We read with interest the article by Zhang et al. regarding suggestions for infection
prevention and control in digestive endoscopy during the COVID-19 pandemic in Wuhan,
China [1]. All upper gastrointestinal endoscopy procedures are aerosol generating, hence there
is risk of spreading infection [2]. Endoscopic retrograde cholangiopancreatography (ERCP) carries the additional risk
of duodenoscope-related infection [3]
[4].
Recently, the US Food and Drug Administration approved the use of disposable duodenoscopes.
We report performing an ERCP in a patient with cholangitis who was under investigation
for COVID-19 using a single-use duodenoscope. The patient was a 90-year-old woman
with a past history of congestive heart failure and dementia who presented with fever
and abdominal pain. She was tested for COVID-19. Physical examination, laboratory
data, and imaging were suggestive of acute cholangitis and septic shock. Therefore,
we proceeded to ERCP and, given that the patient was under investigation for COVID-19,
she was presumed to be positive as per our institutional guidelines. A single-use
duodenoscope (EXALT Model D; Boston Scientific Corporation, Marlborough, Massachusetts,
USA) was used for the procedure. The common bile duct was cannulated and a cholangiogram
revealed choledocholithiasis. A biliary sphincterotomy was deferred because of clopidogrel
use and a plastic biliary stent was placed. The procedure duration was 15 minutes.
Upon completion, the endoscope was disconnected from the processor and sealed in a
bag, and the room was disinfected.
Single-use duodenoscopes could be beneficial in COVID-19 patients requiring ERCP.
The use of a disposable duodenoscope eliminates the low risk of patient-to-patient
transmission related to inadequate disinfection. There is also a risk of aerosolization
during disinfection and therefore of exposure risk to personnel performing endoscope
cleaning, both in the procedure room and in the endoscope cleaning room. Furthermore,
the theoretical risk of aerosolization during endoscope disinfection could contaminate
the cleaning room necessitating its closure for 30 minutes after the disinfection
process is completed.
Using a single-use duodenoscope for ERCP in patients with confirmed or suspected COVID-19
may be a reasonable strategy to reduce the risk of infection among patients and healthcare
personnel. However, more data are needed to see if SARS-Cov-2 specifically is difficult
to remove from duodenoscopes, and considerations of cost-effectiveness and environmental
impact of single-use devices will need to be taken into account.