A 69-year-old man with unresectable cancer of the pancreatic head presented to our
hospital with fever and jaundice. A laser-cut fully covered self-expandable metal
stent (LC-SEMS; X-Suit NIR covered biliary metal stent; Olympus Medical Systems, Tokyo,
Japan) and a duodenal metal stent (Niti-S pyloric/duodenal uncovered stent; Taewoong
Medical, Seoul, South Korea) had been placed 12 months previously for malignant biliary
and duodenal strictures, respectively.
A computed tomography scan revealed LC-SEMS occlusion ([Fig. 1]). We therefore attempted to remove the stent using a balloon catheter (Extractor
ProRX; Boston Scientific, Marlborough, Massachusetts, USA) and a snare (SD-5U-1; Olympus
Medical Systems); however, the stent could not be removed and broke, resulting in
its dislocation to the duodenal side ([Fig. 2]). We therefore abandoned stent removal and attempted to trim the LC-SEMS using a
circumferential argon plasma coagulation (APC) probe (FiAPC probe; 80 W with a flow
rate of 1.2 L/min; ERBE Elektromedizin, Tübingen, Germany) ([Fig. 3]). This APC probe has an insulator ball at the tip, similar to the insulated-tip
knife used for endoscopic submucosal dissection [1]. We hooked the insulator ball onto the mesh of the LC-SEMS, and cut the stent by
pulling the probe. The LC-SEMS was very easily trimmed, and the procedure was completed
within 10 minutes without any complications ([Video 1]). Later, a new LC-SEMS was placed over the previous one.
Fig. 1 Computed tomography image showing the occluded laser-cut fully covered self-expandable
metal stent (yellow arrowheads).
Fig. 2 Duodenoscopy image showing the extension of the stent into the duodenum after its
attempted removal.
Fig. 3 Photographs of the FiAPC probe (ERBE Elektromedizin, Tübingen, Germany).
Video 1 A laser-cut fully covered self-expandable metal stent is trimmed using a circumferential
argon plasma coagulation probe.
The application of APC to trim various SEMSs has been previously reportedly [2]
[3]
[4]. In the present case, the insulating ball allowed us to avoid damage to the duodenal
mucosa and safely cut the LC-SEMS by pulling the probe. Moreover, the shortening of
the procedure time can be expected to prevent mucosal damage of the bile duct from
heat conduction during the APC [5]. The circumferential APC probe enables safe and easy trimming of biliary metal stents.
Endoscopy_UCTN_Code_CPL_1AK_2AD
Endoscopy E-Videos is an open access online section, reporting on interesting cases and new techniques
in gastroenterological endoscopy. All papers include a high quality video and all
contributions are freely accessible online. Processing charges apply (currently EUR
375), discounts and wavers acc. to HINARI are available.
This section has its own submission
website at
https://mc.manuscriptcentral.com/e-videos