CC BY 4.0 · Endosc Int Open 2025; 13: a25368061
DOI: 10.1055/a-2536-8061
Original article

Novel water-based automated endoscope cleaning process vs conventional manual cleaning for reducing duodenoscope contamination

1   Department of Gastroenterology and Hepatology, Erasmus MC, Rotterdam, Netherlands (Ringgold ID: RIN6993)
2   Department of Medical Microbiology and Infectious Diseases, Erasmus MC, Rotterdam, Netherlands (Ringgold ID: RIN6993)
,
Juliëtte A. Severin
2   Department of Medical Microbiology and Infectious Diseases, Erasmus MC, Rotterdam, Netherlands (Ringgold ID: RIN6993)
,
Margreet C. Vos
2   Department of Medical Microbiology and Infectious Diseases, Erasmus MC, Rotterdam, Netherlands (Ringgold ID: RIN6993)
,
3   Department of Biostatistics, Erasmus MC, Rotterdam, Netherlands (Ringgold ID: RIN6993)
,
Adriana J.C. Bulkmans
4   Quality Assurance and Regulatory Affairs office Medical Technology, Erasmus MC, Rotterdam, Netherlands (Ringgold ID: RIN6993)
,
Marco Bruno
1   Department of Gastroenterology and Hepatology, Erasmus MC, Rotterdam, Netherlands (Ringgold ID: RIN6993)
,
Bibi C.G.C. Mason-Slingerland
2   Department of Medical Microbiology and Infectious Diseases, Erasmus MC, Rotterdam, Netherlands (Ringgold ID: RIN6993)
› Author Affiliations
 

Abstract

Background and study aims

Duodenoscope contamination remains a persistent problem, exposing patients to infection risks. Automation in reprocessing may limit human error, reduce workload, and increase uniformity and traceability. However, its effectiveness should be evaluated before implementation. This study assessed the impact of implementing a novel water-based automated endoscope cleaning process on duodenoscope contamination.

Methods

This before-and-after intervention study compared duodenoscope cleaning methods. From January 2022 to December 2023, conventional manual cleaning was used. From January 2024 to June 2024, the AquaTYPHOON system (AT) replaced manual cleaning. Cultures from Pentax ED34-i10T2 patient-ready duodenoscopes were collected. The main outcome was the contamination rate with microorganisms of gut or oral origin (MGO). Secondary outcomes included contamination with solely gut bacteria. Non-inferiority of the AT was tested using a generalized estimating equation with a non-inferiority margin of 5%.

Results

During the manual cleaning period, 333 duodenoscope cultures of eight duodenoscopes were collected; during the AT period, 100 cultures were collected. Pre-introduction of the AT, the contamination rate with MGO was 21.6%, which fell to 16% post-introduction (risk difference: –5.6%, upper bound 90% confidence interval [CI] 6.8%). For gut bacteria, the contamination rate decreased from 14.4% to 9% (risk difference: –5.4%, upper bound 90% CI 3.9%), indicating non-inferiority.

Conclusions

AT reduced the contamination rate with MGO, but non-inferiority was not demonstrated. For gut bacteria, AT was non-inferior to manual cleaning. These results are promising. However, future studies should confirm these findings in larger samples and explore other advantages of using the AT in duodenoscope cleaning.


#

Introduction

During endoscopic retrograde cholangiopancreatography (ERCP) procedures, duodenoscopes are heavily exposed to gastrointestinal bacteria. Consequently, these devices must undergo comprehensive reprocessing to prevent patient-to-patient pathogen transmission [1] [2]. However, reprocessing may fall short, with microorganisms persisting in up to 15% of cases [3]. Manual cleaning, crucial for removing organic debris before automated cleaning and disinfection, involves brushing and flushing duodenoscope channels with water and a detergent solution [4] [5]. However, healthcare workers often face time pressure, complex protocols, and physical discomfort during manual cleaning [6] [7]. Surveillance studies by the U.S. Food and Drug Administration showed that more than 60% of manual cleaning tasks are improperly executed [8].

Automated reprocessing could improve endoscope cleaning pre-disinfection by standardizing practices, alleviating staff workload, and minimizing human errors [9]. The recently introduced AquaTYPHOON system (AT, Plasmabiotics; Pentax Medical), which employs pulsating high-velocity water and compressed air instead of brushes and detergent, shows promising preliminary validation results in clinical settings [10]. Gastrointestinal endoscopes of all types were cleaned with the AT and validated by demonstrating compliance with the defined target level of a residual bacterial load of ≤ 6 colony-forming units (CFU) after reprocessing, as well as protein removal in accordance with ISO standards 15883–4 and 15883–5 [10]. However, guidelines recommend that any CFUs of gastrointestinal microorganisms warrant endoscope quarantine [11] [12]. Because the types of microorganisms detected were not mentioned, true validation of the AT has not been demonstrated. Moreover, the studies report that the AT has been validated across multiple endoscope types from major endoscope manufacturers. However, the number of duodenoscopes included in the validation studies is unclear. Because most reported endoscope-associated outbreaks originate from contaminated duodenoscopes, this must be confirmed prior to implementation [10] [13]. This study aimed to show non-inferiority of the AT in patient-ready duodenoscopes compared with conventional manual cleaning methods.


#

Methods

Setting

This retrospective-prospective before-and-after single-center intervention study was conducted between January 2022 and June 2024 at the Erasmus University Medical Center (Erasmus MC), a large Dutch tertiary care center that performs approximately 750 ERCP procedures annually.


#

Data collection

Between January 2022 and December 2023, duodenoscope cultures from Pentax ED34-i10T2 duodenoscopes with disposable endcaps were collected as part of the PREVENT study (unpublished). During this period, reprocessing was performed according to manufacturer instructions, and the protocols have been described previously [14] [15]. The cultures collected during this period were stored in a database. For the purpose of this study, we excluded cultures from loaner duodenoscopes that were no longer in active use. The remaining cultures from this database were used to represent the first study period, during which conventional manual cleaning methods, including flushing and brushing, were employed. These data were compared with cultures collected in the subsequent prospective period, during which the AT was implemented, as described below.


#

Intervention

The AT includes the AquaTYPHOON device, a barcode scanner, a water pistol (AquaJET), and a label printer ([Fig. 1]). The automated process includes a leak test and uses high-velocity air and water to create a turbulent flow in the endoscope channels to remove organic material and debris. During the cleaning process, the duodenoscope is not submerged in water and no detergents are used. The AquaJET is used by reprocessing personnel to clean the endoscope externally. The AquaTYPHOON device displays the current phase of the process and its percentage of completion. The time needed for the AT to complete automated cleaning is 5 minutes. If an error occurs, such as low air pressure, the cycle is interrupted and the error is displayed. Data records are printed and stored on the device. In this study, the printer was not used.

Zoom Image
Fig. 1 Illustrations of the AquaTYPHOON system, showing the device's front and side views (top left and center), the barcode scanner and AquaJET (top right), color-coded endoscope connection tubes (bottom left), and the cleaning sink (bottom right).

After the AT received its CE mark in March 2023, we developed the protocol for this study and the non-inferiority margin, endpoints, and analysis plan were specified (detailed below). In January 2024, all duodenoscopes were inspected using a borescope and those requiring maintenance were sent to the manufacturer for repairs. When the duodenoscopes returned from the manufacturer, they were included in the study. Before transitioning to the AT period, reprocessing personnel received training from the manufacturer on use and maintenance of the system. In addition, all duodenoscopes underwent four consecutive reprocessing cycles using the AT, followed by automated cleaning and disinfection as a wash-out measure. These four cycles were performed sequentially, without any intermittent use in a patient, and the duodenoscopes were confirmed to be culture-negative before the start of the AT period. Furthermore, all duodenoscopes that returned from the manufacturer were cultured prior to reintroduction for clinical use. The process of bedside precleaning, automated cleaning, high-level disinfection (HLD), and drying remained unchanged [15]. Duodenoscope cultures during the AT period were collected between January 2024 and June 2024.


#

Sampling

Sampling of the duodenoscope was performed shortly before its use for an ERCP procedure. Initially, the distal tip of the duodenoscope was swabbed using a dry Copan Liquid Amies Elution Swab (eSwab Copan). To neutralize any residual disinfectants, 1 mL of a neutralizer (Dey-Engley broth) was introduced into the container. Subsequently, the suction and biopsy channels were subjected to a combined flush-brush-flush procedure that encompassed the entire length of the duodenoscope, from the umbilical connector (processor end) to the distal tip. This involved flushing each channel with 20 mL of sterile water, which was collected in a sterile container containing 40 mL of neutralizer. A single-use endoscope cleaning brush (CS5522A, Pentax) was then passed through the entire length of the channels and the distal tip of the brush was severed using sterile scissors and added to the container. Flushing of both channels was then repeated to complete the sampling process.


#

Microbiological protocols and interpretation

Microbiological methods used were as previously reported [4]. Contamination was defined as: 1) ≥ 1 CFU of microorganisms of gut or oral origin (MGO), including Pseudomonas aeruginosa, Staphylococcus aureus, and yeasts; or 2) ≥ 20 CFU/20 mL of any microorganism of other origin (AM20) [16]. Culture results used in this study were collected separately from routine microbiological surveillance conducted monthly at Erasmus MC. Consequently, duodenoscopes contaminated with MGO were not quarantined. However, if cultures from routine surveillance tested positive for MGO, the duodenoscopes were quarantined according to Dutch guidelines [12]. These duodenoscopes were cleared for clinical use only after subsequent negative culture results.


#

Outcomes

The primary endpoint was the proportion of duodenoscope cultures positive for MGO. Secondary endpoints included the proportion of duodenoscope cultures positive for gut, oral, and AM20 microorganisms.


#

Sample size determination

Sample size calculation was based on the database of duodenoscope cultures from the PREVENT study, which at that time contained 270 culture results from the period using conventional cleaning. The contamination rate for conventional cleaning was approximately 20%. Given implementation of the AT and considering outcomes from validation studies, we anticipated a contamination rate of 10%, making a 10% reduction in contamination a realistic expectation. We conducted a power analysis using simulation, with data simulated through a random intercept logistic mixed model. This model was based on assumptions about contamination rates from logistic mixed models fitted to the available duodenoscope culture results from conventional cleaning. We assumed that the random intercept standard deviation would remain consistent across different phases of the study.

The risk-difference of contamination with MGO between conventional cleaning and AT cleaning was assessed using a generalized estimating equation (GEE) to consider clustering due to repeated measurements of the same duodenoscopes. MGO-positivity served as the response variable, the study phase was the sole covariate, and the model used a Gaussian distribution with identity link. Our simulations suggested that with an expected contamination rate for the AT period of 10%, a sample of 100 duodenoscope cultures in the AT period would achieve approximately 80% statistical power.

Because there was a possibility that the contamination rate might increase after implementation of the AT, an interim analysis was performed after 50 duodenoscope cultures were collected during the AT period. To ensure patient safety, if use of the AT had resulted in significantly higher contamination rates compared with the conventional cleaning method, the study would have been terminated prematurely. Specifics regarding the analysis and threshold for early termination are described below. Because the interim analysis was conducted solely for futility monitoring without formal hypothesis testing for non-inferiority, no adjustments for multiplicity were deemed necessary.


#

Statistical analyses

Statistical analyses were performed in R version 4.1.3 [17]. Categorical variables are presented as counts or proportions (%), whereas continuous variables are described using the median with the first and third quartiles (Q1, Q3) or the mean and standard deviation (SD). The analysis was conducted on an intention-to-treat basis, including all duodenoscope cultures collected during the study periods, except those from loaner duodenoscopes no longer in active use. The primary analysis used a GEE model to assess differences in contamination rates between study phases. This model was the same as that used for sample size calculation. To determine non-inferiority of the AT, the upper limit of the two-sided 90% confidence interval (CI) for the parameter estimate of the study phase was compared with a margin of 5%. We chose a 90% CI, so that, due to its one-sided use in the non-inferiority test, the corresponding significance level is 0.05.

A futility interim analysis was conducted after collecting 50 duodenoscope cultures during the AT period. The same GEE model described above was utilized, except for a one-sided inferiority test that compared the lower bound of the 90% CI for the risk difference with a specified threshold for early termination of 5% in case the AT performed considerably worse than conventional manual cleaning.


#
#

Results

Culture characteristics

A total of 433 cultures were collected from eight Pentax ED34-i10T2 duodenoscopes. During the period when conventional manual cleaning was employed (January 2022 to December 2023), 333 duodenoscope cultures (76.9%) were collected, and 100 cultures (23.1%) were collected during the AT period (January 2024 to May 2024). [Table 1] presents an overview of culture characteristics. During the period employing conventional cleaning methods, 21.6% of cultures (72/333) were contaminated with MGO, of which 48 (14.4%) contained gut bacteria and 26 (7.8%) contained oral bacteria. Of 100 cultures collected during the AT period, 16% (16/100) were contaminated with MGO, of which nine (9%) contained gut flora and eight (8%) contained oral flora. Supplementary Table 1 in the supplementary appendix presents contamination rates of individual duodenoscopes according to each contamination definition.

Table 1 Culture characteristics of periods employing conventional cleaning methods compared with the AquaTYPHOON system.

Conventional cleaning

n = 333 (100%)

AquaTYPHOON system

n = 100 (100%)

AM20, 20 CFU/mL of any other microorganism; SD, standard deviation.

Microorganisms of gut or oral origin, n (%)

72 (21.6%)

16 (16.0%)

Gut, n (%)

48 (14.4%)

9 (9.0%)

Oral, n (%)

26 (7.8%)

8 (8.0%)

AM20, n (%)

303 (91.0%)

94 (94.0%)

Sampled Pentax ED34-i10T2 duodenoscopes, n (%)

A110077

35 (10.5%)

16 (16.0%)

A110095

45 (13.5%)

10 (10.0%)

A110096

54 (16.2%)

6 (6.0%)

A110098

33 (9.9%)

21 (21.0%)

A110100

48 (14.4%)

22 (22.0%)

A110280

36 (10.8%)

8 (8.0%)

A110377

39 (11.7%)

17 (17.0%)

A110409

43 (12.9%)

0 (0.0%)


#

Interim analysis

After 50 duodenoscope cultures were collected during the AT period, an interim analysis was performed. Six of 50 duodenoscope cultures (12%) were positive for MGO. The risk difference between the two periods was -9.3% (lower bound 90% CI -24.3%) in favor of the AT. Because the lower bound of the 90% CI was -24.3%, and thus less than the futility margin of 5%, the study was continued.


#

Cultured microorganisms

[Table 2] shows all cultured MGOs and their frequencies and all other cultured microorganisms are shown in the supplementary appendix (Supplementary Table 2 and Supplementary Table 3). No contamination with P. aeruginosa, Klebsiella pneumoniae, or Enterobacter cloacae complex occurred during the AT period. However, contamination with Stenotrophomonas maltophilia occurred more frequently, seven times versus once in the conventional manual cleaning group. In the primary analysis, the risk difference in MGO-positivity between the AT and conventional manual cleaning methods was -5.6% (upper bound 90% CI 6.8%) ([Fig. 2]). For contamination with gut bacteria, the risk difference was -5.4% (upper bound 90% CI 3.9%), and for oral flora, it was 0.2% (upper bound 90% CI 5.7%). For contamination with AM20, the risk difference was 3% (upper bound 90% CI 8.1%) in favor of conventional manual cleaning. Non-inferiority could not be demonstrated for contamination with MGO, oral flora, or AM20 because the upper bound of the 90% CI exceeded the 5% non-inferiority margin. However, non-inferiority could be demonstrated for contamination with gut bacteria.

Table 2 Cultured microorganisms of gut or oral origin.

Conventional cleaning

n = 333 (100%)

AquaTYPHOON system

n = 100 (100%)

Gut bacteria, n (%)

  • Pseudomonas aeruginosa

23 (6.9%)

0 (0.0%)

  • Klebsiella pneumoniae

16 (4.8%)

0 (0.0%)

  • Stenotrophomonas maltophilia

1 (0.3%)

7 (7.0%)

  • Enterobacter cloacae complex

5 (1.5%)

0 (0.0%)

  • Escherichia coli

2 (0.6%)

1 (1.0%)

  • Klebsiella oxytoca

2 (0.6%)

0 (0.0%)

  • Hafnia alvei

2 (0.6%)

0 (0.0%)

  • Enterococcus faecium

1 (0.3%)

0 (0.0%)

  • Citrobacter koseri

1 (0.3%)

0 (0.0%)

  • Enterococcus faecalis

1 (0.3%)

0 (0.0%)

  • Acinetobacter iwoffi

0 (0.0%)

1 (1.0%)

  • Candida species

1 (0.3%)

0 (0.0%)

  • Paeniclostridium species

1 (0.3%)

0 (0.0%)

  • Staphylococcus aureus

1 (0.3%)

0 (0.0%)

Oral bacteria, n (%)

  • Moraxella osloensis

16 (4.8%)

4 (4.0%)

  • Moraxella species

6 (1.8%)

2 (2.0%)

  • Actinomyces oris

2 (0.6%)

0 (0.0%)

  • Neisseria species

1 (0.3%)

0 (0.0%)

  • Neisseria subflava

1 (0.3%)

0 (0.0%)

  • Rothia mucilaginosa

1 (0.3%)

0 (0.0%)

  • Rothia species

1 (0.3%)

0 (0.0%)

  • Streptococcus mitis

0 (0.0%)

1 (1.0%)

  • Streptococcus sanguinis

1 (0.3%)

0 (0.0%)

  • Rothia dentocariosa

0 (0.0%)

1 (1.0%)

  • Streptococcus cristatus

0 (0.0%)

1 (1.0%)

Zoom Image
Fig. 2 Forest plot with risk differences in contamination between the AquaTYPHOON system and conventional manual cleaning methods according to the different definitions. AM20, microbial growth with ≥ 20 colony-forming units/20 mL of water and/or skin type microorganism; MGO, presence of microorganisms of gut or oral origin.

#

Environmental sampling

Because of the unexpected increase in contamination with S. maltophilia during the AT period, 13 environmental samples from the endoscope reprocessing department were collected to check for a possible common source of contamination. The AquaTYPHOON device connectors, touchscreen, and water pistol were swabbed, as well as the automated endoscope reprocessors (AERs) and AER connectors, drying cabinet connectors, and a hand alcohol dispenser. All samples were negative for S. maltophilia. The drain of the sink where the endoscopes were cleaned was positive for S. maltophilia. However, this was not considered a likely source of the contamination.


#
#

Discussion

After introduction of the AT to replace conventional manual cleaning, the MGO contamination rate of patient-ready duodenoscopes declined by 5.6% (from 21.6% to 16%). This finding illustrates that automated cleaning methods can improve the outcome of duodenoscope reprocessing in a real-world clinical setting. This effect was predominantly caused by a reduction in gut bacteria isolated from duodenoscope samples, which are known to be notorious for causing duodenoscope-associated infections. Total absence of P. aeruginosa and K. pneumoniae during the AT period is particularly promising because these bacteria are frequently involved in outbreaks caused by contaminated duodenoscopes [13].

Enhancing automation in endoscope reprocessing offers numerous benefits, including improved efficiency, prevention of human error, standardization of procedures, and full traceability. Introduction of AERs in cleaning and HLD has been shown to improve guideline adherence, reduce physical discomfort associated with reprocessing, and decrease need for manual labor [9]. The same benefits may apply to automating manual cleaning. In addition, automated cleaning could significantly reduce the cognitive load imposed by manual reprocessing protocols [6].

Manual cleaning and the AT system differ notably in their cleaning methodologies. Unlike manual cleaning, the AT system relies solely on water, without use of detergents or single-use brushes. In addition, the duodenoscope is not submerged during cleaning. These differences may reduce the environmental impact of reprocessing by eliminating detergent use and single-use components while potentially simplifying the cleaning process. However, incorporating submersion in water with detergents could further enhance AT effectiveness by breaking down organic residues on the outside of the endoscope. Future research should explore this as a potential modification of the cleaning process. Moreover, a comprehensive life cycle analysis and environmental impact assessment are necessary to determine the duration of AT use required to achieve a meaningful environmental benefit.

During the AT period, incidence of cultures contaminated with S. maltophilia increased, making it the most prevalent MGO contaminant. S. maltophilia is an opportunistic pathogen present in the gut, but also widely distributed in various environments, including water and soil. Gut colonization with S. maltophilia can particularly occur after patients have used broad-spectrum antibiotics. Given its resistance profile and ability to cause severe infections in vulnerable populations, S. maltophilia is considered an important nosocomial pathogen and it has been involved in outbreaks linked to contaminated bronchoscopes [18] [19] [20]. Bacteria that thrive in moist environments and readily form biofilms are prone to contaminating the AT. Therefore, the AT should be regularly monitored for contamination. Although the sudden increase in contamination prevalence with S. maltophilia suggested a common source of contamination other than a patient, such as contaminated AquaTYPHOON device connectors or AER, the environmental cultures did not show an external contamination source. However, S. maltophilia was cultured from the sink drain at the location where the AT was use. Because the duodenoscope was not submerged in water during cleaning with the AquaJET, we cannot completely exclude the possibility that splashes from the sink drain contaminated the duodenoscopes during cleaning.

Implementation of the AT led to an increase in contamination with AM20 by 3% (upper bound 90% CI 8.1%). In our previous research, we hypothesized that the high contamination rate with AM20 might be partly explained by the design of the single-use brush with a distal sweeper used in our center during manual cleaning [4]. However, even absent any materials introduced into the duodenoscope channels during precleaning, the contamination rate with AM20 did not decrease. The biomatrix of non-MGO bacteria might protect MGOs during HLD, preventing proper disinfection [21]. Thus, the source of such persistent contamination by AM20 remains unknown and requires further investigation.

This study is subject to certain limitations that could have impacted the results. This was a single-center study employing a single type of duodenoscope, which limits generalizability of our findings. In addition, due to our limited sample size, the CI of the main outcome was quite large, preventing us from showing non-inferiority despite a clear difference in the contamination rate with MGO. Furthermore, the order of the periods (conventional cleaning versus AT) was not randomized and no control group was available; therefore, we cannot claim causality between introduction of the AT and reduction in contamination with MGO. Finally, sampling and culturing of the air/water channel was not performed, leaving the effectiveness of the AT in cleaning this channel unestablished.


#

Conclusions

In conclusion, this study found that implementation of the AT led to a reduction in contamination with MGO in Pentax ED34-i10T2 duodenoscopes, although non-inferiority was not demonstrated. However, the AT was non-inferior to conventional manual cleaning in reducing contamination with gut microorganisms. Therefore, the AT could offer an alternative to conventional manual cleaning methods. Future larger studies are necessary to confirm our findings, demonstrate generalizability, and investigate other potential benefits regarding reprocessing efficiency, labor intensity, and environmental impact.


#
#

Conflict of Interest

KvdP, AJCB, NSE, BCGCS and JAS have no conflicts of interest to disclose. MCV has received support from Boston Scientific, 3M, and Pentax Medical. MJB has received research support from Boston Scientific, Cook Medical, Pentax Medical, Mylan, and ChiRoStim and acted as a consultant/lecturer for Boston Scientific, Cook Medical, Pentax Medical, and AMBU.

Acknowledgement

The authors extend their gratitude to the endoscopy and reprocessing staff at Erasmus MC for their invaluable assistance in executing this study. This study received financial and material support from Pentax Medical. The company was not involved in the design, execution of the study, or preparation of the manuscript in any way.

Supplementary Material

  • References

  • 1 Day LW, Muthusamy VR, Collins J. et al. Multisociety guideline on reprocessing flexible GI endoscopes and accessories. Gastrointest Endosc 2021; 93: 11-33 e16
  • 2 Balan GG, Sfarti CV, Chiriac SA. et al. Duodenoscope-associated infections: a review. Eur J Clin Microbiol Infect Dis 2019; 38: 2205-2213
  • 3 Rauwers AW, Voor In 'T Holt AF, Buijs JG. et al. High prevalence rate of digestive tract bacteria in duodenoscopes: A nationwide study. Gut 2018; 67: 1637-1645
  • 4 van der Ploeg K, Haanappel CP, Voor In 't Holt AF. et al. Effect of a novel endoscope cleaning brush on duodenoscope contamination. Endoscopy 2024; 56: 198-204
  • 5 Beilenhoff U, Biering H, Blum R. et al. Reprocessing of flexible endoscopes and endoscopic accessories used in gastrointestinal endoscopy: Position Statement of the European Society of Gastrointestinal Endoscopy (ESGE) and European Society of Gastroenterology Nurses and Associates (ESGENA) – Update 2018. Endoscopy 2018; 50: 1205-1234
  • 6 Sivek AD, Davis J, Tremoulet P. et al. Healthcare worker feedback on duodenoscope reprocessing workflow and ergonomics. Am J Infect Control 2022; 50: 1038-1048
  • 7 Ofstead CL, Hopkins KM, Buro BL. et al. Challenges in achieving effective high-level disinfection in endoscope reprocessing. Am J Infect Control 2020; 48: 309-315
  • 8 US Food and Drug Administration. 522 Postmarket Surveillance Studies Database: Human Factors Study. https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pss.cfm
  • 9 Ofstead CL, Wetzler HP, Snyder AK. et al. Endoscope reprocessing methods: a prospective study on the impact of human factors and automation. Gastroenterol Nurs 2010; 33: 304-311
  • 10 PlasmaBiotics. AquaTYPHOON Verification and Validation Overview (Unpublished manufacturer data). 2023 .
  • 11 FDA/CDC/ASM Working Group on Duodenoscope Culturing. Duodenoscope surveillance sampling & culturing: Reducing the risks of infection. https://www.fda.gov/media/111081/download
  • 12 Nederlandse Vereniging voor Medische Microbiologie. Richtlijn controle op microbiologische veiligheid van thermolabiele flexibele hoog-risico gastrointestinale endoscopen. 2018.
  • 13 Ofstead CL, Buro BL, Hopkins KM. et al. Duodenoscope-associated infection prevention: A call for evidence-based decision making. Endosc Int Open 2020; 8: E1769-E1781
  • 14 van der Ploeg K, Haanappel CP, Voor In 't Holt AF. et al. Unveiling 8 years of duodenoscope contamination: insights from a retrospective analysis in a large tertiary care hospital. Gut 2024; 73: 613-621
  • 15 van der Ploeg K, Vos MC, Erler NS. et al. Impact of duodenoscope reprocessing factors on duodenoscope contamination: a retrospective observational study. J Hosp Infect 2024; 154: 88-94
  • 16 Rauwers AW, Voor in 't Holt AF, Buijs JG. et al. Nationwide risk analysis of duodenoscope and linear echoendoscope contamination. Gastrointest Endosc 2020; 92: 681-691 e681
  • 17 R Core Team. R: A language and Environment for Statistical Computing. R Foundation for Statistical Computing, Vienna, Austria. 2023.
  • 18 Botana-Rial M, Leiro-Fernández V, Núñez-Delgado M. et al. A pseudo-outbreak of Pseudomonas putida and Stenotrophomonas maltophilia in a bronchoscopy unit. Respiration 2016; 92: 274-278
  • 19 Guy M, Vanhems P, Dananché C. et al. Outbreak of pulmonary Pseudomonas aeruginosa and Stenotrophomonas maltophilia infections related to contaminated bronchoscope suction valves, Lyon, France, 2014. Euro Surveill 2016; 21
  • 20 Maraolo AE, Licciardi F, Gentile I. et al. Stenotrophomonas maltophilia Infections: A systematic review and meta-analysis of comparative efficacy of available treatments, with critical assessment of novel therapeutic options. Antibiotics (Basel) 2023; 12: 910
  • 21 Alfa MJ, Singh H. Impact of wet storage and other factors on biofilm formation and contamination of patient-ready endoscopes: a narrative review. Gastrointest Endosc 2020; 91: 236-247

Correspondence

Koen van der Ploeg, MD.
Department of Gastroenterology and Hepatology, Erasmus MC
Rotterdam
Netherlands   

Publication History

Received: 09 October 2024

Accepted after revision: 04 February 2025

Accepted Manuscript online:
10 February 2025

Article published online:
14 March 2025

© 2025. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/).

Georg Thieme Verlag KG
Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany

Bibliographical Record
Koen van der Ploeg, Juliëtte A. Severin, Margreet C. Vos, Nicole S Erler, Adriana J.C. Bulkmans, Marco Bruno, Bibi C.G.C. Mason-Slingerland. Novel water-based automated endoscope cleaning process vs conventional manual cleaning for reducing duodenoscope contamination. Endosc Int Open 2025; 13: a25368061.
DOI: 10.1055/a-2536-8061
  • References

  • 1 Day LW, Muthusamy VR, Collins J. et al. Multisociety guideline on reprocessing flexible GI endoscopes and accessories. Gastrointest Endosc 2021; 93: 11-33 e16
  • 2 Balan GG, Sfarti CV, Chiriac SA. et al. Duodenoscope-associated infections: a review. Eur J Clin Microbiol Infect Dis 2019; 38: 2205-2213
  • 3 Rauwers AW, Voor In 'T Holt AF, Buijs JG. et al. High prevalence rate of digestive tract bacteria in duodenoscopes: A nationwide study. Gut 2018; 67: 1637-1645
  • 4 van der Ploeg K, Haanappel CP, Voor In 't Holt AF. et al. Effect of a novel endoscope cleaning brush on duodenoscope contamination. Endoscopy 2024; 56: 198-204
  • 5 Beilenhoff U, Biering H, Blum R. et al. Reprocessing of flexible endoscopes and endoscopic accessories used in gastrointestinal endoscopy: Position Statement of the European Society of Gastrointestinal Endoscopy (ESGE) and European Society of Gastroenterology Nurses and Associates (ESGENA) – Update 2018. Endoscopy 2018; 50: 1205-1234
  • 6 Sivek AD, Davis J, Tremoulet P. et al. Healthcare worker feedback on duodenoscope reprocessing workflow and ergonomics. Am J Infect Control 2022; 50: 1038-1048
  • 7 Ofstead CL, Hopkins KM, Buro BL. et al. Challenges in achieving effective high-level disinfection in endoscope reprocessing. Am J Infect Control 2020; 48: 309-315
  • 8 US Food and Drug Administration. 522 Postmarket Surveillance Studies Database: Human Factors Study. https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pss.cfm
  • 9 Ofstead CL, Wetzler HP, Snyder AK. et al. Endoscope reprocessing methods: a prospective study on the impact of human factors and automation. Gastroenterol Nurs 2010; 33: 304-311
  • 10 PlasmaBiotics. AquaTYPHOON Verification and Validation Overview (Unpublished manufacturer data). 2023 .
  • 11 FDA/CDC/ASM Working Group on Duodenoscope Culturing. Duodenoscope surveillance sampling & culturing: Reducing the risks of infection. https://www.fda.gov/media/111081/download
  • 12 Nederlandse Vereniging voor Medische Microbiologie. Richtlijn controle op microbiologische veiligheid van thermolabiele flexibele hoog-risico gastrointestinale endoscopen. 2018.
  • 13 Ofstead CL, Buro BL, Hopkins KM. et al. Duodenoscope-associated infection prevention: A call for evidence-based decision making. Endosc Int Open 2020; 8: E1769-E1781
  • 14 van der Ploeg K, Haanappel CP, Voor In 't Holt AF. et al. Unveiling 8 years of duodenoscope contamination: insights from a retrospective analysis in a large tertiary care hospital. Gut 2024; 73: 613-621
  • 15 van der Ploeg K, Vos MC, Erler NS. et al. Impact of duodenoscope reprocessing factors on duodenoscope contamination: a retrospective observational study. J Hosp Infect 2024; 154: 88-94
  • 16 Rauwers AW, Voor in 't Holt AF, Buijs JG. et al. Nationwide risk analysis of duodenoscope and linear echoendoscope contamination. Gastrointest Endosc 2020; 92: 681-691 e681
  • 17 R Core Team. R: A language and Environment for Statistical Computing. R Foundation for Statistical Computing, Vienna, Austria. 2023.
  • 18 Botana-Rial M, Leiro-Fernández V, Núñez-Delgado M. et al. A pseudo-outbreak of Pseudomonas putida and Stenotrophomonas maltophilia in a bronchoscopy unit. Respiration 2016; 92: 274-278
  • 19 Guy M, Vanhems P, Dananché C. et al. Outbreak of pulmonary Pseudomonas aeruginosa and Stenotrophomonas maltophilia infections related to contaminated bronchoscope suction valves, Lyon, France, 2014. Euro Surveill 2016; 21
  • 20 Maraolo AE, Licciardi F, Gentile I. et al. Stenotrophomonas maltophilia Infections: A systematic review and meta-analysis of comparative efficacy of available treatments, with critical assessment of novel therapeutic options. Antibiotics (Basel) 2023; 12: 910
  • 21 Alfa MJ, Singh H. Impact of wet storage and other factors on biofilm formation and contamination of patient-ready endoscopes: a narrative review. Gastrointest Endosc 2020; 91: 236-247

Zoom Image
Fig. 1 Illustrations of the AquaTYPHOON system, showing the device's front and side views (top left and center), the barcode scanner and AquaJET (top right), color-coded endoscope connection tubes (bottom left), and the cleaning sink (bottom right).
Zoom Image
Fig. 2 Forest plot with risk differences in contamination between the AquaTYPHOON system and conventional manual cleaning methods according to the different definitions. AM20, microbial growth with ≥ 20 colony-forming units/20 mL of water and/or skin type microorganism; MGO, presence of microorganisms of gut or oral origin.