CC BY-NC-ND 4.0 · Thromb Haemost 2025; 125(05): 405-420
DOI: 10.1055/a-2407-1400
Review Article

Handheld Point-of-Care Devices for Snakebite Coagulopathy: A Scoping Review

1   Centre for Snakebite Research and Interventions, Liverpool School of Tropical Medicine, Liverpool, United Kingdom
,
Amy E. Marriott
1   Centre for Snakebite Research and Interventions, Liverpool School of Tropical Medicine, Liverpool, United Kingdom
2   Department of Infection Biology and Microbiomes, Institute of Infection, Veterinary and Ecological Sciences, University of Liverpool, Liverpool, United Kingdom
,
Emma Stars
1   Centre for Snakebite Research and Interventions, Liverpool School of Tropical Medicine, Liverpool, United Kingdom
,
Dianne P. Kitchen
3   UK National External Quality Assessment Scheme for Blood Coagulation (UK NEQAS BC), Sheffield, United Kingdom
,
Steve Kitchen
3   UK National External Quality Assessment Scheme for Blood Coagulation (UK NEQAS BC), Sheffield, United Kingdom
,
Tim A. L. Woods
3   UK National External Quality Assessment Scheme for Blood Coagulation (UK NEQAS BC), Sheffield, United Kingdom
,
Benno Kreuels
4   Department of Implementation Research, Bernhard Nocht Institute for Tropical Medicine, Hamburg, Germany
,
John H. Amuasi
4   Department of Implementation Research, Bernhard Nocht Institute for Tropical Medicine, Hamburg, Germany
5   Department of Global Health, School of Public Health, Kwame Nkrumah University of Science and Technology, Kumasi, Ghana
,
Wuelton M. Monteiro
6   College of Health Sciences, University of the State of Amazonas, Manaus, Brazil
7   Department of Teaching and Research, Fundação de Medicina Tropical Dr Heitor Vieira Dourado, Manaus, Brazil
,
Ymkje Stienstra
1   Centre for Snakebite Research and Interventions, Liverpool School of Tropical Medicine, Liverpool, United Kingdom
8   Department of Internal Medicine/Infectious Diseases, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands
,
Subramanian Senthilkumaran
9   Department of Emergency and Critical Care, Manian Medical Centre, Erode, Tamil Nadu, India
,
Geoff K. Isbister
10   Clinical Toxicology Research Group, University of Newcastle, Newcastle, NSW, Australia
,
David G. Lalloo
1   Centre for Snakebite Research and Interventions, Liverpool School of Tropical Medicine, Liverpool, United Kingdom
,
Stuart Ainsworth
1   Centre for Snakebite Research and Interventions, Liverpool School of Tropical Medicine, Liverpool, United Kingdom
2   Department of Infection Biology and Microbiomes, Institute of Infection, Veterinary and Ecological Sciences, University of Liverpool, Liverpool, United Kingdom
,
Nicholas R. Casewell
1   Centre for Snakebite Research and Interventions, Liverpool School of Tropical Medicine, Liverpool, United Kingdom
› Institutsangaben
 


Abstract

Venom-induced consumption coagulopathy (VICC) is a common complication of snakebite that is associated with hypofibrinogenemia, bleeding, disability, and death. In remote tropical settings, where most snakebites occur, the 20-minute whole blood clotting test is used to diagnose VICC. Point-of-care (POC) coagulation devices could provide an accessible means of detecting VICC that is better standardized, quantifiable, and more accurate. In this scoping review, the mechanistic reasons that previously studied POC devices have failed in VICC are considered, and evidence-based recommendations are made to prioritize certain devices for clinical validation studies. Four small studies have evaluated a POC international normalized ratio (INR) device in patients with Australian Elapid, Daboia russelii, and Echis carinatus envenoming. The devices assessed in these studies either relied on a thrombin substrate endpoint, which is known to underestimate INR in patients with hypofibrinogenemia, have been recalled due to poor accuracy, or have since been discontinued. Sixteen commercially available POC devices for measuring INR, activated clotting time, activated partial thromboplastin time, fibrinogen, D-dimer, and fibrin(ogen) degradation products have been reviewed. POC INR devices that detect fibrin clot formation, as well as a novel POC device that quantifies fibrinogen were identified, which show promise for use in patients with VICC. These devices could support more accurate allocation of antivenom, reduce the time to antivenom administration, and provide improved clinical trial outcome measurement instruments. There is an urgent need for these promising POC coagulation devices to be validated in prospective clinical snakebite studies.


#

Introduction

Snakebite causes at least 94,000 deaths each year and is a World Health Organization (WHO)-listed priority–neglected tropical disease.[1] [2] Coagulopathy is a hallmark of systemic envenoming by diverse medically important snake species distributed across all snakebite endemic regions, including North and South American rattlesnakes and lanceheads (Crotalus and Bothrops spp.), Asian pit vipers and true vipers (e.g., the Russell's viper Daboia russelii and the Malayan pit viper Calloselasma rhodostoma), the African saw-scaled vipers and boomslang (Echis spp. and Dispholidus typus), and several Australian Elapids (e.g., the brown snake Pseudonaja textilis and the taipan Oxyuranus scutellatus).

Some snake species can cause venom-induced consumption coagulopathy (VICC) through the following activities: activation of the coagulation pathway (usually at the common pathway level); fibrinogenolysis; and fibrinolysis.[3] [4] VICC has been associated with a high risk of complications[5] and represents an urgent indication for antivenom.[6] [7] Many of the snake species that induce VICC also disrupt small blood vessels through snake venom metalloproteinase (SVMP) toxin mediated hydrolysis of basement membrane proteins.[8] The resultant state of fibrinogen depletion and destabilized endothelial extracellular structures amplifies the risk of spontaneous bleeding, typically resulting in bite site bleeding, venepuncture site bleeding, and systemic bleeding (gingival, hematuria, conjunctival, epistaxis, etc.).[9] Serious bleeding events can also occur, including intracerebral hemorrhage, which is almost always fatal.[10] [11]

The WHO snakebite management guidelines for Africa[7] and Asia[6] recommend that antivenom is administered to patients with an abnormal 20-minute whole blood clotting test (20WBCT). The 20WBCT, a variation of the Lee–White Clotting Time, which was itself developed a century ago,[12] entails the collection of 2 mL of whole blood into a clean glass tube and, after 20 minutes, a visual inspection for the presence or absence of a clot.[13] This test has been used in low- and middle-income country (LMIC) settings as it is inexpensive (only requiring a glass tube) and can be conducted in remote rural settings without the need for laboratory facilities. However, there are certain limitations to the 20WBCT. Accuracy to detect coagulopathy has been good but imperfect (84% sensitivity and 91% specificity in a recent meta-analysis[14]), which can delay antivenom administration for some patients.[15] Although the versatility of being able to use a range of glass tubes to measure the 20WBCT is an advantage, this lack of standardization can be problematic. There have been anecdotal reports of glass tubes being unknowingly coated with substances, such as silicone, which prevent activation of the coagulation pathway by the glass surface of the tube.[14]

A debate over which is the best method to detect VICC recently arose in the development of a global snakebite core outcome measurement set.[16] Although consensus to include laboratory international normalized ratio (INR) as a core outcome measure was reached, stakeholders were justifiably concerned that this could be challenging in LMIC settings and recommended that point-of-care (POC) devices be prioritized for assessment in validation studies.

Various POC coagulation devices exist, although these have largely been developed to suit the requirements of therapeutic anticoagulation services in high-income settings, and have not been validated for use in snakebite. The WHO South-East Asia snakebite guidelines state that “point-of-care devices for measuring INR and D-dimer [are] unreliable in snakebite victims,”[6] although evidence to support this statement has been limited.

This scoping review aims to (1) review the published literature on the use of POC coagulation devices in snakebite, (2) define the potential roles of POC coagulation devices in snakebite endemic settings, and (3) shortlist the most promising POC coagulation devices for evaluation in preclinical and clinical validation studies. The PRISMA extension for scoping reviews checklist was followed[17] ([Supplementary Appendix], available in the online version).


#

Mechanism of Action of Snake Venom Toxins on the Coagulation Pathway

Several snake venom toxin families can activate and consume key components of the coagulation pathway, resulting in VICC ([Fig. 1A]). As these toxins can vary substantially between snake species, it is important to consider their mechanism of action, as this may inform which POC coagulation devices are likely to be informative. Often, there may be several coagulant toxins present within a venom, each targeting different components of the coagulation pathway. There have been several in-depth reviews on VICC,[18] [19] [20] [21] [22] and therefore only a focused summary of this topic will be presented here.

Zoom Image
Fig. 1 (A) Overview of the site of action of procoagulant snake venom toxins on the coagulation pathway. The red boxes detail the major groups of coagulopathic venom toxins, their enzyme classification, and a nonexhaustive list of medically important snake species that utilize the toxin group. The dashed arrows indicate the site of the coagulation pathway where the major groups of coagulopathic venom toxins act. Adapted from “Coagulation Cascade,” by BioRender.com (2023). Retrieved from: https://app.biorender.com/biorender-templates. (B) Levels of measurement of each point-of-care device on the coagulation pathway. The green boxes detail the names and assay types of the point-of-care devices that have been included in this review. The tails of the green arrows indicate where each assay's activator acts on the coagulation pathway. The arrow heads highlight where the endpoint read of each assay exists on the coagulation pathway. Adapted from “Coagulation Cascade,” by BioRender.com (2023). Retrieved from https://app.biorender.com/biorender-templates. Ca2+, calcium; FDPs, fibrin degradation products; PL, phospholipid; TF, tissue factor.

Factor X Activators

Factor X is a major constituent of the prothrombinase complex. Snake venom toxins that specifically activate factor X consist of either SVMPs (such as the highly characterized RVV-X from the venom of D. russelii) or snake venom serine proteases (SVSP), which have been described either in viper venoms or Elapid venoms, respectively.[19] [23] [24] Both these toxins activate factor X in a similar manner to that of physiological tenases, through cleavage of an inhibitory peptide resulting in the activation of factor X's proteolytic activity.


#

Factor V Activators

Factor V is activated to factor Va by thrombin or factor Xa. Factor Va is the active form and is a cofactor that significantly enhances thrombin formation.[25] Venom toxin factor V activators are serine protease enzymes that act to cleave factor V, usually at a different site to thrombin, but nevertheless to an active procoagulant form. Examples include the serine protease RVV-V in Russell's viper (D. russelii) venom, or the thrombocytin toxin found in Bothrops atrox venom.[25]


#

Prothrombin Activators

Prothrombin activators exist in a diverse range of venoms and have been classified into four groups (A, B, C, and D) based on their toxin families and functional characteristics.[26] Group A and B prothrombin activators are SVMP toxins which convert prothrombin into the catalytically active thrombin intermediate, meizothrombin, which subsequently spontaneously converts to thrombin.[19] Ecarin, present in the venom of Echis carinatus (Indian saw scaled viper), is a well-described example of a group A activator,[27] while carinactivase-1, found in the venom of white-bellied carpet viper (Echis leucogaster), is a well-characterized group B activator.[28] The primary difference between group A and B prothrombin activators is in the requirement of cofactors; group A activators act independent of the need of factor Va, Ca2+, or phospholipids,[29] while group B similarly do not require factor Va or phospholipids but are reliant on Ca2+.[28]

Group C prothrombin activators belong to the SVSP family and are only found in the venoms of certain Australian Elapid snakes,[20] such as the brown snake P. textilis [30] and the taipan O. scutellatus.[31] These prothrombin activators consist of both a factor Xa-like and a factor Va-like toxin, which form large multi-subunit enzymes, effectively mimicking the prothrombinase complex. This toxin efficiently converts prothrombin to thrombin in the presence of Ca2+ and phospholipids,[20] and is remarkably similar to the serine protease human clotting factors, in contrast to the SVMP prothrombin activators in viper venoms.

Group D prothrombin activators, similar to group C activators, are SVSP toxins only found in Australian Elapids,[32] [33] such as tiger snakes (Notechis spp.). This group consists of enzymes which resemble factor Xa only, and thus also require activated factor V, in addition to Ca[2] and phospholipids, to effectively cleave prothrombin into thrombin.[33]


#

Thrombin-Like Enzymes

Numerous thrombin-like enzymes (TLEs), sometimes termed fibrin(ogen)olytic enzymes, have been identified in snake venoms, and these can be categorized as fibrinogenases, and are typically members of the SVSP or SVMP toxin families.[34] The α-fibrinogenases tend to preferentially cleave the α-chain of fibrin(ogen), while the β-fibrinogenases preferentially cleave the β-chain of fibrin(ogen).[35] The term fibrin(ogen) highlights that these enzymes cleave both fibrinogen and fibrin clot. Unlike thrombin, these proteinases do not produce fibrinopeptides A or B and result in the formation of a soft fibrin clot that is not crosslinked.


#
#

Search Strategy for This Review

The following terms were used to search PubMed for articles published until November 2023 that described the use of POC coagulation devices in snakebite: (snake OR venom) AND (“point of care” OR bedside) AND (coagulopathy OR coagulation OR clotting). Additional articles were sought from the reference lists of included papers, and through recommendations by the review authors. There were no language restrictions. The PubMed search identified 40 articles, of which four were included that assessed a handheld POC device to detect coagulopathy in humans with snakebite[36] [37] [38] [39] [40] ([Table 1]). No additional articles were identified from reference lists or author recommendations. One article was a short report that included limited data,[36] but the corresponding author was contacted and they shared the unpublished underlying data, which have been included in this scoping review.

Table 1

Summary of published literature on the use of POC devices in VICC

Author (year)

Geographic location

Sample size

Biting genus

POC device (manufacturer); assay type

Findings

Celenza (2011)

Australia

1

Pseudonaja

CoaguChek XS Plus (Roche); INR

The routine use of POC INR in an emergency department, for any disease, was evaluated against a laboratory INR gold standard. Of the 293 patients studied, 1 had snakebite. This was the only occasion where the POC INR produced a false-negative result. The laboratory INR was >10; the paired POC INR was 1.4.

Cubitt (2013)

Australia

1

Notechis

CoaguChek (Roche); INR

INRatio2 (Alere); INR

Cobas h232 system (Roche); D-dimer

This case report of Notechis bite described initial presentation to a small rural hospital where the INRatio2 POC device result was 1.3 (this was not confirmed by a laboratory INR). After transfer to a regional hospital, paired samples demonstrated a laboratory INR of >12 and a POC INR (CoaguChek) of 0.9. POC D-dimer (Cobas H232) was 0.38 mg/L, whereas the laboratory D-dimer was reported as 812 mg/L.

O'Rourke (2013)

Australia

15

Pseudonaja, Oxyuranus, Notechis, Pseudechis, and Acanthophis

i-STAT (Abbott); INR

This prospective study compared the i-STAT POC device with laboratory INR in 15 participants with snakebite. i-STAT POC substantially underestimated the INR in all cases with VICC. Three of the seven participants with VICC had a normal POC INR, and 1 of the 7 patients without laboratory-confirmed VICC had an abnormal POC INR.

Senthilkumaran (2014)

India

18

Daboia russelii, and

Echis carinatus

CoaguChek (Roche); INR

INRatio2 (Alere); INR

i-STAT (Abbott); INR

This letter to the editor provided a narrative report highlighting that POC INR devices produce false-negative results in patients with snakebite envenoming. The underlying data were made available for this scoping review ([Supplementary Appendix], available in the online version). All three POC INR devices substantially underestimated the laboratory INR in patients with Daboia russelii envenoming (n = 15), particularly when the laboratory INR was >4. The devices were accurate in patients with Echis carinatus envenoming, but there were only three such cases, all of whom had a laboratory INR ≤4.3.

Abbreviations: INR, international normalized ratio; POC, point of care; VICC, venom-induced consumption coagulopathy.


A list of handheld POC devices for measuring coagulation was prepared by searching the published literature, contacting commercial suppliers, and by accessing information of devices evaluated by the UK Blood Coagulation National External Quality Assessment Service (NEQAS). To be eligible for inclusion, devices needed to be handheld, battery-powered, fully automated, and suitable for use at the bedside. Furthermore, devices needed to be compatible with measurement in whole blood, to overcome the need for a centrifuge. Thromboelastography and thromboelastometry devices were excluded, as were devices that had been withdrawn from the market.

The following characteristics of each device were considered: parameter(s) measured, sample types accepted (blood ± plasma), sample collection (venipuncture or capillary), sample volume, cost of device, cost of cartridge, limit of detection, whether battery operated, and the method of measurement. Devices have been presented according to the following categories, based on the aspect of the coagulation pathway that they measure: prothrombin time (PT; or INR); activated partial thromboplastin time (aPTT); activated clotting time (ACT); fibrinogen quantification; and fibrin(ogen) degradation products (FDPs), including D-dimer.

The search strategy identified 50 devices, from which 16 were eligible for inclusion in this scoping review ([Table 2]). The reasons for the exclusions were: not handheld (19), no longer commercially available (8), and only compatible with centrifuged plasma (1). A further six devices for measuring D-dimer were excluded as they were qualitative and used a cut-off that was deemed too low for diagnosing VICC, the rationale for which has been described below.

Table 2

Overview of handheld point-of-care coagulopathy devices

Device name (manufacturer)

Blood or plasma (citrated or native)

Sample collection

Sample volume

Cost device (cost per test cartridge) USD[a]

Limit of detection

Confirmed battery operated

Method of measurement

Promising for snakebite

INR

CoaguChek Pro II (Roche)

Blood (native)

Capillary or venous

8 µL

$1,900 ($4)

INR 0.8–8.0

Yes

Activator: human recombinant tissue factor

Endpoint: electrochemical detection of thrombin-dependent cleavage of an electrochemical substrate.

No

(thrombin endpoint prone to false-negative results)

Hemochron Signature Elite (Accriva Diagnostics)

Blood (citrated or native)

Capillary or venous

15 µL

$21,400 ($11)

INR 0.8–10

Yes

Activator: thromboplastin

Endpoint: optical detection of the rate of flow of the blood sample.

No

(prohibitively expensive)

Xprecia Stride (Siemens)

Blood (native)

Capillary or venous

6 µL

NA

INR 0.8–8.0

Yes

Activator: thromboplastin

Endpoint: electrochemical detection of thrombin-dependent cleavage of an electrochemical substrate.

No

(thrombin endpoint prone to false-negative results)

microINR (iLine Microsystems)

Blood (native)

Capillary or venous

3 µL

$500 ($2)

INR 0.8–8.0

Yes

Activator: thromboplastin

Endpoint: optical detection of the rate of flow of the blood sample.

Yes

Coag-Sense (CoaguSense Inc)

Blood (native)

Capillary or venous

12 µL

$1,100 ($1.5)

INR 0.8–8.0

Yes

Activator: thromboplastin

Endpoint: optical detection of the rate of flow of the blood sample.

Yes

i-STAT (Abbott)

Blood (native)

Capillary or venous

20 µL

$3,500 ($6)

INR 0.9–8.0

Yes

Activator: human recombinant thromboplastin

Endpoint: electrochemical detection of thrombin-dependent cleavage of an electrochemical substrate.

No

(thrombin endpoint prone to false-negative results)

LumiraDx INR (LumiraDx)

Blood (native)

Capillary or venous

8 µL

$5,900 ($4)

INR 0.8–7.5

Yes

Activator: recombinant thromboplastin

Endpoint: thrombin-dependent cleavage of a quenched fluorescent substrate. The fluorescent signal is detected and quantified.

No

(thrombin endpoint prone to false-negative results)

Activated partial thromboplastin time (aPTT)

CoaguChek Pro II (Roche)

Blood (native)

Capillary or venous

8 µL

$1,900 ($4)

20–130 seconds

Yes

Activator: celite and phospholipids.

Endpoint: electrochemical detection of thrombin-dependent cleavage of an electrochemical substrate.

No

(thrombin endpoint prone to false-negative results)

Hemochron Signature Elite (Accriva Diagnostics)

Blood (citrated or native)

Venous

45 µL

$21,400 ($11)

20–400 seconds

Yes

Activator: kaolin and phospholipids.

Endpoint: optical detection of the rate of flow of the blood sample.

No

(prohibitively expensive)

Activated clotting time (ACT)

i-STAT (Abbott)

Blood (native)

Venous

40 µL

$3,500 ($14)

50–1000 seconds

Yes

Activator: celite or kaolin cartridges available.

Endpoint: electrochemical detection of thrombin-dependent cleavage of an electrochemical substrate.

No

(thrombin endpoint prone to false-negative results)

Hemochron Signature Elite (Accriva Diagnostics)

Blood (native)

Venous

200 µL

$21,400 ($11)

0–1005 seconds

Yes

Activator: celite or kaolin cartridges available.

Endpoint: optical detection of the rate of flow of the blood sample.

No

(prohibitively expensive)

Fibrinogen

qLabs FIB system (Stago)

Blood (citrated or native) or plasma (citrated)

Venous

15 µL

$1,900 ($21)

1.0–4.0 g/L

Yes

Activator: thrombin-coated channels

Endpoint: detection of the rate of flow of the blood sample by an electrometer

Yes

D-Dimer and other fibrinogen degradation products

Cobas h232 system D-dimer (Roche)

Blood (Heparinized)

Venous

150 µL

$2,400 ($12)

0.1–4.0 µg/mL

Yes

Test strips contain one gold-labeled and one biotinylated monoclonal antibody against fibrin degradation products. In the presence of D-dimer a sandwich complex accumulates and produces a red colored line on the strip, the intensity of which is quantified by an optical measuring system.

No

(false-negative result in a case report of Notechis envenoming[39]).

PL mini handheld D-dimer (PremaLabs)

Blood

Capillary or venous

30 µL

$2,100 ($8)

0.1–10.0 µg/mL

Yes

The PL mini strips contain antibodies fluorescently labeled with chelated lanthanide, which are used to quantify particular biomarkers. When UV light is shone on the test strip, a fluorescence signal is produced that is converted to a digital value, which quantifies the amount of antigen–antibody complexes on the test strip.

Yes

(not in snake venoms predominant in TLEs)

PL mini handheld FDP (PremaLabs)

Blood

Capillary or venous

30 µL

$2,100 ($14)

2.5–80.0 µg/mL

Yes

Yes

LumiraDx D-Dimer

Blood or plasma (citrated)

Capillary or venous

15 µL

$5,900 ($5)

0.19–4.0 µg/mL

Yes

Test strips contain antibodies against fibrin degradation products that are bound to magnetic particles and fluorescent latex particles. In the presence of D-dimer, these form complexes that are held in place by a magnet, and the fluorescence signal is quantitatively measured.

Yes

(not in snake venoms predominant in TLEs)

Abbreviations: INR, international normalized ratio; TLE, thrombin-like enzyme.


a Device and cartridge costs are estimates based on quotes provided by commercial providers. True costs are likely to vary between commercial providers and by geographic location.



#

Evidence for Using POC Devices in VICC

Three studies that have evaluated POC coagulopathy devices in snakebite were conducted in Australia,[37] although two of these only described a single case of snakebite envenoming[38] [39] ([Table 1]). The paper by O'Rourke et al[37] was a prospective observational study that evaluated the i-STAT POC device for measuring INR in 15 participants with snakebite, which found that it consistently, and substantially, underestimated the laboratory INR, although it should be kept in mind that there can be variability in the accuracy of laboratory INR results. The sensitivity of the i-STAT device to detect an INR >1.4 was 57%, and the specificity was 88%. The other two studies, of single cases with VICC in Australia, demonstrated similar findings for the CoaguChek device, with false negative POC INR results.[38] [39]

The underlying data from a study conducted in India by Senthilkumaran et al,[36] were made available for this scoping review ([Supplementary Appendix], available in the online version). This study assessed the following POC INR devices in three patients with E. carinatus envenoming and 15 patients with D. russelii envenoming: Alere INRatio2, CoaguChek S, and i-STAT. All three POC devices grossly underestimated the laboratory assay, particularly when the INR was elevated above 4. The sensitivity of the Alere INRatio2, CoaguChek S, and i-STAT devices to detect a laboratory INR >1.4 was 50, 56, and 56%, respectively. Specificity could not be calculated as all the patients had a laboratory INR value >1.4.

The POC INR and laboratory INR values from the studies by O'Rourke and Senthilkumaran are depicted in [Fig. 2]. The theory as to why so many false-negative results occurred has been described in the next section.

Zoom Image
Fig. 2 Comparison of laboratory and POC device INR by participant from previous studies (O'Rourke et al[37] and Senthilkumaran et al[37]). (A) The laboratory, i-STAT, CoaguChek S, and Alere INRatio2 INR values for 15 participants with Daboia russelii envenoming: the participants are ordered according to increasing laboratory INR values. (B) The laboratory, i-STAT, CoaguChek S, and Alere INRatio2 INR values for three participants with Echis carinatus envenoming. (C) The laboratory and i-STAT INR values for 15 participants with Australian Elapid envenoming. INR, international normalized ratio.

The case report by Cubitt et al is the only published report describing the use of a POC D-dimer assay to detect VICC.[39] This Cobas H232 POC device produced a borderline raised D-dimer measurement of 0.38 mg/L (upper limit of normal 0.36 mg/L), whereas the laboratory assay result was substantially higher than this.

No studies were identified on the use of a POC INR device to detect VICC in Africa, Europe, South America, or North America.


#

Prothrombin Time and International Normalized Ratio

The PT assay was originally defined by Arnand J. Quick in 1935, as the time taken in seconds for a fibrin clot to be formed following activation of the extrinsic pathway (or tissue factor pathway) by thromboplastin.[41] This test is widely used in modern medicine whereby the sample is collected in a blood tube containing citrate, which is subsequently centrifuged to produce platelet-free plasma before adding an excess of calcium (to reverse the anticoagulant effect of the citrate) and thromboplastin (a mixture of tissue factor and phospholipid). In the laboratory environment, the PT assay is typically performed using an automated analyzer, and the endpoint is based on either optical or mechanical detection of a fibrin clot.[42]

As many different forms of thromboplastin are available, which can influence the PT, the WHO introduced the INR. The INR is calculated by dividing the test PT by the mean normal PT (the mean PT of the healthy adult population), using a reagent and method which has been calibrated using the recommended WHO reference thromboplastin reagent and procedure.[43] In clinical practice, the PT is typically used as a routine screen for coagulopathy in hospitalized patients, as a measure of hepatic synthetic function, and to detect disseminated intravascular coagulopathy (DIC). The INR was developed for therapeutic drug monitoring of warfarin and related drugs.

In VICC, the PT and INR tend to be severely prolonged—often to the point of being unrecordable.[44] Because the PT is based on activation of the extrinsic pathway with onward activation of the common pathway to produce a fibrin clot, it is variably affected by any change along this pathway, including low fibrinogen (common pathway) or factor II, V, VII, or X deficiencies. In VICC, fibrinogen is cleaved by TLEs, or by endogenous thrombin that has been produced in response to venom prothrombin or factor X activators.[22] Once circulating fibrinogen has been severely depleted, the PT and INR endpoint, at least for methods based on the detection of a fibrin clot, cannot be reached, and an unrecordable result is produced ([Fig. 1B]).

The two studies that have clinically assessed the i-STAT device, in Australia and India, both reported multiple false-negative results.[36] [37] In the Australian study, of the five participants with a laboratory INR value above the upper limit of detection (INR >10), the paired i-STAT INRs were in the range of 1.2 to 3.2. The study in India found that amongst nine patients with a laboratory INR >10, the paired i-STAT INR values were between 1.1 and 1.6 ([Supplementary Appendix], available in the online version).[36]

The i-STAT INR cartridges contain a thrombin substrate that produces an electrochemical signal in response to the accumulation of thrombin enzyme. Thus, the false-negative results most likely occurred because the patients with VICC had severely depleted fibrinogen (resulting in a prolonged laboratory INR) but still had sufficient amounts of prothrombin,[45] which provided the capacity to produce thrombin (resulting in a near-normal POC-INR). O'Rourke et al highlighted this issue in their manuscript, and further confirmed this hypothesis by demonstrating that a similar false-negative result was produced when a sample from a patient with congenital afibrinogenemia was tested.[37]

The INRatio2 device tested by Senthilkumaran et al in India[36] initiates coagulation using a recombinant human thromboplastin reagent, and measures the change in electrical impedance, across the forming blood clot, between two electrodes. Theoretically, the INRatio2 device may have maintained accuracy in patients with depleted fibrinogen, as it detects the formation of clot, rather than the production of thrombin. However, in 2016 the Food and Drug Administration announced a class I recall of the INRatio2 device and test strips, as it was found to produce false low INR readings that posed a “risk of major or fatal bleeding.” Certain medical conditions were found to be associated with falsely low INRatio2 readings including: anemia (hematocrit <30%), acute inflammatory conditions, bleeding, or bruising[46]; all of which frequently occur in patients with VICC.

The CoaguChek S device cartridge contains iron particles that are mixed with the sample and subject to alternating magnetic fields, which detects reduced movement of the particles as the sample clots. Thus, as this device is measuring clot formation, prolonged INR due to hypofibrinogenemia would be detected, and the cause of the false low INR readings is uncertain. The CoaguChek S device is no longer commercially available as it has been superseded by the CoaguChek Pro II, which relies on a thrombin substrate-based endpoint, and thus would likely be inaccurate in patients with hypofibrinogenemia.

Future studies should restrict clinical assessment of PT/INR POC devices for VICC to those devices that rely on a fibrin endpoint. As outlined in [Table 2], there are three devices that fulfill this requirement: Hemochron Signature Elite, microINR, and Coag-Sense.


#

Activated Partial Thromboplastin Time

The aPTT is used to investigate the intrinsic pathway of the coagulation pathway. Similar to the PT, the sample is collected in a citrated blood tube, centrifuged to platelet-free plasma, and the time until clot formation is reported in seconds. The difference is that the reagents include contact activators such as kaolin,[47] a powdered clay mineral which activates factor XII,[48] silica, or ellagic acid. The aPTT is used in clinical practice for therapeutic monitoring of unfractionated heparin infusions, and to detect deficiencies in the intrinsic pathway associated with certain bleeding disorders, such as von Willebrand disease, hemophilia A and B, and acquired hemophilia.[49]

As the endpoint of laboratory measured aPTT is fibrin clot formation, the aPTT, like the PT, is severely prolonged when fibrinogen is depleted ([Fig. 1B]). In a study of Australian Elapid envenoming, the aPTT produced very similar results to the PT, and was inversely associated with fibrinogen concentration, suggesting that aPTT and PT indirectly measure coagulopathy induced by fibrinogen depletion.[50]

The aPTT assay has a limited POC commercial market because therapeutic monitoring of unfractionated heparin is rarely necessary following the increasing clinical use of longer half-life fractionated heparins. Consequently, few POC devices are available. The CoaguChek Pro II can measure aPTT, but relies on a thrombin substrate endpoint that would likely produce false-negative results for VICC if compared with a laboratory-based aPTT (i.e., with a fibrin clot formation-based endpoint). The Hemochron Signature Elite may be accurate, as it relies on optical detection of the rate of flow of blood. A potential advantage of the Hemochron Signature Elite is that the same device can measure the aPTT and PT (through use of different cartridges), although these parameters are likely to provide similar results in the context of fibrinogen depletion, and the device is prohibitively expensive for snakebite (approximately $21,400 USD for the device and $11 USD per test strip).


#

Activated Clotting Time

Similar to the aPTT, the ACT is based on the time taken for a clot to be formed following the activation of the intrinsic pathway by a factor XII activator (either celite or kaolin; [Fig. 1B]). The ACT was developed by Paul Hattersley in 1966 as a means to evaluate coagulopathy in whole blood, rather than in plasma.[51] As the ACT can provide rapid and reliable results at the bedside, it has become the gold standard for monitoring heparin anticoagulation during cardiopulmonary bypass surgery.[52]

A high surface area powder of celite or kaolin acts as a more potent activator in ACT, rather than the glass of the tube in 20WBCT. Thus, a potential advantage of the ACT is that a result can be read earlier, with samples tending to clot within 2 minutes under normal conditions.[51] An ACT between 180 and 240 seconds has been found to correspond to a Lee–White clotting time of 35 to 45 minutes, which was a key factor in the ACT gaining favor for monitoring of heparin therapy over the Lee–White method.[53]

The ACT tends to be measured in theater using desktop devices (which typically record clot formation through measuring changes in the movement of a magnet within the sample), which is not as practical as a handheld device. The i-STAT and Hemochron Signature Elite devices can measure ACT, but are unsuitable for use in VICC as the former relies on a thrombin cleavage endpoint and the latter may be too expensive. However, it should be considered that the ACT can be manually measured in a blood tube (that contains either celite or kaolin), with visual detection of clot formation—as was the original technique described by Hattersley.[51] This technique may offer a standardized, rapid, and quantifiable assay for diagnosing VICC, which could have utility for both low- and high-income settings and obviate the need for an electronic device.

In a published (although not peer reviewed) dissertation by Shenoy, adults with VICC in Vellore, India underwent measurement of 20WBCT and ACT, which were compared with a laboratory INR gold standard.[54] An ACT threshold of 4 minutes was 94% sensitive and 97% specific for predicting an INR above 1.2 at admission, while the 20WBCT had 81% sensitivity and 90% specificity.[54] In LMIC settings, the ACT might offer an affordable, better standardized, more sensitive, and more specific alternative to the 20WBCT. In high-income settings, the ACT could provide a more rapid result than laboratory-based coagulation studies, which would reduce the time to antivenom administration.


#

Fibrinogen Quantification

Although much less commonly performed than the PT and aPTT, assays to quantify circulating fibrinogen are available. The most widely used approach is the Clauss fibrinogen assay, which is based on adding a standard amount of thrombin to a diluted plasma sample and then measuring the time until clot formation.[55] Through direct cleavage of fibrinogen by thrombin, other components of the coagulation pathway are bypassed, and the concentration of fibrinogen can be estimated ([Fig. 1B]). In clinical practice, fibrinogen is typically measured in specialist circumstances, such as the investigation of congenital fibrinogen disorders, massive trauma, DIC, or obstetric hemorrhage.[56]

Fibrinogen has theoretical value in diagnosing VICC, particularly for TLE predominant snake venoms that cleave fibrinogen without affecting other factors in the coagulation pathway.[40] Clinical data are limited, but a study in Australia that compared the time to improvement of different clotting assays found that fibrinogen was of limited additional benefit for measuring response to antivenom treatment,[50] with median times of recovery for PT, aPTT, and fibrinogen of 9.2, 5.2, and 8.8 hours, respectively.[50] A study conducted in patients with Echis pyramidum envenoming in Djibouti demonstrated a similar pattern with median recovery times of 25, 9, and 40 hours for the PT, aPTT, and fibrinogen assays, respectively.[44] However, time to recovery of these assays in response to antivenom therapy is not necessarily clinically meaningful, and further studies comparing coagulation dynamics against clinically significant outcomes, such as the risk of bleeding events,[57] are urgently needed.

One eligible POC device for measuring fibrinogen was identified; the qLabs FIB system (Stago). This device was recently compared with a Clauss-derived laboratory gold standard and demonstrated 93.5% sensitivity and 100% specificity to predict a fibrinogen cut-off of 2.0 g/L.[58] Furthermore, the accuracy of the qLabs FIB system was maintained at low fibrinogen levels, which are typical in VICC, to the lower limit of detection of 1.0 g/L.[58]


#

Fibrinogen Degradation Products, Fibrin Degradation Products, and Crosslinked Fibrin Degradation Products

Following formation of a normal fibrin clot, plasmin acts to degrade: (1) soluble fibrinogen to FDPs; (2) soluble fibrin to fibrin degradation products (also abbreviated to FDPs), and (3) crosslinked insoluble fibrin to crosslinked fibrin degradation products (XDPs). D-dimer is an XDP that has gained favor as a target for diagnostic testing because it is more specific than FDPs for the investigation of venous thromboembolism (VTE)[59] and it is relatively stable (due to its resistance to further digestion by plasmin).[60]

Because D-dimer measurement is used to investigate VTE in emergency and ambulatory care facilities in high-income settings, many D-dimer POC devices are available. However, these devices are often qualitative, in that they provide a binary result based on a threshold optimized for VTE (typically 0.5 mg/L). This threshold is likely to have poor specificity in snakebite as D-dimer is elevated by inflammatory disorders and trauma,[61] which are caused by local envenoming. In a recent study of patients with VICC in Australia, a D-dimer threshold of 2.5 mg/L was used,[62] which demonstrated 95% sensitivity and specificity to identify VICC (defined as INR >1.4 or fibrinogen <2.0 mg/L).[62] It should be noted that D-dimer results vary between assays and, therefore, it may be necessary develop device specific cut-off values.

There is concern that certain snake venoms do not lead to production of crosslinked fibrin, which is a necessary precursor of D-dimer. TLEs cleave fibrinogen to produce a soft fibrin clot, that is not crosslinked[40] and will not be degraded to form D-dimer through the action of plasmin ([Fig. 1B]). For snake venoms predominant in TLEs, it may thus be preferable to rely on a POC device that can measures FDP, such as the PL mini handheld device (PremaLabs). Few clinical studies of snakebite envenoming have included measurement of both D-dimer and FDPs,[57] and the comparative accuracy of these assays to predict clinically meaningful endpoints is unknown.


#

Potential Role of Point-of-Care Devices in Clinical Practice

Snakebite predominantly occurs in remote rural settings[63] where laboratory infrastructure is limited. Even in high-income settings such as Australia, health care centers in rural areas, where snakebites tend to occur, do not necessarily have onsite laboratory coagulation testing.[37] POC devices could offer potential benefit to patients by providing an earlier diagnosis of VICC in rural health care settings, and reducing the time to antivenom treatment. In LMIC settings, where the current standard is to measure the 20WBCT or Lee–White clotting time, a POC device or the manual ACT method may improve accuracy to diagnose VICC. Increased sensitivity to detect VICC would ensure that more patients with systemic envenoming receive antivenom, while improved specificity would reduce the number of patients that unnecessarily receive antivenom, which is costly and often associated with a high risk of allergic reaction.

It should also be considered that the current thresholds for defining coagulopathy and administering antivenom, such as the 20WBCT in LMIC settings, or the coagulation thresholds of the snakebite severity score in the United States,[64] have not been optimized through clinical validation studies. For example, in a study of patients with VICC in Sri Lanka, a WBCT measurement interval of 15 minutes, rather than 20 minutes, was found to improve accuracy to detect VICC.[65] POC devices provide standardized and quantifiable measures of coagulation that can be used in future clinical studies to define evidence-based thresholds for administering antivenom, which may ultimately translate into better outcomes for patients. These studies must use clinically meaningful endpoints, to ensure that antivenom is administered to those in whom the benefits outweigh the risks.


#

Potential Role of Point-of-Care Devices in Clinical Research

Many of the snakebite clinical trials that have been conducted in LMIC settings have not included any outcome measures based on laboratory coagulation assays.[66] Such assays require access to a reliable electricity supply, a centrifuge, a coagulation analyzer, and appropriately trained laboratory technicians. There are substantial costs associated with setting up a laboratory facility for the purposes of a clinical trial, particularly if multiple laboratories are required across several recruitment sites, while funding for snakebite research has historically been limited.[67] Even if such capacity strengthening were to take place, it is unlikely that the facilities for undertaking coagulation studies would be sustainable in the light of ongoing maintenance costs.

These challenges have resulted in a situation where snakebite clinical trials based in high-income settings rely on gold standard laboratory-based coagulation study endpoints,[68] while those in LMIC settings tend to use the 20WBCT.[9] POC devices could be the key to enabling the use of formal coagulation study endpoints in clinical trials, regardless of their setting.

Before POC tests can be advocated for use in snakebite clinical trials, their accuracy urgently needs to be assessed in prospective clinical observational studies. As demonstrated by O'Rourke et al in Australia and Senthilkumaran et al in India, accuracy can be evaluated by analyzing blood samples in duplicate on a POC device and a validated laboratory-based analyser.[37] Ideally a range of POC devices that measure PT, ACT, and fibrinogen should be evaluated. Furthermore, it would be important to assess POC devices in diverse geographic settings including Africa, Asia, North and South America, and Australia. Variations in snake venom toxin compositions between biting species will mean that different POC devices might be more or less useful in different geographic locations. Furthermore, it is important for future studies to go further than simply assessing whether POC devices can predict their corresponding laboratory-based assay. Of greater importance is whether the use of a POC device can meaningfully impact upon clinical and patient-centered outcomes,[69] such as time to treatment, antivenom use, or frequency of bleeding events.


#

Potential Role of Point-of-Care Devices for In Vivo Preclinical Research

Because clinical trials are costly and time-consuming to complete, preclinical murine models of envenoming have been heavily relied upon to assess the efficacy of antivenoms before they are used in humans.[70] Ideally, antivenoms with promising preclinical efficacy should subsequently be assessed in human clinical trials, but many of the antivenoms marketed in resource-limited settings have only been assessed preclinically.[71]

The WHO Expert Committee on Biological Standardization recommends that antivenom efficacy should be confirmed in murine median effective dose (ED50) models prior to human use.[72] This assay involves co-administering a rapidly lethal dose of venom with antivenom to mice, with the binary measure of mortality used as the outcome measure. Despite several limitations, including growing recognition that this model does not reflect clinical snakebite,[73] or the pharmacokinetics of envenoming,[74] for now, such assays are the gold standard.[75]

Coagulopathy-based outcome measures are rarely used in preclinical research.[76] Literature on the use of handheld POC devices in rodent preclinical models is limited, with benchtop devices and ELISAs for measuring coagulation only occasionally being employed.[77] [78] [79] These devices require large volumes of whole blood or plasma, which typically necessitate sacrificing the animal. As handheld POCs require a reduced volume, multiple blood samples could be taken from each animal. This is known as “microsampling,” which is an animal experimentation refinement technique encouraged by the National Centre for the Replacement, Refinement, and Reduction of animals in Research (NC3Rs) that reduces the numbers of animals required for a study while increasing the amount of data generated.[80] [81]

By using low sample volumes, POC devices could be used to measure improvement in VICC, rather than simply rescue from lethal effect. Quantifiable assays, such as INR or fibrinogen, that can be recorded at multiple time points, would provide a more sensitive outcome measure, and provide kinetic information on the time to therapeutic effect and the occurrence of recurrent coagulopathy. As well as providing important efficacy readouts that align with clinical trial outcome measures,[16] these approaches could reduce the number of mice needed to power a preclinical study, and lessen suffering through administration of lower doses of venom.


#

Cost Considerations

The affordability and potential cost–benefit of POC coagulation devices will be discussed in turn for high-income, middle-income, and low-income settings.

In high-income snakebite endemic settings, such as the United States and Australia, these devices are likely to be affordable, yet their costs would need to be justified given that laboratory-based coagulation studies are likely to be available. However, the study by O'Rourke et al, which was conducted in Australia, highlighted that many of the health care facilities based in the rural areas where snakebites tend to occur do not have access to on-site laboratory coagulation testing.[37] A POC device could streamline care by preventing unnecessary onward referral for coagulation testing of patients with snakebite that do not have VICC.

In middle-income settings, such as Brazil and India, certain POC devices may be affordable. The cost–benefit ratio may be particularly favorable given the high incidence of snakebite in these settings.[1] [82] The most affordable POC device identified in this review was the microINR system, which costs 500 USD for the device and 2 USD per test cartridge. Considering that a course of antivenom treatment typically costs over 50 USD,[83] [84] and that a POC device could facilitate more effective allocation of antivenom, it is possible that the cost of the device would be off-set. Formal cost-effectiveness studies would be important to understand the indirect cost savings of implementing POC tests in these settings.

Unfortunately, in low-income settings access to essential diagnostics for a range of diseases of public health importance, including human immunodeficiency virus and tuberculosis, is severely limited.[85] Thus, it is unlikely that funding would be allocated to rolling out POC devices for snakebite envenoming in these settings. A notable exception identified in this review is the manual ACT method. The consumables for this assay include a plastic or glass tube containing a small amount of either kaolin or celite, which are simply ground up rock or diatomaceous earth, respectively. One kilogram of celite costs ∼20 USD, which would provide enough material to produce thousands of ACT tubes. Thus, the manual ACT method could be produced at a negligible additional cost to the 20WBCT or Lee–White clotting test, while potentially offering improved accuracy and a more rapid result.[54]

A final important consideration about the feasibility of using POC devices in low-income settings is their potential role in clinical research, including clinical trials. Although funding for snakebite research has historically been limited,[67] the costs of these devices are within the limits of modest clinical research budgets, and they are more affordable than establishing laboratory facilities, which require dedicated personnel and a reliable electricity supply. Furthermore, the improved accuracy of POC devices, and their provision of a continuous outcome measure, as opposed to the binary readout of the 20WBCT, is likely to allow for a reduction in the sample size of a clinical trial, which could substantially reduce costs. The need to improve outcome measure reporting in snakebite clinical trials has recently been advocated for,[66] and a group of leading global snakebite experts reached consensus that INR should ideally be measured in all future snakebite clinical trials, including those based in low-income settings.[16]


#

Conclusion

This review has identified a range of POC devices that show promise for diagnosing VICC without the need for laboratory facilities. Various devices that rely on a thrombin substrate–based endpoint should not be considered for evaluation, as existing mechanistic and clinical data suggest they will produce false-negative results. However, POC INR devices that detect fibrin clot formation by measuring the rate of flow of the blood sample, and a recently marketed POC fibrinogen device, are particularly promising. Although POC devices remain prohibitively expensive for routine use in most low-income settings, they could be used as outcome measurement instruments in clinical trials. An exception is the manual ACT method, which may offer an affordable, rapid, and accurate bedside test for diagnosing VICC. POC devices could also support microsampling in preclinical studies, which would enable the development of clinically representative models associated with less animal suffering and greater richness of data. The devices shortlisted in this review should be prioritized for formal validation studies in preclinical and clinical studies of VICC, to better understand the extent of their utility in the field of snakebite envenoming.


#
#

Conflict of Interest

None declared.

Supplementary Material

  • References

  • 1 Kasturiratne A, Wickremasinghe AR, de Silva N. et al. The global burden of snakebite: a literature analysis and modelling based on regional estimates of envenoming and deaths. PLoS Med 2008; 5 (11) e218
  • 2 The Lancet. Snake-bite envenoming: a priority neglected tropical disease. Lancet 2017; 390 (10089): 2
  • 3 Isbister GK, Currie BJ, Little M, Daly FFS, Isbister JP. Coagulopathy from tiger snake envenoming and its treatment. Pathology 2002; 34 (06) 588-590 , author reply 590
  • 4 Macêdo JKA, Fox JW. Biological activities and assays of the snake venom metalloproteinases (SVMPs). In: Gopalakrishnakone P, Calvete JJ. eds. Venom Genomics and Proteomics. Dordrecht: Springer Netherlands; 2016: 211-238
  • 5 Gopalakrishnan M, Saurabh S, Sagar P, Bammigatti C, Dutta TK. A simple mortality risk prediction score for viper envenoming in India (VENOMS): a model development and validation study. PLoS Negl Trop Dis 2022; 16 (02) e0010183
  • 6 WHO. Guidelines for the management of snakebites, 2nd edition. World Health Organization, Regional Office for South-East Asia: New Delhi. 2016 [cited August 16, 2023]. Accessed September 16, 2024 at: https://www.who.int/publications-detail-redirect/9789290225300
  • 7 WHO Regional Office for Africa. Guidelines for the prevention and clinical management of snakebite in Africa: Mauritius. 2010 [cited August 16, 2023]. Accessed September 16, 2024 at: https://apps.who.int/iris/handle/10665/204458
  • 8 Escalante T, Rucavado A, Fox JW, Gutiérrez JM. Key events in microvascular damage induced by snake venom hemorrhagic metalloproteinases. J Proteomics 2011; 74 (09) 1781-1794
  • 9 Abubakar IS, Abubakar SB, Habib AG. et al; Nigeria-UK EchiTab Study Group. Randomised controlled double-blind non-inferiority trial of two antivenoms for saw-scaled or carpet viper (Echis ocellatus) envenoming in Nigeria. PLoS Negl Trop Dis 2010; 4 (07) e767
  • 10 Naveen A, Sahu MR, Mohanty MK, Padhi KS, Patnaik A. Fatal intracranial bleedings in a viper bite: a case report. Chin J Traumatol 2023; 26 (02) 121-124
  • 11 Berling I, Brown SGA, Miteff F, Levi C, Isbister GK. Intracranial haemorrhages associated with venom induced consumption coagulopathy in Australian snakebites (ASP-21). Toxicon 2015; 102: 8-13
  • 12 Lee RI, White PD. A clinical study of the coagulation time of blood. Am J Med Sci 1913; 145: 495-503
  • 13 Warrell DA, Davidson NMcD, Greenwood BM. et al. Poisoning by bites of the saw-scaled or carpet viper (Echis carinatus) in Nigeria. Q J Med 1977; 46 (181) 33-62
  • 14 Lamb T, Abouyannis M, de Oliveira SS. et al. The 20-minute whole blood clotting test (20WBCT) for snakebite coagulopathy-a systematic review and meta-analysis of diagnostic test accuracy. PLoS Negl Trop Dis 2021; 15 (08) e0009657
  • 15 Isbister GK, Maduwage K, Shahmy S. et al. Diagnostic 20-min whole blood clotting test in Russell's viper envenoming delays antivenom administration. QJM 2013; 106 (10) 925-932
  • 16 Abouyannis M, Esmail H, Hamaluba M. et al; Snakebite Global Core Outcome Set Study Group. A global core outcome measurement set for snakebite clinical trials. Lancet Glob Health 2023; 11 (02) e296-e300
  • 17 Tricco AC, Lillie E, Zarin W. et al. PRISMA Extension for Scoping Reviews (PRISMA-ScR): checklist and explanation. Ann Intern Med 2018; 169 (07) 467-473
  • 18 Slagboom J, Kool J, Harrison RA, Casewell NR. Haemotoxic snake venoms: their functional activity, impact on snakebite victims and pharmaceutical promise. Br J Haematol 2017; 177 (06) 947-959
  • 19 Kini RM, Koh CY. Metalloproteases affecting blood coagulation, fibrinolysis and platelet aggregation from snake venoms: definition and nomenclature of interaction sites. Toxins (Basel) 2016; 8 (10) 1-27
  • 20 Bos MHA, Camire RM. Procoagulant adaptation of a blood coagulation prothrombinase-like enzyme complex in australian elapid venom. Toxins (Basel) 2010; 2 (06) 1554-1567
  • 21 Larréché S, Chippaux J-P, Chevillard L. et al. Bleeding and thrombosis: insights into pathophysiology of Bothrops venom-related hemostasis disorders. Int J Mol Sci 2021; 22 (17) 9643
  • 22 Maduwage K, Isbister GK. Current treatment for venom-induced consumption coagulopathy resulting from snakebite. PLoS Negl Trop Dis 2014; 8 (10) e3220
  • 23 Takeya H, Nishida S, Miyata T. et al. Coagulation factor X activating enzyme from Russell's viper venom (RVV-X). A novel metalloproteinase with disintegrin (platelet aggregation inhibitor)-like and C-type lectin-like domains. J Biol Chem 1992; 267 (20) 14109-14117
  • 24 Khan SU, Al-Saleh SS. Biochemical characterization of a factor X activator protein purified from Walterinnesia aegyptia venom. Blood Coagul Fibrinolysis 2015; 26 (07) 772-777
  • 25 Rosing J, Govers-Riemslag JW, Yukelson L, Tans G. Factor V activation and inactivation by venom proteases. Haemostasis 2001; 31 (3–6): 241-246
  • 26 Manjunatha Kini R, Morita T, Rosing J. Registry of Exogenous Hemostatic Factors of the Scientific and Standardization Committee of the International Society on Thrombosis and Haemostasis. Classification and nomenclature of prothrombin activators isolated from snake venoms. Thromb Haemost 2001; 86 (02) 710-711
  • 27 Kornalik F, Blombäck B. Prothrombin activation induced by Ecarin - a prothrombin converting enzyme from Echis carinatus venom. Thromb Res 1975; 6 (01) 57-63
  • 28 Isolation and Characterization of Carinactivase, a Novel Prothrombin Activator in Echis carinatus Venom with a Unique Catalytic Mechanism (*) - Journal of Biological Chemistry. [cited May 15, 2023]. Accessed September 16, 2024 at: https://www.jbc.org/article/S0021-9258(18)82570-4/fulltext
  • 29 Rosing J, Tans G. Structural and functional properties of snake venom prothrombin activators. Toxicon 1992; 30 (12) 1515-1527
  • 30 Rao VS, Kini RM. Pseutarin C, a prothrombin activator from Pseudonaja textilis venom: its structural and functional similarity to mammalian coagulation factor Xa-Va complex. Thromb Haemost 2002; 88 (04) 611-619
  • 31 Speijer H, Govers-Riemslag JW, Zwaal RF, Rosing J. Prothrombin activation by an activator from the venom of Oxyuranus scutellatus (Taipan snake). J Biol Chem 1986; 261 (28) 13258-13267
  • 32 Joseph JS, Chung MCM, Jeyaseelan K, Kini RM. Amino acid sequence of trocarin, a prothrombin activator from Tropidechis carinatus venom: its structural similarity to coagulation factor Xa. Blood 1999; 94 (02) 621-631
  • 33 Kini RM. The intriguing world of prothrombin activators from snake venom. Toxicon 2005; 45 (08) 1133-1145
  • 34 Isbister GK. Procoagulant snake toxins: laboratory studies, diagnosis, and understanding snakebite coagulopathy. Semin Thromb Hemost 2009; 35 (01) 93-103
  • 35 Swenson S, Markland Jr FS. Snake venom fibrin(ogen)olytic enzymes. Toxicon 2005; 45 (08) 1021-1039
  • 36 Senthilkumaran S, David SS, Jena NN, Thirumalaikolundusubramanian P. Limitations and consumer aspects of point-of-care in snake envenomation. Emerg Med Australas 2014; 26 (02) 208
  • 37 O'Rourke KM, Correlje E, Martin CL, Robertson JD, Isbister GK. Point-of-care derived INR does not reliably detect significant coagulopathy following Australian snakebite. Thromb Res 2013; 132 (05) 610-613
  • 38 Celenza A, Skinner K. Comparison of emergency department point-of-care international normalised ratio (INR) testing with laboratory-based testing. Emerg Med J 2011; 28 (02) 136-140
  • 39 Cubitt M, Armstrong J, McCoubrie D, White J, Williams V, Isbister GK. Point-of-care testing in snakebite: an envenomed case with false negative coagulation studies. Emerg Med Australas 2013; 25 (04) 372-373
  • 40 Ullah A, Masood R, Ali I. et al. Thrombin-like enzymes from snake venom: Structural characterization and mechanism of action. Int J Biol Macromol 2018; 114: 788-811
  • 41 Quick AJ. On various properties of thromboplastin (aqueous tissue extracts). Am J Physiol Legacy Content 1935; 114: 282-296
  • 42 Yang R, Moosavi L. Prothrombin Time. StatPearls. Treasure Island, FL: StatPearls Publishing; 2023
  • 43 World Health Organization. WHO Expert Committee on Biological Standardization, sixty-seventh report. Geneva: World Health Organization; 2017 (WHO technical report series ; no. 1004)
  • 44 Mion G, Larréché S, Benois A, Petitjeans F, Puidupin M. Hemostasis dynamics during coagulopathy resulting from Echis envenomation. Toxicon 2013; 76: 103-109
  • 45 Isbister GK, Scorgie FE, O'Leary MA, Seldon M, Brown SG, Lincz LF. ASP Investigators. Factor deficiencies in venom-induced consumption coagulopathy resulting from Australian elapid envenomation: Australian Snakebite Project (ASP-10). J Thromb Haemost 2010; 8 (11) 2504-2513
  • 46 Patel MR, Hellkamp AS, Fox KAA. ROCKET AF Executive Committee and Investigators. Point-of-care warfarin monitoring in the ROCKET AF trial. N Engl J Med 2016; 374 (08) 785-788
  • 47 CLSI. One-Stage Prothrombin Time (PT) Test and Activated Partial Thromboplastin Time (APTT) Test; Approved Guideline—Second Edition. CLSI document H47–A2. Wayne, PA: Clinical and Laboratory Standards Institute; 2008
  • 48 Smith SA, Travers RJ, Morrissey JH. How it all starts: initiation of the clotting cascade. Crit Rev Biochem Mol Biol 2015; 50 (04) 326-336
  • 49 Rountree KM, Yaker Z, Lopez PP. Partial Thromboplastin Time. StatPearls. Treasure Island, FL: StatPearls Publishing; 2023
  • 50 Isbister GK, Williams V, Brown SGA, White J, Currie BJ. Australian Snakebite Project Investigators. Clinically applicable laboratory end-points for treating snakebite coagulopathy. Pathology 2006; 38 (06) 568-572
  • 51 Hattersley PG. Activated coagulation time of whole blood. JAMA 1966; 196 (05) 436-440
  • 52 Puis L, Milojevic M, Boer C. et al; EACTS/EACTA/EBCP Committee Reviewers. 2019 EACTS/EACTA/EBCP guidelines on cardiopulmonary bypass in adult cardiac surgery. Interact Cardiovasc Thorac Surg 2020; 30 (02) 161-202
  • 53 Hill JD, Dontigny L, De Leval M, Mielke Jr CH. A simple method of heparin management during prolonged extracorporeal circulation. Ann Thorac Surg 1974; 17 (02) 129-134
  • 54 Shenoy RK. Clinical and laboratory spectrum of Venom-induced consumption coagulopathy (VICC) and bedside diagnostic tests in haemotoxic snakebite in a tertiary care centre in South India: ProTIV Study. masters, Christian Medical College, Vellore. 2019 ; available at: https://www.academia.edu/122221075/Clinical_and_laboratory_spectrum_of_Venom_induced_consumption_coagulopathy_VICC_and_bedside_diagnostic_tests_in_haemotoxic_snakebite_in_a_tertiary_care_centre_in_South_India_ProTIV_Study
  • 55 Stang LJ, Mitchell LG. Fibrinogen. Methods Mol Biol 2013; 992: 181-192
  • 56 Kaur J, Jain A. Fibrinogen. StatPearls. Treasure Island, FL: StatPearls Publishing; 2023
  • 57 Silva de Oliveira S, Campos Alves E, Dos Santos Santos A. et al. Bothrops snakebites in the Amazon: recovery from hemostatic disorders after Brazilian antivenom therapy. Clin Toxicol (Phila) 2020; 58 (04) 266-274
  • 58 Sanfilippo S, Buisson L, Rouabehi H. et al. The qLabs® FIB system, a novel point-of-care technology for a rapid and accurate quantification of functional fibrinogen concentration from a single drop of citrated whole blood. Thromb Res 2023; 226: 159-164
  • 59 Whitaker AN, Elms MJ, Masci PP. et al. Measurement of cross linked fibrin derivatives in plasma: an immunoassay using monoclonal antibodies. J Clin Pathol 1984; 37 (08) 882-887
  • 60 Francis CW, Marder VJ, Barlow GH. Plasmic degradation of crosslinked fibrin. Characterization of new macromolecular soluble complexes and a model of their structure. J Clin Invest 1980; 66 (05) 1033-1043
  • 61 Konstantinides SV, Meyer G, Becattini C. et al; The Task Force for the diagnosis and management of acute pulmonary embolism of the European Society of Cardiology (ESC). 2019 ESC Guidelines for the diagnosis and management of acute pulmonary embolism developed in collaboration with the European Respiratory Society (ERS): The Task Force for the diagnosis and management of acute pulmonary embolism of the European Society of Cardiology (ESC). Eur Respir J 2019; 54 (03) 1901647
  • 62 Isbister GK, Noutsos T, Jenkins S, Isoardi KZ, Soderstrom J, Buckley NA. D-dimer testing for early detection of venom-induced consumption coagulopathy after snakebite in Australia (ASP-29). Med J Aust 2022; 217 (04) 203-207
  • 63 Longbottom J, Shearer FM, Devine M. et al. Vulnerability to snakebite envenoming: a global mapping of hotspots. Lancet 2018; 392 (10148): 673-684
  • 64 Dart RC, Hurlbut KM, Garcia R, Boren J. Validation of a severity score for the assessment of crotalid snakebite. Ann Emerg Med 1996; 27 (03) 321-326
  • 65 Wedasingha S, Silva A, Siribaddana S, Seneviratne K, Isbister GK. Comparison of bedside clotting tests for detecting venom-induced consumption coagulopathy following Sri Lankan viper envenoming. Clin Toxicol (Phila) 2022; 60 (12) 1328-1335
  • 66 Abouyannis M, Aggarwal D, Lalloo DG, Casewell NR, Hamaluba M, Esmail H. Clinical outcomes and outcome measurement tools reported in randomised controlled trials of treatment for snakebite envenoming: a systematic review. PLoS Negl Trop Dis 2021; 15 (08) e0009589
  • 67 Global funding for snakebite envenoming research. [cited April 6, 2020]. Accessed September 16, 2024 at: https://wellcome.ac.uk/sites/default/files/global-funding-for-snakebite-envenoming-research-2007-2018.pdf
  • 68 Isbister GK, Buckley NA, Page CB. et al; ASP Investigators. A randomized controlled trial of fresh frozen plasma for treating venom-induced consumption coagulopathy in cases of Australian snakebite (ASP-18). J Thromb Haemost 2013; 11 (07) 1310-1318
  • 69 Drain PK, Hyle EP, Noubary F. et al. Diagnostic point-of-care tests in resource-limited settings. Lancet Infect Dis 2014; 14 (03) 239-249
  • 70 Ainsworth S, Menzies SK, Casewell NR, Harrison RA. An analysis of preclinical efficacy testing of antivenoms for sub-Saharan Africa: inadequate independent scrutiny and poor-quality reporting are barriers to improving snakebite treatment and management. PLoS Negl Trop Dis 2020; 14 (08) e0008579
  • 71 Potet J, Smith J, McIver L. Reviewing evidence of the clinical effectiveness of commercially available antivenoms in sub-Saharan Africa identifies the need for a multi-centre, multi-antivenom clinical trial. PLoS Negl Trop Dis 2019; 13 (06) e0007551-e0007551
  • 72 World Health Organization, WHO Expert Committee on Biological Standardization (2016: Geneva Switzerland). WHO Expert Committee on Biological Standardization, sixty-seventh report. Geneva: World Health Organization; 2017. . Accessed September 16, 2024 at: https://apps.who.int/iris/handle/10665/255657
  • 73 Marriott AE, Casewell NR, Lilley E, Gutiérrez J-M, Ainsworth S. Improving in vivo assays in snake venom and antivenom research: a community discussion. F1000 Res 2024; 13: 192
  • 74 Knudsen C, Casewell NR, Lomonte B, Gutiérrez JM, Vaiyapuri S, Laustsen AH. Novel snakebite therapeutics must be tested in appropriate rescue models to robustly assess their preclinical efficacy. Toxins (Basel) 2020; 12 (09) 528
  • 75 Target product profiles for animal plasma-derived antivenoms: antivenoms for treatment of snakebite envenoming in sub-Saharan Africa. Geneva: World Health Organization; 2023. [cited August 15, 2023]. Accessed September 16, 2024 at: https://www.who.int/publications-detail-redirect/9789240074569
  • 76 Camacho E, Ramírez-Vargas G, Vargas K. et al. Neutralization, by a polyspecific antivenom, of the coagulopathy induced by the venom of Bothrops asper: assessment by standard coagulation tests and rotational thromboelastometry in a murine model. Toxicon 2023; 234: 107301
  • 77 Emeis JJ, Jirouskova M, Muchitsch E-M, Shet AS, Smyth SS, Johnson GJ. A guide to murine coagulation factor structure, function, assays, and genetic alterations. J Thromb Haemost 2007; 5 (04) 670-679
  • 78 De Pablo-Moreno JA, Liras A, Revuelta L. Standardization of coagulation factor V reference intervals, prothrombin time, and activated partial thromboplastin time in mice for use in factor V deficiency pathological models. Front Vet Sci 2022; 9: 846216
  • 79 Foerch C, Arai K, Jin G. et al. Experimental model of warfarin-associated intracerebral hemorrhage. Stroke 2008; 39 (12) 3397-3404
  • 80 Coleman D, Smith G, Lawrence R, McManus D, Diaram S, Edwards J. Capillary microsampling in nonclinical safety assessment: practical sampling and bioanalysis from a CRO perspective. Bioanalysis 2017; 9 (10) 787-798
  • 81 Chapman K, Chivers S, Gliddon D. et al. Overcoming the barriers to the uptake of nonclinical microsampling in regulatory safety studies. Drug Discov Today 2014; 19 (05) 528-532
  • 82 Mohapatra B, Warrell DA, Suraweera W. et al; Million Death Study Collaborators. Snakebite mortality in India: a nationally representative mortality survey. PLoS Negl Trop Dis 2011; 5 (04) e1018
  • 83 Brown NI. Consequences of neglect: analysis of the sub-Saharan African snake antivenom market and the global context. PLoS Negl Trop Dis 2012; 6 (06) e1670
  • 84 Whitaker R, Whitaker S. Venom, antivenom production and the medically important snakes of India. Curr Sci 2012; 103: 635-643
  • 85 Fleming KA, Horton S, Wilson ML. et al. The Lancet Commission on diagnostics: transforming access to diagnostics. Lancet 2021; 398 (10315): 1997-2050

Address for correspondence

Michael Abouyannis, PhD
Centre for Snakebite Research and Interventions, Liverpool School of Tropical Medicine
Liverpool L3 5QA
United Kingdom   

Publikationsverlauf

Eingereicht: 15. Mai 2024

Angenommen: 17. August 2024

Accepted Manuscript online:
30. August 2024

Artikel online veröffentlicht:
19. November 2024

© 2024. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany

  • References

  • 1 Kasturiratne A, Wickremasinghe AR, de Silva N. et al. The global burden of snakebite: a literature analysis and modelling based on regional estimates of envenoming and deaths. PLoS Med 2008; 5 (11) e218
  • 2 The Lancet. Snake-bite envenoming: a priority neglected tropical disease. Lancet 2017; 390 (10089): 2
  • 3 Isbister GK, Currie BJ, Little M, Daly FFS, Isbister JP. Coagulopathy from tiger snake envenoming and its treatment. Pathology 2002; 34 (06) 588-590 , author reply 590
  • 4 Macêdo JKA, Fox JW. Biological activities and assays of the snake venom metalloproteinases (SVMPs). In: Gopalakrishnakone P, Calvete JJ. eds. Venom Genomics and Proteomics. Dordrecht: Springer Netherlands; 2016: 211-238
  • 5 Gopalakrishnan M, Saurabh S, Sagar P, Bammigatti C, Dutta TK. A simple mortality risk prediction score for viper envenoming in India (VENOMS): a model development and validation study. PLoS Negl Trop Dis 2022; 16 (02) e0010183
  • 6 WHO. Guidelines for the management of snakebites, 2nd edition. World Health Organization, Regional Office for South-East Asia: New Delhi. 2016 [cited August 16, 2023]. Accessed September 16, 2024 at: https://www.who.int/publications-detail-redirect/9789290225300
  • 7 WHO Regional Office for Africa. Guidelines for the prevention and clinical management of snakebite in Africa: Mauritius. 2010 [cited August 16, 2023]. Accessed September 16, 2024 at: https://apps.who.int/iris/handle/10665/204458
  • 8 Escalante T, Rucavado A, Fox JW, Gutiérrez JM. Key events in microvascular damage induced by snake venom hemorrhagic metalloproteinases. J Proteomics 2011; 74 (09) 1781-1794
  • 9 Abubakar IS, Abubakar SB, Habib AG. et al; Nigeria-UK EchiTab Study Group. Randomised controlled double-blind non-inferiority trial of two antivenoms for saw-scaled or carpet viper (Echis ocellatus) envenoming in Nigeria. PLoS Negl Trop Dis 2010; 4 (07) e767
  • 10 Naveen A, Sahu MR, Mohanty MK, Padhi KS, Patnaik A. Fatal intracranial bleedings in a viper bite: a case report. Chin J Traumatol 2023; 26 (02) 121-124
  • 11 Berling I, Brown SGA, Miteff F, Levi C, Isbister GK. Intracranial haemorrhages associated with venom induced consumption coagulopathy in Australian snakebites (ASP-21). Toxicon 2015; 102: 8-13
  • 12 Lee RI, White PD. A clinical study of the coagulation time of blood. Am J Med Sci 1913; 145: 495-503
  • 13 Warrell DA, Davidson NMcD, Greenwood BM. et al. Poisoning by bites of the saw-scaled or carpet viper (Echis carinatus) in Nigeria. Q J Med 1977; 46 (181) 33-62
  • 14 Lamb T, Abouyannis M, de Oliveira SS. et al. The 20-minute whole blood clotting test (20WBCT) for snakebite coagulopathy-a systematic review and meta-analysis of diagnostic test accuracy. PLoS Negl Trop Dis 2021; 15 (08) e0009657
  • 15 Isbister GK, Maduwage K, Shahmy S. et al. Diagnostic 20-min whole blood clotting test in Russell's viper envenoming delays antivenom administration. QJM 2013; 106 (10) 925-932
  • 16 Abouyannis M, Esmail H, Hamaluba M. et al; Snakebite Global Core Outcome Set Study Group. A global core outcome measurement set for snakebite clinical trials. Lancet Glob Health 2023; 11 (02) e296-e300
  • 17 Tricco AC, Lillie E, Zarin W. et al. PRISMA Extension for Scoping Reviews (PRISMA-ScR): checklist and explanation. Ann Intern Med 2018; 169 (07) 467-473
  • 18 Slagboom J, Kool J, Harrison RA, Casewell NR. Haemotoxic snake venoms: their functional activity, impact on snakebite victims and pharmaceutical promise. Br J Haematol 2017; 177 (06) 947-959
  • 19 Kini RM, Koh CY. Metalloproteases affecting blood coagulation, fibrinolysis and platelet aggregation from snake venoms: definition and nomenclature of interaction sites. Toxins (Basel) 2016; 8 (10) 1-27
  • 20 Bos MHA, Camire RM. Procoagulant adaptation of a blood coagulation prothrombinase-like enzyme complex in australian elapid venom. Toxins (Basel) 2010; 2 (06) 1554-1567
  • 21 Larréché S, Chippaux J-P, Chevillard L. et al. Bleeding and thrombosis: insights into pathophysiology of Bothrops venom-related hemostasis disorders. Int J Mol Sci 2021; 22 (17) 9643
  • 22 Maduwage K, Isbister GK. Current treatment for venom-induced consumption coagulopathy resulting from snakebite. PLoS Negl Trop Dis 2014; 8 (10) e3220
  • 23 Takeya H, Nishida S, Miyata T. et al. Coagulation factor X activating enzyme from Russell's viper venom (RVV-X). A novel metalloproteinase with disintegrin (platelet aggregation inhibitor)-like and C-type lectin-like domains. J Biol Chem 1992; 267 (20) 14109-14117
  • 24 Khan SU, Al-Saleh SS. Biochemical characterization of a factor X activator protein purified from Walterinnesia aegyptia venom. Blood Coagul Fibrinolysis 2015; 26 (07) 772-777
  • 25 Rosing J, Govers-Riemslag JW, Yukelson L, Tans G. Factor V activation and inactivation by venom proteases. Haemostasis 2001; 31 (3–6): 241-246
  • 26 Manjunatha Kini R, Morita T, Rosing J. Registry of Exogenous Hemostatic Factors of the Scientific and Standardization Committee of the International Society on Thrombosis and Haemostasis. Classification and nomenclature of prothrombin activators isolated from snake venoms. Thromb Haemost 2001; 86 (02) 710-711
  • 27 Kornalik F, Blombäck B. Prothrombin activation induced by Ecarin - a prothrombin converting enzyme from Echis carinatus venom. Thromb Res 1975; 6 (01) 57-63
  • 28 Isolation and Characterization of Carinactivase, a Novel Prothrombin Activator in Echis carinatus Venom with a Unique Catalytic Mechanism (*) - Journal of Biological Chemistry. [cited May 15, 2023]. Accessed September 16, 2024 at: https://www.jbc.org/article/S0021-9258(18)82570-4/fulltext
  • 29 Rosing J, Tans G. Structural and functional properties of snake venom prothrombin activators. Toxicon 1992; 30 (12) 1515-1527
  • 30 Rao VS, Kini RM. Pseutarin C, a prothrombin activator from Pseudonaja textilis venom: its structural and functional similarity to mammalian coagulation factor Xa-Va complex. Thromb Haemost 2002; 88 (04) 611-619
  • 31 Speijer H, Govers-Riemslag JW, Zwaal RF, Rosing J. Prothrombin activation by an activator from the venom of Oxyuranus scutellatus (Taipan snake). J Biol Chem 1986; 261 (28) 13258-13267
  • 32 Joseph JS, Chung MCM, Jeyaseelan K, Kini RM. Amino acid sequence of trocarin, a prothrombin activator from Tropidechis carinatus venom: its structural similarity to coagulation factor Xa. Blood 1999; 94 (02) 621-631
  • 33 Kini RM. The intriguing world of prothrombin activators from snake venom. Toxicon 2005; 45 (08) 1133-1145
  • 34 Isbister GK. Procoagulant snake toxins: laboratory studies, diagnosis, and understanding snakebite coagulopathy. Semin Thromb Hemost 2009; 35 (01) 93-103
  • 35 Swenson S, Markland Jr FS. Snake venom fibrin(ogen)olytic enzymes. Toxicon 2005; 45 (08) 1021-1039
  • 36 Senthilkumaran S, David SS, Jena NN, Thirumalaikolundusubramanian P. Limitations and consumer aspects of point-of-care in snake envenomation. Emerg Med Australas 2014; 26 (02) 208
  • 37 O'Rourke KM, Correlje E, Martin CL, Robertson JD, Isbister GK. Point-of-care derived INR does not reliably detect significant coagulopathy following Australian snakebite. Thromb Res 2013; 132 (05) 610-613
  • 38 Celenza A, Skinner K. Comparison of emergency department point-of-care international normalised ratio (INR) testing with laboratory-based testing. Emerg Med J 2011; 28 (02) 136-140
  • 39 Cubitt M, Armstrong J, McCoubrie D, White J, Williams V, Isbister GK. Point-of-care testing in snakebite: an envenomed case with false negative coagulation studies. Emerg Med Australas 2013; 25 (04) 372-373
  • 40 Ullah A, Masood R, Ali I. et al. Thrombin-like enzymes from snake venom: Structural characterization and mechanism of action. Int J Biol Macromol 2018; 114: 788-811
  • 41 Quick AJ. On various properties of thromboplastin (aqueous tissue extracts). Am J Physiol Legacy Content 1935; 114: 282-296
  • 42 Yang R, Moosavi L. Prothrombin Time. StatPearls. Treasure Island, FL: StatPearls Publishing; 2023
  • 43 World Health Organization. WHO Expert Committee on Biological Standardization, sixty-seventh report. Geneva: World Health Organization; 2017 (WHO technical report series ; no. 1004)
  • 44 Mion G, Larréché S, Benois A, Petitjeans F, Puidupin M. Hemostasis dynamics during coagulopathy resulting from Echis envenomation. Toxicon 2013; 76: 103-109
  • 45 Isbister GK, Scorgie FE, O'Leary MA, Seldon M, Brown SG, Lincz LF. ASP Investigators. Factor deficiencies in venom-induced consumption coagulopathy resulting from Australian elapid envenomation: Australian Snakebite Project (ASP-10). J Thromb Haemost 2010; 8 (11) 2504-2513
  • 46 Patel MR, Hellkamp AS, Fox KAA. ROCKET AF Executive Committee and Investigators. Point-of-care warfarin monitoring in the ROCKET AF trial. N Engl J Med 2016; 374 (08) 785-788
  • 47 CLSI. One-Stage Prothrombin Time (PT) Test and Activated Partial Thromboplastin Time (APTT) Test; Approved Guideline—Second Edition. CLSI document H47–A2. Wayne, PA: Clinical and Laboratory Standards Institute; 2008
  • 48 Smith SA, Travers RJ, Morrissey JH. How it all starts: initiation of the clotting cascade. Crit Rev Biochem Mol Biol 2015; 50 (04) 326-336
  • 49 Rountree KM, Yaker Z, Lopez PP. Partial Thromboplastin Time. StatPearls. Treasure Island, FL: StatPearls Publishing; 2023
  • 50 Isbister GK, Williams V, Brown SGA, White J, Currie BJ. Australian Snakebite Project Investigators. Clinically applicable laboratory end-points for treating snakebite coagulopathy. Pathology 2006; 38 (06) 568-572
  • 51 Hattersley PG. Activated coagulation time of whole blood. JAMA 1966; 196 (05) 436-440
  • 52 Puis L, Milojevic M, Boer C. et al; EACTS/EACTA/EBCP Committee Reviewers. 2019 EACTS/EACTA/EBCP guidelines on cardiopulmonary bypass in adult cardiac surgery. Interact Cardiovasc Thorac Surg 2020; 30 (02) 161-202
  • 53 Hill JD, Dontigny L, De Leval M, Mielke Jr CH. A simple method of heparin management during prolonged extracorporeal circulation. Ann Thorac Surg 1974; 17 (02) 129-134
  • 54 Shenoy RK. Clinical and laboratory spectrum of Venom-induced consumption coagulopathy (VICC) and bedside diagnostic tests in haemotoxic snakebite in a tertiary care centre in South India: ProTIV Study. masters, Christian Medical College, Vellore. 2019 ; available at: https://www.academia.edu/122221075/Clinical_and_laboratory_spectrum_of_Venom_induced_consumption_coagulopathy_VICC_and_bedside_diagnostic_tests_in_haemotoxic_snakebite_in_a_tertiary_care_centre_in_South_India_ProTIV_Study
  • 55 Stang LJ, Mitchell LG. Fibrinogen. Methods Mol Biol 2013; 992: 181-192
  • 56 Kaur J, Jain A. Fibrinogen. StatPearls. Treasure Island, FL: StatPearls Publishing; 2023
  • 57 Silva de Oliveira S, Campos Alves E, Dos Santos Santos A. et al. Bothrops snakebites in the Amazon: recovery from hemostatic disorders after Brazilian antivenom therapy. Clin Toxicol (Phila) 2020; 58 (04) 266-274
  • 58 Sanfilippo S, Buisson L, Rouabehi H. et al. The qLabs® FIB system, a novel point-of-care technology for a rapid and accurate quantification of functional fibrinogen concentration from a single drop of citrated whole blood. Thromb Res 2023; 226: 159-164
  • 59 Whitaker AN, Elms MJ, Masci PP. et al. Measurement of cross linked fibrin derivatives in plasma: an immunoassay using monoclonal antibodies. J Clin Pathol 1984; 37 (08) 882-887
  • 60 Francis CW, Marder VJ, Barlow GH. Plasmic degradation of crosslinked fibrin. Characterization of new macromolecular soluble complexes and a model of their structure. J Clin Invest 1980; 66 (05) 1033-1043
  • 61 Konstantinides SV, Meyer G, Becattini C. et al; The Task Force for the diagnosis and management of acute pulmonary embolism of the European Society of Cardiology (ESC). 2019 ESC Guidelines for the diagnosis and management of acute pulmonary embolism developed in collaboration with the European Respiratory Society (ERS): The Task Force for the diagnosis and management of acute pulmonary embolism of the European Society of Cardiology (ESC). Eur Respir J 2019; 54 (03) 1901647
  • 62 Isbister GK, Noutsos T, Jenkins S, Isoardi KZ, Soderstrom J, Buckley NA. D-dimer testing for early detection of venom-induced consumption coagulopathy after snakebite in Australia (ASP-29). Med J Aust 2022; 217 (04) 203-207
  • 63 Longbottom J, Shearer FM, Devine M. et al. Vulnerability to snakebite envenoming: a global mapping of hotspots. Lancet 2018; 392 (10148): 673-684
  • 64 Dart RC, Hurlbut KM, Garcia R, Boren J. Validation of a severity score for the assessment of crotalid snakebite. Ann Emerg Med 1996; 27 (03) 321-326
  • 65 Wedasingha S, Silva A, Siribaddana S, Seneviratne K, Isbister GK. Comparison of bedside clotting tests for detecting venom-induced consumption coagulopathy following Sri Lankan viper envenoming. Clin Toxicol (Phila) 2022; 60 (12) 1328-1335
  • 66 Abouyannis M, Aggarwal D, Lalloo DG, Casewell NR, Hamaluba M, Esmail H. Clinical outcomes and outcome measurement tools reported in randomised controlled trials of treatment for snakebite envenoming: a systematic review. PLoS Negl Trop Dis 2021; 15 (08) e0009589
  • 67 Global funding for snakebite envenoming research. [cited April 6, 2020]. Accessed September 16, 2024 at: https://wellcome.ac.uk/sites/default/files/global-funding-for-snakebite-envenoming-research-2007-2018.pdf
  • 68 Isbister GK, Buckley NA, Page CB. et al; ASP Investigators. A randomized controlled trial of fresh frozen plasma for treating venom-induced consumption coagulopathy in cases of Australian snakebite (ASP-18). J Thromb Haemost 2013; 11 (07) 1310-1318
  • 69 Drain PK, Hyle EP, Noubary F. et al. Diagnostic point-of-care tests in resource-limited settings. Lancet Infect Dis 2014; 14 (03) 239-249
  • 70 Ainsworth S, Menzies SK, Casewell NR, Harrison RA. An analysis of preclinical efficacy testing of antivenoms for sub-Saharan Africa: inadequate independent scrutiny and poor-quality reporting are barriers to improving snakebite treatment and management. PLoS Negl Trop Dis 2020; 14 (08) e0008579
  • 71 Potet J, Smith J, McIver L. Reviewing evidence of the clinical effectiveness of commercially available antivenoms in sub-Saharan Africa identifies the need for a multi-centre, multi-antivenom clinical trial. PLoS Negl Trop Dis 2019; 13 (06) e0007551-e0007551
  • 72 World Health Organization, WHO Expert Committee on Biological Standardization (2016: Geneva Switzerland). WHO Expert Committee on Biological Standardization, sixty-seventh report. Geneva: World Health Organization; 2017. . Accessed September 16, 2024 at: https://apps.who.int/iris/handle/10665/255657
  • 73 Marriott AE, Casewell NR, Lilley E, Gutiérrez J-M, Ainsworth S. Improving in vivo assays in snake venom and antivenom research: a community discussion. F1000 Res 2024; 13: 192
  • 74 Knudsen C, Casewell NR, Lomonte B, Gutiérrez JM, Vaiyapuri S, Laustsen AH. Novel snakebite therapeutics must be tested in appropriate rescue models to robustly assess their preclinical efficacy. Toxins (Basel) 2020; 12 (09) 528
  • 75 Target product profiles for animal plasma-derived antivenoms: antivenoms for treatment of snakebite envenoming in sub-Saharan Africa. Geneva: World Health Organization; 2023. [cited August 15, 2023]. Accessed September 16, 2024 at: https://www.who.int/publications-detail-redirect/9789240074569
  • 76 Camacho E, Ramírez-Vargas G, Vargas K. et al. Neutralization, by a polyspecific antivenom, of the coagulopathy induced by the venom of Bothrops asper: assessment by standard coagulation tests and rotational thromboelastometry in a murine model. Toxicon 2023; 234: 107301
  • 77 Emeis JJ, Jirouskova M, Muchitsch E-M, Shet AS, Smyth SS, Johnson GJ. A guide to murine coagulation factor structure, function, assays, and genetic alterations. J Thromb Haemost 2007; 5 (04) 670-679
  • 78 De Pablo-Moreno JA, Liras A, Revuelta L. Standardization of coagulation factor V reference intervals, prothrombin time, and activated partial thromboplastin time in mice for use in factor V deficiency pathological models. Front Vet Sci 2022; 9: 846216
  • 79 Foerch C, Arai K, Jin G. et al. Experimental model of warfarin-associated intracerebral hemorrhage. Stroke 2008; 39 (12) 3397-3404
  • 80 Coleman D, Smith G, Lawrence R, McManus D, Diaram S, Edwards J. Capillary microsampling in nonclinical safety assessment: practical sampling and bioanalysis from a CRO perspective. Bioanalysis 2017; 9 (10) 787-798
  • 81 Chapman K, Chivers S, Gliddon D. et al. Overcoming the barriers to the uptake of nonclinical microsampling in regulatory safety studies. Drug Discov Today 2014; 19 (05) 528-532
  • 82 Mohapatra B, Warrell DA, Suraweera W. et al; Million Death Study Collaborators. Snakebite mortality in India: a nationally representative mortality survey. PLoS Negl Trop Dis 2011; 5 (04) e1018
  • 83 Brown NI. Consequences of neglect: analysis of the sub-Saharan African snake antivenom market and the global context. PLoS Negl Trop Dis 2012; 6 (06) e1670
  • 84 Whitaker R, Whitaker S. Venom, antivenom production and the medically important snakes of India. Curr Sci 2012; 103: 635-643
  • 85 Fleming KA, Horton S, Wilson ML. et al. The Lancet Commission on diagnostics: transforming access to diagnostics. Lancet 2021; 398 (10315): 1997-2050

Zoom Image
Fig. 1 (A) Overview of the site of action of procoagulant snake venom toxins on the coagulation pathway. The red boxes detail the major groups of coagulopathic venom toxins, their enzyme classification, and a nonexhaustive list of medically important snake species that utilize the toxin group. The dashed arrows indicate the site of the coagulation pathway where the major groups of coagulopathic venom toxins act. Adapted from “Coagulation Cascade,” by BioRender.com (2023). Retrieved from: https://app.biorender.com/biorender-templates. (B) Levels of measurement of each point-of-care device on the coagulation pathway. The green boxes detail the names and assay types of the point-of-care devices that have been included in this review. The tails of the green arrows indicate where each assay's activator acts on the coagulation pathway. The arrow heads highlight where the endpoint read of each assay exists on the coagulation pathway. Adapted from “Coagulation Cascade,” by BioRender.com (2023). Retrieved from https://app.biorender.com/biorender-templates. Ca2+, calcium; FDPs, fibrin degradation products; PL, phospholipid; TF, tissue factor.
Zoom Image
Fig. 2 Comparison of laboratory and POC device INR by participant from previous studies (O'Rourke et al[37] and Senthilkumaran et al[37]). (A) The laboratory, i-STAT, CoaguChek S, and Alere INRatio2 INR values for 15 participants with Daboia russelii envenoming: the participants are ordered according to increasing laboratory INR values. (B) The laboratory, i-STAT, CoaguChek S, and Alere INRatio2 INR values for three participants with Echis carinatus envenoming. (C) The laboratory and i-STAT INR values for 15 participants with Australian Elapid envenoming. INR, international normalized ratio.