Keywords
diathermy - electrosurgery - prevention - surgical trainees
Since its invention by Bovie and first use by Cushing in the removal of intracranial
tumors in 1927, electrosurgery is now an integral part of modern surgery.[1] Injuries and equipment loss resulting from diathermy use are much more common than
what we might think.[2] The true incidence is probably under reported and also difficult to determine. It
is currently estimated that around 500 to 600 surgical fires occur annually in the
United States.[3] In spite of this, not all the surgical trainees or even consultants have a good
understanding of electrosurgery.[4] Approximately, two-thirds of the injuries may not be detected during the procedure.[5] Unfortunately, this has not only implications on the patient care but also immensely
increases the financial burden on the health institutes costing them millions of dollars
every year.[3]
Biophysics of Diathermy and Its Effects on Cells
Biophysics of Diathermy and Its Effects on Cells
To understand the etiology and pathophysiology behind electrosurgery, it is initially
important to understand the working of an electrosurgical unit and its different components.
This will aid us to effectively manage and more importantly prevent electrosurgical
injuries. An electrosurgical circuit includes electrosurgical unit (ESU), active electrode,
patient, and the dispersive electrode. The ESU receives alternative current (AC) from
the wall supply (50–60 Hz) and converts it into electromagnetic current in excess
of 200 kilohertz (KHz).[6] Most modern ESUs are capable of producing current up to 850 KHz[7] that is transferred to the patient via an active electrode (handheld device). The
dispersive electrode (diathermy plate) then carries the current from the patient back
to ESU, thereby forming an isolated circuit, which is the present-day model without
requiring any grounding. This is in contrast to the initial models, which were ground-referenced
circuits. Thus, the radiofrequency electrosurgery is always bipolar, it is only the
active electrode that is either monopolar or bipolar. An active electrode does not
develop heat by itself in contrary to what many believe. The heat is produced in the
cells and tissues, which is then transferred back to the electrode that makes it hot.
The dispersive electrode must be at least 70 cm2 in area to carry sufficient current back to ESU to avoid inadvertent thermal injury
at the site. The dispersive electrode has two electrodes on a single plate, each individually
acting as a sensor in case of nonuniform application of the pad. There is a sequential
conversion of electromagnetic energy to kinetic energy to thermal energy within the
cells and tissues. This whole process is referred to as and actually is “diathermy.”[8] Cauterization on the other hand is a mere transfer of heat from one conducting source
to another.
The ESU has primarily two different outputs or waveforms. These include low-voltage
continuous output and high-voltage interrupted output. These two outputs bring about
vaporization, and desiccation and coagulation of the tissues. The low voltage is commonly
referred to and many a times labeled on the ESU as “cut” mode (yellow button or pad
on ESU) and is a misnomer. Similarly, high-voltage output is commonly labeled as “coagulation”
mode (blue button or pad) that is also a misnomer. There are various other settings
to modulate these two outputs as needed. Some of the common modulations include “cut,”
“blend,” “desiccate and coagulate,” and “fulgurate or spray.” Cut is linear vaporization
caused by moving the electrode in a linear or curvilinear manner. Blend is modulating
the duty cycle on low-voltage output. Duty cycle is the percentage of time occupied
by the waveform or output. In low-voltage continuous waveform, the duty cycle is 100%.
By modulating the duty cycle, voltage could be slightly increased or decreased.[9] In any given ESU, there are levels, whereby one can set a duty cycle. The higher
the level (numerically, e.g., level 3 compared with level 1), shorter the duty cycle
and higher the voltage. The low-voltage output irrespective of modulation is 600 to
700 V producing temperatures up to 1000°C. In contrary, high-voltage output, which
has a duty cycle of approximately 6%, is 3 to 8 kV producing low temperatures.[10]
The energy generated within the cells result, essentially, in two types of reaction
depending on the temperature that the energy achieves. Tissue reaction varies at different
frequencies. At a high frequency of more than 100 KHz, the muscles and nerves do not
depolarize. Desiccation refers to efflux of water molecules out of the cells resulting
in their shrinkage. Coagulation refers to protein denaturation. Desiccation and coagulation
occur simultaneously at 60 to 90°C. Vaporization occurs above 100°C and refers to
conversion of water to gas resulting in rapid expansion of cell volume and rupture
of cell membrane and wall. Both vaporization and desiccation and coagulation are results
of application of low-voltage continuous current. The difference lies in the manner
the active electrode is applied to the tissues. If the electrode is used in a noncontact
manner close to the tissues, it results in vaporization due to development of higher
temperatures. On the other hand, if the electrode is used in a contact manner with
the tissue, it results in desiccation and coagulation.
Fulguration refers to superficial coagulation over a wider surface area. It is also
known as spray mode. Here, the active electrode is used in a noncontact manner and
is a result of high-voltage interrupted current with a very low duty cycle. In practice,
this is the mode employed to desiccate and coagulate tissues in general using a monopolar
electrode; the electrode is used in a contact manner. The reason is that it leads
to superficial desiccation over a wider area that is often sufficient for nonvascular
tissues or very small blood vessels. To produce a uniform seal in a blood vessel,
low voltage must be used, which unlike high-voltage waveform, coagulates the vessel
slowly and produces deeper desiccation with higher temperatures. This effect is observed
well with the use of bipolar active electrode or direct coupling technique.
Complications of Electro Surgery
Complications of Electro Surgery
There exist a variety of injuries caused by electrosurgery. Various authors and guidelines
group them in different ways. We present here a more practical approach and have grouped
all the possible complications into four sections: thermal-, smoke-, fire-, and explosion-related
injuries. In addition, patients with implantable electromagnetic devices need a special
mention due to the fact that these patients are increasing in number, and the electrosurgical
devices interact differently and cause specific injuries in the presence of these
devices.
As with any mechanical or electrical device, when an issue arises in relation to the
equipment, there are various approaches one could apply to solve it, especially if
the equipment comprises of various elements. However, it is crucial to have a strong
foundation and understanding of not only the biophysics of the equipment but also
the pathophysiology and mechanism of the injury or adverse event. Henceforth, we have
detailed not just the injury related to electrosurgery here, but also the mechanism
of injury along with an approach to solve both the issue and prevent the injury ([Table 1]).
Table 1
Mechanism and solutions to various electrosurgery-related injuries
Injury
|
Mechanism and physics
|
Solution/Preventative strategies
|
Thermal
|
From active electrode
|
|
|
Inappropriate contact (too much power or using wrong setting) with target or neighboring
tissue leading to direct injury
|
Correct usage of electrode with recommended/proper settings. Use the lowest possible
power (W)
Use audible activation to avoid unnecessary prolonged contact with tissues or requesting
power to be increased
Complete the communication loop between the surgeon and the scout nurse changing
the settings
|
|
Inadvertent activation of the active electrode
Plastic sleeves may melt and cause thermal damage to underlying tissue deep to the
drapes
Wetness in the diathermy holder may increase resistance and cause diversion of current
|
Avoid accidental stepping on the foot pedal (in laparoscopic surgery)
Remove electrode from within patient's body when not using (in laparoscopic surgery)
Use rigid plastic holder (not plastic sleeves) for active electrode, thereby ensuring
it is not wet and there are no other instruments in the holder that may also contribute
to injury from diversion of current
|
|
Inappropriate placement of hot electrode on the drapes leading to transmission of
heat in patient's skin
|
Place active electrodes in the plastic holder
|
|
Lateral extension of current on wet skin prepared with skin preparation containing
alcohol
|
Wait for skin preparation to dry before using diathermy
|
|
Vertical extension of current along a vascular pedicle or adhesions
|
Relax on the pedicle prior to activating the electrode
Use of shorter periods of activation (1–2 s)
|
|
From dispersive electrode
|
|
|
Loss of contact of the electrode with the patient
|
Ensure uniform contact of the electrode (both the split electrodes)
Ensure skin is clean, dry, and shaved
Replace the electrode with a new one if it needs repositioning
|
|
Inappropriate application of electrode causing diversion of current
|
Avoid bony prominences
Do not place adjacent to ECG leads
|
|
From diversion of current
|
|
|
Insulation failure leading to dispersion of high-density current. Smaller the break
in insulation, higher the density of current dispersed
|
Regular check of instruments and their insulation by industry engineers using active
electrode monitoring systems
|
|
Direct coupling injury: Direct coupling is transfer of current from one conductive
source to another. Type of indirect injury: Injury could be to the patient or the
surgeon.
Inadvertent transfer of current through small breaks that may happen in gloves, particularly
an issue for surgeons wearing single glove. The gloves and hand in addition increase
impedance encouraging lateral flow of current (15% of new and 50% of used gloves have
holes)
Instrument in contact with the active electrode may not be completely in view (in
laparoscopic surgery) and in contact with a nontarget tissue
Increased risk in SILS due to close proximity of instruments
|
Avoid unnecessary contact of the active electrode with another conductive source
(e.g., retractors)
When using direct coupling intentionally (as with sealing a blood vessel), ensure
that the conductive source is not in contact with a nontarget tissue or another conductive
source. Make sure that the active electrode is brought into contact with the conductive
source (as in the case of forceps) below the level of the handgrip
Surgeons need to wear double gloves
Tip of the electrode and the conductive source must always be in view
Extra care need to be taken in SILS while crossing the instruments and ensuring that
the working segment of all the active instruments are always in view
|
|
Capacitive coupling injury: Capacitive coupling is transfer of current along the
length of the active electrode by means of displacement (e.g., when the insulator
becomes dielectric), most commonly in laparoscopic surgery
Factors increasing the risk: High voltage (coagulation mode), open activation (electrode
not in contact with tissue), and factors that increase impedance (wrapping of cable
around a metal clamp, such as towel clip and activation over previously coagulated
tissue)
|
Use the lowest power setting possible
Activate electrode only after making contact with tissue
Avoid factors that increase impedance to flow of current
|
|
Alternate site injury when patient is in contact with another high conductance element
such as metal (rare nowadays with isolated generators)
|
Ensure that patient is not in contact with any metal that he is not supposed to be
Extra care need to be taken while using fixed metal retraction system to ensure that
there is no contact of the active electrode with any part of that retractor
|
Smoke
|
Noxious gases irritating the respiratory tract
|
Use smoke evacuators and suction
|
|
Carcinogenic elements potentially increasing the risk of cancers
|
Use smoke evacuators and suction
|
Fire
|
Inappropriate use of diathermy near an oxygen source or alcohol skin preparation.
This could even cause an explosion
|
Avoid using diathermy near oxygen source
Allow alcohol-based skin preparations to dry
Stop the flow of all airway gases to the patients, may need temporal removal of the
airway itself
Remove the drapes and burnt material from the patient
Extinguish fire
In extreme situations, follow local fire evacuation protocol
|
|
Inadvertent activation of electrode placed over the drapes or careless placement
of hot electrodes over the drapes (exacerbated by nearby oxygen source) resulting
drapes catching fire
|
Place active electrodes in the plastic holder
|
Explosion
|
Inappropriate use of active electrode in the vicinity of inflammable gas under pressure
(e.g., opening an obstructed colon containing methane)
|
Use of alternative cutting/dissecting instrument (e.g., scissors)
|
PPM, ICD and other pacemakers (e.g., sacral nerve stimulator, spinal stimulators)
|
Interference form electromagnetic current. Rare with modern devices
Possible effects include reprograming, damage to PPM or ICD, inhibition, reversion
to back up mode and myocardial thermal injury
|
Use bipolar electrode or advanced bipolar devices when possible
When using monopolar electrode, use in short burst for the shortest time period possible
Apply dispersive electrode as far as possible from PPM, ICD, and other pacemakers
Monitor PPM for inhibition with ECG. Reprogram postoperatively if there is inhibition
during surgery or if any major adverse events occur during surgery
Deactivate ICD immediately preoperatively and active postoperatively. If deactivation
could not be organized with the cardiologist, then a magnet could be placed over the
ICD for the entire period of operation
In case of cardiac arrest, follow standard ALS guidelines
|
Abbreviations: ALS, advanced life support; ECG, electrocardiogram; ICD, implantable
cardioverter defibrillator; PPM, permanent pacemaker; SILS, single-incision laparoscopic
surgery.
Discussion
Most of the injuries caused by electrosurgery are related to thermal energy and so
commonly referred to as thermal injuries. There are other types of injuries with electrosurgery
as mentioned above. Thermal injuries can be classified as direct and indirect injuries.
Direct injuries are those that are caused by inadvertent use of active electrode on
any part of the body apart from the intended organ or tissue. Indirect injuries are
those that occur as a result of contact of the tip or any part of the active electrode
with any other metal instrument, which in turn, is in contact with the tissue or that
injury that occurs out of the operating field due to spread of current from the shaft
of the active electrode to nearby tissue.
Smoke-related injuries are commonly due to impaired vision and almost exclusively
seen in laparoscopic surgeries. Impaired vision leads to inappropriate contact of
the active electrode with tissue, thereby causing thermal injury. The smoke released
upon contact of active electrode with tissue contains various noxious elements.[11] These include gases, such as benzene, nitriles, hydrocyanide and other hydrocarbons,
and formaldehyde.[12] Some of these are potentially carcinogenic. The smoke also contains bodily products,
such as blood, fatty acids, and viruses, which could transmit infections. The thermal
energy also converts part of the blood to methemoglobin and carboxyhemoglobins.
Although there are no documented cases of cancers directly related to exposure of
smoke produced from diathermy, there have been several cases of minor respiratory
illness and ocular irritation.[13]
[14] These particles could easily be inhaled in spite of wearing masks. The average pore
size of a surgical mask is 5 µm, whereas most of the particles contained in the smoke
measure between 1.0 and 1.5 µm.[15] It is a current recommendation, although not mandatory, by various work place health
and safety organizations in many countries to use suction devices in operating theatres
to evacuate such smoke.
Fires in operating theaters are rare but devastating when occur. There are approximately
100 fires cases reported in the United States every year; 85% of them are minor and
result in minimal or no injuries with only a few disabling or disfiguring injuries.[16] More than 50% of these injuries occur in the head and neck region. A major contributor
is the oxygen supply around the oxygen mask, endotracheal tube, or laryngeal mask
airway.[17]
[18] Explosion-related injuries are rarely seen and if so, are again, most often related
to head and neck surgeries.[19] The confined space along with inflammable gas contributes to more severe burns.
In the general surgical setting, performing an enterotomy on a dilated colon poses
increased risk of explosion, and alternative methods such as scissors are safer.[20]
Electromagnetic devices, such as permanent pacemaker, implantable defibrillator, and
pacing devices, such as spinal stimulator and sacral nerve stimulators, are increasingly
used and tend to interfere with electrosurgical devices.[21]
[22] Although these are rare with modern electromagnetic devices, they may not be able
to be assessed, especially during an emergency operation. The interference caused
by diathermy could lead to arrhythmias and potentially be life threatening.[22] A safer approach would be to utilize bipolar electrosurgical methods. Alternatively,
it is safe to deactivate the device temporarily before the operation for the entire
duration of the operation or to use a magnet intraoperatively. Care must be taken
to activate and get the device checked by cardiologist or an industry representative
accordingly.
Laparoscopic cholecystectomy and appendectomy are some of the common general surgical
operations performed by the junior surgical trainees. Even though the complication
rates related to electrosurgery in these operations are low, these could be significant
if occur.[23]
[24] Injuries related to diathermy may not be identified intraoperatively. The injuries
often present late with strictures of the bile duct or peritonitis from bile leak
or bowel perforation. When using diathermy near any of these structures, a safer technique
is to keep away from bile duct and bowel and stimulate the active electrode only for
a short period of time to avoid pedicle effect and serosal injury.[2]
[25]
With increasing expertise in endoscopy, therapeutic procedures, such as polypectomy
and endoscopic mucosal resections, are becoming more common. Surgical trainees are
increasingly getting exposed and trained in these procedures. Diathermy forms a vital
part of the practice to achieve hemostasis and maintain a clear filed while performing
these procedures. Injury in colon is more likely to happen on the right side, which
is thinner.[26] The rates are low at 0.1 to 0.2% and are often recognized only 24 hours postoperatively.
Majority of them could be managed nonoperatively with good outcomes.[27]
[28] On the other hand, endoscopic mucosal resection utilized for excision of early gastrointestinal
cancers has a slightly higher rate of perforation at 4 to 6% and may need operative
measures to address the perforation.[29]
[30]
Irrespective of the procedure, a thorough knowledge of the electrosurgical unit and
patient forms the basis of prevention of diathermy-related injuries for any surgeon.
Having a framework to set the patient up and the related necessary equipment before
the start of the procedure in such a way that there is minimal interference of the
diathermy with the patient are vital. There is no substitute to using the diathermy
appropriately and safely.
Conclusion
There are varied complications that may be encountered with the use of electromagnetic
current. Thankfully, almost all of them are preventable and the key lies in taking
those preventative measures and following appropriate, safe surgical techniques at
every step of the operation. The operator should always exercise caution and good
communication to improve situational awareness, ultimately improving the surgical
outcome.