Keywords Pancreatobiliary (ERCP/PTCD) - ERC topics - Quality and logistical aspects - Training
- Quality management - Performance and complications
Introduction
Endoscopic sphincterotomy is the cornerstone of therapeutic endoscopic retrograde
cholangiopancreatography (ERCP), enabling access to the biliary and pancreatic ductal
systems. A sphincterotome is generally used for this purpose, and various types are
commercially available. The pull type, which facilitates incisions by pushing the
cutting wire, is commonly selected [1 ]. Among the various types of sphincterotomes, one is equipped with a rotational function
[2 ]. However, clear evidence of its effectiveness remains elusive. The primary reason
for the limited use of sphincterotomes with rotational function in patients with normal
anatomy is that experienced professionals can perform precise curved incisions by
manually pulling the cutting wire and manipulating the therapeutic duodenoscope. However,
endoscopic sphincterotomy poses difficulties in patients with surgically altered anatomy,
such as those who have undergone Billroth II and Roux-en-Y reconstructions. Although
ERCP experts can access the papilla in 85% to 95% of these cases [3 ]
[4 ]
[5 ] by using balloon-assisted enteroscopy, performing endoscopic sphincterotomy as intended
is challenging because the direction of the scope and cutting wire against the papilla
is inconsistent.
Because of its difficulty, endoscopic balloon dilation is sometimes chosen by default.
In a randomized controlled trial (RCT) by Bergman et al. [6 ] comparing sphincterotomy and balloon dilation in patients with Billroth II reconstruction,
no difference was observed in procedure time or incidence of early adverse events
(AEs). However, previous RCTs [7 ]
[8 ]
[9 ]
[10 ] comparing sphincterotomy and balloon dilation revealed a higher incidence of post-ERCP
pancreatitis in balloon dilation and more severe bleeding in sphincterotomy. In a
multicenter clinical trial in the United States [10 ], post-ERCP pancreatitis occurred in 15.4% of patients who underwent common bile
duct (CBD) stone removal with balloon dilation, including two (1.7%) deaths; however,
pancreatitis occurred in only 0.8% of those who underwent sphincterotomy (P < 0.001). Based on these findings, sphincterotomy has become the preferred method
for stone extraction in recent years.
If sphincterotomy is considered safe and appropriate even in difficult cases, clinicians
will likely choose this procedure. An endoscopic rotatable sphincterotome (r-sphincterotome)
may be useful for these patients. To the best of our knowledge, two r-sphincterotomes
are currently available. However, their rotational function is inadequate, and they
are not widely used, even for patients with surgically altered anatomy. In our experience,
these r-sphincterotomes do not adequately rotate under the condition of a bent endoscope.
A sphincterotome consists of an operating handle, catheter, and cutting wire. The
cutting wire is tapered at its distal end for proper incision. The torque caused by
rotating the operating handle (proximal rotational force) may not always be transmitted
to the distal side through the catheter.
To address this issue, a novel r-sphincterotome (AIMINGTOME; Asahi Intecc Co., Ltd.,
Aichi, Japan) with a newly devised cutting wire has been developed. The aim of this
study was to compare the feasibility of this novel r-sphincterotome (“N”) with that
of the two existing r-sphincterotomes (“A” and “B”) in an experimental model.
Methods
Experimental procedure
[Fig. 1 ] shows an actual model of the experimental procedure and the input and output degrees
of the r-sphincterotomes evaluated in this study. As in clinical treatment, the r-sphincterotome
was inserted into a side-viewing duodenoscope (TJF290V; Olympus Medical Systems, Tokyo,
Japan). The proximal end of the r-sphincterotome was attached to a servomotor (XU8
K112501; Oriental Motor, Tokyo, Japan), which rotated the sphincterotome at a constant
speed. The duodenoscope was manipulated to mimic the curvature of the alimentary tract.
With the r-sphincterotome inserted into the duodenoscope, it was rotated at a constant
speed (input; 1.5 revolutions per minute) for three complete rotations, while the
degree of rotation at the distal end was recorded over time (output; every 2 seconds).
Ideal rotation of the sphincterotome was defined as input rotation directly reflecting
output in a uniform manner (i.e. rotation in one-to-one correspondence). Deviations
from the expected ideal rotation that collected integral value from the base axis
were noted. Each measurement was conducted 10 times for each r-sphincterotome.
Fig. 1 An actual model of the experimental procedure show input and output degrees of the
r-sphincterotomes evaluated in this study. The r-sphincterotome was inserted into
a side-viewing duodenoscope, that was manipulated to mimic the curvature of the alimentary
tract. The proximal end of the r-sphincterotome was attached to a servomotor, which
rotated the sphincterotome at a constant speed (input; 1.5 revolution per minute),
while the degree of rotation at the distal end was recorded over time (output; every
2 seconds).
Actual endoscopic r-sphincterotomes and cutting wires
[Fig. 2 ] shows the actual r-sphincterotomes and cutting wires of the r-sphincterotomes. [Table 1 ] summarizes detailed information about the cutting wire and the manufacturers of
each sample.
Fig. 2 The actual r-sphincterotomes and cutting wires of the r-sphincterotomes.
Table 1 Detailed information about the cutting wire for each device.
Distal diameter, mm
Length of tapered area, mm
Proximal diameter, mm
Manufacturer
All the devices were made of stainless steel.
N
0.21
125
0.64
Asahi Intecc
A
0.26
270
0.45
Boston Scientific Corporation
B
0.26
175
0.45
Micro-Tech
The proximal diameter of the cutting wire for the newly designed r-sphincterotome
was thicker than that of the others (N vs. A/B: 0.64 vs. 0.45 mm), whereas the distal
end was thinner (0.21 vs. 0.26 mm). The length of the tapered area of the cutting
wire was shorter in the newly designed r-sphincterotome ([Table 1 ] and [Fig. 3 ]). The material used for each inner structure device was stainless steel.
Fig. 3 Components of the endoscopic rotatable sphincterotome and comparison of cutting wire
configurations. The sphincterotome consists of an operating handle, a catheter, and
a cutting wire. The cutting wire is tapered at its distal end to facilitate proper
incision. The length of the tapered area and the proximal and distal diameters of
the cutting wire are unique to each sphincterotome.
Statistical analysis
Data are expressed as the mean ± 2 × standard error. Differences in quantitative data
were compared using the Mann–Whitney U test. All statistical analyses were two-sided,
and statistical significance was defined as P < 0.05. All statistical analyses were performed using SPSS software, version 23 (IBM
Corp., Armonk, New York, United States).
Results
Measurement of degree of endoscopic r-sphincterotomes
[Fig. 4 ] summarizes the input and output degrees of the r-sphincterotomes during rotation
with the servomotor. Deviation from the ideal value was significantly lower in the
newly designed r-sphincterotome than in the conventional ones (N vs. A: 44.9 ± 27.8
vs. 73.7 ± 44.6, P < 0.001; N vs. B: 44.9 ± 27.8 vs. 130.1 ± 71.4 degrees, P < 0.001). [Video 1 ] demonstrates the actual movement of each r-sphincterotome. The newly designed r-sphincterotome
rotated in the same manner as the input rotation, smoothly and at a constant speed.
Fig. 4 Schema illustrating the relationship between input and output degrees of the experimental
procedure. The endoscopic rotatable sphincterotome was inserted into a side-viewing
duodenoscope. The proximal end of the sphincterotome was fastened to a servomotor,
which rotated the sphincterotome at a constant speed (input). The duodenoscope was
held in a bent position to mimic the alimentary tract. The degree of rotation at the
distal end over time was recorded (output).
Feasibility of newly designed rotatable sphincterotome for endoscopic sphincterotomy.Video
1
Discussion
The present experimental study demonstrated that use of the novel r-sphincterotome
was feasible and exhibited excellent rotational function. Although the rotational
function of a sphincterotome is not always necessary for ERCP experts treating patients
with normal anatomy, it might be useful for patients with surgically altered anatomy.
This newly designed r-sphincterotome may change the current paradigm and represent
a new era in endoscopic sphincterotomy.
Endoscopic sphincterotomy is performed to treat various pancreatobiliary diseases,
such as CBD stones, cholangitis, and obstructive jaundice. For biliary sphincterotomy,
many experts recommend cutting against the papilla in the 11- to 12-o'clock direction,
which provides clean and bloodless opening of the sphincter of Oddi [1 ]
[11 ]
[12 ]. However, in some patients with surgically altered anatomy, the papilla cannot be
reached using the same therapeutic duodenoscope. Furthermore, even when access to
the papilla is possible, the view does not originate from the same direction, resulting
in an unfeasible position for precise sphincterotomy. Surgically altered anatomy is
broadly classified as Billroth I reconstruction, jejunal interposition, Billroth II
reconstruction, and Roux-en-Y reconstruction. In cases of Billroth I reconstruction
and jejunal interposition, endoscopic sphincterotomy can be performed in almost the
same manner as in patients with normal anatomy using a therapeutic duodenoscope. However,
in cases of Billroth II reconstruction and Roux-en-Y reconstruction, the procedure
cannot be executed in the same manner. The main reason for this difficulty, as mentioned
above, is the altered orientation. Unconventional endoscopic sphincterotomy may cause
unexpected AEs such as hemorrhage and/or perforation, even when performed by experts.
Furthermore, under technically challenging conditions, hemostasis or stent placement
becomes more difficult.
A sphincterotome with rotational function may facilitate an appropriate cutting incision
and rescue treatment. However, previously designed r-sphincterotomes could not rotate
through a bent endoscope. Whether the cutting wire rotates as intended by the clinician
depends on how the rotational force is directly transmitted from the operating handle
(proximal side) to the distal end. The cutting wire of the newly designed r-sphincterotome
consists of a thick proximal side and a short distal end, which easily transmits the
rotational force. This novel r-sphincterotome is equipped with a well-conceived cutting
wire that uniformly transmits the rotational force from the proximal to the distal
end (as shown in Video 1). If the rotational force is not transmitted uniformly, the
force retained in the cutting wire can cause twisted rotation in an unintended direction
or wire breakage. Ideally, the rotational force on the proximal side should be transmitted
directly to the distal end. However, achieving this ideal condition is challenging,
particularly within the constraints of a bent endoscope and working channel. As depicted
in Fig. 4, the newly designed r-sphincterotome reflects the input rotation to output
with higher accuracy and precision compared with previous r-sphincterotomes (P < 0.001).
A limitation of this study is that all measurements were performed in vitro using
laboratory equipment. In addition, because the study was conducted under strictly
defined conditions, the numeric data may not represent absolute values; rather, they
serve as useful relative values. A large-scale, multicenter, prospective study under
various clinical conditions is necessary to establish the practicality of the r-sphincterotome.
Conclusions
In conclusion, we evaluated the feasibility of the newly designed r-sphincterotome
using an experimental model. We believe that the findings from these experiments may
contribute to easier and more precise endoscopic sphincterotomy.
Feasibility of newly designed rotatable sphincterotome for endoscopic sphincterotomy
(with video)
Yasuki Hori, Kazuki Hayashi, Itaru Naitoh et al. Endoscopy International Open 2024; 12: E1374–E1378. DOI: 10.1055/a-2422-2425 In the above-mentioned article institution 2 was corrected. This was corrected in
the online version on 03.12.2024.