Endoscopy 2022; 54(04): 384-388
DOI: 10.1055/a-1518-7039
Innovations and brief communications

Peroral cholecystoscopy using a multibending ultraslim endoscope through a lumen-apposing metal stent for endoscopic ultrasound-guided gallbladder drainage: a feasibility study

1   Digestive Disease Center and Research Institute, Department of Internal Medicine, SoonChunHyang University School of Medicine, Bucheon, South Korea
,
1   Digestive Disease Center and Research Institute, Department of Internal Medicine, SoonChunHyang University School of Medicine, Bucheon, South Korea
,
Yun Nah Lee
1   Digestive Disease Center and Research Institute, Department of Internal Medicine, SoonChunHyang University School of Medicine, Bucheon, South Korea
,
Yeon Han Song
1   Digestive Disease Center and Research Institute, Department of Internal Medicine, SoonChunHyang University School of Medicine, Bucheon, South Korea
,
Jae Kook Yang
2   Digestive Disease Center and Research Institute, Department of Internal Medicine, SoonChunHyang University School of Medicine, Cheonan, South Korea
,
Tae Hoon Lee
2   Digestive Disease Center and Research Institute, Department of Internal Medicine, SoonChunHyang University School of Medicine, Cheonan, South Korea
,
Sang-Woo Cha
3   Digestive Disease Center and Research Institute, Department of Internal Medicine, SoonChunHyang University School of Medicine, Seoul, South Korea
,
Young Deok Cho
3   Digestive Disease Center and Research Institute, Department of Internal Medicine, SoonChunHyang University School of Medicine, Seoul, South Korea
,
Sang-Heum Park
2   Digestive Disease Center and Research Institute, Department of Internal Medicine, SoonChunHyang University School of Medicine, Cheonan, South Korea
› Author Affiliations


Abstract

Background Endoscopic ultrasound-guided gallbladder drainage (EUS-GBD) using a lumen-apposing metal stent (LAMS) has rendered endoscopic gallbladder assessment through a stent possible, but an appropriate endoscope is required. We assessed the feasibility of peroral cholecystoscopy (POCC) using a multibending ultraslim endoscope passed through a LAMS after EUS-GBD.

Methods 14 patients with a LAMS for EUS-GBD who consecutively underwent POCC were included. POCC was performed using a multibending endoscope inserted through the LAMS. The primary outcome was the technical success rate, defined as complete endoscopic examination of the gallbladder from the orifice of the cystic duct to the fundus. The types of intervention and adverse events were recorded.

Results 17 POCCs were performed in 14 patients. Of the 17 POCCs, 15 (88.2 %) were technically successful. Narrow-band imaging endoscopy was performed in 12 procedures (70.6 %), and cholecystoscopy-guided target biopsies were obtained in six. Gallstone extraction was performed in two patients. After POCC, all LAMSs remained stable and no adverse events were observed.

Conclusions POCC using a multibending ultraslim endoscope can be effectively and safely performed through a LAMS after EUS-GBD.

Table 1s, Figs. 1s, 2s



Publication History

Received: 01 December 2020

Accepted after revision: 05 May 2021

Article published online:
06 July 2021

© 2021. Thieme. All rights reserved.

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany

 
  • References

  • 1 Chan JHY, Teoh AYB. The development of peroral cholecystoscopy and advanced gallbladder interventions. Endosc Ultrasound 2018; 7: 85-88
  • 2 Jang JW, Lee SS, Park DH. et al. Feasibility and safety of EUS-guided transgastric/transduodenal gallbladder drainage with single-step placement of a modified covered self-expandable metal stent in patients unsuitable for cholecystectomy. Gastrointest Endosc 2011; 74: 176-181
  • 3 Chan SM, Teoh AYB, Yip HC. et al. Feasibility of per-oral cholecystoscopy and advanced gallbladder interventions after EUS-guided gallbladder stenting (with video). Gastrointest Endosc 2017; 85: 1225-1232
  • 4 Ge N, Sun S, Sun S. et al. Endoscopic ultrasound-assisted transmural cholecystoduodenostomy or cholecystogastrostomy as a bridge for per-oral cholecystoscopy therapy using double-flanged fully covered metal stent. BMC Gastroenterol 2016; 16: 9
  • 5 Itoi T, Itokawa F, Tsuchiya T. et al. Transgastric large gallstone extraction through a lumen-apposing metal stent in a patient with acute cholecystitis. Gastrointest Endosc 2014; 79: 547
  • 6 Teoh AY, Chan AW, Chiu PW. et al. In vivo appearances of gallbladder carcinoma under magnifying endoscopy and probe-based confocal laser endomicroscopy after endosonographic gallbladder drainage. Endoscopy 2014; 46: E13-E14
  • 7 Lee YN, Moon JH, Lee TH. et al. Prospective randomized trial of a new multibending versus conventional ultra-slim endoscope for peroral cholangioscopy without device or endoscope assistance (with video). Gastrointest Endosc 2020; 91: 92-101
  • 8 Yoo HW, Moon JH, Jo SJ. et al. A novel electrocautery-enhanced delivery system for one-step endoscopic ultrasound-guided drainage of the gallbladder and bile duct using a lumen-apposing metal stent: a feasibility study. Endoscopy 2020; DOI: 10.1055/a-1301-1526.
  • 9 Cotton PB, Eisen GM, Aabakken L. et al. A lexicon for endoscopic adverse events: report of an ASGE workshop. Gastrointest Endosc 2010; 71: 446-454
  • 10 Inui K, Nakazawa S, Yoshino J. et al. Percutaneous cholecystoscopy. Endoscopy 1989; 21 Suppl 1: 361-364
  • 11 Fujita R, Hirata N, Fujita Y. Peroral cholecystoscopy. Endoscopy 1989; 21: 378-380
  • 12 Takahara N, Kawakubo K, Isayama H. et al. Cholesterolosis of the gallbladder visualized by peroral cholecystoscopy using a SpyGlass probe. Endoscopy 2012; 44: E145-E146
  • 13 Gutkin E, Hussain SA, Kim SH. The successful treatment of chronic cholecystitis with SpyGlass cholangioscopy-assisted gallbladder drainage and irrigation through self-expandable metal stents. Gut Liver 2012; 6: 136-138
  • 14 Barkay O, Bucksot L, Sherman S. Endoscopic transpapillary gallbladder drainage with the SpyGlass cholangiopancreatoscopy system. Gastrointest Endosc 2009; 70: 1039-1040
  • 15 Walter D, Teoh AY, Itoi T. et al. EUS-guided gall bladder drainage with a lumen-apposing metal stent: a prospective long-term evaluation. Gut 2016; 65: 6-8