Common uses
Abdominal handheld ultrasound
Abdominal ultrasound examination covers a range of different diseases and organs and
is performed in various clinical settings ranging from acute examinations to routine
and follow-up examinations. In addition, physicians from a wide range of specialties
including pediatrics, emergency care medicine, gastroenterology, urology and gastrointestinal
surgery perform abdominal ultrasound. This stresses the need for systematic and well-considered
implementation of HHUSDs in abdominal ultrasound to achieve comparable and reliable
information between different patients, diseases and physicians.
The clinical trials published on abdominal HHUSDs generally fall into one of four
groups: gastroenterology, gynecology, urology and abdominal aortic ultrasound. Most
of the articles on HHUSDs in gastroenterology and related fields focus on tentative
diagnosis and triage of acutely ill patients or on assessments of ascites for paracentesis
[8]
[9]
[11]
[12].
In a study performed by residents in 199 recently admitted medical patients [8], the diagnosis of ascites, liver metastases, abdominal tumors, hydronephrosis and
cholelithiasis could be confirmed but not ruled out by HHUSD, when compared to standard
ultrasound examination performed by an experienced radiologist. This finding is confirmed
by a study on 28 medical ward patients [13], which suggests that even though the specificity of HHUSDs is high for hepatic,
biliary, renal and ascites evaluation, the sensitivity is insufficient when conventional
ultrasound devices are used as the gold standard. One feasibility study on known focal
lesions with a focus on HHUSD image quality and ability to reliably evaluate abdominal
lesion size and vascularization showed satisfactory results and overall high diagnostic
accuracy and value for determining the puncture site for paracentesis among other
things [7]. In 216 trauma patients, HHUSD FAST-scanning had an NPV of 99.5 %, signifying that
HHUSDs could be used to rule out free abdominal fluid [12]. Concordantly, in patients clinically suspected of ascites, HHUSDs successfully
confirmed the diagnosis in 96 % of cases, while the diagnosis was only missed in mild
cases as assessed by conventional ultrasound or computed tomography (CT) [14]. In the same study HHUSD-guided paracentesis of ascites was feasible in 38 of 40
patients, with no severe complications. In another study performed in an outgoing
palliative care unit, an HHUSD was successfully used to both diagnose ascites and
guide paracentesis when necessary in terminally ill patients [15]. In a study of patients in a cardiology ward, experienced physicians successfully
used HHUSD examination to screen for abdominal aortic aneurysms [16].
In a study of 196 medical patients [9], an HHUSD was used to confirm cardiac, lung and abdominal findings, which resulted
in a changed primary diagnosis in 18.4 % of patients, a confirmed diagnosis in 19.4 %
of patients, and an additional diagnosis in 9.2 % of patients. These results are similar
to another study showing either a confirmed or altered primary diagnosis or an important
secondary diagnosis following HHUSD examination in roughly 40 % of patients admitted
to the emergency ward [17]. In another study focusing mainly on abdominal ultrasound performed by general practitioners
and specialists in internal medicine and gastroenterology, there was agreement between
the initial clinical hypothesis and the latter HHUSD examination in 66 % of cases.
In the remaining 34 %, the initial hypothesis was not confirmed by the HHUSD examination.
Gastroenterologists had higher agreement between their clinical hypothesis and HHUSD
examinations (92.4 %) than specialists in internal medicine (67.1 %), who again had
higher agreement than general practitioners (47.0 %) [18].
HHUSDs have also been applied for urologic US evaluations. Different hand-held bladder
scan devices have been commercially available for bladder volume evaluation for many
years [19]. Most of these devices are dedicated to this purpose and exceed the scope of this
review. With respect to general purpose HHUSDs, a few studies have been published
concerning kidney and bladder diagnostics. One study in 36 patients found that the
evaluation of quantitative parameters such as kidney length, renal-pelvis length,
renal cyst size and post-void bladder and prostate volume was feasible with only small
variations between observers and compared with a standard US device [20]. In addition, correlation with standard US was moderate to substantial for evaluation
of the presence of hydronephrosis, the number of renal cysts and the presence of the
ureter jet sign. In another study concerning the evaluation and grading of hydronephrosis
in 200 kidneys and with high-end US as the gold standard, HHUSDs were found to have
an NPV between 96 % and 98 % depending on the threshold for hydronephrosis [21]. In a recent study, a modified HHUSD device was successfully used for ultrasound-guided
percutaneous nephrolithotomy in 31 patients with kidney stones verified by CT scans
[22].
A few studies have been performed using HHUSDs in gynecology and obstetrics. Studies
in women with vaginal bleeding or pelvic pain in first-trimester pregnancies showed
good ability of HHUSDs to triage for further treatment, except in cases of ectopic
pregnancy [23]
[24]. For routine third-trimester or antenatal examinations, HHUSDs have been successfully
applied to assess fetal growth and wellbeing, placental location and fetus presentation
[24]
[25]. In gynecological patients, HHUSDs could assess the presence or absence of pelvic
masses either using abdominal transducers [24] or a modified transvaginal transducer [26], while the latter also allows for reliable assessment of focal mass sizes and vascularization
compared with conventional ultrasound.
The published studies vary considerably in size and study setup. The number of included
patients range from small cohorts to a few hundred patients [7]
[11]
[12]
[13]. The published studies involve physicians from a great variety of medical specialties
and with experience ranging from medical students to highly experienced sonographers,
making interpolation between studies difficult. Also, the quantity of pre-study ultrasound
training varies between studies.
The published studies also vary with respect to their design as some evaluate the
impact on clinical decision making, while others compare the diagnostic accuracy of
HHUSDs with conventional ultrasound or other imaging modalities. For all of the above-mentioned
reasons, the overall level of evidence for HHUSDs in abdominal ultrasound is still
modest, but the amount of data is steadily growing.
Position Comment
Handheld ultrasound devices are rapidly becoming a part of everyday clinical practice
for the evaluation of abdominal disease. HHUSDs should primarily be used in POCUS
with few and clear examination objectives, such as the assessment of the presence
of ascites, guidance of paracentesis, and the assessment of hydronephrosis or gall-bladder
calculi. Furthermore, the ability of HHUSDs to alter or support clinical decision
making at a very early point during patient hospitalization should be further explored.
Echocardiography and Lung Ultrasound
The use of HHUSDs in the initial bedside evaluation of the acute patient is increasing
viewed as a natural extension of the traditional physical examination of the cardiovascular
system [27].
Several studies have evaluated the performance of handheld devices for the assessment
of specific clinical questions, either in the daily assessment of the acute patient
in a variety of environments (admissions ward, emergency room, intensive care) or
in the follow-up of patients with an established diagnosis at outpatient clinics.
Many of these studies have determined the performance of a HHUSD cardiac scan by measuring
the agreement with an independently performed high-resolution echocardiogram. Left
ventricular (LV) size, LV systolic function, regional wall motion abnormalities and
pericardial effusion have all been detected using HHUSDs with good to high levels
of accuracy [27]. In one study with trained cardiologists performing bedside ultrasound with HHUSDs
in 104 acute cardiac care patients, the level of agreement for systolic function and
pericardial effusion had a kappa of 0.89 and 0.81, respectively [28]. Agreement was less robust for assessing aortic, mitral and tricuspid valve function
(kappa: 0.55 – 0.66). In a more recent study, 82 patients admitted with acute myocardial
infarction were initially scanned with a, HHUSD by an experienced sonographer, who
tried to perform as comprehensive examination as possible, including the use of color
Doppler (average time 10 minutes). Correlation coefficients between the initial scan
and a subsequent standard echo were good for LV ejection fraction and global wall
movement (the wall motion index) score of 0.75 and 0.69, respectively. The levels
of agreement for left ventricular enlargement, right ventricular dysfunction, pericardial
effusion and inferior vena cava (IVC) distention were less robust with highly variable
agreement in the assessment of segmental regional wall motion, dependent upon which
part of the myocardium was being evaluated. This was assumed to be due to the limited
field of view offered by the HHUSD that was used [29]. The evaluation of filling pressures using HHUSDs by measuring IVC diameter during
the respiratory cycle has been evaluated in a number of studies again with variable
agreement with respect to the reference scan [27]. As with the previous study, a feasibility study of 108 inpatients initially scanned
with an HHUSD by an experienced cardiologist followed by a departmental echo found
reduced correlation between exams for IVC diameter compared to that for LV and RV
function, valvular function, atrial size and pericardial effusion [30]. Although the authors speculated that this may reflect the time interval between
the two scans, satisfactory visualization of the IVC was scored at only 85 % and this
again may be due to the limited field range of HHUSDs [27].
Despite some of the current HHUSDs lacking either color or spectral Doppler, the assessment
of valve disease, both stenosis and regurgitation (only those with color Doppler),
is possible with these devices. In a study of 130 patients with suspected or known
aortic stenosis (AS), an additive scoring system was applied based upon a grayscale
assessment of the level of restriction in the opening of each valve cusp using an
HHUSD. This was then compared with aortic valve area indices calculated using high-end
echocardiography. The receiver operating curves for diagnosing severe and moderate
to severe stenosis using the calculated AS score were 0.946 and 0.936, respectively.
Although not significantly better than skilled clinical examination, this performed
better than aortic calcium scores obtained using the HHUSD [31]. Using an HHUSD equipped with color Doppler in the assessment of acute coronary
patients, trained cardiologists were able to identify aortic stenosis and regurgitation
to a good level of agreement compared with subsequent high-end echocardiography. However,
agreement was only moderate for mitral and tricuspid regurgitation (k = 0.55 – 0.56),
with a tendency to overestimate regurgitation using the HHUSD [28]. In another study comparing inter- and intra-observer agreement between HHUSDs and
standard echocardiography in 320 patients, the level of agreement for grading mitral
and aortic regurgitation was at least moderate (r > 0.6), while the detection of mitral
regurgitation was less favorable, with milder cases being missed by the HHUSD [32].
With reference to some of the above studies, the perceived limitations of HHUSDs for
cardiac ultrasound when compared to even portable echo machines were highlighted in
a position statement by the European Association for Echocardiography in 2011. Based
on the technical performance of units at that time, the listed indications were: complement
to clinical examination, screening tool in the emergency room or ambulance, initial
evaluation in outreach clinics, triage tool for determining who should undergo fully
comprehensive echocardiogram and teaching tool [33]. One study evaluated the spontaneous use (not part of a clinical trial) of HHUSDs
by non-specialist hospital residents who had undergone appropriate training [34]. The use of an HHUSD by 24 residents admitting 542 patients overnight was retrospectively
analyzed. The residents elected to scan 42 % of patients as part of their initial
assessment and not to scan 58 % of patients, with lack of clinical indication, time
constraints and poor patient cooperation as the main reasons for the latter choice.
There was, however, a wide variation in the percentage use of the HHUSD by individual
residents (17 – 85 %), despite all receiving the same level of training. A number
of factors may account for this variation, including comfort with use of the device,
time and efficient data collection, i. e., information already available from previous
patient tests and imaging.
The identification of fluid overload within the lungs by the visualization of multiple
ring-down artifacts in the image known as B-lines or “lung rockets” has been shown
to be reliably achievable with HHUSDs and can be taught to other members of the patient
care team [35]. In a prospective study using HHUSDs for the assessment of 185 outpatients with
established heart failure (HF), 8 chest zones were scanned by skilled investigators
for a median of 2 minutes, with all recorded clips reviewed offline. 32 % of patients
had ≥ 3 B-lines, yet 82 % had no findings at auscultation. At follow-up these patients
had a four-fold increase in hospital admission for heart failure and death than patients
with < 3 B-lines. HHUSD assessment may therefore allow optimization of patients with
HF [36]. In another study, however, a potential discrepancy with HHUSDs in the assessment
of B-lines in patients with HF was highlighted due to their limited clip store capacity.
When comparing an HHUSD with 2-second clip store with a high-end system able to record
at least 6 seconds, there was a significant difference in the number of B-lines observed
during a standardized scanning protocol, with a higher number being seen using the
longer recorded clips [37].
HHUSDs have also been successfully employed for the detection of B-lines in other
interstitial lung conditions. In a study of 39 patients with rheumatoid arthritis
comparing HRCT with standardized lung ultrasound, a subset of 29 patients were also
scanned using an HHUSD by a physician with limited training in B-line recognition
at ultrasound. In this subset, the sensitivity and specificity for interstitial lung
disease with respect to HRCT were 89 % and 50 %, respectively, while the concordance
with standard ultrasound was good (k = 0.78) [38]. The value of lung ultrasound performed by operators with limited experience using
HHUSDs has also been demonstrated in the acute assessment of patients with dyspnea.
Five junior doctors performed 69 scans using HHUSDs combined with clinical assessment
of patients with breathlessness. The final diagnosis was determined by the senior
admitting physician at discharge and this was used to calculate receiver operator
curves for clinical and ultrasound performance. The resultant area under the curve
was significantly higher for ultrasonic diagnosis (chronic obstructive lung disease,
pulmonary edema, pneumonia, pleural effusion) than clinical assessment: 0.87 vs. 0.81
[39]. Notably, two of the doctors who underwent extended training demonstrated increased
diagnostic accuracy compared to their peers. In the further management of pleural
effusion (PE), HHUSDs may offer an ideal solution for bedside quantification and guidance
of thoracentesis for the resolution of respiratory compromise. In a study of 73 people
with an abnormal chest radiograph suggestive of PE, a single experienced ultrasound
operator performed pleural ultrasound with an HHUSD. Significant PE was diagnosed
by single-point measurement in 46 patients in whom successful aspiration of the effusion
was performed using ultrasound guidance. On review of the measured inter-pleural distance,
the authors determined that a value of > 6.3 cm predicted an effusion volume > 1000 ml
with a sensitivity of 91.7 % and a specificity of 99.9 % [40].
Incorporating both lung and cardiac ultrasound into the initial assessment of the
acutely dyspneic patient with an HHUSD can enhance the diagnostic performance of this
point of care approach. In 68 patients presenting at an emergency department with
breathlessness, an HHUSD was used to evaluate the lung to look for B-lines and pleural
effusions, the heart to look for pericardial effusion and assess ejection fraction,
and the IVC to look for increased distention. Patients were divided into cardiogenic
and non-cardiogenic causes of dyspnea, with the final diagnosis determined by a senior
physician reviewing all investigations and patient response to treatment. Two-by-two
contingency tables were used to calculate sensitivity and specificity data for differing
combinations of the ultrasound exam compared with the final diagnosis. As with other
studies, lung ultrasound alone showed good sensitivity and specificity of 92.6 % and
80.5 %, respectively, for detecting cardiogenic edema, but the overall accuracy was
best (90 %) when this was combined with either the cardiac or IVC findings [41].
Position Comment
There are now a number of good quality studies with a reasonable number of patients,
comparing HHUSDs with high-quality echocardiograms, thereby establishing the value
of these devices in the initial assessment of both the acute and non-acute patient.
Initial assessment of LV enlargement, LV function, pericardial effusion and valve
function (to some extent) is possible, with good levels of agreement with departmental
echocardiograms. However, some limitations have been highlighted: namely the field
of view offered by these units, which can prevent reliable evaluation of the IVC.
HHUSDs can also reliably be used to determine the presence of both pulmonary edema
and pleural effusions and guide thoracentesis in the latter. This can be performed
without the need for departmental ultrasound and there is good evidence to indicate
that this can be reliably performed by clinicians with limited ultrasound training.
Handheld Devices for Pediatric Ultrasound
The use of handheld ultrasound is particularly attractive when it comes to pediatric
applications given that the body habitus of children is well-suited for ultrasound.
The low proportion of fat and lower examination depth allow for acquisition of excellent
quality images, requiring no high-end technology. Consequently, pediatric ultrasound
applications could be potentially performed using less sophisticated devices, such
as portable devices or even HHUSDs. Moreover, the reduced size of such devices may
be better tolerated by the pediatric patient, allowing for a calmer and more effective
examination. The notion of point-of-care ultrasound has been endorsed and efforts
have been made to incorporate this into pediatric practice [6]
[42]
[43]. As seen in all aspects of US, HHUSDs can be used for initial evaluation and diagnosis,
follow-up and procedural guidance in pediatric imaging.
There is limited evidence regarding HHUSDs in general pediatric applications, but
there are some reports of use for cardiac applications. The added value of an HHUSD
in a neonatal intensive care unit was illustrated in a pictorial paper and includes
the early identification of intraventricular hemorrhage, hydrocephalus and ovarian
cysts. The availability of an HHUSD allowed serial monitoring of an intraventricular
thrombus over multiple frequent examinations, demonstrating thrombus evolution to
a cystic area following central liquefaction. This interpretation would have been
problematic if the patient was only examined once in the radiology department [44]. A major part of the available literature deals with cardiac applications. A study
enrolled neonates and children < 6 years old, comparing the quality of cardiac ultrasound
images acquired with an HHUSD and a conventional device. The diagnoses in question
included various heart abnormalities: patent ductus arteriosus, atrio-ventricular
canal, peripheral pulmonary valve stenosis, aortic coarctation, atrial septal defect,
ventricular septal defect, preoperative or postoperative tetralogy of Fallot and mitral
regurgitation. It was concluded that the image quality of cardiac anatomic structures
did not differ significantly based on the readers’ evaluation between the HHUSD and
the conventional device. Only 9 % of findings, none of which were critical, were missed
by the HHUSD, showing the value of such a device, especially in remote areas with
poor access to medical care [45]. In keeping with these results, another study investigated the diagnostic accuracy
of HHUSDs handled by critical care physicians for the diagnosis of pericardial effusion,
decreased cardiac function and left ventricular enlargement in pediatric patients.
The reference method used in this study was a standard echocardiogram and the examined
population included patients aged 3 months to 20 years. It was found that HHUSDs accurately
detected effusions in more than 90 % of cases and correctly calculated left ventricular
size and systolic function in 96 % of patients [46].
Rheumatic heart disease (RHD) is another field where HHUSDs have been evaluated in
pediatric patients [47]
[48]
[49]. In a field study taking place in a deprived region and involving more than 1000
children, HHUSDs had a sensitivity of 78.9 % and a specificity of 87.2 % for the diagnosis
RHD, although the sensitivity was even higher for definite RHD. The inter- and intra-observer
agreement ranged from 66 % to 83 % and 71.4 % to 94.1 %, respectively. In light of
these results and the low cost of HHUSDs, these devices can be used as an initial
screening tool for RHD, especially in remote areas, but a standard ultrasonographic
device should be used to confirm the findings in suspicious cases [47]. Similarly, another team investigated the diagnostic accuracy of an HHUSD used by
trained nurses for diagnosing RHD with a sensitivity of more than 77 % and a specificity
of 90 %. The percentage of agreement between the different observers was 91.4 % [48]. Another study concluded that HHUSD echocardiography has a sensitivity of 90.2 %
and a specificity of 92.9 % for RHD and even higher diagnostic accuracy for definite
RHD [49]. A similar approach of screening for RHD with HHUSDs has also been tested in school
students, revealing a significant number of children with previously undetected abnormalities
[50].
HHUSDs have also been tested for the detection of patent ductus arteriosus in premature
infants. When the images were evaluated by a neonatologist after short training, the
examination had a sensitivity of 69 % and specificity of 88 %, but when interpreted
by an experienced cardiologist, the sensitivity increased to 87 %. As a result, such
applications should only be endorsed after proper training or if there are no specialists
available [51]. The same hypothesis was tested in a study where the HHUSD exams were evaluated
by pediatric cardiologists. It was found that HHUSDs could detect patent ductus arteriosus
in neonates with false-positive rates around 11 % and false-negative rates of less
than 6 %. The results were better for neonates weighing more than 1000 g and having
a gestational age > 37 weeks [52].
Another longstanding application of HHUSDs in children is US-guided urine collection
by suprapubic aspiration. Initial results have shown that suprapubic aspiration is
more frequently successful if guided by an HHUSD and fewer attempts are required [53]. One recent study showed that using an HHUSD to determine urine bladder volume prior
to urethral catheterization increased the success rate of the procedure and avoided
unsuccessful repeated attempts [54].
A case series presents 3 patients with head trauma where portable ocular ultrasound
detected increased intracranial pressure by measuring the diameter of the optic nerve
sheath [55]. Another proposed application for portable ultrasonography, but with limited evidence,
is the measurement of thyroid volume in children in an attempt to define the prevalence
of goiters and thus detect the status of iodine deficiency disorders [56].
Position Comment
There is currently limited evidence regarding the incorporation of HHUSDs in pediatrics.
However, HHUSDs have provided promising results in the detection of RHD and other
cardiac applications and are strongly recommended as an initial screening tool, especially
in remote areas or developing countries. Although there are limited reports available
regarding other applications, it is expected that HHUSDs could greatly assist minimally
invasive procedures, like vein cannulation, which are commonly performed under the
guidance of a portable ultrasonographic machine in intensive care units or emergency
departments. Given the lower quality of HHUS images compared to conventional ultrasonographic
machines, it remains to be addressed whether it is better for the examination to be
performed by an experienced radiologist or a clinician.
Handheld Devices for Medical Students
Compared with standard ultrasound devices, the lower costs and handiness make HHUSDs
an attractive training tool for medical students in the preclinical and clinical setting.
The need and demand for ultrasound training will grow with the expanded use of these
devices. The training of medical students will therefore be of outmost importance
to meet these needs. In the preclinical setting, ultrasound training will also improve
the understanding of anatomy [57] and allow a more rapid diagnosis and decision-making process during clinical training.
There are few studies reporting on the use of handheld devices for educational purposes,
with most of them focusing on intrathoracic organs. One randomized trial assessed
the efficacy of different ultrasound training programs involving the use of HHUSDs
for teaching focused cardiac ultrasound. 45 third-year medical students were allocated
to one of three educational programs: (a) lecture-based approach with scan training
by a sonographer, or (b) coupled electronic education modules with sonographer scan
training, or (c) self-directed program, combining electronic modules with scan training
on a high-fidelity ultrasound simulator. Image interpretation skills and scanning
technique were evaluated after each program. It was found that all three programs
were associated with a significant improvement in image interpretation (mean improvement
1.3-fold), but the quality scores of self-directed students were lower than those
of students taught by sonographers [58].
At Mayo Medical School, 42 first-year students participated in 3 weeks of echocardiography
training, using each other as training subjects. The aim of the program, which was
implemented in their standard anatomy curriculum, was to identify cardiovascular structures
at a parasternal long-axis plane. The rate of correct identification of anatomical
structures increased from 3.7 % pre-training to 91.0 % at the completion of the program
[59], suggesting that teaching on handheld devices is effective.
A Hong Kong study assessed the use of handheld devices by fourth-year medical students
as means for their introduction to point-of-care transthoracic echocardiography during
their 2-week anesthesiology rotation [60]. The students were allocated to groups of 8 – 9 subjects. Each student received
a booklet on basic transthoracic echocardiography before training, which included
lectures and practice. The aim was to teach identification of cardiac structures using
the parasternal long axis, parasternal short axis, apical and subcostal views. At
the end of their training, students achieved a mean success rate of 82 %.
At the University of California, Irvine, a study assessed the feasibility of point-of-care
diagnostic ultrasound training in the four-year medical school curriculum using handheld
ultrasound devices. The program was introduced to first-year students in 2010 and
involved web-based lectures, peer-to-peer training supervised by faculty members,
and performance assessments. Although this study did not provide efficacy data, it
postulates that early training of medical students with handheld devices will enhance
their understanding of anatomy and physiology, and may promote the growth of predictive,
preventative, participatory and personalized medicine [57].
A systematic review of the literature in 2017 found it possible to teach medical students
to use HHUSDs for a limited number of pathologies mainly cardiac. There was no consensus
on the protocols best-suited for the educational needs of medical students and no
data on long-term skills retention [61].
Position comment
To date, a limited number of studies have been published, indicating that the use
of handheld devices in medical education may be feasible as an adjunct in teaching
anatomy and clinical examination. However, teaching ultrasound in general to undergraduate
students is still not officially included in the standard curriculum of many medical
schools and the evidence on handheld devices comes mostly from retrospective studies.
We expect the need and demand for ultrasound training will grow with the use of handheld
devices and early training will be of outmost importance. Further investigation of
the optimal learning scenarios and implementation methods with a critical eye on study
design and educational research method is needed [62].
Education and practical training
As with any other medical training, the use of HHUSDs needs dedicated education and
practical training [63]
[64]
[65]
[66]. To our knowledge, there are currently no prospective trials investigating handheld
ultrasound training.
The importance of appropriate training in POCUS was addressed by a statement of the
American College of Radiology (ACR) which recognized point-of-care ultrasound as an
adjunct to the physical examination [67]. The ACR also pointed out that training and credentialing as well as quality assurance
are of paramount importance. There is a clear risk that without adequate training
POCUS can be harmful to patient care [67].
To address these concerns, many large medical societies request formal training in
POCUS. US is now an integral part of training in many medical specialties. Likewise,
emergency ultrasound training from the American College of Emergency Physicians (ACEP)
requires formal didactic and experiential components in POCUS training [68]. Formal training can be accomplished either by attending a residency curriculum
covering emergency ultrasound or by attending an introductory emergency ultrasound
course [68]. Residents are also required to gain practical experience in emergency ultrasound,
since the certification concluding training must follow the guidelines of the ACEP.
Similar training courses and numbers of supervised examinations are also requested
by other societies.
So far, little is known regarding the adequacy of short introductory courses and limited
examinations for achieving the necessary expertise. Most studies investigating learning
curves in POCUS suffer from methodological issues [64]. The impact of a 16-hour course including 8 hours of hands-on training on POCUS
skills was investigated in a study by Mandavia et al. [70]. During a 10-month follow-up period, they found that the 18 second-year emergency
medicine residents were able to perform POCUS studies considered to be adequate in
96.1 % of cases and the diagnosis was accurate in 94.6 % [70].
The number of required examinations to reach a level of expertise was investigated
in a retrospective study by Blehar et al. [71]. Based on 52 408 US examinations, they investigated the learning curve of 191 emergency
physicians to learn 18 predefined examination types (i. e., FAST, aorta, right upper
quadrant) [71]. The authors concluded that a range of 50 to 75 examinations sufficed to reach excellent
interpretation skills for most of the examination types [3]. This opinion was not shared by Jang et al., who prospectively investigated the
learning curve of 127 emergency physicians to reach a sufficient expertise level [72]. This study showed no major impact on the accuracy of US examinations of the right
upper quadrant after performance of the first 50 examinations [72]. The authors concluded that ”rather than simply requiring an arbitrary number of
examinations, another method of competency assessment may be necessary” [72].
One method to improve US skills could be the use of simulators. However, the data
on the optimal use of simulation-based training is still sparse [73]. A systematic review performed by Østergaard et al. could only identify 17 studies
investigating simulation-based abdominal US training [74]. However, among these 17 studies, no studies used tests with established evidence
of validity and 11 studies were identified to be at high risk for bias [69]. Based on the present literature, it was concluded that simulation appears to be
“equally good or better than no training or than existing training methods” [74]. Recently the same group published the first simulation-based test for abdominal
ultrasound skills with a pass/fail standard [75].
Position Comment
POCUS education and training is needed no matter what equipment is used. There is
no consensus on the number of cases to be performed or the time required for reaching
a safe and acceptable level. Possibly competency-based assessment and training methods
such as ultrasound simulation training and simulation-based tests may be preferred.