Introduction
Zenkerʼs diverticulum (ZD) is a rare condition with an unknown overall prevalence
given many patients are likely asymptomatic [1]. ZD are pseudodiverticula that manifest as mucosal and submucosal pulsion through
Killianʼs triangle. Symptoms most frequently include dysphagia, regurgitation, and,
in severe cases, weight loss and malnutrition [2]. Historically, ZD were treated by an open surgical approach but this has largely
been abandoned due to development of transoral interventions using both rigid and
flexible endoscopes [2]
[3].
The conventional flexible endoscopic approach involves complete division of the cricopharyngeal
(CP) septum (including the mucosa) to the base of the diverticulum. This approach
can be performed using a variety of endoscopic accessories [4]. Recently, Li et al. described a novel technique to divide the CP septum while preserving
the mucosa via creation of a short submucosal tunnel. The procedure has been termed
Zenkerʼs peroral endoscopic myotomy (POEM) or Z-POEM as the technique is similar to
esophageal POEM, an evolving technique in the treatment of achalasia [5]
[6].
Outcomes following flexible endoscopic intervention for ZD have largely relied on
clinical follow-up, which is subjective and not standardized. Post-myotomy esophagram
findings do not correlate with symptoms [2]
[7]. With the evolution of Functional Luminal Imaging Probe (FLIP) technology (EndoFLIP,
Medtronic, Minneapolis, Minnesota, United States), clinicians are now easily able
to objectively measure luminal parameters such as cross-sectional area (CSA), diameter,
and distensibility index (DI) using a catheter that is 8 cm in length and contains
16 sensors within an electrode-containing balloon. The catheter attaches to a console,
which executes calculations to perform high-resolution impedance planimetry [8]. Impedance planimetry gained initial traction in predicting response to POEM for
achalasia, and has been used to evaluate the pylorus in gastroparesis patients undergoing
gastric POEM [9]. We aimed to determine if impedance planimetry measurements of ZD improve with flexible
endoscopic diverticulotomy and predict clinical response as determined by the previously
validated Eating Assessment Tool (EAT-10) [10]
[11].
Patients and methods
The study protocol was approved by the Michigan Medicine Institutional Review Board
(HUM00160186). Patients were included if all of the following criteria were met: ≥ 18
years of age, upper esophageal symptoms with compatible esophagram and/or esophagogastroduodenoscopy
(EGD) findings demonstrating ZD, and who underwent EGD with impedance planimetry before
and after flexible endoscopic CP myotomy. The EAT-10 questionnaire was completed by
each patient prior to the procedure ([Table 1]). All procedures were performed by a single endoscopist (RL). All patients underwent
general anesthesia with endotracheal intubation and placed in the supine position.
The impedance planimetry balloon was then positioned across the upper esophageal sphincter
(UES) under direct visualization. The balloon was inflated to 40 mL and 50 mL for
30 seconds at each volume and then deflated. Diameter and DI were recorded at each
inflation. At the discretion of the endoscopist, the ZD was then treated with a conventional
flexible endoscopic CP myotomy or the Z-POEM approach. The decision of which approach
to employ was based upon the size of the diverticulum and overall amenability to a
POEM approach. A clear cap was attached to the scope and advanced to the diverticulum.
In select cases, a transnasal feeding tube was placed into the esophageal lumen to
serve as a landmark during division of the CP septum. In the conventional flexible
endoscopic CP myotomy approach, an electrosurgical knife was used to incise the CP
septum, including the mucosal layer. In contrast, when using the Z-POEM approach,
we utilized the same techniques as previously published by Li et al [5]. Prior to hemostatic clip closure of the defect, post-myotomy impedance planimetry
data were acquired with the same pre-myotomy methods. The post-procedure EAT-10 questionnaire
was obtained within 1 month and then repeated during follow up. Data were analyzed
using descriptive statistics.
Table 1
Eating Assessment Tool-10.[1]
To what extent are the following scenarios problematic for you?[2]
|
My swallowing problem has caused me to lose weight.
|
0 1 2 3 4
|
My swallowing problem interferes with my ability to go out for meals.
|
0 1 2 3 4
|
Swallowing liquids takes extra effort.
|
0 1 2 3 4
|
Swallowing solids takes extra effort.
|
0 1 2 3 4
|
Swallowing pills takes extra effort.
|
0 1 2 3 4
|
Swallowing is painful.
|
0 1 2 3 4
|
The pleasure of eating is affected by my swallowing.
|
0 1 2 3 4
|
When I swallow, food sticks in my throat.
|
0 1 2 3 4
|
I cough when I eat.
|
0 1 2 3 4
|
Swallowing is stressful.
|
0 1 2 3 4
|
1 Adapted from Belafsky [10]
2 0 = no problem. 4 = severe problem.
Results
Thirteen patients underwent ZD myotomy (mean age 79 years; 6 patients were female).
This included one patient previously published as a single case report [12]. Patient demographics, endoscopic approaches, and follow-up time are described in
[Table 2]. Ten patients (77 %) underwent myotomy via the Z-POEM approach. After the 1-month
mark, patients were periodically called to assess for durability of clinical response.
Mean follow-up time after the myotomy was 94 days (range 2–277 days). No patient reported
a recurrence of symptoms. No intra-procedural or post-procedural adverse events (AEs)
occurred during the follow-up period. Pre-myotomy impedance planimetry data and EAT-10
scores are described in [Table 3]. Post-myotomy data are described in [Table 4]. The changes in impedance planimetry and EAT-10 scores due to the myotomy are described
in [Table 5]. All patients reported subjective clinical improvement post-myotomy. Representative
pre-myotomy and post-myotomy impedance planimetry topographic readings are shown in
[Fig. 1] and [Fig. 2], respectively. All patientsʼ EAT-10 scores improved, with a mean improvement of
15.3 points. The mean diameter increased by 2.3 mm (–0.8 to 3.7) at 40 mL, which correlated
with a mean diameter increase of 18.4 % (–7.3 to 31.8). The mean diameter increased
by 2.6 mm (–0.7 to 6.1) at 50 mL, which correlated with a mean diameter increase of
18.8 % (–6.2 to 45.2). The mean DI increased by 1.9 mm2/mmHg at 40 mL (0.3–7), which correlated with a mean DI increase of 71.7 % (6.8–66.7).
The mean DI increased by 1.8 mm2/mmHg (0.4–5.3) at 50 mL, which correlated with a mean DI increase of 83.9 % (26.7–265).
We did not perform impedance planimetry during follow-up. As this was only an exploratory
study, we could not justify the additional cost to the patient and minimal risk of
performing follow-up impedance planimetry, particularly in patients who were doing
clinically well.
Table 2
Patient demographics, endoscopic approaches, follow-up time.
Patient age, years (mean)
|
No. female patients (%)
|
No. Z-POEM (%)
|
No. CP myotomy (%)
|
Follow-up time, days (mean)[1]
|
54–94 (79.6)
|
6 (46)
|
10 (77)
|
3 (23)
|
2–277 (94)
|
Z-POEM, Zenker's peroral endoscopic myotomy; CP, cricopharyngeal.
1 The patient with 2 days of follow-up died several days after the procedure due to
an unrelated medical illness.
Table 3
Pre-myotomy impedance planimetry data and EAT-10 scores.
Metric
|
40 mL (mean)
|
50 mL (mean)
|
Diameter (mm)
|
11–15.7 (12.8)
|
11.3–18.2 (13.6)
|
DI (mm2/mmHg)
|
1.3–4.4 (2.4)
|
1.1–6.1 (2.2)
|
EAT-10 (mean)
|
7–31 (18.3)
|
EAT-10, Eating Assessment Tool.
Table 4
Post-myotomy in impedance planimetry data and EAT-10 scores.
Metric
|
40 mL (mean)
|
50 mL (mean)
|
Diameter (mm)
|
10.2–19.0 (15.2)
|
10.6–22.0 (16.2)
|
DI (mm2/mmHg)
|
1.9–12.7 (5.1)
|
1.8–10.5 (4.0)
|
EAT-10 (mean)
|
0–8 (3)
|
EAT-10, Eating Assessment Tool
Table 5
Change in impedance planimetry data and EAT-10 scores.[1]
Metric
|
40 mL (mean)
|
50 mL (mean)
|
Diameter (mm)
|
–0.8 to + 3.7 (+ 2.3)
|
–0.7 to + 6.1 (+ 2.6)
|
DI (mm2/mmHg)
|
+ 0.3–7 (+ 1.9)
|
+ 0.4–5.3 (+ 1.8)
|
EAT-10 (mean)
|
–31 to –2 (–15.3)
|
EAT-10, Eating Assessment Tool.
1 A “ + ” indicated the value increased or changed to a greater absolute value. A “-”
indicated the value decreased or changed to a lesser absolute value.
Fig. 1 Representative pre-myotomy impedance planimetry topographic reading.
Fig. 2 Representative post-myotomy impedance planimetry topographic reading.
Discussion
To date, no reliable objective intra-procedural marker of success following Zenkerʼs
diverticulotomy has been identified. Available data suggest impedance planimetry correlates
with outcomes in achalasia after POEM [13]
[14] and surgical intervention [15]. Based on these findings, we evaluated impedance planimetry technology to assess
its predictive ability following flexible endoscopic Zenkerʼs repair.
Our study evaluated 13 patients who underwent successful intervention for the treatment
of ZD. All patients demonstrated clinical improvement with endoscopic intervention
without AEs. The mean luminal diameter, DI, and EAT-10 all improved. One patientʼs
diameter appeared to decrease given measurements were obtained after hemostatic clip
placement, which created a narrowed luminal caliber and likely artifactual decrease
in the diameter. In this patient, clip placement was performed prior to impedance
planimetry measurements given the patient had an adjacent esophageal diverticulum
such that balloon inflation risked increasing the mucosotomy size, potentially resulting
in difficult endoscopic closure and tissue apposition.
Notably, the impedance planimetry data and EAT-10 scores improved at 40-mL and 50-mL
balloon volumes. We identified that all patients with a DI improvement of at least
26.7 % at 50 mL experienced a clinical response. This suggests there may be a degree
of relative improvement in the DI that could predict clinical success. This is of
particular use to the endoscopist as intra-procedural impedance planimetry data could
guide when to cease further myotomy. In addition, the percent change between pre-
and post-myotomy may be more valuable than a specific diameter, distensibility, or
CSA threshold.
In this study, we considered the change in pre-myotomy and post-myotomy impedance
planimetry data (i. e. DI, diameter, CSA) to be more valuable than the absolute numbers,
as the threshold for technical success when utilizing absolute numbers requires standardized
values to characterize normal and abnormal. There are limited UES impedance planimetry
data available in the current literature, and only in healthy controls. Regan et al.
evaluated 13 subjects without any esophageal pathology and used metrics similar to
but not the same as DI [16]. Generalizing their data is difficult given balloon volumes were low at 20-mL inflation
and correlating manometric data was not obtained. Pyloric and lower esophageal sphincter
interventions that utilize impedance planimetry monitoring have noted clinical success
based on a change in DI values, even when the post-intervention measurement did not
reach the normal value [9]
[15]. This suggests normal values, while important for diagnostic purposes, may not play
as important of a role in predicting symptom response to intervention. With these
considerations in mind, we believe the percent change between pre-myotomy and post-myotomy
impedance planimetry provides a more suitable parameter to assess technical success.
Several limitations of this pilot study should be noted, including a small sample
size from a single tertiary care institution. In addition, all procedures were performed
by a single endoscopist. This may limit generalizability. Strengths of this study
include prospective approach, maintenance of consistent endoscopic techniques during
the myotomy procedures, and use of the validated EAT-10 symptom tool.
Myotomy length was not measured, given that measurements are subjective and likely
inaccurate, particularly in Zenkerʼs patients where differences are in millimeters,
or centimeters, at best. In addition, identification of the junction of esophageal
circular muscle and the cricopharyngeus is difficult to identify in the submucosal
tunnel approach. Thus, measurements using a POEM approach may be difficult to interpret.
Conversely, in the patients undergoing conventional endoscopic CP myotomy, the risk
of AEs may lead to an incomplete myotomy. Thus, the meaning of the actual myotomy
length is difficult to interpret. Septotomy length was not compared to DI, given concerns
regarding accuracy in measuring septotomy length. Impedance planimetry may be able
to guide septum division in the future, but we need to have a reliable target to aim
for (i. e. normative post-myotomy data in patients with endoscopically/surgically
treated ZD).
Conclusions
At present, the role of impedance planimetry in patients with ZD remains unclear.
The results of our pilot study suggest this technology could be useful in providing
objective data. Further research in a larger sample size is required to corroborate
these results, evaluate for a possible minimum change in impedance planimetry to achieve
clinical response, and compare the efficacy of conventional flexible endoscopic CP
myotomy versus the Z-POEM approach.