RSS-Feed abonnieren
DOI: 10.1055/s-0031-1272547
© Thieme Medical Publishers
Successful Retrieval of a Coronary Stent Dislodged in the Brachial Artery by Means of Improvised Snare and Guiding Catheter
Publikationsverlauf
Publikationsdatum:
14. März 2011 (online)

ABSTRACT
This is a case report regarding the retrieval, by means of an improvised snare and guiding catheter, of a stent dislodged in the brachial artery during a transradial coronary intervention. A full-length guiding catheter could not be used to approach the lost stent, which was a mere 30 to 35 cm away from the sheath insertion site at the radial artery, and a commercial snare was not available at the time. Thus, we had to improvise a shortened guiding catheter and a snare, which was formed by folding an angioplasty Whisper guide wire (Abbott Laboratories, Abbott Park, IL) and was used successfully to snare the stent and retrieve it.
KEYWORDS
Percutaneous coronary intervention - transradial - stent - dislodgement - complication
REFERENCES
- 1
Brilakis E S, Best P J, Elesber A A et al..
Incidence, retrieval methods, and outcomes of stent loss during percutaneous coronary
intervention: a large single-center experience.
Catheter Cardiovasc Interv.
2005;
66
333-340
Reference Ris Wihthout Link
- 2 Yang Soon C, Chong E, Sangiorgi G M. A challenging case of dislodged stent retrieval with the use of Goose neck snare kit. Catheter Cardiovasc Interv. 2010; 75 630-633
- 3 Ahmar W, Malaiapan Y, Meredith I T. Transradial retrieval of a dislodged stent from the left main coronary artery. J Invasive Cardiol. 2008; 20 545-547
Georgios GiannopoulosM.D.
154 Mesogeion Avenue
Athens, Greece
eMail: ggiann@med.uoa.gr