CC BY-NC-ND 4.0 · Rev Bras Ortop (Sao Paulo) 2019; 54(03): 261-267
DOI: 10.1055/s-0039-1688756
Artigo Original | Original Article
Sociedade Brasileira de Ortopedia e Traumatologia. Published by Thieme Revnter Publicações Ltda Rio de Janeiro, Brazil

Biomechanical Analysis of Two Types of Humerus Supracondylar Fracture Fixation in Anatomical Model[*]

Artikel in mehreren Sprachen: português | English
1   Serviço de Ortopedia Pediátrica, Hospital São Vicente de Paulo, Passo Fundo, RS, Brasil
,
Ticiano Dozza Posser
1   Serviço de Ortopedia Pediátrica, Hospital São Vicente de Paulo, Passo Fundo, RS, Brasil
,
Charles Leonardo Israel
2   Laboratório de Bioengenharia, Universidade de Passo Fundo, Passo Fundo, RS, Brasil
,
Leandro de Freitas Spinelli
2   Laboratório de Bioengenharia, Universidade de Passo Fundo, Passo Fundo, RS, Brasil
3   Serviço de Ortopedia e Traumatologia, Hospital Santa Casa de Misericórdia de Porto Alegre, Porto Alegre, RS, Brasil
,
Luis Gustavo Calieron
1   Serviço de Ortopedia Pediátrica, Hospital São Vicente de Paulo, Passo Fundo, RS, Brasil
,
Jung Ho Kim
1   Serviço de Ortopedia Pediátrica, Hospital São Vicente de Paulo, Passo Fundo, RS, Brasil
› Institutsangaben
Weitere Informationen

Address for correspondence

Marcos Ceita Nunes, MD
Rua Uruguai, 2050, Passo Fundo, RS, Brasil. Cep: 99010-112

Publikationsverlauf

28. September 2017

02. Juli 2018

Publikationsdatum:
27. Juni 2019 (online)

 

Abstract

Objective To analyze the stability of humerus supracondylar fracture fixation with Kirschner wires comparing intramedullary and lateral (Fi), and two parallel lateral wires (FL) fixation in experimental models, to define which configuration presents greater stability.

Methods A total of 72 synthetic humeri were cross-sectioned to simulate the fracture. These bones were divided into two equal groups and the fractures were fixed with parallel Kirschner wires (FL) and with a lateral and intramedullary (Fi) wire. Then, the test specimens were subjected to stress load tests on a universal test machine, measured in Newtons (N). Each group was subdivided into varus load, valgus, extension, flexion, external rotation and internal rotation. An analysis of the data was performed comparing the subgroups of the FL group with their respective subgroups of the Fi group through the two-tailed t test.

Results The two-tailed t test showed that in 4 of the 6 evaluated conditions there was no significant statistical difference between the groups (p > 0.05). We have found a significant difference between the group with extension load with a mean of 19 N (FL group) and of 28.7 N (Fi group) (p = 0.004), and also between the groups with flexural load with the mean of the forces recorded in the FL group of 17.1 N and of 22.9 N in the Fi group (p = 0.01).

Conclusion Fixation with one intramedullary wire and one lateral wire, considering loads in extension and flexion, presents greater stability when compared to a fixation with two lateral wires, suggesting similar clinical results.


#

Introduction

Supracondylar fracture is more common in the 4- to 7-year-old age group,[1] corresponding to two-thirds of children hospitalized for elbow trauma and to 30% of the fractures in this population.[2] Due to the particular anatomy of the elbow and to the ossification order in the growth nuclei, the supracondylar fracture virtually always behaves in extension patterns, with posterior medial, posterior lateral and flexion displacements. The displacement degree is defined by the direction of the deforming force, by the position of the limb during the trauma, and by the magnitude of this force.[3] [4]

Gartland classified this fracture in three types according to the displacement degree; it is agreed that grade 1 fractures require conservative treatment.[5] Some papers describe conservative techniques, that is, reduction and immobilization, in grade 2 and 3 fractures.[6] However, many authors describe reduction and percutaneous fixation as the gold standard for displaced fractures.[7] [8] As such, there is no consensus as to the best positioning of Kirschner wires in the stabilization of this fracture.[9] Fixation with a cross-wired configuration provides better stability, but there is a risk of iatrogenic injury of the ulnar nerve. The configuration with two lateral wires showed lower stability of the crossed wires and lower incidence of ulnar nerve lesions; in addition, it is technically more challenging, since the space for wire placement both in divergent and parallel directions is small. However, both configurations have similar clinical results.[10] [11] [12] [13]

In 1991, Bertol et al[14] published the technique of supracondylar fractures fixation with posterior medial deviation using an intramedullary Kirschner wire inserted just lateral to the olecranon and another one lateral at the epicondyle entry, in a presumably easier technique, since it optimizes the lateral spine space.

Numerous biomechanical studies compare different positional configurations of Kirchner wires in the stabilization of supracondylar humerus fracture,[13] [15] [16] but there are no reports analyzing the configuration with an intramedullary and a lateral wire. The present study aims to compare fixation techniques using an intramedullary wire or two parallel lateral wires.


#

Materials and Methods

The test specimens of the present study were 72 synthetic humeri (model 3022B – left humerus with medullary canal and spongy material) (Nacional Ossos, Jaú, SP, Brazil), which were equally cross-sectioned, parallel to the articular surface in the coronal plane, with a distal guided saw; the section passed into the olecranon fossa at 3 centimeters from the distal humeral edge, simulating a supracondylar fracture ([Fig. 1]). The cross-section was selected because 80% of the supracondylar fractures have a transverse pattern in lateral radiographs[17]; moreover, the fracture obliquity causes instability.[18]

Zoom Image
Fig. 1 Test specimens: cross-section of a synthetic humerus at its distal portion, through the olecranon fossa at 3 centimeters from the distal edge of the bone, simulating a supracondylar fracture.

The sectioned synthetic humeri were divided into two groups according to fixation: a group fixed with two lateral wires (FL) and a group with fixed with an intramedullary wire and a lateral wire (Fi). All of the models were submitted to the anatomical reduction and fixation with 2.0 mm Kirschner wires. Each group had a standard fixation model to assure that the fixations were identical. In the FL group, the fixation was performed with 2 2.0 mm Kirschner wires entering laterally at the epicondyle, with the most distal wire at the lower edge of the lateral epicondyle and the proximal wire 1 cm above the former, parallel to the axis of the humeral shaft, fixed on the opposite cortical layer, 3 cm above the fracture line.

In the Fi group, the fixation was also performed with 2 2.0 mm Kirschner wires, with the 1st wire entering 2 mm lateral to the lateral border of the trochlea, at the trochlear groove, thus becoming intramedullary and introduced up to the transition between the middle third and the distal third of the humerus, 11 cm from the distal humeral end, and the 2nd wire inserted in the center of the lateral epicondyle at a 30° angle to the humeral axis, crossing the 1st wire at 2 cm from the fracture line and fixed in the opposite cortical layer at 3 cm from the fracture line.

Fixations were aided by a perforator and fluoroscopy. All of the specimens were compared to their respective standardized models by fluoroscopy, complying with the previously mentioned fixation criteria, and assuring the similarity between them ([Fig. 2]).

Zoom Image
Fig. 2 Test specimens compared with their respective standardized models at fluoroscopy, complying with the fixation criteria and ensuring the similarity between them.

Specimens that did not comply with the fixation criteria were excluded. After the fixation, the humeri from both groups were sent to the Engineering Laboratory, where, together with a collaborating engineer, each group was divided into subgroups according to the performed load tests: subgroup 1, varus load; subgroup 2, valgus load; subgroup 3, load in extension; subgroup 4, load in flexion; subgroup 5, load in internal rotation; and subgroup 6, load in external rotation.

Load tests were performed on a universal tensile testing machine, model UPM 200 (3022B, left humerus with medullary canal and spongy material, Nacional Ossos, Jaú, SP, Brazil), and an HBM U9B (3022B, left humerus with medullary canal and spongy material, Nacional Ossos, Jaú, SP, Brazil) load cell (20KN = 1mV/V). The test measures the load generated in Newtons (N) during the continuous displacement promoted by the traction test machine at a speed of 1 mm/s, with a maximum established displacement of 10 cm, which promotes an angulation of up to 45° in the specimen with rotation fulcrum at the fracture line ([Fig. 3]).

Zoom Image
Fig. 3 Universal tensile testing machine model UPM 200 and an HBM U9B load cell (20 KN = 1 mV/V)

A support for the anatomical coupling of the distal humerus was developed, allowing the application and measurement of loads in bone models at a point 10 cm proximal to the fracture line up to a 45° of angulation and/or material failure ([Fig. 4]). A mechanism to create rotational forces from the load established by the tensile testing machine was also developed ([Fig. 5]). Rotational loads were applied until the breakage of the bone models.

Zoom Image
Fig. 4 Applied and measured loads on the specimens, at a point 10 cm proximal to the fracture line, until reaching a 45° angle and/or material failure.
Zoom Image
Fig. 5 Mechanism for the application of rotational loads.

The data generated by the load cell in each bone model show that, during displacement, the force in N initially increases until it reaches a plateau (which is related to the higher recorded forces); next, the applied force decreases, which is related to the bone model breakage and/or to a 45° displacement. In this way, the force in N when reaching this plateau was defined as the variable to be analyzed, that is, the maximum force recorded during the displacement, at the end of the linear region of the graph.

The sample size was calculated in the PEPI (Programs for Epidemiologists) software, version 4.0, and based on the study by Bloom et al. [19] For a significance level of 5%, 90% power, and an estimated standard deviation [SD] of 3.5 with a mean difference of 8N, a minimum total of 6 parts per subgroup was obtained, totaling 36 per group.

The data analysis was performed with Microsoft Office Excel 2010 software (Microsoft Corporation, Redmond, WA, USA), comparing FL subgroups to their respective Fi subgroups through two-tailed t tests. The present study does not have conflicts of interests.


#

Results

Loading tests results to compare the stability of the two wire configurations are represented in N in [Table 1]. The two-tailed t test showed that there was no significant statistical difference in 4 of the 6 loads applied (p < 0.05) ([Table 1]).

Table 1

Group (FL)

Group (Fi)

Two-tailed t test

Varus (N)

28.7 ± 3.5

30.7 ± 4.9

p = 0.230

Valgus (N)

20.6 ± 5.2

22.9 ± 3.4

p = 0.240

Extension (N)

19.0 ± 3.4

28.7 ± 4.0

p = 0.004

Flexion (N)

17.1 ± 1.2

22.9 ± 4.0

p = 0.015

Internal Rotation (N)

12.55 ± 1.2

11.7 ± 2.6

p = 0.256

External Rotation (N)

11.2 ± 1.8

11.6 ± 1.0

p = 0.292

In the bone models submitted to the varus load, the mean of the highest recorded forces during displacement in FL group was of 28.7 N, with a SD of 3.5 N. In the Fi group, the mean force was 30.7 N, with a SD of 4.9 N. Thus, the two-tailed t test did not reveal a statistically significant difference between these groups (p = 0.23) ([Fig. 6]).

Zoom Image
Fig. 6 Graphic representation of test results of loads in varus, valgus, extension and flexion and internal and external rotation.

In the models submitted to valgus load, the mean of the highest recorded forces in the FL group was of 20.6 N, with a SD of 5.2 N. In the Fi group, the mean value was of 22.9 N, with a SD of 3.4 N. As with the varus load, the two-tailed t test did not reveal a statistically significant difference between the groups (p = 0.24) ([Fig. 6]).

In addition, there was no statistically significant difference between the groups at the load tests in internal and external rotation (p = 0.25 and p = 0.24, respectively) ([Fig. 6]).

There was a statistically significant difference between the 2 groups in the extension load tests (p = 0.004), with a mean of the highest recorded forces in the FL group of 19.0 N and a SD of 3.4 N, whereas the Fi group presented a value of 28.7 Newtons and a SD of 4.0 N ([Fig. 6]). Thus, during the constant displacement established by the test machine, a greater force was generated and recorded by the load cell in the Fi group compared with the FL group. As such, we can also suggest that the configuration with an intramedullary wire and a lateral wire provides greater stability in extension loads when compared with the configuration with two parallel lateral wires.

Models submitted to the flexion load also showed a significant statistical difference between the groups (p = 0.01), with the mean of the highest recorded forces in the FL group of 17.1 N and a SD of 1.2 N, and a mean value of 22.9 N and a SD of 4 N in the Fi group ([Fig. 6]).


#

Discussion

The main goals of the treatment of displaced supracondylar fractures are anatomic reduction and a secure fixation with no angular deformities. This is usually achieved with closed reduction and percutaneous fixation.[20] [21] [22] Fixation requires full attention to the clinical and radiological examination of the contralateral elbow, in addition to true orthogonal projections at fluoroscopy and the consideration of well-described radiographic parameters for total correction of the deformity.[23] Defective consolidations are also related to inadequate fixations and technical errors during the procedure.[19]

Several biomechanical studies have demonstrated that the cross-wire fixation has a greater rotational stability than the lateral wiring fixations,[15] but with a higher risk of iatrogenic injury of the ulnar nerve.[12] Bloom et al[16] reported that three lateral divergent pins provides the same resistance as two crossed wires, which are more resistant than two lateral wires, but, in most cases, there is not enough space for lateral pinning.[19] In a prospective randomized clinical trial comparing lateral and crossed fixation techniques for the treatment of type III humeral supracondylar fractures, Kocher et al[11] did not find a significant difference between both groups regarding radiographic and clinical outcome. In another prospective randomized study, Blanco et al[24] found no significant radiological differences between crossed and lateral wiring fixation.

Our study shows that the technique with intramedullary wire presents a greater resistance under flexion and extension loads than the technique with lateral wires; at other loads, the results are similar. In the former technique, the first step after achieving a suitable reduction is the introduction of the intramedullary wire,[14] which blocks the forces in axial direction, mainly flexion and extension, safely allowing the correction of the remaining rotational deformities, that is, this technique tolerates a rotational adjustment after the precise reduction in the axial direction, which is not possible with lateral wiring. As such, the intramedullary wire fixation facilitates anatomical reduction, which maximizes the stability of all fixation configurations.[19] In addition, intramedullary wire fixation maintains a greater lateral space for wire placement.

Since the clinical results of the two crossed wires technique are similar to those obtained with two lateral wires,[1] [24] [25] [26] we can assume that, according to the present mechanical study, the clinical results of the fixation with an intramedullary wire are equivalent to those provided by these techniques; however, a randomized clinical trial is required to confirm this assumption.

Some limitations of the present study should be recognized. Although the use of synthetic models for mechanical analysis of fracture reduction techniques is common in the literature, these investigations do not consider the variability in fracture patterns nor the anatomy with the surrounding periosteum that may contribute to fragment stability.[15] [27] Furthermore, the physiological loads acting on the elbow are certainly more complex than the single axis of the load test directions used in the present study. In addition, the pins were placed in an ideal situation, without considering the difficulty of intraoperative insertion, which cannot be simulated. The design of the present study does not allow direct comparisons of the applied loads in models with organic bones, allowing only the comparison between the fixation techniques for these fractures.


#

Conclusion

In the present study, the intramedullary wire fixation provides a greater stability under flexion and extension loads when compared with the lateral wiring fixation, with similar results under other applied loads, suggesting acceptable clinical results, as already proven by Bertol et al.[14] As such, it is an excellent option for the configuration of Kirschner wires when treating these fractures.


#
#

Conflitos de Interesses

Os autores declaram não haver conflitos de interesses.

* Worked performed at the Hospital São Vicente de Paulo, Passo Fundo, RS, Brazil.


  • Referências

  • 1 Cheng JC, Shen WY. Limb fracture pattern in different pediatric age groups: a study of 3,350 children. J Orthop Trauma 1993; 7 (01) 15-22
  • 2 Kasser JR, Beaty JH. Supracondylar fractures of the distal humerus. In: Beaty JH, Kasser JR, Wilkins KE, Rockwood CE. , eds. Rockwood and Wilkins' fractures in children. 6th ed. Philadelphia: Lippincott Williams and Wilkins; 2006: 543-89
  • 3 Omid R, Choi PD, Skaggs DL. Supracondylar humeral fractures in children. J Bone Joint Surg Am 2008; 90 (05) 1121-1132
  • 4 Mahan ST, May CD, Kocher MS. Operative management of displaced flexion supracondylar humerus fractures in children. J Pediatr Orthop 2007; 27 (05) 551-556
  • 5 Barton KL, Kaminsky CK, Green DW, Shean CJ, Kautz SM, Skaggs DL. Reliability of a modified Gartland classification of supracondylar humerus fractures. J Pediatr Orthop 2001; 21 (01) 27-30
  • 6 Izadpanah M. [Closed treatment of supracondylar fractures of the humerus: a modification of Blounts technique (author's transl)]. Arch Orthop Unfallchir 1973; 77 (04) 348-358
  • 7 Wilkins K, Beaty J. Fractures in children. 4th ed. Philadelphia: Lippincott-Raven; 1996
  • 8 Flynn JC, Zink WP. Fractures and dislocations of the elbow. In: MacEwen GD, Kasser JR, Heinrich SD. , eds. Pediatric fractures. A practical approach to assessment and treatment. Baltimore: Willians &Wilkins; 1993: 133-64
  • 9 Kasser JR, Beaty JH. Supracondylar fractures of the distal humerus. In: Beaty JH, Kasser JR, Wilkins KE, Rockwood CE. , eds. Rockwood and Wilkins' fractures in children. 5th ed. Philadelphia: Lippincott Williams and Wilkins; 2004: 594-95
  • 10 Topping RE, Blanco JS, Davis TJ. Clinical evaluation of crossed-pin versus lateral-pin fixation in displaced supracondylar humerus fractures. J Pediatr Orthop 1995; 15 (04) 435-439
  • 11 Kocher MS, Kasser JR, Waters PM, Bae D, Snyder BD, Hresko MT. , et al. Lateral entry compared with medial and lateral entry pin fixation for completely displaced supracondylar humeral fractures in children. A randomized clinical trial. J Bone Joint Surg Am 2007; 89 (04) 706-712
  • 12 Rasool MN. Ulnar nerve injury after K-wire fixation of supracondylar humerus fractures in children. J Pediatr Orthop 1998; 18 (05) 686-690
  • 13 Brauer CA, Lee BM, Bae DS, Waters PM, Kocher MS. A systematic review of medial and lateral entry pinning versus lateral entry pinning for supracondylar fractures of the humerus. J Pediatr Orthop 2007; 27 (02) 181-186
  • 14 Bertol P, Monteggia GM, Paula MD. Fixação percutânea das fraturassupracondilianas do úmero na criança. Rev Bras Ortop 1991; 26 (03) 48-51
  • 15 Larson L, Firoozbakhsh K, Passarelli R, Bosch P. Biomechanical analysis of pinning techniques for pediatric supracondylar humerus fractures. J Pediatr Orthop 2006; 26 (05) 573-578
  • 16 Bloom T, Robertson C, Mahar A, Pring M, Newton PO. Comparison of supracondylar humerus fracture pinning when the fracture is not anatomically reduced. In: The Annual Meeting of the Pediatric Orthopaedic Society of North America, Hollywood, FL, 2007 May 23-26
  • 17 Nand S. Management of supracondilar fractures in children. Int Surg 1972; 177: 203-209
  • 18 Bahk MS, Srikumaran U, Ain MC, Erkula G, Leet AI, Sargent MC. , et al. Patterns of pediatric supracondylar humerus fractures. J Pediatr Orthop 2008; 28 (05) 493-499
  • 19 Bloom T, Robertson C, Mahar AT, Newton P. Biomechanical analysis of supracondylar humerus fracture pinning for slightly malreduced fractures. J Pediatr Orthop 2008; 28 (07) 766-772
  • 20 Kurer MH, Regan MW. Completely displaced supracondylar fracture of the humerus in children. A review of 1708 comparable cases. Clin Orthop Relat Res 1990; (256) 205-214
  • 21 Royce RO, Dutkowsky JP, Kasser JR, Rand FR. Neurologic complications after K-wire fixation of supracondylar humerus fractures in children. J Pediatr Orthop 1991; 11 (02) 191-194
  • 22 Ito N, Eto M, Maeda K, Rabbi ME, Iwasaki K. Ultrasonographic measurement of humeral torsion. J Shoulder Elbow Surg 1995; 4 (03) 157-161
  • 23 Pring M, Rang M, Wenger D. Elbow-distal humerus. In: Pring M, Rang M, Wenger D. , eds. Rang's Children's Fractures. Philadelphia, PA: Lippincott Williams & Wilkins; 2005: 95-118
  • 24 Blanco JS, Gaston G, Cates T, Busch MT, Schmitz ML, Schrader T. , et al. Lateral pin versus crossed pin fixation in type 3 supracondylar humerus fractures: a randomized prospective study. In: The Annual Meeting of the Pediatric Orthopaedic Society of North America, Hollywood, FL, 2007 May 23-26.
  • 25 Foead A, Penafort R, Saw A, Sengupta S. Comparison of two methods of percutaneous pin fixation in displaced supracondylar fractures of the humerus in children. J Orthop Surg (Hong Kong) 2004; 12 (01) 76-82
  • 26 Gordon JE, Patton CM, Luhmann SJ, Bassett GS, Schoenecker PL. Fracture stability after pinning of displaced supracondylar distal humerus fractures in children. J Pediatr Orthop 2001; 21 (03) 313-318
  • 27 Schwartz A, Oka R, Odell T, Mahar A. Biomechanical comparison of two different periarticular plating systems for stabilization of complex distal humerus fractures. Clin Biomech (Bristol, Avon) 2006; 21 (09) 950-955

Address for correspondence

Marcos Ceita Nunes, MD
Rua Uruguai, 2050, Passo Fundo, RS, Brasil. Cep: 99010-112

  • Referências

  • 1 Cheng JC, Shen WY. Limb fracture pattern in different pediatric age groups: a study of 3,350 children. J Orthop Trauma 1993; 7 (01) 15-22
  • 2 Kasser JR, Beaty JH. Supracondylar fractures of the distal humerus. In: Beaty JH, Kasser JR, Wilkins KE, Rockwood CE. , eds. Rockwood and Wilkins' fractures in children. 6th ed. Philadelphia: Lippincott Williams and Wilkins; 2006: 543-89
  • 3 Omid R, Choi PD, Skaggs DL. Supracondylar humeral fractures in children. J Bone Joint Surg Am 2008; 90 (05) 1121-1132
  • 4 Mahan ST, May CD, Kocher MS. Operative management of displaced flexion supracondylar humerus fractures in children. J Pediatr Orthop 2007; 27 (05) 551-556
  • 5 Barton KL, Kaminsky CK, Green DW, Shean CJ, Kautz SM, Skaggs DL. Reliability of a modified Gartland classification of supracondylar humerus fractures. J Pediatr Orthop 2001; 21 (01) 27-30
  • 6 Izadpanah M. [Closed treatment of supracondylar fractures of the humerus: a modification of Blounts technique (author's transl)]. Arch Orthop Unfallchir 1973; 77 (04) 348-358
  • 7 Wilkins K, Beaty J. Fractures in children. 4th ed. Philadelphia: Lippincott-Raven; 1996
  • 8 Flynn JC, Zink WP. Fractures and dislocations of the elbow. In: MacEwen GD, Kasser JR, Heinrich SD. , eds. Pediatric fractures. A practical approach to assessment and treatment. Baltimore: Willians &Wilkins; 1993: 133-64
  • 9 Kasser JR, Beaty JH. Supracondylar fractures of the distal humerus. In: Beaty JH, Kasser JR, Wilkins KE, Rockwood CE. , eds. Rockwood and Wilkins' fractures in children. 5th ed. Philadelphia: Lippincott Williams and Wilkins; 2004: 594-95
  • 10 Topping RE, Blanco JS, Davis TJ. Clinical evaluation of crossed-pin versus lateral-pin fixation in displaced supracondylar humerus fractures. J Pediatr Orthop 1995; 15 (04) 435-439
  • 11 Kocher MS, Kasser JR, Waters PM, Bae D, Snyder BD, Hresko MT. , et al. Lateral entry compared with medial and lateral entry pin fixation for completely displaced supracondylar humeral fractures in children. A randomized clinical trial. J Bone Joint Surg Am 2007; 89 (04) 706-712
  • 12 Rasool MN. Ulnar nerve injury after K-wire fixation of supracondylar humerus fractures in children. J Pediatr Orthop 1998; 18 (05) 686-690
  • 13 Brauer CA, Lee BM, Bae DS, Waters PM, Kocher MS. A systematic review of medial and lateral entry pinning versus lateral entry pinning for supracondylar fractures of the humerus. J Pediatr Orthop 2007; 27 (02) 181-186
  • 14 Bertol P, Monteggia GM, Paula MD. Fixação percutânea das fraturassupracondilianas do úmero na criança. Rev Bras Ortop 1991; 26 (03) 48-51
  • 15 Larson L, Firoozbakhsh K, Passarelli R, Bosch P. Biomechanical analysis of pinning techniques for pediatric supracondylar humerus fractures. J Pediatr Orthop 2006; 26 (05) 573-578
  • 16 Bloom T, Robertson C, Mahar A, Pring M, Newton PO. Comparison of supracondylar humerus fracture pinning when the fracture is not anatomically reduced. In: The Annual Meeting of the Pediatric Orthopaedic Society of North America, Hollywood, FL, 2007 May 23-26
  • 17 Nand S. Management of supracondilar fractures in children. Int Surg 1972; 177: 203-209
  • 18 Bahk MS, Srikumaran U, Ain MC, Erkula G, Leet AI, Sargent MC. , et al. Patterns of pediatric supracondylar humerus fractures. J Pediatr Orthop 2008; 28 (05) 493-499
  • 19 Bloom T, Robertson C, Mahar AT, Newton P. Biomechanical analysis of supracondylar humerus fracture pinning for slightly malreduced fractures. J Pediatr Orthop 2008; 28 (07) 766-772
  • 20 Kurer MH, Regan MW. Completely displaced supracondylar fracture of the humerus in children. A review of 1708 comparable cases. Clin Orthop Relat Res 1990; (256) 205-214
  • 21 Royce RO, Dutkowsky JP, Kasser JR, Rand FR. Neurologic complications after K-wire fixation of supracondylar humerus fractures in children. J Pediatr Orthop 1991; 11 (02) 191-194
  • 22 Ito N, Eto M, Maeda K, Rabbi ME, Iwasaki K. Ultrasonographic measurement of humeral torsion. J Shoulder Elbow Surg 1995; 4 (03) 157-161
  • 23 Pring M, Rang M, Wenger D. Elbow-distal humerus. In: Pring M, Rang M, Wenger D. , eds. Rang's Children's Fractures. Philadelphia, PA: Lippincott Williams & Wilkins; 2005: 95-118
  • 24 Blanco JS, Gaston G, Cates T, Busch MT, Schmitz ML, Schrader T. , et al. Lateral pin versus crossed pin fixation in type 3 supracondylar humerus fractures: a randomized prospective study. In: The Annual Meeting of the Pediatric Orthopaedic Society of North America, Hollywood, FL, 2007 May 23-26.
  • 25 Foead A, Penafort R, Saw A, Sengupta S. Comparison of two methods of percutaneous pin fixation in displaced supracondylar fractures of the humerus in children. J Orthop Surg (Hong Kong) 2004; 12 (01) 76-82
  • 26 Gordon JE, Patton CM, Luhmann SJ, Bassett GS, Schoenecker PL. Fracture stability after pinning of displaced supracondylar distal humerus fractures in children. J Pediatr Orthop 2001; 21 (03) 313-318
  • 27 Schwartz A, Oka R, Odell T, Mahar A. Biomechanical comparison of two different periarticular plating systems for stabilization of complex distal humerus fractures. Clin Biomech (Bristol, Avon) 2006; 21 (09) 950-955

Zoom Image
Fig. 1 Corpos de prova: úmeros sintéticos seccionados transversalmente em sua porção distal, com a secção passando na fossa do olécrano a 3 cm da borda distal do úmero, simulando uma fratura supracondiliana.
Zoom Image
Fig. 2 Corpos de prova comparados com seus respectivos modelos padronizados através de fluoroscopia, seguindo os critérios de fixação e assegurando a semelhança entre eles.
Zoom Image
Fig. 3 Máquina de ensaio universal de tração modelo UPM 200 e uma célula de carga HBM U9B (20 KN = 1 mV/V).
Zoom Image
Fig. 4 Cargas aplicadas e mensuradas nos corpos de prova, em um ponto 10 cm proximal ao traço de fratura, até que atingissem 45° de angulação e/ou quebra do material.
Zoom Image
Fig. 5 Mecanismo para aplicação das cargas torcionais.
Zoom Image
Fig. 1 Test specimens: cross-section of a synthetic humerus at its distal portion, through the olecranon fossa at 3 centimeters from the distal edge of the bone, simulating a supracondylar fracture.
Zoom Image
Fig. 2 Test specimens compared with their respective standardized models at fluoroscopy, complying with the fixation criteria and ensuring the similarity between them.
Zoom Image
Fig. 3 Universal tensile testing machine model UPM 200 and an HBM U9B load cell (20 KN = 1 mV/V)
Zoom Image
Fig. 4 Applied and measured loads on the specimens, at a point 10 cm proximal to the fracture line, until reaching a 45° angle and/or material failure.
Zoom Image
Fig. 5 Mechanism for the application of rotational loads.
Zoom Image
Fig. 6 Gráfico mostrando os resultados dos testes das cargas em varo, valgo, extensão e flexão e rotação interna e externa.
Zoom Image
Fig. 6 Graphic representation of test results of loads in varus, valgus, extension and flexion and internal and external rotation.