Artificial Intelligence-Assisted Colonoscopy With or Without Mucosal Exposure Device for Detection of Colorectal Adenomas: A Meta-Analysis
Authors
Abstract
Background and study aims
Artificial intelligence (AI) and mucosal exposure devices like Endocuff have independently improved the adenoma detection rate (ADR) during colonoscopy. This meta-analysis evaluated the combined effect of Endocuff and AI versus AI alone on colorectal neoplasia detection rates.
Methods
We searched PubMed, Embase, and Web of Science for randomized controlled trials (RCTs) comparing the impact of Endocuff+AI versus AI alone on colorectal neoplasia detection. Primary outcome was ADR; secondary outcomes included advanced adenoma detection rate (AADR), sessile serrated lesion detection rate (SSLDR), cecal intubation time, and withdrawal time. Pooled risk ratios (RRs) and mean differences (MDs) with 95% confidence intervals (CIs) were calculated using a random-effects model.
Results
Three RCTs with 2404 subjects were included (n = 1198 Endocuff+AI vs. n = 1206 AI alone). ADR was significantly higher in the Endocuff+AI group than in the AI alone group (54% vs. 48%, respectively) (RR 1.12, 95% CI 1.03–1.21, P = 0.01, I2 = 0%). There was a trend toward higher AADR (12.3% vs. 10%, RR 1.23, 95% CI 0.96–1.59, P = 0.10, I2 = 17%) and SSLDR (17.6% vs. 15.5%, RR 1.16, 95% CI 0.96–1.40, P = 0.13, I2 = 0%) in the Endocuff+AI group compared with the AI alone group, but it did not reach statistical significance. Both cecal intubation time (MD -0.61 minutes; 95% CI -1.54–0.33; P = 0.20; I2 = 87%) and withdrawal time (MD -0.42 minutes; 95% CI -1.01–0.17; P = 0.17, I2 = 60%) were similar between the two groups.
Conclusions
Endocuff combined with AI was superior to AI alone in improving the adenoma detection rate without increasing intubation or withdrawal times.
Keywords
Endoscopy Lower GI Tract - Polyps / adenomas / ... - Quality and logistical aspects - Quality management - CRC screeningIntroduction
Colorectal cancer (CRC) is the third most common cancer in the United States and the second leading cause of cancer-related deaths [1]. Colonoscopy is the gold standard for CRC screening, enabling detection and removal of precancerous polyps through polypectomy, which is the cornerstone of CRC prevention [2] [3]. Adenoma detection rate (ADR) is a key quality metric for screening colonoscopy, with rates varying significantly among endoscopists due to its operator-dependent nature [4] [5]. Reported ADRs range from 10% to 50%, with higher ADRs associated with reduced incidence and mortality of interval CRC [5] [6] [7]. Missed adenomas are often attributed to challenges in recognizing subtle lesions and incomplete mucosal inspection [8].
Artificial intelligence (AI), or computer-assisted detection (CADe), is increasingly utilized for real-time colorectal polyp detection. These systems enhance endoscopic accuracy by highlighting mucosal regions that exhibit characteristics suggestive of polyps. Several randomized controlled trials (RCTs) have demonstrated that AI improves ADR [9] [10] [11] and reduces adenoma miss rates [12]. However, about 10% to 20% of adenomas remain undetected even with AI, often due to polyps being obscured by colonic folds or other blind spots [12] [13]. Mucosal exposure devices, such as the Endocuff (Olympus, Tokyo, Japan), have proven effective in increasing mucosal exposure during colonoscopy withdrawal, thereby improving ADR [14] [15]. RCTs have shown that Endocuff is effective in increasing ADR [16] [17]. In this meta-analysis, we aimed to investigate the effect of a combination of Endocuff and AI versus AI alone during colonoscopy on detection rates of colorectal neoplasia.
Methods
We adhered to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines [18] in presenting this systematic review. The protocol for our review was preregistered with the International Prospective Register of Systematic Reviews (PROSPERO).
Data sources and search strategy
We searched PubMed, Embase, and Web of Science databases from inception through November 30, 2024. The following MeSH terms were used: (“Endocuff” or “mucosal exposure device”) and (“artificial intelligence” or “computer-aided detection”) and (“colonoscopy”). The search was limited to English language, because the language filter did not seem to affect the initial search results. Supplementary Table 1 describes the full search terms used in each database searched. Two reviewers (AB and TN) independently screened and selected the potentially included studies. Discrepancies were handled by a third reviewer (DR).
Inclusion and exclusion criteria
Only published RCTs comparing the effect of a combination of Endocuff and AI versus AI alone on detection rates of colorectal neoplasia were eligible for inclusion. Observational studies were excluded. Conference abstracts were also excluded, because this filter did not seem to affect the initial search results.
Data extraction
Data on baseline study and subject characteristics and outcome measures were extracted by two independent reviewers (AB and TN). Extracted study characteristics included country of origin, sample size, AI system, and reported outcomes: ADR, advanced adenoma detection rate (AADR), sessile serrated lesions detection rate (SSLDR), cecal intubation time, and withdrawal time. Extracted subject characteristics included mean age, male sex percentage, mean Boston Bowel Preparation Scale (BBPS) score, cecal intubation time, withdrawal time, and indications for colonoscopy.
Outcomes and definitions
The primary outcome was ADR, defined as the proportion of patients with at least one detected adenoma. Secondary outcomes were AADR, SSLDR, cecal intubation time (in minutes), and withdrawal time (in minutes). Advanced adenoma was defined as an adenoma either ≥ 10 mm in size, with villous component, or high-grade dysplasia. SSLDR was defined as the proportion of patients with at least one detected SSL.
Statistical analysis
Pooled risk ratios (RRs) with corresponding confidence intervals (CI) were calculated using the random-effects model, incorporating the Mantel-Haenszel method as a foundation. To account for anticipated between-study heterogeneity, we applied a random-effects extension, as described by van Houwelingen et al. [19], which models 2×2 tables using non-central hypergeometric distributions. P < 0.05 was considered statistically significant. Heterogeneity was evaluated using the I2 statistic, as outlined in the Cochrane handbook for systematic reviews, and I2 value ≥ 50% was considered significant statistical heterogeneity [20]. All statistical analyses were conducted via Review Manager 5.4 and Comprehensive Meta-Analysis 3.3.
Bias assessment
The revised Cochrane risk of bias tool was used to assess quality of the RCTs. Two reviewers (AB and DR) independently assessed each study for bias. Discrepancies were resolved by a third reviewer (JG). The Grading of Recommendations, Assessment, Development, and Evaluation (GRADE) approach was used to rate certainty of evidence (CoE) for outcomes. CoE is rated high for estimates derived from RCTs. CoE is rated down for methodological limitations of the included studies (i.e., risk of bias), imprecision (using a minimally contextualized approach, for ADR, CoE was rated down once if the CI of the absolute difference in detection rate overlapped a preset threshold of 180 per 1000 derived from Lui et al [21]. For AADR and SSLDR, CoE was rated down once if the absolute difference in detection rate overlapped a preset threshold of 25 and 39 per 1000, respectively, corresponding with a relative risk difference of 25%, indirectness, inconsistency of the results (by visual inspection of the variability of the results in the forest plots and based on an arbitrary threshold of I2. 50%), or publication bias [22] [23] [24] [25] [26] [27] [28]. The CoE for each outcome is summarized in Supplementary Table 2 based on the GRADE approach.
Results
Study selection
Among the 100 studies initially screened, five met the eligibility criteria for our systematic review. However, two of these were excluded because they were observational or pilot studies. Ultimately, three studies [21] [29] [30] met the eligibility criteria and were included in the meta-analysis. The selection process is summarized in the PRISMA diagram in [Fig. 1].


Study and subject characteristics
Study and subject characteristics are summarized in [Table 1]. The three studies were published between March 2023 and July 2024, with one conducted in China, another in Thailand, and the third in Europe (Italy and Switzerland). A total of 2,404 subjects were included (n = 1198 Endocuff+AI vs. n = 1206 AI alone). Mean age was 62.4 ± 9.4 years, and males represented 45.9% of patients.
Primary outcome: Adenoma detection rate
All three studies [21] [29] [30] (n = 2404) reported ADR. ADR was significantly higher in the Endocuff+AI group than in the AI alone group (54% vs. 48%, respectively) (RR 1.12, 95% CI 1.03–1.21, P = 0.01, I2 = 0%, [Fig. 2]) (CoE was moderate, Supplementary Table 2).


Secondary outcomes: AADR, SSLDR, cecal intubation time, and withdrawal time
All three studies [21] [29] [30] (n = 2404) reported AADR and two studies [21] [30] reported SSL detection rate. There was a trend toward higher AADR (12.3% vs. 10%, RR 1.23, 95% CI 0.96–1.59, P = 0.10, I2 = 17%, moderate CoE, [Fig. 3] a, Supplementary Table 2) and SSL detection rate (17.6% vs. 15.5%, RR 1.16, 95% CI 0.96–1.40, P = 0.13, I2 = 0%, moderate CoE, [Fig. 3] b, Supplementary Table 2) in the Endocuff+AI group compared with the AI alone group, but it did not reach statistical significance.


All three studies [21] [29] [30] reported cecal intubation and withdrawal time. Both cecal intubation time (6.6 ± 4.8 vs. 7 ± 4.8 minutes, MD -0.61 minutes; 95% CI -1.54 to 0.33; P = 0.20; I2 = 87%; low CoE; [Fig. 3] c, Supplementary Table 2) and withdrawal time (8.8 ± 3.9 vs. 9.2 ± 3.9 minutes, MD -0.42 minutes; 95% CI -1.01 to 0.17; P = 0.17, I2 = 60%, moderate CoE, Fig. 3d, Supplementary Table 2) were similar between the two groups.
Bias assessment
Risk of bias was assessed and is summarized in Supplementary Table 3. All three included studies [21] [29] [30] were low risk of bias (Supplementary Table 3). Given the limited number of studies (< 10), we could not evaluate publication bias.
Discussion
In this systematic review and meta-analysis of all RCTs investigating the combination of Endocuff with AI versus AI alone during colonoscopy, we found that combining Endocuff with AI during colonoscopy was associated with significant improvement in ADR without increasing intubation or withdrawal times compared with AI alone.
Our study revealed that combining AI with Endocuff device improved the ADR by 6% compared with AI alone. The results are consistent with the largest RCT by Spadaccini et al. [30] in Europe, which showed improvement in ADR from 44% with AI alone to 49.6% with Endocuff and AI. The European RCT used Fujifilm CAD EYE as AI system. However, the other two trials used different AI systems: Medtronic GI Genius [29] and Olympus OIP-1 [21]. Despite the variation in AI systems, all three trials used the same mucosal exposure device (Olympus Endocuff Vision) and consistently showed that the combined Endocuff+AI group achieved higher ADR compared with AI alone group. This improvement is likely attributable to the ability of the Endocuff to optimize mucosal exposure by reducing blind spots caused by colonic folds, enabling AI systems to detect polyps that might otherwise be missed.
Our study observed a trend toward a higher AADR in the Endocuff+AI group compared with the AI-alone group; however, this difference did not reach statistical significance. Notably, none of the three studies were able to establish the superiority of the combined Endocuff+AI approach over AI alone in detecting advanced adenomas. This may be attributed to the relatively larger size and lower incidence rates of these advanced lesions, which could result in insufficient statistical power to detect a significant difference. Although only one RCT by Lui et al. [21] demonstrated improved detection of SSLs in the Endocuff+AI group compared with the AI-alone group, our study showed a trend toward a higher SSLDR with the Endocuff+AI group, although it was not statistically significant. A critical factor in achieving a high SSLDR is quality of bowel preparation. In the Lui et al study [21], mean BBPS score in the Endocuff+AI group was relatively higher than in the AI-alone group (7.3 ± 1.7 vs. 6.9 ± 1.7, P = 0.06). Conversely, in the RCT by Spadaccini et al. [30], both groups had comparable mean BBPS scores (7.14 vs. 7.13, P = 0.88). In addition, AI sometimes fails to detect flat lesions, especially large flat SSLs [31]. Moreover, only two studies were included in the analysis of SSLDR in our meta-analysis, potentially limiting the overall statistical power to detect a meaningful difference.
Although a previous RCT reported that Endocuff increased polyp detection with a shorter withdrawal time compared with high-definition colonoscopy without AI [15], we found no significant differences in cecal intubation time (6.6 ± 4.8 vs. 7 ± 4.8 minutes, respectively) or withdrawal time (8.8 vs. 9.2 minutes, respectively) between the Endocuff+AI and AI-alone groups. Lack of a significant reduction in withdrawal time with the Endocuff+AI group compared with the AI-alone group may be explained by the increased inspection time required when using AI, partly due to the higher number of false positives flagged by the system [32].
Adoption and utility of the Endocuff Vision device have varied across healthcare settings. Its use should be carefully individualized, because its use may be limited in patients with anatomical constraints, such as narrowed sigmoid colons due to diverticular disease, where removal of the device might be necessary. However, only a small percentage of patients (3%) require device removal [33].
Our study has certain limitations. First, lack of blinding of endoscopists to the intervention may have introduced operator bias. Nonetheless, this limitation is common and often difficult to avoid in interventional studies. However, the blinding of pathology and outcome assessors in some of the included studies [30] may have helped mitigate this bias. Second, the trials included in this meta-analysis were conducted in only two regions (Asia and Europe), which may limit generalizability of our findings to other populations. However, the trials that previously evaluated the role of Endocuff Vision in enhancing endoscopist detection performance compared with standard colonoscopy (without AI) were conducted across multiple continents [34]. In addition, patient populations varied across studies, with different indications for colonoscopy (screening, surveillance, or diagnostic), introducing clinical heterogeneity. To account for this, we used a random-effects model. Third, all the included RCTs utilized the same mucosal exposure device, leaving it uncertain whether these findings are applicable to other mucosal exposure devices. However, our aim was not to demonstrate that mucosal exposure devices are universally effective in enhancing endoscopist performance during AI-assisted colonoscopy. Instead, our analysis—supported by the three individual studies—emphasized that AI alone may not be sufficient to address all detection challenges. Proper mucosal exposure, achieved through optimal colonoscopy techniques, remains fundamental. Notably, use of a device with well-established efficacy in improving detection performance, namely Endocuff Vision, served to reinforce this key concept [34].
Conclusions
In conclusion, Endocuff combined with AI was superior to AI alone in improving ADR without increasing intubation or withdrawal times. This combined approach could be an effective strategy to improve adenoma detection in colorectal cancer screening.
Conflict of Interest
JJG: Travel Support - Boston Scientific, Olympus Corporation, Ovesco Endoscopy AG DKR: Consultant - Olympus Corporation, Boston Scientific, Braintree Laboratories, Norgine, Medtronic, Acacia Pharmaceuticals; Research Support - Olympus Corporation, Medivators, Erbe USA Inc, Braintree Laboratories; Shareholder - Satisfai Health. The remaining authors declare that there is no conflict of interest.
-
References
- 1
Shaukat A,
Kahi CJ,
Burke CA.
et al.
ACG Clinical Guidelines: Colorectal Cancer Screening 2021. Am J Gastroenterol 2021;
116: 458-479
Reference Ris Wihthout Link
- 2
Brenner H,
Chang-Claude J,
Seiler CM.
et al.
Protection from colorectal cancer after colonoscopy: a population-based, case-control
study. Ann Intern Med 2011; 154: 22-30
Reference Ris Wihthout Link
- 3
Zauber AG,
Winawer SJ,
O'Brien MJ.
et al.
Colonoscopic polypectomy and long-term prevention of colorectal-cancer deaths. N Engl
J Med 2012; 366: 687-696
Reference Ris Wihthout Link
- 4
Rex DK,
Anderson JC,
Butterly LF.
et al.
Quality indicators for colonoscopy. Am J Gastroenterol 2024; 119: 1754-1780
Reference Ris Wihthout Link
- 5
Aniwan S,
Vanduangden K,
Kerr SJ.
et al.
Usefulness of mean number of adenomas per positive screenee for identifying meticulous
endoscopists among those who achieve acceptable adenoma detection rates. Endoscopy
2021; 53: 394-401
Reference Ris Wihthout Link
- 6
Corley DA,
Jensen CD,
Marks AR.
et al.
Adenoma detection rate and risk of colorectal cancer and death. N Engl J Med 2014;
370: 1298-1306
Reference Ris Wihthout Link
- 7
Kaminski MF,
Wieszczy P,
Rupinski M.
et al.
Increased rate of adenoma detection associates with reduced risk of colorectal cancer
and death. Gastroenterology 2017; 153: 98-105
Reference Ris Wihthout Link
- 8
Pedersen L,
Valori R,
Bernstein I.
et al.
Risk of post-colonoscopy colorectal cancer in Denmark: time trends and comparison
with Sweden and the English National Health Service. Endoscopy 2019; 51: 733-741
Reference Ris Wihthout Link
- 9
Wang P,
Berzin TM,
Glissen Brown JR.
et al.
Real-time automatic detection system increases colonoscopic polyp and adenoma detection
rates: a prospective randomised controlled study. Gut 2019; 68: 1813-1819
Reference Ris Wihthout Link
- 10
Wang P,
Liu X,
Berzin TM.
et al.
Effect of a deep-learning computer-aided detection system on adenoma detection during
colonoscopy (CADe-DB trial): a double-blind randomised study. Lancet Gastroenterol
Hepatol 2020; 5: 343-351
Reference Ris Wihthout Link
- 11
Repici A,
Badalamenti M,
Maselli R.
et al.
Efficacy of real-time computer-aided detection of colorectal neoplasia in a randomized
trial. Gastroenterology 2020; 159: 512-520 e517
Reference Ris Wihthout Link
- 12
Wang P,
Liu P,
Glissen Brown JR.
et al.
Lower adenoma miss rate of computer-aided detection-assisted colonoscopy vs routine
white-light colonoscopy in a prospective tandem study. Gastroenterology 2020; 159:
1252-1261 e1255
Reference Ris Wihthout Link
- 13
Lui TKL,
Hui CKY,
Tsui VWM.
et al.
New insights on missed colonic lesions during colonoscopy through artificial intelligence-assisted
real-time detection (with video). Gastrointest Endosc 2021; 93: 193-200 e191
Reference Ris Wihthout Link
- 14
Rex DK,
Sagi SV,
Kessler WR.
et al.
A comparison of 2 distal attachment mucosal exposure devices: a noninferiority randomized
controlled trial. Gastrointest Endosc 2019; 90: 835-840 e831
Reference Ris Wihthout Link
- 15
Rex DK,
Slaven JE,
Garcia J.
et al.
Endocuff Vision reduces inspection time without decreasing lesion detection: A clinical
randomized trial. Clin Gastroenterol Hepatol 2020; 18: 158-162 e151
Reference Ris Wihthout Link
- 16
Rameshshanker R,
Tsiamoulos Z,
Wilson A.
et al.
Endoscopic cuff-assisted colonoscopy versus cap-assisted colonoscopy in adenoma detection:
randomized tandem study-DEtection in Tandem Endocuff Cap Trial (DETECT). Gastrointest
Endosc 2020; 91: 894-904 e891
Reference Ris Wihthout Link
- 17
Triantafyllou K,
Polymeros D,
Apostolopoulos P.
et al.
Endocuff-assisted colonoscopy is associated with a lower adenoma miss rate: a multicenter
randomized tandem study. Endoscopy 2017; 49: 1051-1060
Reference Ris Wihthout Link
- 18
Page MJ,
McKenzie JE,
Bossuyt PM.
et al.
The PRISMA 2020 statement: an updated guideline for reporting systematic reviews.
BMJ 2021; 372: n71
Reference Ris Wihthout Link
- 19
Van Houwelingen HC,
Zwinderman KH,
Stijnen T.
A bivariate approach to meta-analysis. Stat Med 1993; 12: 2273-2284
Reference Ris Wihthout Link
- 20
Higgins JP,
Thompson SG,
Deeks JJ.
et al.
Measuring inconsistency in meta-analyses. Bmj 2003; 327: 557-560
Reference Ris Wihthout Link
- 21
Lui TK,
Lam CP,
To EW.
et al.
Endocuff With or Without Artificial Intelligence-Assisted Colonoscopy in Detection
of Colorectal Adenoma: A Randomized Colonoscopy Trial. Am J Gastroenterol 2024; 119:
1318-1325
Reference Ris Wihthout Link
- 22
Guyatt GH,
Oxman AD,
Vist G.
et al.
GRADE guidelines: 4. Rating the quality of evidence--study limitations (risk of bias).
J Clin Epidemiol 2011; 64: 407-415
Reference Ris Wihthout Link
- 23
Guyatt GH,
Oxman AD,
Kunz R.
et al.
GRADE guidelines 6. Rating the quality of evidence--imprecision. J Clin Epidemiol
2011; 64: 1283-1293
Reference Ris Wihthout Link
- 24
Guyatt GH,
Oxman AD,
Kunz R.
et al.
GRADE guidelines: 8. Rating the quality of evidence--indirectness. J Clin Epidemiol
2011; 64: 1303-1310
Reference Ris Wihthout Link
- 25
Guyatt GH,
Oxman AD,
Kunz R.
et al.
GRADE guidelines: 7. Rating the quality of evidence--inconsistency. J Clin Epidemiol
2011; 64: 1294-1302
Reference Ris Wihthout Link
- 26
Guyatt GH,
Oxman AD,
Montori V.
et al.
GRADE guidelines: 5. Rating the quality of evidence--publication bias. J Clin Epidemiol
2011; 64: 1277-1282
Reference Ris Wihthout Link
- 27
Zeng L,
Brignardello-Petersen R,
Hultcrantz M.
et al.
GRADE guidelines 32: GRADE offers guidance on choosing targets of GRADE certainty
of evidence ratings. J Clin Epidemiol 2021; 137: 163-175
Reference Ris Wihthout Link
- 28
Zeng L,
Brignardello-Petersen R,
Hultcrantz M.
et al.
GRADE Guidance 34: update on rating imprecision using a minimally contextualized approach.
J Clin Epidemiol 2022; 150: 216-224
Reference Ris Wihthout Link
- 29
Aniwan S,
Mekritthikrai K,
Kerr SJ.
et al.
Computer-aided detection, mucosal exposure device, their combination, and standard
colonoscopy for adenoma detection: a randomized controlled trial. Gastrointest Endosc
2023; 97: 507-516
Reference Ris Wihthout Link
- 30
Spadaccini M,
Hassan C,
Rondonotti E.
et al.
Combination of mucosa-exposure device and computer-aided detection for adenoma detection
during colonoscopy: A randomized trial. Gastroenterology 2023; 165: 244-251 e243
Reference Ris Wihthout Link
- 31
Rex DK,
Mori Y,
Sharma P.
et al.
Strengths and weaknesses of an artificial intelligence polyp detection program as
assessed by a high-detecting endoscopist. Gastroenterology 2022; 163: 354-358 e351
Reference Ris Wihthout Link
- 32
Hassan C,
Badalamenti M,
Maselli R.
et al.
Computer-aided detection-assisted colonoscopy: classification and relevance of false
positives. Gastrointest Endosc 2020; 92: 900-904 e904
Reference Ris Wihthout Link
- 33
Triantafyllou K,
Gkolfakis P,
Tziatzios G.
et al.
Effect of Endocuff use on colonoscopy outcomes: A systematic review and meta-analysis.
World J Gastroenterol 2019; 25: 1158-1170
Reference Ris Wihthout Link
- 34
Patel HK,
Chandrasekar VT,
Srinivasan S.
et al.
Second-generation distal attachment cuff improves adenoma detection rate: meta-analysis
of randomized controlled trials. Gastrointest Endosc 2021; 93: 544-553 e547
Reference Ris Wihthout Link
Correspondence
Publication History
Received: 27 March 2025
Accepted after revision: 14 June 2025
Accepted Manuscript online:
04 August 2025
Article published online:
29 August 2025
© 2025. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/).
Georg Thieme Verlag KG
Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany
Azizullah Beran, Tarek Nayfeh, Daryl Ramai, Almaza Albakri, Nasir Saleem, Marco Spadaccini, Cesare Hassan, Alessandro Repici, John J. Guardiola, Douglas K. Rex. Artificial Intelligence-Assisted Colonoscopy With or Without Mucosal Exposure Device for Detection of Colorectal Adenomas: A Meta-Analysis. Endosc Int Open 2025; 13: a26764144.
DOI: 10.1055/a-2676-4144
-
References
- 1
Shaukat A,
Kahi CJ,
Burke CA.
et al.
ACG Clinical Guidelines: Colorectal Cancer Screening 2021. Am J Gastroenterol 2021;
116: 458-479
Reference Ris Wihthout Link
- 2
Brenner H,
Chang-Claude J,
Seiler CM.
et al.
Protection from colorectal cancer after colonoscopy: a population-based, case-control
study. Ann Intern Med 2011; 154: 22-30
Reference Ris Wihthout Link
- 3
Zauber AG,
Winawer SJ,
O'Brien MJ.
et al.
Colonoscopic polypectomy and long-term prevention of colorectal-cancer deaths. N Engl
J Med 2012; 366: 687-696
Reference Ris Wihthout Link
- 4
Rex DK,
Anderson JC,
Butterly LF.
et al.
Quality indicators for colonoscopy. Am J Gastroenterol 2024; 119: 1754-1780
Reference Ris Wihthout Link
- 5
Aniwan S,
Vanduangden K,
Kerr SJ.
et al.
Usefulness of mean number of adenomas per positive screenee for identifying meticulous
endoscopists among those who achieve acceptable adenoma detection rates. Endoscopy
2021; 53: 394-401
Reference Ris Wihthout Link
- 6
Corley DA,
Jensen CD,
Marks AR.
et al.
Adenoma detection rate and risk of colorectal cancer and death. N Engl J Med 2014;
370: 1298-1306
Reference Ris Wihthout Link
- 7
Kaminski MF,
Wieszczy P,
Rupinski M.
et al.
Increased rate of adenoma detection associates with reduced risk of colorectal cancer
and death. Gastroenterology 2017; 153: 98-105
Reference Ris Wihthout Link
- 8
Pedersen L,
Valori R,
Bernstein I.
et al.
Risk of post-colonoscopy colorectal cancer in Denmark: time trends and comparison
with Sweden and the English National Health Service. Endoscopy 2019; 51: 733-741
Reference Ris Wihthout Link
- 9
Wang P,
Berzin TM,
Glissen Brown JR.
et al.
Real-time automatic detection system increases colonoscopic polyp and adenoma detection
rates: a prospective randomised controlled study. Gut 2019; 68: 1813-1819
Reference Ris Wihthout Link
- 10
Wang P,
Liu X,
Berzin TM.
et al.
Effect of a deep-learning computer-aided detection system on adenoma detection during
colonoscopy (CADe-DB trial): a double-blind randomised study. Lancet Gastroenterol
Hepatol 2020; 5: 343-351
Reference Ris Wihthout Link
- 11
Repici A,
Badalamenti M,
Maselli R.
et al.
Efficacy of real-time computer-aided detection of colorectal neoplasia in a randomized
trial. Gastroenterology 2020; 159: 512-520 e517
Reference Ris Wihthout Link
- 12
Wang P,
Liu P,
Glissen Brown JR.
et al.
Lower adenoma miss rate of computer-aided detection-assisted colonoscopy vs routine
white-light colonoscopy in a prospective tandem study. Gastroenterology 2020; 159:
1252-1261 e1255
Reference Ris Wihthout Link
- 13
Lui TKL,
Hui CKY,
Tsui VWM.
et al.
New insights on missed colonic lesions during colonoscopy through artificial intelligence-assisted
real-time detection (with video). Gastrointest Endosc 2021; 93: 193-200 e191
Reference Ris Wihthout Link
- 14
Rex DK,
Sagi SV,
Kessler WR.
et al.
A comparison of 2 distal attachment mucosal exposure devices: a noninferiority randomized
controlled trial. Gastrointest Endosc 2019; 90: 835-840 e831
Reference Ris Wihthout Link
- 15
Rex DK,
Slaven JE,
Garcia J.
et al.
Endocuff Vision reduces inspection time without decreasing lesion detection: A clinical
randomized trial. Clin Gastroenterol Hepatol 2020; 18: 158-162 e151
Reference Ris Wihthout Link
- 16
Rameshshanker R,
Tsiamoulos Z,
Wilson A.
et al.
Endoscopic cuff-assisted colonoscopy versus cap-assisted colonoscopy in adenoma detection:
randomized tandem study-DEtection in Tandem Endocuff Cap Trial (DETECT). Gastrointest
Endosc 2020; 91: 894-904 e891
Reference Ris Wihthout Link
- 17
Triantafyllou K,
Polymeros D,
Apostolopoulos P.
et al.
Endocuff-assisted colonoscopy is associated with a lower adenoma miss rate: a multicenter
randomized tandem study. Endoscopy 2017; 49: 1051-1060
Reference Ris Wihthout Link
- 18
Page MJ,
McKenzie JE,
Bossuyt PM.
et al.
The PRISMA 2020 statement: an updated guideline for reporting systematic reviews.
BMJ 2021; 372: n71
Reference Ris Wihthout Link
- 19
Van Houwelingen HC,
Zwinderman KH,
Stijnen T.
A bivariate approach to meta-analysis. Stat Med 1993; 12: 2273-2284
Reference Ris Wihthout Link
- 20
Higgins JP,
Thompson SG,
Deeks JJ.
et al.
Measuring inconsistency in meta-analyses. Bmj 2003; 327: 557-560
Reference Ris Wihthout Link
- 21
Lui TK,
Lam CP,
To EW.
et al.
Endocuff With or Without Artificial Intelligence-Assisted Colonoscopy in Detection
of Colorectal Adenoma: A Randomized Colonoscopy Trial. Am J Gastroenterol 2024; 119:
1318-1325
Reference Ris Wihthout Link
- 22
Guyatt GH,
Oxman AD,
Vist G.
et al.
GRADE guidelines: 4. Rating the quality of evidence--study limitations (risk of bias).
J Clin Epidemiol 2011; 64: 407-415
Reference Ris Wihthout Link
- 23
Guyatt GH,
Oxman AD,
Kunz R.
et al.
GRADE guidelines 6. Rating the quality of evidence--imprecision. J Clin Epidemiol
2011; 64: 1283-1293
Reference Ris Wihthout Link
- 24
Guyatt GH,
Oxman AD,
Kunz R.
et al.
GRADE guidelines: 8. Rating the quality of evidence--indirectness. J Clin Epidemiol
2011; 64: 1303-1310
Reference Ris Wihthout Link
- 25
Guyatt GH,
Oxman AD,
Kunz R.
et al.
GRADE guidelines: 7. Rating the quality of evidence--inconsistency. J Clin Epidemiol
2011; 64: 1294-1302
Reference Ris Wihthout Link
- 26
Guyatt GH,
Oxman AD,
Montori V.
et al.
GRADE guidelines: 5. Rating the quality of evidence--publication bias. J Clin Epidemiol
2011; 64: 1277-1282
Reference Ris Wihthout Link
- 27
Zeng L,
Brignardello-Petersen R,
Hultcrantz M.
et al.
GRADE guidelines 32: GRADE offers guidance on choosing targets of GRADE certainty
of evidence ratings. J Clin Epidemiol 2021; 137: 163-175
Reference Ris Wihthout Link
- 28
Zeng L,
Brignardello-Petersen R,
Hultcrantz M.
et al.
GRADE Guidance 34: update on rating imprecision using a minimally contextualized approach.
J Clin Epidemiol 2022; 150: 216-224
Reference Ris Wihthout Link
- 29
Aniwan S,
Mekritthikrai K,
Kerr SJ.
et al.
Computer-aided detection, mucosal exposure device, their combination, and standard
colonoscopy for adenoma detection: a randomized controlled trial. Gastrointest Endosc
2023; 97: 507-516
Reference Ris Wihthout Link
- 30
Spadaccini M,
Hassan C,
Rondonotti E.
et al.
Combination of mucosa-exposure device and computer-aided detection for adenoma detection
during colonoscopy: A randomized trial. Gastroenterology 2023; 165: 244-251 e243
Reference Ris Wihthout Link
- 31
Rex DK,
Mori Y,
Sharma P.
et al.
Strengths and weaknesses of an artificial intelligence polyp detection program as
assessed by a high-detecting endoscopist. Gastroenterology 2022; 163: 354-358 e351
Reference Ris Wihthout Link
- 32
Hassan C,
Badalamenti M,
Maselli R.
et al.
Computer-aided detection-assisted colonoscopy: classification and relevance of false
positives. Gastrointest Endosc 2020; 92: 900-904 e904
Reference Ris Wihthout Link
- 33
Triantafyllou K,
Gkolfakis P,
Tziatzios G.
et al.
Effect of Endocuff use on colonoscopy outcomes: A systematic review and meta-analysis.
World J Gastroenterol 2019; 25: 1158-1170
Reference Ris Wihthout Link
- 34
Patel HK,
Chandrasekar VT,
Srinivasan S.
et al.
Second-generation distal attachment cuff improves adenoma detection rate: meta-analysis
of randomized controlled trials. Gastrointest Endosc 2021; 93: 544-553 e547
Reference Ris Wihthout Link