Keywords
single-sided deafness - cochlear implant - sound localization
Introduction
Cochlear implantation (CI) is a globally accepted surgical procedure for an increasing
number of patients with severe to profound hearing loss. The criteria for CI have
expanded over the past decades. Single-sided deafness (SSD) is characterized by severe
to profound sensorineural hearing loss in one ear and normal hearing in the contralateral
ear. CI is a well-established option for hearing rehabilitation in patients with SSD.[1] Nowadays, patients with SSD make up a nonnegligible percentage of all CI recipients
in countries with cost coverage for this group.
Beside CI treatment, the options for SSD include conventional hearing aids, conventional
contralateral routing of sound (CROS) hearing aids, and bone-anchored hearing aids
(BAHAs). The latter offer patients pseudo-binaural hearing but cannot restore a profoundly
deaf ear. Compared with pseudo-binaural hearing solutions (BAHA or CROS), CIs restore
binaural hearing and can improve hearing abilities significantly.[1] As with adults, children with SSD have objectively and subjectively binaural and
bilateral hearing benefits from CI.[2]
Difficult hearing situations, such as hearing in noise or hearing at greater distances,
are facilitated by binaural hearing.[3] Improvements in hearing abilities with a CI are especially beneficial for open set
speech recognition in noise and for sound localization.[4]
Sound localization is a fundamental auditory skill. Localization abilities are based
on interaural time differences (ITDs) and interaural level differences (ILDs), both
of which require the use of binaural hearing, which people with SSD do not have. However,
to make full advantage of ITD cues, artificial electrical stimulation by the CI needs
to be translated into neural activity.
Nevertheless, before CI, several candidacy considerations have to be obtained. Friedmann
et al point out absolute indications (late-stage unilateral Menière's disease, an
“at risk” only hearing ear, and pediatric progressive hearing loss) for CI.[5] Several other factors, such as age and duration of hearing loss, are well known
and have to be considered, as the benefits from a CI might not be satisfactory.
Since these factors influence the outcome overtime, based on patient experiences with
bilateral hearing loss and CIs, it is of special interest to look at the group of
patients with SSD. It can be assumed that, based on the asymmetric hearing situation,
different factors than those in the previous group influence the outcome.
The aim of this study was to investigate a group of patients who presented themselves
after unilateral CI for SSD with nonsatisfactory hearing ability in our department.
Materials and Methods
The study was reviewed and supported by the institutional review board (IRB-ukb-HNO-2016/04).
Preoperatively, computed tomography (CT) scan of the temporal bone and magnetic resonance
imaging (MRI) of the neurocranium were performed. Cases with any malformation were
excluded. In cases of long-term hearing loss (> 20 years), promontorial testing was
performed. CI included a mastoidectomy and a posterior tympanotomy as well as a round
window or modified round window approach to access the cochlea. The senior author
performed all surgeries. Out of a series of 65 patients with SSD, seven patients with
subjective poor performance were included in this study. All patients met the criteria
for unilateral CI in an SSD situation with normal hearing in the contralateral ear.
Audiologic testing included speech perception testing in a sound field using the German
Freiburger test for monosyllabic words at 65 dB sound pressure level (SPL) with a
signal coming from the front. The Freiburger test for monosyllabic words was measured
at the initial fitting and after 3, 6, and 12 months. All patients underwent audiologic
rehabilitation after the initial fitting. In some patients, localization testing was
performed in a room with 47 loudspeakers in a semicircle with a 2.35-m distance to
the patient; all speakers were covered. Two different stimuli with 750-milliseconds
noise bursts were applied. Low-frequency bursts were between 300 and 1,200 Hz and
high-frequency bursts between 2,000 and 8,000 Hz at 65 dB. Sound was presented from
a determined angle. The patient had to point in the exact direction of the sound origin.
Results
Seven patients (three females and four males) were included in this study. Mean age
at implantation was 60.82 ± 16.75 years with a mean duration of hearing loss of 6.98
± 11.25 years. Six of seven patients had a speech recognition test available up to
1 year after implantation. The Freiburger monosyllabic word score (FWS) showed mean
values of 14.29 ± 18.58 at the initial fitting, 23.57 ± 19.30 at 3 months, 24.17 ±
24.58 at 6 months, and 31.67 ± 31.09 at 12 months. Analysis of variance (ANOVA) for
repeated measures showed no significant difference in FWS over time with p = 0.285. Three patients completed the lateralization testing.
-
Patient 1 underwent CI surgery 2 years after idiopathic sudden hearing loss. He presented
with subjective poor performance in the right ear after CI. FWS was 45% at the initial
fitting and dropped to 25% at the 3-month visit. Over the first year, the patient
improved to 55% at 6 months and 75% at 12 months. Retrospective analysis of cerebral
MRI before implantation revealed several microembolisms in the area of the right middle
cerebral artery. Sound localization testing at high and low frequencies showed for
the right (CI)-side lateralization to the left side at 10, 30, 60, and 90 degrees.
In summary, lateralization ability did not improve with the CI ([Fig. 1A, B]).
-
Patient 2 had a progressive hearing loss on the right side after multiple middle ear
surgeries due to a cholesteatoma. He underwent implantation of a vibrant sound bridge,
which improved the hearing ability temporarily. After 7 years of hearing loss, he
underwent CI. FWS improved from 0% at the initial fitting to 15% at 3 months, 30%
at 6 months, and 35% at 12 months. Because of an unsatisfactory hearing outcome, the
patient wore the audio processor only occasionally, even during auditory rehabilitation.
-
Patient 3 underwent CI on the left side 18 months after sudden hearing loss due to
an assumed rupture of the round window membrane. Preoperative CT and MRI scans were
inconspicuous. Postoperative radiologic examination revealed that the electrode array
translocated from the scala tympani into the scala vestibuli. FWS increased from 0%
at the initial fitting to 35% at 3 months but decreased to 20% after 12 months. During
auditory rehabilitation, the patient claimed adaption problems (voices sounded like
hammer blows) and started slowly to retreat socially.
-
Patient 4 was referred to our department with bacterial meningitis due to acute mastoiditis
on the right side. Four days after mastoidectomy, a placeholder was implanted as the
patient reported substantial hearing loss on the right side evidenced by pure-tone
audiometry (PTA). Two months after meningitis, a CI was implanted without any complications.
After the initial fitting and during the year of rehabilitation, the patient did not
report any hearing sensation. An MRI revealed cerebral residuals after meningitis
with cerebral microangiopathia, which led to an impairment of the central auditory
pathway and an additional cerebral ischemic insult ([Fig. 2A, B]).
-
Patient 5 had 4 years of unilateral hearing loss following a sudden hearing loss before
CI. The preoperative cerebral magnet response imaging (cMRI) showed a lacunar infarct
on the right side of the frontal lobe. FWS increased from 35% at the initial fitting
to 60% at 3 months and remained rather steady until 12 months. Sound localization
testing in low frequencies on the right (CI) side showed a deviation to the left side
at 5 degrees. At 30 degrees, localization testing on the right side was out of the
normal distribution. In the high-frequency range, at all tested frequencies, sound
provided to the right ear was lateralized to the left side, indicating the inability
to localize with the CI ([Fig. 3A, B]).
-
Patient 6 had a history of unilateral hearing loss for 2 years after a car accident,
which resulted in a temporal bone fracture and cerebral bleeding, resulting in a lesion
in the left pontine area. She reported poor performance and showed an FWS score of
15% at the initial fitting and at 3 months, which decreased to 10% at 6 months and
0% at 12 months. Localization testing in the low- and high-frequency ranges showed
abrogated localization ability in both the ears; most presented sounds were localized
in the portion between 50 and −50 degrees ([Fig. 4A, B]).
-
Patient 7 was implanted after progressive hearing loss over 15 years. She reported
low speech perception at the initial fitting and only wore the audio processor 2 hours
a day. FWS scores were 15% at the initial fitting and 5% at 3 months. The preoperative
MRI showed severe subcortical white matter lesions in both the hemispheres and linear
signal accentuations periventricularly.
Fig. 1 Localization testing in the (A) low (LF) and (B) high frequencies (HF) in patient 1. The patient has a cochlear implant on the right
side.
Fig. 2 MRI scan before (A) and after (B) cochlear implantation on the right side. T2-weighted images after cochlear implantation
include the artifact caused by the magnet. The lacunar infarct is seen in the pontine
area.
Fig. 3 Localization testing in the (A) low (LF) and (B) high frequencies (HF) in patient 5. The patient has a cochlear implant on the right
side.
Fig. 4 Localization testing in the (A) low (LF) and (B) high frequencies (HF) in patient 6. The patient has a cochlear implant on the right
side.
Discussion
Cochlear implantation for patients with severe to profound hearing loss has become
a standard surgical procedure in recent years. However, the criteria and patient populations
have changed. Hearing abilities can certainly be improved by a CI, especially for
open set speech recognition in noise and for sound localization.[1]
[4] In contrast to the pseudo-binaural benefits of different treatment options, adults
with SSD significantly benefitted from a CI.[2] Zhang et al showed that the bilateral integration of acoustic and electric auditory
stimulation via the normal hearing ear and the contralateral CI improved performance
in speech perception.[6]
We present a small case series of nonsatisfying outcomes and poor performance after
CI implantation in patients with unilateral deafness. The etiologies of the poor CI
performance were variable and could be separated into three types: preoperative, operative,
and postoperative.
Preoperative reasons included patients 1, 5, 6, and 7. Patient 1 showed several microembolisms
preoperatively on MRI scans in the area of the right middle cerebral artery. Although
scoring satisfactorily on the FWS test, the sound localization test revealed a missing
ability to localize sound with the normal hearing ear and the CI. Cerebral microembolisms
can directly influence central auditory pathway transmission and may influence the
CI outcome. Patients 5, 6, and 7 showed intracerebral lesions as well. Patient 5 had
a lacunar infarct area on the right side of the frontal lobe and patient 6 had a lesion
in the left pontine area. The pons contains the cochlear nuclei ascending fibers,
which are a part of the auditory pathway. Disturbance of these fibers can result in
defective transmission, which may lead to limited speech perception and limited sound
localization. Lesions in the frontal lobe should not influence speech perception and
sound localization, but it could not be completely excluded that further lesions (although
not seen on MRI) impaired the auditory pathway. Patient 7 had white matter lesions
in the subcortical frontoparietal part of the brain and early-stage degenerative cerebral
microangiopathy. These focal lesions may impair brain connectivity, which has recently
been termed “connectional diaschisis” and reflects changes that occur to brain regions
distant from the location of the lesion.[7] While connections to and from lesioned areas are likely to suffer owing to lost
afferents,[8] connectional diaschisis can also involve connectivity changes between distal regions
that are not directly connected to the lesion site. Therefore, the outcome remains
unclear and is not easily predictable. As the CI has more advantages than CROS or
bone conduction devices in patients with unilateral hearing loss,[1] the indication for the CI should be favorable in those cases with potential structural
damage.
Operative reasons included patient 3. This patient showed a scalar dislocation of
the electrode array on postoperative radiologic images. Electrode extrusion into the
scala vestibuli causes disappointing audiologic outcomes as word scores correlate
negatively with the number of electrodes in the scala vestibuli.[9]
[10] Speech recognition results, up to 1 year after implantation, were poor in this patient.
As preoperative MRI scans revealed no cerebral pathology and the duration of deafness
was short, the scalar rearrangement of the electrode array can be regarded as crucial.
Postoperative reasons for poor performance were found with patient 4. Bacterial meningitis
can result in cochlear obstruction secondary to labyrinthitis ossificans.[11] Labyrinthitis ossificans can impede CI and result in incomplete insertion.[12] Patient 4 suffered from bacterial meningitis due to acute mastoiditis. He received
a placeholder and underwent CI with full electrode insertion and no sign of labyrinthitis
ossificans after 3 months. Nevertheless, the patient did not have any speech perception
due to lesions in the central auditory pathway and stopped using the audio processor
after 1 year. Further diagnostics included a cMRI, which revealed cerebral residuals
after meningitis with cerebral microangiopathy and a recent postoperative cerebral
ischemic insult.
Various reasons were found to impair the functional outcome after CI in this study
group. To provide a sufficient and high-quality CI program, all factors surrounding
implantation should be assessed. These especially include pre- and postoperative radiologic
evaluations as well as audiologic tests. The magnet within the CI induces an artifact.
Minimization of the artifact can be achieved by careful placement of the device at
the right angle and distance to the external auditory canal and by selecting specific
technical scanning sequences on MRI scans.[13]
This study has some limitations. Despite the anatomic and pathophysiologic conditions,
psychologic and psychophysical reasons may influence outcomes, benefits, and quality
of life with a CI.[14] This study addresses only anatomic and pathophysiologic conditions.
The majority of CI recipients benefit from the implant. As the indication for CI has
been extended, patients with SSD have become a certain part of the patient collective.
The benefits for patients with SSD include sound localization and speech comprehension
in the majority of subjects.[15] The authors present here a small selected group of individuals who did not benefit
from a CI in terms of speech recognition and/or sound localization. For the first
time, the authors have shown in single cases, presumably related to specific defects
in the auditory pathway, that although improved speech perception can be achieved,
localization abilities do not develop after CI surgery in SSD.
Conclusion
Cochlear implantation for the treatment of patients with SSD has become a standard
procedure. Despite some cases with poor performance after CI, the majority of patients
perform well with the implant. Nevertheless, careful selection of patients and pre-
and postoperative radiologic evaluations are needed.
Ethical Approval
All procedures performed in studies involving human participants were in accordance
with the ethical standards of the institutional and/or national research committee
and with the 1964 Helsinki declaration and its later amendments or comparable ethical
standards.
Informed Consent
Informed consent was obtained from all individual participants included in the study.