Keywords
pelvic fractures - canine - skeletal traction device
Introduction
Pelvic fractures are common in small animals constituting 16% of all fractures in
dogs and 25% of all fractures in cats.[1] Although the cause of pelvic fractures is often not known, road traffic accidents
are considered the most common reason in both species. Other high-energy traumatic
incidents such as crush injury or falling from a height account for some pelvic injuries.[2] Young animals are overrepresented and this may be associated with roaming behavior.[2]
[3] Anatomically the pelvis may be thought of as a ‘box-like structure’ with struts
running through its interior and supported by a large muscle mass.[4] Because of this rigid structure, displacement of fragments is not usually seen unless
there are at least two or often three fractures, which commonly occur in specific
locations and follow several predictable patterns.[5] Fractures involving the pubic symphysis are the exception and may result in avulsion/fracture
of the ischial tuberosity or iliac wing, or non-displaced stress fractures.[4] The nature of pelvic fractures necessitates thorough evaluation of organs and organ
systems. In a study of dogs involved in road traffic accidents, 71% had injuries to
multiple organs, including pulmonary, soft tissue, spinal and urinary tract trauma,
as well as haemoabdomen and cardiac dysrhythmias.[6] In particular, cranial or craniomedial displacement of the iliac bone has been associated
with damage to the L6 and L7 nerve roots and injury to the lumbosacral trunk. Thus,
the timing for repair of pelvic fractures depends largely on the degree of collateral
injuries, and surgery often has to be delayed until the animal is stabilized and able
to undergo general anaesthesia.[7] In animals with life-threatening injuries, surgical repair may be delayed for up
to 7 to 10 days after injury.[4] After this time, muscle contraction and early fibrosis at the fracture site render
adequate fracture reduction difficult or impossible.[4] The aim of this case report was to introduce and evaluate the use of a skeletal
traction device for the delayed repair of pelvic fractures that would not have been
difficult to reduce using conventional techniques.
Case Report
Two dogs were referred for pelvic fractures; case 1 suffered a road traffic accident,
whereas in case 2, non-unions were present as complications of previous bilateral
double pelvic osteotomy. Case 1 was a 9-year-old, mixed-breed dog weighing 30 kg and
case 2 was a 6-month-old, Bernese Mountain dog weighing 35 kg. At the time of referral,
the fractures had been present for 12 days in case 1 and 14 days in case 2. The results
of physical and neurological examinations, haematological and serum biochemical analyses,
of thoracic and abdominal radiography and ultrasonography were within normal limits.
Ventrodorsal and lateral radiographic views of the pelvis were obtained in both cases.
Case 1 had bilateral fractures, which included a long oblique fracture of the left
wing of the ilium, fracture of the pubis and transverse fracture of the right iliac
body. Case 2 had undergone bilateral double pelvic osteotomy and had in addition a
fracture of the right ischial table; the complications were associated with implant
failure due to the use of undersized screws and also the fact that the screw orientation
was perpendicular to the plate and not divergent to resist the pull out.
Surgical Technique
The anaesthetic protocols were based on individual patient requirements and included
methadone (Semfortan, Dechra, Italy) (0.3 mg/kg, intramuscularly]) and meloxicam (Metacam,
Boehringer Ingelheim, Italy) (0.2/mg/kg, intramuscularly) for perioperative analgesia.
Cefazoline (Teva, Italy) (22 mg/kg, intravenously) was administrated 30 minutes preoperatively
and every 90 minutes thereafter for a total of three doses during the surgical procedure.
General anaesthesia was induced with propofol (Propovet, Zoetis, Italy) (2–4 mg/kg,
intravenously) and maintained with 2% isoflurane (Piramal Critical Care, West Drayton,
UK) in oxygen. Intraoperative analgesia was provided with methadone. Both dogs were
placed in lateral recumbency with the forelimbs secured to the operating table. A
hanging limb technique was used, and the surgical site was aseptically prepared for
surgery.
In case 1, the most displaced fragment was treated first to facilitate reduction on
the opposite site. Fluoroscopy was used intraoperatively to guide and confirm reduction
in the fractures. The skeletal traction device, commercially available (AdMaoira,
Reggio Emilia, Italy: https://www.ad-maiora.eu/en/), is L-shaped, and the long component has a micrometric movement that allowed stand
elongation. One end of the stand was secured to the operating table rail with clamps.
The short component consisted of a stainless-steel tube for attachment of the tie-band
([Fig. 1]). A small stab incision was made over the ischial tuberosity on both sides, and
a 3 mm Steinman pin (Alcyon Srl, Cherasco, Italy) was carefully introduced on each
side perpendicular to the long axis of the pelvis while avoiding the nerves and vascular
supply of the region. The pins were inserted to a distance of ∼2 cm from the caudal
edge of the ischial tuberosity to avoid possible fracture at this site, which was
considered the anchorage point ([Fig. 2]). A sterilized tie-band was looped at the base of each of the pin–skin interfaces
to evaluate the feasibility of traction and to apply equal traction on both sides
of the pelvis ([Fig. 3C]). The forelimbs were secured to the edge of the operating table in a cranial direction
to counteract the traction forces and to avoid translation; injury to the patient
was avoided as described for long bone traction, paying attention that during the
traction the pins do not bend and create soft tissue damage.[8]
[9] This attachment was considered the opposition point. After a second aseptic preparation
of the surgical field, a standard lateral approach to the iliac body was done. The
load required for distraction of the fractured segment was applied by an operating
room assistant who increased the load at the surgeon's request with 5-minute rest
intervals between each load increase until visual and palpable reduction was evident
and confirmed by fluoroscopy ([Fig. 3D]). Distraction force to the two bilateral Steinmann pins was applied simultaneously
to avoid translation of the opposite side and deformation of pelvic conformation.
Load application for reduction was performed by turning the handle on the traction
device in a clockwise direction, increasing the length of the long component of the
L-shaped stand. During the procedure, contamination of the surgical field was prevented
because the caudal end of the device and the assistant were outside the surgical field
and the portion of the device closest to the surgical field was covered by the surgical
drapes. Once the fracture on the right side had been reduced, a 3.5-mm locking bone
plate (Fixin, InTrauma, Rivoli, Italy) was applied; after wound closure, the procedure
was repeated on the left side after re-positioning the dog in right lateral recumbency.
Postoperative radiographs were obtained to score alignment and apposition as anatomical,
near-anatomical, good, fair or poor[10] and confirmed proper implant positioning ([Fig. 3E, F]), also the alignment of the bone segment was considered good to near-anatomical
without evidence of pelvic canal narrowing.
Fig. 1 The L-shaped skeletal traction device mounted on the operating table (A, B). The device has a long (*) and a short component (#) and a clamp (^) for attaching
the device to the table.
Fig. 2 Bone model showing the position of the Steinmann pin, on the ischiatic tuberosity,
in lateral, ventral and dorsal views (A–C). The connection between the pin and the skeletal traction device is also shown (D–F).
Fig. 3 Ventrodorsal and lateral radiographic views of the pelvis in case 1 showing a long
oblique fracture of the left ilial wing, fracture of the pubis and transverse fracture
of the right iliac body with fracture of the right ischiatic table (A, B). Preoperative bilateral Steinmann pin placement in the ischial tuberosities and
their connection to the skeletal traction device (C). Perioperative fluoroscopic image used to estimate the degree of distraction of
the pelvis (D). Postoperative radiographic view showing proper implant positioning with no narrowing
of the pelvic canal (E, F).
In case 2, a staged surgery, 2 days apart, was performed to minimize surgical time
and the risk of infection. For the first intervention, the patient was placed in left
lateral recumbency to reduce and repair the iliac body fracture on the right. All
surgical steps involving placement of the 3-mm Steinman pins (Alcyon Srl) were done
using the technique described in case 1. After removal of the double pelvic osteotomy
plate and screws, traction was applied until reduction was confirmed visually and
by means of palpation and fluoroscopic examination. The fracture was stabilized with
two 3.5-mm locking plates (Fixin) placed parallel to one another. Postoperative radiographs
showed proper implant placement and near-anatomical bone alignment ([Fig. 4E, F]). The iliac body fracture on the left side was repaired 2 days later using the same
surgical technique, with postoperative radiographs revealing good alignment and no
narrowing of the pelvic canal. Both dogs had methadone and meloxicam for 48 hours
postoperatively and received tramadol ( Formevet, Italy), gabapentin (Pfizer, Italy)
and meloxicam orally for 14 days after the discharge.
Fig. 4 Ventrodorsal and lateral radiographic views of the pelvis in case 2 showing implant
failure on the right side, with plate pull-out and failure of the plate screws on
the left side (A, B). Preoperative placement of a Steinmann pin and its connection to the skeletal traction
device (C, D). Postoperative radiographic view of successful reduction and fixation of the left
and right iliac body fracture using locking plates (E, F).
Discussion
Fractures of the pelvis occur most commonly in the ilium,[11] and sacroiliac fracture/luxation is a significant source of pain and instability
and the most common pelvic injury associated with neurological deficits.[12] In human as well as in veterinary surgery, pelvic fractures require adequate reduction
and stabilization to control pain and to reduce blood loss.[7]
[13]
[14] Pelvic fractures can pose two potential problems requiring surgical intervention:
significant narrowing of the pelvic canal by the caudal fragment, sometimes with compromise
of the urinary bladder and large intestines; and trauma to the lumbosacral nerve trunk,
which branches into the sciatic nerve just medial to the midbody of the ilium.[15] Early aggressive fracture management using a combination of internal and external
fixation techniques has become the gold standard in human medicine. Although attempts
to provide fracture fixation within 24 hours of injury is a first-line approach, particularly
for femur fractures, this strategy has been questioned for pelvic fractures, which
are often accompanied by multiple injuries.[16]
[17] It currently appears that, in human medicine, the safest window for internal fixation
is 6 days to 2 weeks after injury when the systemic inflammatory response has subsided
and before callus formation limits reduction.[18] Traction tables are commonly used in human trauma patients and standardized reproducible
techniques are routinely used for fracture reduction. These techniques include proper
patient positioning, specific instrumentation and application of intraoperative skeletal
traction. The rationale behind the use of a skeletal traction device is to counteract
muscle contraction to facilitate reduction in the bone fragments and regain normal
limb length. The entire application and reduction in the described fractures using
a skeletal traction device are based on the ligamentotaxis technique with the goal
of using tension for alignment of bone fragments. This technique could possibly be
used to facilitate specific minimally invasive surgical procedures. We found that
the use of a skeletal traction device intraoperatively greatly facilitated reduction
in the bone fragments before application of a plate. To our knowledge, there are no
studies on the use of a skeletal traction device for the treatment of pelvic fractures
in veterinary medicine. Rovesti et al reported excellent results after using a skeletal
traction device for long bone fracture reduction and repositioning of misaligned bone
fragments associated with the fracture pattern or bone segment traction in dogs.[8] In the present study, reduction and realignment of the bone fragments were achieved
through coordination by both, the primary surgeon working intraoperatively to manipulating
the bone fragments and the assistant moving the short segment of the traction device.
Complications such as impairment of vascular and nerve integrity attributable to excessive
traction were not encountered. In human medicine, neurological damage associated with
application of skeletal traction has been reported in patients with no neurological
signs before traction.[19] However, the interval between the initial injury and surgical intervention appeared
to impact on the duration and force of traction needed to realign the bone fragments.
Although the application of anchorage and opposition point appears to be straightforward,
slippage of those points could potentially result in tissue trauma. Moreover, if the
procedure requires prolonged traction, it may be advisable to temporarily use bone
clamps to maintain reduction thereby allowing reduction in the load exerted by the
skeletal traction device and preservation of tissue perfusion. Once reduction and
stabilization are achieved with bone plates, the traction load may be modified by
the surgeon. We suppose that young patients with delayed fracture repair may require
more force during skeletal traction because of rapid callus formation at the fracture
site. Preparation and installation of the traction system may appear time-consuming,
but the reduction in surgical time during bone apposition and osteosynthesis was comparable
to the time needed for a conventional surgical procedure time of ∼60 to 90 minutes.
In human medicine, a dual pelvic sling used to treat pelvic fractures provided effective
suspension, traction and compression when applied to the iliac crest and trochanteric
regions. The force of traction can be varied by adjusting the degree of lateral displacement
of each sling, the direction of pull and the pulley-weight arrangement.[20] However, this type of skeletal traction depends on the nature of the fracture-dislocation
and the degree and direction of displacement determined on anteroposterior, inlet
and tangential radiographic views.[21] In our cases, the only parameter of substantial importance was the presence of an
intact ischial tuberosity connected to the iliac body, which provided the anchorage
point for pin insertion and traction. The indications for using a skeletal traction
device in our cases differed from those cited in a previous study[8]; we used the device to overcome muscle contraction, which resulted from the delayed
nature of the fracture repair, to facilitate open reduction and internal fixation.
Our findings emphasize that longer standing pelvic fractures in dogs constitute complex
injuries that are often difficult to manage surgically. The limitations of this study
are on one hand inherent to a two-case report and on the other, to the relative lack
of experience of the authors with regard to the forces required for skeletal traction.
Evaluations are ongoing by the authors to better estimate the traction force, times
and intervals necessary to reduce and align segments in various pelvic fracture patterns,
as well as overall surgery times. In conclusion, this report shows that using a skeletal
traction device greatly facilitated reduction and stabilization of longer standing
pelvic fractures in two dogs.