester mesh is sewn into each disc and mesh to induce stasis and thrombosis.
On CXRs, they appear as one or two discs or two dots (corresponding to the ends of
the waist of the device)[4] ([Fig. 17]). The position of the device is described with respect to the thoracic vertebral
bodies on both frontal and lateral radiographs and with respect to the hilar–caval
line (drawn from anterior margin of right hilum to posterior margin of inferior vena
cava on the lateral view—expected position of the interatrial septum). In children,
these are generally projected over T7 to T9 vertebrae with the centers at or just
to the right of the vertebral spinous process.[4] On the lateral radiograph, the waist of the occluder is usually seen on the hilar–caval
line.
Fig. 17 Frontal chest radiograph in two patients with septal defect closure devices. (A) A device used to close an atrial septal defect (white arrow). The two discs of the
device are superimposed and only separate central radiopaque dots are seen. Ventricular
septal defect closure is seen in B with two discs (black arrow) and two central dots
clearly delineated due to the oblique nature of the defect. Also note mediastinal/pleural
drains (black dotted arrows), epicardial pacing leads (white dotted arrows), right
central line (white star), and endotracheal tube (arrowhead).
Ventricular Septal Defect (VSD) Closure Device
VSD is the most common congenital cardiac defect (it can be seen as an isolated anomaly
or in association with various other complex cardiac anomalies). Here again, catheter-based
closure devices can be used besides direct surgical closure. On CXR, VSD occluder
appears similar to ASD closure device (one or two radiopaque discs or two dots) but
in the region of the interventricular septum ([Fig. 17]).
Prosthetic Cardiac Valves
Prosthetic cardiac valves are implanted surgically or via transcutaneous route to
treat valvular stenosis or regurgitant pathologies. In younger patients, valve-sparing
surgeries are preferred wherever possible. Correct identification of prosthetic cardiac
valves is easy when clinical details or echocardiography reports are available. In
emergency, correct identification of the valve (without any prior history) on a CXR
may be useful as aortic valves are more susceptible to endocarditis and the mitral
valves to thrombotic complications.[9] On a posteroanterior CXR, an imaginary line is drawn from right cardiophrenic angle
to inferior aspect of left hilum and the aortic valve should lie above and the mitral
valve below this line. This however is not very reliable on an anteroposterior CXR.[9] The aortic valves are often oriented horizontally and valve orifice is often in
profile. In comparison, the mitral valve is vertical in location and the valve orifice
appears en face. The pulmonary valve prosthesis is much higher and lateral to the
other valves along the outflow tract. The tricuspid valve prosthesis is also seen
en face on frontal CXR and is seen along the right paravertebral region ([Fig. 18] and [Table 1]). Lateral views are also helpful in differentiating one valve from the other but
are often not performed in routine clinical practice.[10]
Table 1
Various parameters of valves on CXR helping in deciding their anatomical location
Type of valve
|
Valve orifice
|
Location with respect to line from right cardiophrenic angle to left hila
|
Location with respect to spine
|
Abbreviation: CXR, chest radiograph.
|
Aortic
|
In profile
|
Above
|
Left paravertebral
|
Mitral
|
En face
|
Below
|
Left paravertebral
|
Pulmonary
|
Variable
|
Above
|
Left paravertebral
|
Tricuspid
|
En face
|
Variable
|
Right paravertebral
|
Fig. 18 Frontal chest radiograph demonstrating various prosthetic valves. In part A, both aortic (white arrowhead and mitral valve (black arrow) can be seen. Notice
the en-face view of the mitral valve with the aortic valve being more horizontal in
nature. Prosthetic pulmonary valve can be seen at a higher location (dashed arrow)
in part B. Prosthetic tricuspid valve can be seen in parts C and D along the right paravertebral aspect (white arrow). Median sternotomy (white star)
and right central venous line (black dotted arrow) are also seen.
Ventricular Assist Device
Ventricular assist devices (VAD) are used to support the mechanical function of a
failing heart and can be biventricular or only for the left ventricle. They are used
for short term/temporary use (wherein the pump and controller are generally outside
the patient) or long term (in such situations the pump is implanted in a peritoneal
or abdominal wall pocket with external control and power). These devices have inflow
cannulae placed in the cardiac apex/right atrium/atrial appendage and outflow cannulae
placed in the aorta/main pulmonary artery or right pulmonary artery depending on whether
they are left or RV assist devices, respectively ([Fig. 19]). Biventricular assist devices have both sets of cannulae ([Fig. 20]). On CXR, these appear as short radiopaque cylindrical structures (especially the
left ventricular assist device [LVAD] inflow cannula at the LV apex on frontal CXR)
when seen enface or as radiopaque rings when seen end-on (the other mentioned cannulae
on the frontal CXR). Chest/mediastinal drains that mimic these cannulae are long tubes
with breaks in the radiopaque line indicating side holes. The pump, external controls,
and battery pack may or may not be seen on the CXR depending on where they are located.
Computed tomography (CT) may be a better modality for the assessment of complications
related to the LVAD, which include thrombosis and misplacement of the cannula ([Fig. 20]). CXR is useful in assessing the position of the device on serial films and associated
complications such as pleural effusions/pneumothorax.
Fig. 19 Frontal chest radiograph (CXR) in two patients with left ventricular assist device.
The pump can be seen (black arrow) with the inflow cannula (white dotted arrow) in
the left ventricle (A and B). Both patients (A and B) have biventricular pacing defibrillator devices in situ. The outflow tract is not
radiopaque and is often not visualized on a CXR.
Fig. 20 Frontal scannogram from a computed tomography (CT) examination (A) showing two ventricular assist devices, one for each ventricle with outflow and
inflow systems clearly visualized from the pump mechanism (black arrow and white arrowhead).
(B) CT examination of a patient with left ventricular assist device showing thrombus
in the left ventricular apex around the inflow cannula (black dotted arrow).
Vascular Devices
Intraaortic Balloon Pump
Intraaortic balloon pump (IABP) is a simple, cost-effective, and easily implantable
circulatory support device in patients requiring hemodynamic support either in cardiogenic
chock or at a higher risk of hemodynamic decompression during a high-risk coronary
intervention.[11] The pump inflates during cardiac diastole along with aortic valve closure, thereby
displacing the blood from the aorta into the peripheral vessels. Then it rapidly deflates
before the onset of systole. This results in improved diastolic pressure, reduced
systolic aortic pressure, and myocardial oxygen demand.[11] The tip of the IABP should be distal to the left subclavian artery and the proximal
portion of the balloon above the origin of the renal arteries. On CXR, the tip of
the IABP (seen as a linear radiopaque) should be above the carina at the level of
the arch. If the tip is lower than this, then the abdominal visceral branches of the
aorta can be compressed during inflation of the pump leading to visceral hypoxia ([Fig. 21])
Fig. 21 Frontal chest radiograph showing good position of the intraaortic balloon pump tip
(white arrow) in part A and low position of the tip in part B (black arrow). Temporary transvenous pacing wire is seen in part A (white dotted arrow), nasogastric tube (black dotted arrow). In part B, we can also see a coiled nasogastric tube (black dotted arrow) along with Swan-Ganz
catheter (white star) and chest drains (white dotted arrows) and a pacemaker with
a single right ventricular lead (black star). The endotracheal tube is also seen in
both images (arrowhead).
PDA Closure Device
Patent ductus arteriosus (PDA) is one of the causes of pulmonary arterial hypertension.
It contributes to 10% of the congenital cardiac anomalies.[1] It is more common in premature babies and females. Obstructive pulmonary diseases
and heart failure are the complications of untreated PDA. Surgical closure is the
treatment of choice when medical treatment fails. Surgical/interventional closure
can be done either by clips or coils. On frontal CXR, it is seen on left side of the
spine below the aortic arch. Different types of devices may be seen ([Fig. 22]).
Coarctation Stent
Coarctation of aorta accounts for 5 to 8% of congenital cardiac anomalies. Congestive
cardiac failure, stroke, aortic rupture, and infective endocarditis are associated
with late complications in untreated aortic coarctation.[12] Endovascular treatment avails minimally invasive alternative techniques to conventional
open surgeries with shorter hospital stay.
Intravascular stent grafts can be seen on plain radiographs as tubular mesh-like radio-opacity.
We should look for its position, extent, and continuity or any break ([Fig. 23]).
Fig. 22 Frontal chest radiograph showing different patent ductus arteriosus closure devices
(black arrow and dotted arrow) (A and B). The device shown in part B is similar to the atrial septal defect/ventricular septal defect closure device with
two discs and two central dots.
Fig. 23 Frontal chest radiograph with aortic stents in situ for coarctation (white arrow
in parts A and B). One can also notice the notching of the ribs inferiorly in both cases due to long-standing
coarctation and subsequent collaterals.
Pulmonary Artery Stent
Branch pulmonary artery stenosis can be seen in isolation or part of the other congenital
cardiac defects. RV hypertrophy and failure are the delayed complications. RV systolic
pressure >50% of aortic pressure or ipsilateral lung perfusion <35% are the indications
for intervention.[13] Balloon dilatation or stent placement is the treatment option. Early postoperative
pulmonary stenosis, postoperative kinking, external compression, elastic recoil, and
intimal tear are some of the indications for stenting. Stents may also be placed in
the RV outflow tract in cases of Tetralogy of Fallot. These stents can be seen on
plain radiographs as tubular mesh-like radio-opacity ([Fig. 24]). Complications of stenosis or underlying thrombosis are not properly seen on CXR
and CT is better for this.
Fig. 24 Frontal chest radiograph in different patients showing stents (white arrow) in the
right and left pulmonary artery (A) and in the right ventricular outflow tract (B). Both patients had Tetralogy of Fallot.
Arteriovenous Malformation/Embolization Clip
Transcatheter intervention for hemoptysis is now becoming the first line of management
in most cases. Bronchial artery and arteriovenous malformations are often the cause
of hemoptysis and these can be well visualized on CT and/or catheter angiography.[14] Different materials can be instilled into the culprit vessels such as glue (N-butyl
cyanoacrylate) or metallic clips. The glue is not radiopaque but is mixed with Lipiodol,
which is radiopaque and can be seen on CXR. The clips on the other hand are radiopaque
on their own and can be seen as well-defined structures along the path of intervention
([Fig. 25]).
Fig. 25 Frontal chest radiograph in a patient who underwent bronchial artery embolization
for hemoptysis with metal clips (white arrow) (A) and a zoomed-up view of the same (B). Also, note a peripherally inserted central catheter with its tip in right superior
vena cava (black dotted arrow).
Musculoskeletal Implants/Devices
Pectus Bar
Pectus excavatum is the most common congenital deformity of the chest wall accounts
for 1: 400 births. It is four times more common in males. The deformity is more pronounced
during the adolescent stage of life. Many are asymptomatic. Symptoms that can occur
in severe cases include dyspnea, chest pain, exercise intolerance, and psychological
problems.[15] Surgery in most of the cases is for cosmetic reasons. The main aim of surgery is
to correct the chest deformity, which improves the self-acceptance, breathing, and
cardiac function.[16] Two commonly used correction methods are highly modified Ravitch’s repair and Nuss
procedure. In both procedures, metallic bar is placed behind the sternum. Wound infection,
pneumothorax, overcorrection, and bar displacement/migration are some of the associated
complications.[17] Nuss procedure involves the placement of a U-shaped rod with lateral anchor points
and are well recognized. The Ravitch’s bar is a linear metal density bar with no anchor
points ([Fig. 26]). It can be confused as a foreign body if clinical history is not known.
Fig. 26 Ravitch’s bar is seen on the frontal chest radiograph (arrow) in a patient who was
treated for pectus deformity of the chest wall.
Spinal Fixation
Spinal instrumentation is done commonly to restore the spinal stability, correcting
the deformities, in degenerative diseases, trauma, infection, and neoplasms. These
can be anterior spinal fixation devices, posterior or posterolateral spinal fixation
devices, and interbody grafts.[18] These are easily recognized due to their paravertebral location ([Fig. 27]).
Fig. 27 Frontal chest radiograph in two patients with spinal fixation devices. (A)A thoracic spinal fixation device with a disc prosthesis (arrow) and (B) a cervicothoracic
spinal fixation device (arrow). Also note symmetrical well-defined increased density
in both breasts (arrowhead) in part B with clear margins in keeping with bilateral breast implants.
Sternotomy
Standard median sternotomy sutures are seen along the sternum and are well spaced
out. These are seen as a coiled metal wire without any breaks ([Figs. 3 ]and [5]). It is essential to ensure continuity of each of the wire knots and to make note
of any broken wire. In some cases, break in the sternal wires is often ignored, if
the patient has no symptoms. In patients who have infections along the sternum, elaborate
operations are often required for osteomyelitis and different varieties of sternal
fixation may be seen ([Figs. 28 ]and [29]).
Fig. 28 Frontal (A) and lateral (B) chest radiograph of a patient with unusual metal ware (arrow) for sternal fixation
in a patient poststernal osteomyelitis.
Fig. 29 Frontal chest radiograph in a patient (A) who underwent extensive reconstructive surgery of the sternum and the adjoining
costochondral junctions for osteomyelitis. (B) Patient with redo-sternotomy for aortic valve replacement (black arrow) showing
interconnecting sternal wires (white arrow). Also note the presence of right central
line (star), endotracheal tube (arrowhead) and nasogastric tube (black dotted arrow).
Rotator Cuff Anchors
Full-thickness rotator cuff tear repair is performed using suture anchors that produce
secure and effective soft tissue fixation to bone. Both metallic and biodegradable
suture anchors can be used and have similar results.[19] On CXR, they are projected over the humeral head and can be confused with clothing
artifacts ([Fig. 30]).
Fig. 30 Frontal chest radiograph zoomed view, showing metalware in right humeral head in
keeping with rotator cuff anchors (arrow).
Miscellaneous
Tracheobronchial Stents
Airway stenting is performed to relieve airway obstruction/strictures that are not
suitable for resection and reconstruction. They are made of metal or silicon. Some
of the silicon stents are not radiopaque and hence may not be seen on CXRs. Stents
can be covered or uncovered (appears like a mesh with perforations to allow for epithelial
growth). Some stents are expandable for easy insertion. The choice of stent type depends
on the location and type of lesion to be tackled. When seen on CXRs, they appear like
a cylinder within the lumen of trachea/bronchus ([Fig. 31]). It is useful to compare previous radiographs for ensuring the stent has not migrated
from its intended location.
Fig. 31 Frontal chest radiograph in a patient with extensive stents in the trachea and proximal
bronchi (arrow) in a patient with relapsing polychondritis.
Surgical Clips/Suture Material
Radiopaque surgical clips/suture materials can be seen in different locations. These
may be seen in lung parenchyma postthoracic surgery or open lung biopsy ([Fig. 32]). These may also be seen along the mediastinum in patients postcoronary artery bypass
surgeries whereby the clips are put in vessels ([Fig. 5]).
Fig. 32 Frontal chest radiograph in a patient after surgery for emphysematous lung disease.
Surgical suture material is seen along the edge of the right lung (white arrow), seen
better on the zoomed view (B).
Breast Implants
Prosthetic breast implants are increasingly used for breast augmentation surgeries,
postmastectomy, and cosmetic augmentation.[20] On CXR, these are recognized as areas of increased density in the lower part of
the chest and in many cases, they are not apparent. Nipple prosthesis may also be
present and are well recognized on CXR as radiopaque material. On long standing or
following untreated rupture of the prosthesis, there may be calcification of the breast
prosthesis where these can be well seen on CXR ([Figs. 27 ]and [33]).
Fig. 33 Frontal chest radiograph in a patient with bilateral breast implants (arrowhead)
(A) with much higher density than usual. (B) A single left breast implant (arrowhead) with a nipple prosthesis (black arrow).
Extracorporeal Membrane Oxygenation
Extracorporeal membrane oxygenation (ECMO) provides circulatory support to patients
with respiratory, cardiac, or combined failure for days to a week. Two types of ECMO
are used VV (venovenous) system is used to support the lungs and give them time to
recover and VA (venoarterial) system is used for cases of cardiac or cardiopulmonary
failure.[21] On CXR, the ECMO cannulae are different from the normal lines because of their corrugated
appearance/large size. The cannulation for the VV ECMO is usually in the jugular vein
and femoral vein. The cannulation in the VA ECMO can be obtained centrally (right
atrium and ascending aorta) or peripherally (jugular/femoral vein and carotid or femoral
artery) ([Fig. 34]).
Fig. 34 Frontal chest radiograph in patients on extracorporeal membrane oxygenation (ECMO)
support. (A) Central cannulation (black arrow) in the right atrium (right paravertebral location)
and ascending aorta (left paravertebral location) in a venoarterial ECMO system. A
Swan-Ganz catheter can be seen (white dotted arrow) along a left central line (black
dotted arrow). (B) Peripheral cannulation in the jugular and femoral vein (black arrow) in a venovenous
ECMO system. Left central line (black dotted arrow), endotracheal (arrowhead) and
nasogastric tubes (white arrow) are also seen.
Deep Brain Stimulator
Deep brain stimulation is a well-established therapy for movement disorders. In this
technique, thin wires are placed intracranially and are connected to a neurostimulator
(very similar to the pacing box in the pacemaker) via an insulated wire that is implanted
under the skin. The stimulator is placed in the subcutaneous tissue of the chest wall
and can be seen on the CXR with wires leading along the neck superiorly ([Fig. 35]).
Fig. 35 Frontal chest radiograph showing the pulse generator box in the right lower thoracic
cavity (subcutaneous plane) with pacing leads going cephalad in a patient with deep
brain stimulation system (arrow).
Transesophageal Echocardiography Probe
There is increasing use of TOE in routine practice especially in sick patients or
during cardiac surgeries. Some ill patients often have a TOE probe in situ for a longer
time and the probes may be seen in a CXR. It is seen as a linear structure along the
path of the esophagus and is much thicker and denser compared with a nasogastric tube
([Fig. 36]).
Fig. 36 Frontal chest radiograph in a patient postcardiac surgery showing a transesophageal
echocardiography probe in situ (black arrow). Also note the intraaortic balloon pump
tip (black dotted arrow) in appropriate position and the Swan-Ganz line with tip in
right intermediate pulmonary artery (white arrow). Multiple drains (white dotted arrow)
and endotracheal tube (arrowhead) are also seen.
Surgical Gauze
Retained surgical gauze material after cardiothoracic surgery can be a major complication
and needs to be identified at the earliest. It is customary to do a swab count at
the end of the surgery to avoid such a mishap but such mistakes can still happen.
CXR is a good test in depicting these surgical gauze materials as they are embedded
with radiopaque linear structures. These radiopaque materials are irregular and do
not follow any vascular anatomy ([Fig. 37]).
Fig. 37 Frontal chest radiograph in patients with inadvertent retained surgical gauze material
(black arrow). Note radiopaque linear material in keeping with the gauze (A–C). Also note nasogastric tube (white dotted arrow), tracheostomy (star), right central
line (black dotted arrow), and right chest drain (arrowhead) in part B.