Introduction
COVID-19, a respiratory disease that spreads via direct contact or through droplet
transmission, has caused a major disruption in the healthcare system since it was
first identified in 2019 [1 ]. Endoscopic procedures during the pandemic were divided into urgent endoscopy, semi-urgent
endoscopy, and elective endoscopy, with the aim of preventing infection among healthcare
professionals and providing necessary medical services [2 ]. For patients who required urgent gastroscopy, the face-to-face contact with endoscopists,
aggressive suctioning, and coughing [3 ] during examination brought opportunities for transmission. Considering the reduced
endoscopy volume in most endoscopy departments, diagnosis of patients who needed semi-urgent
endoscopy and elective endoscopy may have been delayed [4 ]. Therefore, focusing on how to return to gradually return to provision of routine
endoscopy services while maintaining safety for endoscopists deserves attention [5 ].
Capsule endoscopy provides a great view of the gastrointestinal tract through uyse
of a miniature, pill-sized camera and it is as accurate as other gastrointestinal
endoscopy modalities [6 ]. Its unique advantages include single use, excellent tolerance, and minimal medical
staff requirement, which reduces the risk of exposure to pathogenic microorganisms
and cross-infection [7 ] and makes it an appropriate method for examining and triaging an endoscopy department
during the COVID-19 pandemic [8 ].
However, capsule endoscopy cannot be used to avoid face-to-face contact between endoscopists
and patients who have asymptomatic COVID-19 cases [9 ]. Herein, we describe a novel noncontact magnetically controlled capsule endoscopy
(ncMCE) system using a remote control system to separate endoscopists and patients
physically, which was developed and tested in our medical center [10 ]
[11 ]. The objective of the present study was to evaluate the feasibility and safety of
ncMCE for gastric examination.
Patients and methods
Study design and patients
This study was a pilot, open-label, randomized controlled trial approved by the institutional
review board of Shanghai Changhai Hospital and registered at ClinicalTrials.gov (NCT
04389333). All of the diagnostic and therapeutic modalities were carried out in accordance
with the approved guidelines.
Beginning on March 26, 2020, we consecutively recruited 40 patients in Changhai Hospital
using predefined inclusion criteria and exclusion criteria. After fully understanding
the procedure of ncMCE and this study, patients who gave informed consent were randomly
assigned into ncMCE and MCE groups (in a 1:1 ratio) by using a stratified block randomization
(fixed block size four). Randomization was based on a computer-generated list of random
numbers using SPSS software (version 22.0, SPSS Inc., Chicago, Illinois, United States).
Only after the enrolled participants completed all baseline assessments did the endoscopist
inform them about their allocation. Eligible patients were adults ( ≥ 18 years of
age) undergoing outpatient MCE. Patients with any of the following conditions were
excluded: (1) dysphagia or symptoms of gastric outlet obstruction, suspected or known
intestinal stenosis, overt gastrointestinal bleeding, fistulas and strictures; (2)
history of upper gastrointestinal surgery or suspected delayed gastric emptying; (3)
poor general condition, asthma or claustrophobia; (4) implanted metallic devices such
as pacemakers, defibrillators, artificial heart valves or joint prostheses; (5) pregnancy
or mental illness; (6) current participation in another clinical study; or (7) difficulty
communicating.
Study intervention
The MCE system (Ankon Technologies Co., Ltd. Shanghai, China) consists of a guidance
C-arm magnet robot, a computer workstation with ESNavi software, endoscopic capsule,
capsule locator, and a vest-like data recorder that can receive capsule signal from
both sides [12 ]. The ncMCE system adds a remote control workstation and an audio-visual exchange
system to the original well-established MCE system, which was developed and tested
at our medical center [10 ]. To simplify the preparation process, we embedded the data recorder in the examination
bed. The ncMCE system separates endoscopist and patients in two rooms (control room
for endoscopist and examination room for patients), offering physical isolation for
noninvasive gastric examination during the pandemic ([Fig. 1 ]) [11 ].
Fig. 1 Schematic diagram of noncontact magnetically controlled capsule endoscopy Real-time
communication via audio-visual exchange system and magnetic control precisely by a
remote-controlled workstation ensured the safety of endoscopists and avoided the risk
of cross-infection. Originally published in Liao Z, He S, Sha W, et al. Capsule endoscopy
practice during the COVID-19 pandemic: Recommendations from the Capsule Endoscopy
Group of the Chinese Society of Digestive Endoscopy. Endosc Int Open 2021; 9: E280–E283.
Permission for reuse has been obtained from the authors.
Patients were required to report personal history and have chest computed tomography
or nucleic acid testing to exclude COVID-19 infection within 3 days before their examination.
Patients who are confirmed to have a COVID-19 infection and symptoms of pneumonia
should be treated at special medical institutions first. Only when this population
needs immediate endoscopy should those procedures be done in specific wards that have
COVID-19 infection prevention protocols in place [13 ].
On arrival at the hospital in the morning after an overnight fast (> 8 hours), each
patient took a 400-mg simethicone suspension dissolved in 100 mL water 40 minutes
before they ingested the capsule ingestion according to standardized gastric preparation
[14 ]. Another 800 to 1000 mL of water could be taken 10 minutes before capsule ingestion
to distend the stomach [12 ]. Then, a patient’s hands were disinfected and they were given the data recorder
after entering the examination room. Next, a patient laid down on the examination
bed next to the computer workstation, assumed the supine or left-lateral decubitus
position, and swallowed the capsule, which had already been activated. An endoscopist
controlled capsule movement with the help of a magnet robot using two joysticks in
front of the workstation or automatically using a default mode. To ensure that the
gastric mucosa was completely visualized, the endoscopist directed a patient to change
position to supine and right-lateral decubitus. In the ncMCE group, a patient and
the endoscopist entered two rooms through different routes and gastric preparation
was under the remote guidance [15 ]. The audio-visual exchange system provided a platform for the endoscopist to see
and communicate with a patient during examination. The data recorder was embedded
in the examination bed to avoid cross-infection of COVID-19 among patients and simplify
the disinfection process.
Enhanced personal protective equipment (PPE) including an N95 mask, isolation gown
with water resistance, head cover, eye protection, and face shield [11 ] should be worn by medical staff in the capsule endoscopy examination room when performing
MCE. The steps for donning and doffing PPE were strictly followed, according to the
standard process [16 ]. Medical masks were used as basic PPE for endoscopists in the ncMCE group (Supplementary material , Table 1 ) [17 ]. After each examination, the returned data recorder was disinfected using ultraviolet
light, 75 % alcohol, or ethylene oxide and the sheets on the bed were replaced.
Data collection
Information on the basic characteristics of the enrolled patients was collected prospectively.
Each patient was followed up for 2 weeks by telephone and hospital information system
to confirm capsule excretion and any adverse events (AEs), including COVID-19 infection.
The maneuverability questionnaire for the endoscopist and improved questionnaire on
pre-procedure perception and post-procedure satisfaction for patients were completed
after the examination. To avoid systematic bias, a single endoscopist (W.Z.) performed
each examination and made the diagnoses for all enrolled patients immediately. The
evaluation of completion was made by another blinded researcher (J.H.Z.).
Study outcomes
The primary endpoints were feasibility assessed by completion rate (CR) and safety
evaluated by the occurrence of AE. CR was defined as complete observation of the cardia,
fundus, body, angulus, antrum, and pylorus of the stomach with no technical failure.
Technical failure included incomplete observation, direct contact in the ncMCE group,
and examination discontinuation. The safety of ncMCE was evaluated 2 weeks after the
procedure for any AEs, such as infection with SARS-CoV-2 and capsule-related AEs including
abdominal pain, nausea, vomiting, and capsule impaction or retention. In the current
study, the endoscopist was continuously monitored for infection until 2 weeks after
the last examination.
Secondary endpoints included maneuverability score, pre-procedure perception and post-procedure
satisfaction of patients, gastric examination time (GET), and diagnostic yield (DY).
The maneuverability of the ncMCE system was evaluated with a questionnaire that consisted
of signal transmission quality (stability and fluency), endoscopist comfort (strength
needed to control joysticks and degree of fatigue), visualization of gastric mucosa
(whether the mucus, foam in the stomach, and gastric fullness affected observation
of gastric mucosa) and compliance of patients (patient ability to readily and correctly
follow verbal instructions to change positions to assist in optimizing gastric views).
The score for each index was 1 to 5, with 1 as the worst and 5 as the best, respectively,
and a total ranging from 4 to 20. All patients were asked about the comfort and acceptability
of the procedure using an improved pre-procedure perception and post-procedure satisfaction
questionnaire, which was reviewed ease of swallowing, pain or discomfort experienced
during and after the procedure, overall tolerability and convenience of the procedure,
and knowledge, necessity for, and acceptability of ncMCE (Supplementary material , Questionnaire 1 ) [18 ]. GET was defined as the time taken for gastric examination and determined using
a digital stopwatch in the ESNavi software. DY or detection rate was calculated using
the following formula: number of patients with positive findings divided by the total
number of patients that underwent examination.
Statistical analysis
As a pilot study to evaluate the feasibility and safety of ncMCE, the sample size
was not calculated [19 ]. Considering the capsule endoscopy volume of Changhai Hospital during the pandemic,
we planned to recruit 40 patients into two groups. Basic characteristics were presented
as means with standard deviations for continuous variables such as age, body mass
index (BMI), and percentages for binary variables including sex, medical history,
and symptoms. Quantitative data were summarized with parametric statistics, mean and
standard deviation, or nonparametric statistics, median, and interquartile range (IQR),
whereas categorical data were presented as frequency (percentage).
Under the condition of normal distribution and homogeneity of variance, variance analysis
was used to compare the difference in mean between groups, while a K-W test was used
to compare the difference in distribution between groups and Fisher’s test was used
to compare the two and disordered multi-classification data of subjects under the
condition that homogeneity of variance was not satisfied. All reported P values are two-sided and differences that reached P < 0.05 were considered statistically significant. Statistical analyses were performed
with SPSS (version 22.0, SPSS Inc., Chicago, Illinois, United States).
Results
Patients
From March 26, 2020 to April 26, 2020, 40 patients (21 men and 19 women) were enrolled
and included in the analysis. Of them, 20 accepted MCE and 20 underwent ncMCE according
to the random number table. The mean BMI of patients was 23.27 kg/m2 and the indications in patients were counted (abdominal pain, 12 [30.0 %]; abdominal
distension, 13 [32.5 %]; health checkup, 10 [25.0 %]; acid reflux, 2 [5.0 %]; and
miscellaneous, 3 [7.5 %]). Ten of 40 patients (25 %) had a history of surgery. Baseline
patient characteristics are shown in [Table 1 ].
Table 1
Baseline characteristic.
ncMCE group (n = 20)
MCE group (n = 20)
P value
Male
11
8
0.342
Age, y, mean ± SD
46.1 ± 14.8
53.5 ± 9.2
0.067
Body mass index, kg/m2 , mean ± SD
23.1 ± 3.3
23.5 ± 3.2
0.698
History of abdominal surgery
4 (20 %)
6 (30 %)
0.465
Diabetes
2 (10 %)
1 (5 %)
0.999
Smoking history
3 (15 %)
6 (30 %)
0.449
Drinking history
6 (30 %)
7 (35 %)
0.736
Indication
16 (80 %)
14 (70 %)
0.716
4 (20 %)
6 (30 %)
0.716
Indication
9 (45 %)
3 (15 %)
0.038
7 (35 %)
6 (30 %)
0.736
0 (0 %)
2 (10 %)
0.487
4 (20 %)
6 (30 %)
0.465
0 (0 %)
3(15 %)
0.231
The other indications of MCE group include one with belching, one for clinical review
of gastric ulcer, and one with esophageal pain.
The counting data was described by n (%).
ncMCE, noncontact magnetically controlled capsule endoscopy; MCE, magnetically controlled
capsule endoscopy; SD, standard deviation.
Feasibility and safety
MCE and ncMCE were successfully performed on all patients with 100 % CR in both groups.
No infection and any suspicious symptoms in patients or the endoscopist or capsule-related
AEs were reported during the 2-week follow-up period.
Maneuverability score
The median maneuverability score in the ncMCE group was 19.25 (IQR, 18.6–20.0), which
was not significantly different than in the MCE group (median, 20.0; IQR, 19.0–20.0;
P = 0.179). The transmission quality in the ncMCE group (median, 4.5; range, 4.5–5.0)
was significantly lower (P < 0.001) than in the MCE group (median 5.0; range, 5.0–5.0), which was acceptable
for endoscopists. Whereas the comfortableness, visualization and compliance of patients
were similar in both groups ([Table 2 ]).
Table 2
Maneuverability of ncMCE and MCE.
ncMCE group (n = 20)
MCE group (n = 20)
P value
Maneuverability, median (IQR)
4.5 (4.5–5)
5 (5–5)
0.001
5 (5–5)
5 (5–5)
0.487
5 (5–5)
5 (4–5)
0.716
5 (5–5)
5 (5–5)
0.999
19.25 (18.63–20)
20 (19–20)
0.179
The signals transmission quality includes the stability and fluency of image and the
comfortableness of endoscopist includes the strength to control joysticks and fatigue
degree, which are based on the score of MCE group is 5 (Supplementary material, questionnaire
2).
ncMCE, noncontact magnetically controlled capsule endoscopy; MCE, magnetically controlled
capsule endoscopy; IQR, interquartile range.
Pre-procedure perception and post-procedure satisfaction
There was no significant difference in pre-procedure perception (9 vs. 9, P = 0.626) or post-procedure satisfaction (45 vs. 44, P = 0.999) in patients in the two groups ([Table 3 ]). Time off work for all the patients was less than 6 hours as a result of ncMCE
or MCE. Both ncMCE and MCE were acceptable to them (score for 15th question ≥ 2) with
no inconvenience. Ninety-five percent of patients agreed with the necessity for ncMCE
during the pandemic (score for 13th question ≥ 2).
Table 3
Pre-procedure perception and post-procedure satisfaction.
Item
ncMCE group (n = 20)
MCE group (n = 20)
P value
Pre-procedure patient perception, median (IQR)
4 (3–4)
4 (3–4)
0.802
4 (4–4)
4 (4–4)
0.553
1 (1.0–1.8)
1 (1–1)
0.077
9 (8–9)
9 (8–9)
0.626
Post-procedure patient satisfaction, median (IQR)
4 (4–4)
4 (4–4)
0.595
4 (4–4)
4 (4–4)
0.152
4 (4–4)
4 (4–4)
0.152
4 (4–4)
4 (4–4)
0.317
4 (4–4)
4 (4–4)
0.152
4 (3–4)
3 (3–4)
0.294
3 (3–4)
3.5 (3–4)
0.522
4 (3.3–4.0)
3.5 (3–4)
0.107
3.5 (3–4)
4 (3–4)
0.608
3.5 (3–4)
3 (2.0–3.8)
0.121
4 (3–4)
3.5 (3–4)
0.174
4 (4–4)
4 (4–4)
0.743
45 (42.3–46.8)
44 (42–45)
0.999
ncMCE, noncontact magnetically controlled capsule endoscopy; MCE, magnetically controlled
capsule endoscopy.
GET and diagnostic yield
The GET in the ncMCE group was 12.3 minutes on average (IQR, 7.5–16.3), which was
longer than the 7.2 minutes in the MCE group (IQR, 6.5–8.9; P = 0.003). There was no obvious difference between the two groups in positive findings
(P = 0.465). Four patients (20 %) in the ncMCE group had positive findings as did six
patients (30 %) in the MCE group ([Table 2 ]). Two (10%) and three (15 %) patients were diagnosed with polyps and erosion was
detected in two (10 %) and one (5 %) patient in the ncMCE and MCE groups, respectively.
The other two patients in the MCE group were diagnosed with ulcer ([Fig. 2 ], [Table 4 ]).
Fig. 2 Findings of noncontact magnetically controlled capsule endoscopy. Representative
gastric diseases observed by ncMCE (upper panel) and MCE (lower panel). a Gastric polyp. b Antrum erosion. c Normal antrum. d Normal gastric body. e Gastric polyp. f Gastric ulcer. g Normal gastric angle. h Normal pylorus.
Table 4
Diagnostic yield of ncMCE and MCE.
ncMCE group (n = 20)
MCE group (n = 20)
P value
Diagnostic yield, n (%)
2 (10 %)
3 (15 %)
2 (10 %)
1 (5 %)
0 (0 %)
2 (10 %)
Total
4 (20 %)
6 (30 %)
0.465
ncMCE, noncontact magnetically controlled capsule endoscopy; MCE, magnetically controlled
capsule endoscopy.
Discussion
To the best of our knowledge, this study is the first randomized trial to evaluate
the feasibility and safety of ncMCE compared with MCE during the COVID-19 pandemic.
We observed that ncMCE, which had 100 % CR and safety with excellent acceptance and
tolerance, was as maneuverable as MCE and offered physical isolation.
Since the COVID-19 outbreak, a series of infection prevention measures and guidelines
have been published for endoscopy departments [20 ]. ncMCE is a feasible method for endoscopists and patients, especially for high-risk
gastric examination during the COVID-19 pandemic.
For endoscopists working in critical and uncertain conditions of the pandemic, ncMCE
may protect medical staff from infection with SARS-CoV-2 by ensuring physical isolation.
Medical staff are facing great challenges to their physical and mental health. In
one of the earliest documented cases of infection in Wuhan, 29 % of patients were
healthcare workers [21 ]. A 13.7 % prevalence of SARS-CoV-2 antibodies was reported among health workers
in a large cohort study in New York State [22 ]. These findings remind us that the risk of infection in health workers may higher
than is expected. Standard or enhanced PPE should be used by personnel who are on
the front lines against COVID-19. But the use of PPE leads to decreased work efficiency
and medical resource shortages [23 ], which can be solved by ncMCE. In the current study, a questionnaire was used to
innovatively quantify the device’s maneuverability. We did not detect any significant
difference in the score for maneuverability between the two groups, except for signal
transmission quality. The endoscopist may have had to talk to the patients in the
ncMCE group more frequently, but it did not affect his or her comfortableness with
the technology.
For patients who need gastric examination, ncMCE offers an infection-free method to
avoid delay in diagnosis and monitoring. A study has shown that the number of patients
diagnosed with gastric cancer dropped by 49.1 %, and 4.6 % of gastric cancer patients
would have higher stage with a 6-month shift during the COVID-19 pandemic [24 ]. In our previous study, a cardia gastric cancer was found in 75-year-old man with
ncMCE, which underscores the need for more suitable endoscopy during COVID-19 [10 ]. The use of ncMCE solved the conflict between medical requirement and personal protection,
given the 100 % CR with AEs in our study. Safety was associated with use of strict
personal protection measures by the endoscopist for the MCE group and physical isolation
in the ncMCE group. The 100 % CR was due to the simple preparation and examination
procedure, which could be understood and completed easily by patients through the
remote guidance of the endoscopist. As a method of gastrointestinal examination, the
diagnostic ability of ncMCE was similar to that for MCE. The standardized gastric
preparation and proficient operation of the endoscopist led to good visualization
of the gastric mucosa. Positive findings were present in one-quarter of all our patients
totally, which was consist with the rate of findings in western and eastern countries
[25 ].
Occasionally, the signal was delayed during the examination procedure, which led to
lower transmission quality and longer GET in the ncMCE group. The data recorder in
the ncMCE group was put on the examination bed instead of being worn on the body,
so the signal receiving mode changed from bilateral to dorsal. When patients turned
to the right-lateral decubitus position, the antrum was the highest point in the whole
stomach and the farthest from the data recorder, which theoretically may lead to longer
signal transmission time. In addition, the extra time spent on communication with
patients and instructing them to change their positions also contributed to the longer
GET. However, the time to wear PPE for endoscopists and wear the data recorder for
patients can be saved in the ncMCE group.
Our study is novel in three ways. First, it was an exploration of infection-free endoscopy
for the first time, which is necessary not only during COVID-19, but also in special
cases, such as patients who have influenza, radiation injury, or impaired immunity.
Second, ncMCE provide a new method of using telemedicine in endoscopy. Telemedicine
or virtual visits is a part of the emergency response in some countries [23 ] because of the better view, higher precision of an operation [26 ], fewer outpatient visits [27 ], and complete isolation between patients and doctors [28 ]. The technology is developing rapidly and likely will remain in place after the
pandemic [29 ]. As with the DaVinci robotic system for surgery [26 ], in the field of remote manipulation endoscopy, a robot-assisted system named YunSRobot
was shown to result in no obvious differences in percentage of clear view time or
longer inspection time, compared with traditional manual endoscopy [30 ]. Unlike other telemedicine devices, the ncMCE system can be built easily in a short
time (< 10 minutes) with no other costs except for one data line and two computers
or even a glass partition if conditional, which means there was no extra financial
burden on patients.
Third, we were particularly concerned about the maneuverability for endoscopists during
the pandemic and designed a questionnaire to quantify that innovatively. According
to previous studies, maneuverability is the first of three issues that need to be
resolved for use of the stomach capsule, the other two being location and gastric
preparation [31 ]. Maneuverability should be focused on during the pandemic because of the special
working environment and personal protection requirements, which is different from
the typical situation.
Some limitations of our study should be acknowledged. First, there no patients with
confirmed SARS-CoV-2 infection were enrolled in this study because of the substantial
reductions in COVID-19 transmission in China at that time [32 ], which led to the 0 % infection rate in both groups. However, the technical feasibility
(100 % CR) shown in our study suggests the potential for preventing COVID-19 infection,
which needs confirmation in other centers. Second, the current study was single-center
and had a small sample size with no blinding of patients, which may have caused subjective
bias. Third, to limit subjective bias, we chose the same endoscopist for the two groups,
which made the transmission of COVID-19 between patients and endoscopist untraceable.
Furthermore, as a kind of capsule endoscopy, ncMCE also had a limitation in terms
of taking biopsies and therapeutic interventions compared to esophagogastroduodenoscopy
(EGD) [33 ].
With the development of wireless communication technology, remote surgery can be carried
out, which requires high data and speed capacity and a long operation time over a
reliable network with very low transmission delay [34 ]. With the help of wireless communication technology and an artificial intelligence-based
auxiliary diagnostic system, a “one to many” medical model and high-quality community
health care system will be built, and the shortage of endoscopists can be overcome.
Considering the requirement for epidemic prevention and control worldwide for a long
time, we believe that telemedicine is suitable for triaging patients before they arrive
in the emergency department [35 ].
Conclusions
In conclusion, our novel ncMCE provides a feasible and safe method of gastric examination
during the COVID-19 pandemic, which has the potential for preventing infection between
endoscopists and patients and unintended harm from diagnostic delays to patients.
It might be the first step in use of telemedicine for capsule endoscopy.
Noncontact magnetically controlled capsule endoscopy for infection-free gastric examination
during the COVID-19 pandemic: a pilot, open-label, randomized trial Jia-Hui Zhu, Jun Pan, Xiao-Nan Xu et al. Endoscopy International Open 2022; 10: E163–E170.
DOI: 10.1055/a-1648-2238 In the above-mentioned article the corresponding author was corrected to Zhuan Liao.
His correspondence address is: Prof. Zhuan Liao Department of Gastroenterology Changhai Hospital 168 Changhai Road Shanghai 200433 China Fax: +86-21-55621735 liaozhuan@smmu.edu.cn This was corrected in the online version on 17.03.2022.