Abbreviations
ADD:
automated disinfection device
CRE:
carbapenem-resistant Enterobacteriaceae
CSSD:
central sterilization and supply department
EN:
European Standard
ERCP:
endoscopic retrograde cholangiopancreatography
ESGE:
European Society of Gastrointestinal Endoscopy
ESGENA:
European Society of Gastrointestinal Endoscopy Nurses and Associates
EWD :
endoscope washer-disinfector
GI:
gastrointestinal
IFU:
instructions for use (manufacturer’s)
ISO:
international standard (International Organization for Standardization)
OPA:
orthophthalaldehyde
PAA:
peracetic acid
PPE:
personal protective equipment
PTC:
percutaneous transhepatic cholangiography
RPE:
respiratory protective equipment
Definitions of terms
Automated disinfection devices (ADDs) These are intended to disinfect loads containing flexible endoscopes and their accessories
in a closed system after manual cleaning; thus their cycle includes disinfection and
rinse steps but not cleaning.
Bedside cleaning (Precleaning) Rinsing and flushing of scope channels and wiping of the outer surfaces of the endoscope
insertion tubes with dedicated detergent solution, at the examination site.
Cleaning Removal of blood, secretions, and any other contaminants and residues from endoscopes
and accessories.
Clinical service provider An organization, person, or persons legally responsible for the provision of a clinical
service. This could be an institution (such as a health service), a hospital or department,
or a doctor working in their own premises.
Detergent A compound or a mixture of compounds intended to assist cleaning of medical devices
(e. g. endoscopes).
Disinfection Reduction of microorganisms present on a product to a level previously specified
as appropriate for its intended further handling or use (EN ISO 15883).
Endoscope components Detachable/removable parts of endoscopes (valves, distal caps, balloons for echoendoscopes,
etc.).
Endoscope product family This refers to commercially available thermolabile endoscopes. Selection criteria
for the endoscope product family are based on the principal endoscope characteristics,
including the number, construction, and purpose of the different endoscope channels
and their clinical applications [1 ].
Endoscope washer-disinfector (EWD) Device intended for cleaning and disinfection of flexible thermolabile endoscopes
and their endoscope components within a closed system (according to EN ISO 15883 – 4).
Endoscopes In this Position Statement, the thermolabile flexible endoscopes used in gastroenterology.
Endoscopic accessories All devices used in conjunction with an endoscope to perform diagnosis and treatment,
excluding peripheral equipment.
Compressed air for drying Compressed air for drying purposes with the following minimum specifications:
Process chemicals All chemicals used during reprocessing procedures, including detergents, disinfectants,
etc.
Shelf-life of endoscopes Longest storage time that can safely elapse between the last reprocessing and use
on the next patient without any further reprocessing.
Sterilization Complete destruction of all microorganisms including bacterial spores; also a validated
process used to render a device free from all forms of viable microorganism (EN ISO
11139).
Storage cabinet Equipment designed to provide a controlled environment for the storage of endoscope(s)
and, if specified, drying of the endoscope including the endoscope(s) channels (EN
16442).
Type test Testing to verify conformity of washer-disinfectors or EWDs to standards, and to
establish reference data in subsequent tests (EN ISO 15883).
User Person or department using equipment; organization(s) or persons within those organization(s)
who operate and/or use the equipment.
Validation Documented procedure for obtaining, recording, and interpreting the results required
to establish that a process will consistently yield products/outcomes complying with
predetermined specifications (EN ISO 15883).
Washer-disinfector Device intended to clean and disinfect medical devices within a closed system (EN
ISO 15883); typically applying thermal disinfection methods (e. g. 90 °C).
1. Introduction, and scope of position statement
1. Introduction, and scope of position statement
Endoscopy procedures are well established in gastrointestinal (GI) endoscopy, playing
an integral part in the prevention, diagnosis, and treatment of GI diseases. Endoscopy
has significantly changed over the last 30 years, as technological developments have
established a huge variety of diagnostic and therapeutic options. The increasing number
of invasive procedures entails substantial infrastructure and specialized, trained,
and competent staff.
Flexible endoscopes are reusable sophisticated medical devices with multiple lumens
and narrow channels. Their thermolabile nature and complex design demand a specialized
approach to decontamination. Appropriate reprocessing of flexible endoscopes and endoscopic
accessories are an essential part of patient safety and quality assurance in GI endoscopy.
Since 1994, the Guideline Committee of the European Society of Gastrointestinal Endoscopy
(ESGE) and the European Society of Gastrointestinal Endoscopy Nurses and Associates
(ESGENA) has developed a number of guidelines and position statements focused on hygiene
and infection control in endoscopy [2 ]
[3 ]
[4 ]
[5 ]
[6 ]
[7 ]
[8 ].
The aims of this updated ESGE – ESGENA document are:
To set standards for the reprocessing of endoscopes and endoscopic devices prior to
each individual endoscopic procedure, whether performed in endoscopy centers, hospitals,
private clinics, ambulatory health centers, medical offices, or other areas where
flexible endoscopes are used;
To support individual endoscopy departments/healthcare providers in developing local
standards and protocols for reprocessing of endoscopic equipment;
To support national societies and official bodies in developing national recommendations
and quality assurance programs for hygiene and infection control in GI endoscopy.
This Position Statement focuses only on flexible endoscopes, endoscope components,
and endoscopic accessories used in gastrointestinal endoscopy.
It is important to follow the manufacturer’s instructions for use (IFU) at all times.
The recommendations in this Position Statement should be adapted locally to comply
with local regulations and national law.
2. Method
This ESGE-ESGENA Position Statement is based on a multidisciplinary consensus from
an expert working group, consisting of gastroenterologists, endoscopy nurses, chemists,
microbiologists, and industry representatives, with experience in developing national
and international recommendations for hygiene and infection control.
Most recommendations on reprocessing of endoscopes are based on expert opinions, in
turn based on evaluation of national guidelines available in English, German, and
French [9 ]
[10 ]
[11 ]
[12 ]
[13 ]
[14 ]
[15 ]
[16 ]
[17 ]
[18 ]
[19 ]
[20 ]
[21 ]. Recommendations are also established on the basis of microbiological studies, reviews,
or conclusions from case reports. Clinical trials in the field of endoscope decontamination
are scarce because of the reluctance to expose any control arm patients to a potential
infection risk.
A literature search was carried out that evaluated publications during the period
2008 – 2018. Based on the assessment of the literature reviews and advice from various
official national bodies, this Position Statement reflects expert opinion on what
constitutes good clinical practice [22 ]
[23 ]. The quality of evidence and strength of recommendations were not formally graded
as they were generally low [24 ].
The authors met three times during 2016 – 2018. A consensus document was agreed upon
in 2018. The manuscript was sent to all ESGE and ESGENA member societies and individual
members and to two ESGE Governing Board members for approval, resulting in this final
version, agreed by all authors.
3. Endoscopy-related infections
3. Endoscopy-related infections
Since the late 1970 s there have been sporadic reports of nosocomial infections linked
to endoscopic procedures [25 ]
[26 ]
[27 ]. The majority of documented cases were caused by noncompliance with national and
international guidelines (including inadequate reprocessing, drying, or storage of
endoscopes and endoscopic accessories). Damage, design limitations, contaminated water,
and contaminated endoscope washer-disinfectors (EWDs) were also reported [25 ]
[26 ]
[27 ].
Endoscopy-related infections are categorized as follows:
Endogenous infections from the patient’s own microbial flora;
Exogenous infections caused by inadequately reprocessed equipment. Endoscopes, endoscope
components, and reusable endoscopic accessories can be vehicles for pathogenic or
opportunistic microorganisms that are transmitted from previous patients or water.
Detailed information about endoscopy-related infections is given in Appendix 1 .
4. Classification of endoscopic equipment
4. Classification of endoscopic equipment
Noncritical: According to the Spaulding classification ([Table 1 ]) [28 ], reusable medical devices that come into contact only with the skin and mucosa are
defined as noncritical devices (e. g. mouthguards, blood pressure cuffs, finger tips,
or electrodes), and must undergo cleaning and disinfection but do not need to be sterile.
Table 1
Spaulding classification and reprocessing of medical devices.
Spaulding classification
Examples in GI endoscopy
Reprocessing
Noncritical devices
Semicritical devices
Thorough manual cleaning including brushing is mandatory, followed by: Reprocessing, including cleaning, disinfection (attaining at least a given level of
minimum bactericidal, fungicidal, mycobactericidal, and virucidal activity), and rinsing
Automated reprocessing in an EWD is strongly recommended
Thorough drying before storage in closed cabinets or storage cabinets with a drying
function
Competent staff specially trained in endoscope reprocessing (in line with national
laws and regulations) are required.
Critical devices
Endoscopic accessories, e. g. biopsy forceps, polypectomy snares, ERCP accessories,
etc.
Flexible endoscopes only if medical indication for sterilization is given
For reusable devices, validated and standardized reprocessing, preferably in a CSSD
is strongly recommended, including:
Proof of structured training for reprocessing medical devices (in line with national
laws and regulations)
GI, gastrointestinal; ECG, electrocardiogram; EWD, endoscope washer-disinfector; ERCP,
endoscopic retrograde cholangiopancreatography; CSSD, central sterilization and supply
department.
Semicritical: Most flexible endoscopes used in GI endoscopy are classified as semicritical devices,
as they come into contact with intact mucous membranes and do not ordinarily penetrate
sterile tissue [9 ]
[10 ]
[11 ]
[12 ]
[13 ]
[14 ]
[15 ]
[16 ]
[17 ]
[18 ]
[19 ]
[20 ]
[21 ]. Semicritical devices require cleaning and disinfection with bactericidal, fungicidal,
mycobactericidal, and virucidal activity.
Critical devices: Endoscopic accessories that penetrate the mucosal barrier (e. g. biopsy forceps,
guidewires, polypectomy snares, injection needles, etc.) are classified as critical
devices and must be sterile at the point of use [9 ]
[10 ]
[11 ]
[12 ]
[13 ]
[14 ]
[15 ]
[16 ]
[17 ]
[18 ]
[19 ]
[20 ]
[21 ].
Flexible endoscopes used in sterile body cavities such as laparoscopic endoscopes
should be sterile at the point of use.
Endoscopes inserted through natural orifices into sterile cavities (e. g., during
natural orifice transluminal endoscopic surgery [NOTES], peroral endoscopic myotomy
[POEM], peroral choledochoscopy) enter via naturally colonized body cavities. Endoscopes
used during percutaneous cholangioscopy enter the biliary system via a stable track
previously established during a percutaneous transhepatic cholangiography (PTC). Currently
the minimum requirement is that freshly reprocessed endoscopes should be used for
these purposes. The question of whether these endoscopes should be sterilized has
not yet been answered. National regulations should be followed.
Single-use devices should not be reprocessed at any time [6 ]
[9 ]
[10 ]
[11 ]
[12 ]
[13 ]
[14 ]
[15 ]
[16 ]
[17 ]
[18 ]
[19 ]
[20 ]
[21 ].
5. Preconditions and general issues
5. Preconditions and general issues
Patients undergoing digestive endoscopy should be examined and treated without risks
of transmission of infection or of side effects that may result from inadequate reprocessing
of endoscopes and endoscope components.
5.1 Principles of infection control
As the carrier status of patients is often unknown, all patients should be treated
as potentially infectious.
All endoscopes and reusable endoscopic accessories should be reprocessed with a uniform,
standardized reprocessing procedure following every endoscopic procedure (universal
precautions).
A traceability system should be in place to allow recall of patients in the case of
an outbreak.
The endoscopy department should be informed about the carrier status of the patient,
so any pertinent precautions can be taken.
In daily routine, patients with known infections or special risks are often scheduled
to undergo their procedure at the end of the daily patient list. However, given the
universal endoscope reprocessing regime, which presumes that all patients are potentially
infectious, it is no longer recommended that patients with known infections should
be examined only at the end of the endoscopy list. Nevertheless, infection control
policies often include this recommendation in order to make staff aware and to ensure
appropriate cleaning and disinfection of the working environment.
5.2. Health and safety aspects of endoscope reprocessing
Endoscopy staff should be protected against infectious material during the endoscopic
procedure as well as against direct contact with contaminated equipment or potentially
harmful chemicals during the reprocessing procedures.
A department-specific health and safety policy as well as appropriate equipment should
be available regarding spillages, handling of sharp instruments, chemicals, and body
fluids.
All staff involved in the reprocessing procedure should wear appropriate personal
protective equipment (PPE) including:
Chemically resistant single-use gloves (EN 374);
Protective eyewear (glasses or visors), face masks, and surgical scrub cap-type hair
covering;
Respiratory protective equipment (RPE) when handling chemicals, especially disinfectants
containing respiratory sensitizers;
Long-sleeved, moisture-resistant protection gowns (EN 14126).
Splashing should be avoided throughout the entire reprocessing procedure in order
to avoid contact with infectious material, detergents, and disinfectants.
Regular health surveillance is recommended for all staff working with potentially
sensitizing or allergy-inducing chemicals.
It is recommended that all staff should be offered appropriate vaccination against
infectious agents.
Staff known to be disease carriers should avoid duties that could transmit their disease
to patients. Treatment should be offered if applicable.
Reprocessing staff are exposed to the following health and safety hazards while reprocessing
endoscopic equipment [9 ]
[10 ]
[11 ]
[12 ]
[13 ]
[14 ]
[15 ]
[16 ]
[17 ]
[18 ]
[19 ]
[20 ]
[21 ]
[29 ]
[30 ]
[31 ]
[32 ]:
Biological hazards (direct contact with body fluids, contaminated equipment, and potentially
infectious material);
Chemical hazards (contact with process chemicals as liquids and vapors, drugs, and
potential allergens such as latex);
Ergonomic and physical hazards (e. g. working in standing and bending positions, with
risk of musculoskeletal disorders);
Risk of injuries (e. g. from needles or other sharp instruments);
Psychological hazards (e. g., noise, workload).
The implementation of health and safety policies is as mandatory for endoscopy as
it is for surgery or ambulatory care [29 ]
[30 ]
[31 ]
[32 ]. Regular health checks as well as staff protection measures are essential to ensure
a safe working environment.
General infection prevention principles are essential to maintain a safe environment
and prevent the spread of disease to patients and endoscopy personnel. The ESGE-ESGENA
statement on health and safety issues should be followed [33 ].
5.3. Staff requirements
To ensure appropriate and adequate reprocessing, the following requirements should
be considered:
Sufficient numbers of trained, dedicated, competent staff and sufficient time are
prerequisites for correct reprocessing of endoscopes and endoscopic accessories.
As the design of endoscopes varies depending on the type of endoscope and on manufacturer,
it is essential that staff are familiar with the design and construction of all equipment
used in their departments. This also includes any loan endoscopes.
Endoscopy and reprocessing staff should follow a formal officially recognized endoscopy
reprocessing training program, followed by regular practice and periodically updated
training to maintain competency.
Regular audits should be performed in order to assess compliance with guidelines and
recommendations and to identify any noncompliance or lack of competence at an early
stage. If any bad practice or lack of knowledge is identified, immediate action should
be taken (e. g. practice corrections, additional training) followed by a reassessment
of competence.
Shortage of staff increases the risk of nosocomial infections, as data from hospital
infections and from intensive care units have shown. Hugonnet et al. found that higher
staffing levels were associated with a > 30 % reduction of infection risk [34 ]. In a systematic review Erasmus et al. showed that lower compliance with hand hygiene
guidelines is associated with heavy workload [35 ]. Santos et al. evaluated hand hygiene compliance in endoscopy and showed the positive
effect of staff training in hand hygiene [36 ].
In a survey, 75 % of reprocessing staff reported on time pressure, noncompliance with
guidelines, and occupational health problems related to reprocessing [37 ]. The survey also reported on the positive effect of staff training and regular audits
to ensure compliance with guidelines.
Systematic reviews of endoscopy-related infections showed that the majority of reported
outbreaks originated from noncompliance with existing national and international guidelines
[25 ]
[26 ]
[27 ]. In a recent outbreak of multidrug-resistant Klebsiella pneumoniae related to endoscopic retrograde cholangiopancreatography (ERCP), insufficient cleaning
and drying of endoscopes were identified as the responsible factors [38 ]. Additional training followed by strict adherence to guidelines could stop any such
outbreak.
Reprocessing of endoscopes requires specialized knowledge and skills [9 ]
[10 ]
[11 ]
[12 ]
[13 ]
[14 ]
[15 ]
[16 ]
[17 ]
[18 ]
[19 ]
[20 ]
[21 ]. Formal training has been established in several European countries. ESGENA has
developed a European Curriculum for Reprocessing in GI endoscopy [39 ] based on the European job profile for endoscopy nurses [40 ].
5.4 Design of endoscope reprocessing area
Reprocessing of endoscopic equipment should only be performed in a separate purpose-designed
reprocessing room, in order to:
Minimize the risk of infection and contamination for other personnel and the general
public;
Protect from chemicals used in cleaning and disinfection procedures;
Protect from cross-contamination with potentially infectious material, blood, and
other body fluids.
The room should have:
Appropriate size and lighting, and ventilation and fume extraction in order to minimize
the risks from chemical vapors;
Appropriate technical equipment and protective measures in order to ensure safe reprocessing
following standardized and validated reprocessing procedures;
Strict spatial or at least operational separation of dirty and clean/storage areas,
in order to avoid recontamination of reprocessed endoscopes and endoscopic accessories.
This should be supported by the room architecture and design as well as by the one-way
workflow from dirty to clean areas. Ideally, the standards should comply with those
of the central sterilization and supply department (CSSD) in the particular country.
It is the responsibility of the clinical service provider to ensure that adequate
facilities for reprocessing are available.
Independently of the distance between endoscopy rooms and reprocessing area, the workflow
should ensure immediate reprocessing of used equipment.
The size and design of the reprocessing area depend on several factors. Some of these
are:
Workload (number of patients and procedures managed);
Number and types of endoscopes reprocessed in this area;
Number and types of EWDs/washer-disinfectors, storage, and/or drying cabinets.
Irrespective of the size and design of the reprocessing area, and depending on the
set-up of the reprocessing workflow, the following should be present [9 ]
[10 ]
[11 ]
[12 ]
[13 ]
[14 ]
[15 ]
[16 ]
[17 ]
[18 ]
[19 ]
[20 ]
[21 ]:
Personal protective equipment (PPE);
Separate dedicated hand-washing basins and hand-disinfection facilities in dirty and
clean working areas;
Separate sinks of adequate size for cleaning, disinfection, and rinsing, ideally height-adjustable
(even though an EWD is being used);
Protection lids at sinks and purpose-designed fume extraction facilities in order
to minimize the risks from chemical vapors;
Adequate equipment for manual cleaning steps (e. g., brushes, cleaning adapters, endoscope
leak test units);
EWD;
Appropriate storage of process chemicals;
Compressed air with suitable technical specifications, for drying;
Storage facilities for endoscopes, ideally storage cabinets with/without a drying
function;
Transport facilities between clinical areas and reprocessing, and vice versa, for
endoscopes in closed containers;
Documentation and traceability equipment.
There is a trend from one-room to two-room reprocessing concepts with separate rooms
for dirty and clean work zones and the use of “pass-through” EWDs [12 ]
[14 ].
Centralized reprocessing areas can be either located in the endoscopy units or in
the CSSD. The Dutch and British guidelines provide helpful diagrams and flowcharts
showing the design and organization of reprocessing units, adapted to the available
space and the workload [12 ]
[14 ].
The separation into dirty and clean reprocessing rooms reduces the risks of recontamination
of reprocessed equipment and reduces risks of environmental contamination. The spread
of contaminated aerosols, droplets, and dust particles can be minimized by using negative
pressure ventilation.
Standards for CSSDs are available in all European countries. As endoscopy requires
a level of safety similar to that of a CSSD, the long-term aim is to translate CSSD
standards into those of endoscope reprocessing units. These standards cover the material
used for working surfaces, sinks, and cleaning accessories, the electrical systems,
floors, walls, ceilings, doors, lighting, temperature, humidity, and ventilation [12 ]
[14 ]
[17 ].
5.5 Principles for the use of process chemicals
Process chemicals must be compatible with endoscopes and endoscope components, endoscopic
accessories, and the reprocessing equipment (e. g. EWDs).
Reprocessing should employ single-use chemicals only.
Detergents should be compatible with the applied disinfectant and any detergent residue
carried over into the disinfectant solution should not impair the microbiological
efficacy of the disinfectant.
Deposition of process chemicals should be avoided.
Process chemicals used for endoscope reprocessing are designed, tested, and manufactured
according to the European Medical Device Directive and their claimed activity has
been demonstrated [41 ]:
Detergents are class I medical device products recognized by the CE sign on the label;
Disinfectants are class IIb medical device products recognized by the CE sign plus
a four-digit number on the label.
Material compatibility tests are performed on test pieces or on complete endoscopes
using the detergent and the disinfectant alone and in combination. Manufacturers of
process chemicals, endoscopes, and EWDs should provide information about material
compatibility [41 ]. Slight cosmetic changes with no negative impact on the functionality of the endoscopes
can be accepted.
Any kind of deposition can be of concern for microbiological growth.
5.5.1 Detergents
Detergent solutions applied for manual cleaning should not be reused.
Detergent solutions with a claim of antimicrobial activity (for staff and environment
protection) can be reused, and should be freshly prepared at least on a daily basis.
The frequency of changing these detergent solutions depends on the number of reprocessed
endoscopes. However, if a solution is visibly dirty, it must be changed immediately.
Detergents containing aldehydes should not be used for the manual cleaning step, as
they denature and coagulate protein (fixation).
Detergents can be divided into two main groups (see Appendix 2 ):
Detergents containing antimicrobial active substances are used only for the bedside
and the manual cleaning steps.
5.5.2 Disinfectants
Disinfectants used for reprocessing flexible endoscopes should be tested according
to the European Standard EN 14885. The required disinfection efficacy must be:
Disinfectant activity should be demonstrated under “use” conditions in the presence
of interfering substances, according to EN ISO 15883.
The EN 14885 standard specifies the requirements for disinfection efficacy and the
test protocols that should be applied to prove the efficacy. The EN ISO 15883 standard
requires additional tests under use conditions (e. g. of temperature and time) to
demonstrate that there is no negative effect from carry-over of residues from previous
cycles (residues from the load or from the detergent).
Disinfectants containing oxidizing substances or aldehydes act by chemical reactions
with microorganisms and they are broadly efficacious against them. See more information
about disinfectants in Appendix 2 .
Alcohols, phenols, and quaternary ammonium compounds are not recommended for endoscope
disinfection as they do not show the required efficacy against all relevant microorganisms.
In the United Kingdom and France, national guidelines recommend against using aldehyde-
and alcohol-based disinfectants in endoscope reprocessing because of their protein-fixative
properties [10 ]
[15 ]
[16 ]
[42 ].
5.5.3 Rinse aid
If a rinsing aid is used to improve drying of endoscopes, its toxicological characteristics
should be assessed according to ISO 10993 – 1 (Biological Assessment of Medical Devices) as this substance remains on endoscope surfaces.
5.5.4 Combination of products from different manufacturers
Detergents and disinfectants as well as rinsing aids should only be used and combined
in compliance with the recommendations of the manufacturers of endoscopes, EWDs, and
process chemicals.
The combination of different product groups for cleaning and disinfection could cause
compatibility problems. Therefore, the manufacturers’ recommendations must be followed
at all times. Interactions can cause a change of color of endoscope surfaces and depositions
or sedimentation on surfaces of endoscopes and inside EWDs. For example, the combination
of glutaraldehyde with detergents containing antimicrobial substances based on amine
compounds may cause colored residues as a result of chemical interaction. Any kind
of deposition can be of concern regarding microbiological growth.
5.5.5 Change of process chemicals
If an endoscopy department plans to change detergents and/or disinfectants:
The user should consult the persons/department responsible for infection control and
occupational health, as well as the relevant personnel of the clinical service provider.
Manufacturers of endoscopes, EWDs, and process chemicals must provide compatibility
evidence.
Any necessity for requalification of the reprocessing procedure/EWD must be clarified.
Staff must be trained in the changed reprocessing procedure taking into account the
new products.
Prior to the use of different process chemistry, it is strongly recommended that a
requalification of the process should be performed in order to demonstrate efficacy
[7 ]. The qualification of EWD processes should be performed according to the requirements
of EN ISO 15883-4 [7 ]
[43 ]. Unauthorized use of chemical products may invalidate guarantees and/or service
contracts.
Staff training must include information about contact time, concentration of products,
and personal protection measures [39 ].
6. Reprocessing of endoscopes
6. Reprocessing of endoscopes
6.1 General considerations
Each endoscopy unit should have department-specific standard operating procedures
based on manufacturers’ IFUs.
Detailed instructions should be given for the treatment of each of the different types
of equipment (including endoscopes) used in the department.
The reprocessing staff should be aware of the risks and of the importance of each
reprocessing process step.
Department-specific protocols should periodically be updated and archived.
GI endoscopes can have a normal bacterial load of 108 – 10 (8 – 10 log10 ) [44 ]. Standardized automated reprocessing cycles lead to an 8 – 12 log10 reduction in microorganisms. Consequently, the safety margin is very low, at 0 – 2
log10 . Therefore, it is essential to adhere to the standardized protocols.
The efficacy of endoscope reprocessing depends on the reprocessing staff´s comprehensive
knowledge of the construction and function of the equipment. Hence, it is essential
to have detailed protocols describing the different steps of reprocessing necessary
for each type of endoscope. Reprocessing protocols need to be updated on a regular
basis, taking into account, for example, new equipment, technical modifications, and
updated guidelines and laws/regulations. Reprocessing staff must be informed accordingly
about such changes.
The reprocessing workflow consists of four different phases ([Fig. 1 ]):
Fig. 1 Different methods of endoscope reprocessing. EWD, endoscope washer-disinfector; ADD,
automated disinfection device.
Bedside cleaning;
Manual cleaning at the reprocessing area (including leak testing and brushing of endoscope
channels);
Cleaning and disinfection;
Drying and storage (if required).
For safe and effective reprocessing, it is essential to follow all the steps of the
reprocessing workflow in a thorough and timely manner. The clinical service provider
must document and explain any deviation from their specific reprocessing workflow.
Endoscope reprocessing should always be performed immediately after finishing the
procedure, regardless of where the endoscopic procedure is performed.
The time that elapses between manual cleaning and reprocessing in the EWD should not
exceed the time of one EWD cycle.
Cleaning is the most important step in reprocessing. It is impossible to effectively
disinfect or even sterilize an inadequately cleaned instrument.
Bedside cleaning and the manual cleaning steps with flushing and brushing of the entire
channel systems are the most important steps for the removal of debris, blood, and
body fluids. Remaining protein debris can become fixed by drying or by the use of
inappropriate chemicals. Biofilm formation is possible if the cleaning and rinsing
steps have not been carried out correctly. As some Gram-negative bacteria can undergo
cell division every 20 to 30 minutes, it is essential to complete all reprocessing
steps quickly, before bacterial growth and debris begin to dry on surfaces [45 ]
[46 ]
[47 ]. Microorganisms embedded in biofilms are 10 to 100 times more resistant to process
chemicals than planktonic (free-floating) microorganisms [46 ] and are frequently released from biofilms. Therefore it is important to follow the
IFU of the endoscope manufacturer and the national guidelines.
Some national guidelines recommend performance of all manual reprocessing steps within
30 minutes after completion of the patient examination [8 ]
[21 ]
[47 ] (see [Fig. 1 ]). If endoscope reprocessing is delayed, augmented cleaning steps may be considered.
Endoscopes that are immersed into detergent or disinfecting solutions for several
hours may be damaged.
6.2 Bedside cleaning
Bedside cleaning of the endoscope should start immediately after the endoscope has
been withdrawn from the patient, in order to:
Remove debris from external and internal surfaces;
Prevent any drying of body fluids, blood, or debris;
Reduce any build-up of bio burden or growth of biofilms;
Carry out a first check for correct functioning of the endoscope channels.
The insertion tube and critical components (e. g. the distal end of duodenoscopes
and echoendoscopes) should be wiped externally with cleaning solution, using a soft,
disposable cloth/sponge, and checked for any macroscopic damage.
Typically, air/water channels should be flushed with water from the water bottle.
It is important to consider the use of cleaning valves for the air/water channel,
according to the manufacturer’s IFU.
Before the endoscope is detached from the light source and video processor, detergent
solution should be sucked through the instrument/suction channel. European and national
guidelines recommend flushing with a volume of 200 – 250 mL or for a duration of 10 – 20
seconds as a benchmark [6 ]
[11 ]
[21 ]. Flushing must be continued until clear suction liquid demonstrates the cleanliness
of the channel system.
Additional channels should be rinsed/flushed according to the manufacturer’s IFU.
The presence of any faults, such as blockages or defects, must be communicated to
the reprocessing staff so that they can be addressed appropriately.
6.3 Transport of contaminated equipment
After completion of bedside cleaning, each precleaned endoscope and its components
and accessories should be transported in a closed container, clearly marked as contaminated
equipment, to the reprocessing room.
Such containers should be cleaned and disinfected manually using surface disinfectants
or automatically in CSSDs.
Transport in closed containers avoids contamination of the environment and third parties.
Even if several endoscopes are used during one procedure, each endoscope should be
transported in a separate container, in order to avoid any damage and to enable separation
from other equipment. In the United Kingdom, the endoscope and its valves stay together
as a traceable unique set and the valves should not be used with any other endoscope
[15 ]
[16 ].
6.4 Manual cleaning in the reprocessing area
6.4.1 Leak test
The manual leak test should be performed according to the manufacturer’s IFU, after
bedside cleaning but before starting any further cleaning steps.
The manual leak test should be performed in addition to automated leak tests in the
EWD in order to identify any damage at an early stage.
In the case of any detected leakage, the reprocessing procedure must be interrupted
immediately, and repair of the endoscope should be initiated. In such cases, the user
should clearly mark the endoscope as “Not disinfected” prior to shipment to the nearest
repair center.
Outbreaks in gastroenterological, bronchoscopic, and cardiological settings showed
that damaged parts of endoscopes may become reservoirs for microorganisms that cause
cross-contamination and severe infections [25 ]
[26 ]
[27 ]
[48 ]
[49 ]
[50 ]
[51 ]
[52 ]
[53 ]
[54 ]
[55 ]. Therefore, it is essential that the manual leak test is performed at the start
of each reprocessing cycle.
6.4.2. Equipment for manual cleaning
During manual cleaning stages, only single-use cleaning solutions, brushes and other
cleaning devices (such as sponges and cloths) should be used. This is in order to:
Ensure maximum and standardized effectiveness of cleaning;
Avoid any damage to endoscope components;
Reduce any tissue carry-over and cross-contamination.
The endoscopes should be placed into sinks of appropriate size and fully immersed
in detergent solution before brushing activities are started.
The size (length and diameter) and type of cleaning brush should appropriately match
the size and type of the endoscope channel to ensure contact with channel walls, and
access to all small/narrow lumens.
Purpose-designed brushes should be used for cleaning of critical endoscope components
(such as the elevator mechanism of duodenoscopes and echoendoscopes), according to
the manufacturer’s IFU.
Special connectors and cleaning devices should be available for each type of endoscope
used in a department. Reusable connectors should be cleaned and maintained according
to standardized reprocessing protocols and according to the manufacturer’s IFU.
Single-use brushes ensure a standardized cleaning quality as these have undamaged
bristles without any tissue remaining from previous examinations. Consequently, European
and national guidelines recommend use of single-use brushes only [6 ]
[10 ]
[15 ]
[16 ].
Damage to fragile endoscope components may be caused by damaged cleaning brushes.
Following outbreaks of carbapenem-resistant Enterobacteriaceae (CRE) infections in
the United States, reviews and surveys considered the off-label use of cleaning brushes
that may have promoted the outbreaks [56 ]
[57 ]
[58 ]
[59 ]. The outbreaks stopped when the departments changed to single-use brushes [51 ]
[53 ]. Reusable brushes may carry risks from insufficiently cleaned bristles and from
kinks that may damage internal surfaces of endoscopes. In order to avoid any cross-contamination,
reusable brushes must be reprocessed between each endoscope reprocessing.
Various types of brushes are available for different channel diameters and for special
endoscope components such as valves, ports, or distal tips. The different endoscope
channels and components should be reprocessed according to the manufacturer’s IFU.
All types of duodenoscopes require meticulous manual cleaning, since crevices behind
the elevator cannot easily be reached with conventional brushes. Manufacturers provide
purpose-designed small brushes and reprocessing recommendations, which should be incorporated
into existing department-specific reprocessing protocols [60 ]. In addition, various design improvements for endoscopes have been developed in
recent years, including single-use components for distal tips and removable elevator
mechanisms that can be autoclaved. ESGE and ESGENA [8 ], as well as national bodies and professional societies [60 ]
[61 ]
[62 ]
[63 ], have also published statements focusing on CRE infections and duodenoscope reprocessing.
All endoscopes are supplied with the appropriate cleaning adapters that ensure appropriate
access to and rinsing of all accessible endoscope channels. These cleaning adapters
should be used in manual cleaning steps according to the manufacturer’s IFU.
6.4.3 Manual cleaning steps
Thorough cleaning should cover all external surfaces, critical components (e. g. elevator
mechanism, valves) and all accessible endoscope channels, in line with the manufacturer’s
IFU.
Special attention should be given to complex endoscopes such as duodenoscopes and
echoendoscopes.
Detergent concentrations and contact times of the detergent should follow its manufacturer’s
recommendations.
Thorough manual cleaning with detergent is the most important step of the endoscope
reprocessing procedure as any debris that remains may impair the efficacy of subsequent
reprocessing steps and may support the formation of biofilms.
Cleaning steps for the endoscope include:
Full immersion of the endoscope in detergent solution.
Cleaning of all external surfaces, valve ports, channel openings, and distal tips
(including the elevator mechanism of duodenoscopes or the balloon of echoendoscopes),
using a soft disposable cloth, sponges, and/or purpose-designed brushes.
Brushing of all accessible channels using flexible, purpose-designed single-use brushes,
until there is no visible debris. The direction and order of brushing should be considered,
according to the manufacturer’s IFU.
Flushing of all lumens in order to remove organic material (blood, tissue, stool,
etc.) after brushing. Endoscope type-specific cleaning adapters must be used in order
to access all channels.
Even if they have not been used during the endoscopic procedure, all the auxiliary
water channels, wire channels, and balloon channels (in echoendoscopes and probes)
must be flushed with detergents. Because of the capillary effect, all the endoscope
channels become contaminated and partly filled with fluids/debris even when they have
not been directly used in the endoscopic procedure.
Flushing of the endoscope channels also confirms the correct functioning and patency
of the endoscope channels.
There is a clear trend toward single-use endoscope components (e.g. biopsy ports,
valves, distal caps). If these detachable endoscope components are reusable, they
must be cleaned using dedicated brushes, according to the manufacturer’s IFU.
During manual cleaning it is important to follow the detergent contact time, temperature,
and concentration as recommended by its manufacturer in order to ensure the detergent’s
efficacy. Flushing of endoscope channels can be done manually or can be supported
by automated flushing/rinsing devices.
All guidelines emphasize the thorough cleaning of endoscope channels [9 ]
[10 ]
[11 ]
[12 ]
[13 ]
[14 ]
[15 ]
[16 ]
[17 ]
[18 ]
[19 ]
[20 ]
[21 ]. French guidelines recommend double cleaning [10 ]. Multiple cleaning procedures may show positive reprocessing results [64 ]
[65 ]
[66 ]. However, it is difficult to exactly calculate the optimal number of brushing cycles,
as contamination varies greatly from patient to patient.
6.4.4 Intermediate rinsing
Fresh water (drinking water of defined quality, without any pathogens) should be used
as the rinsing solution for each endoscope.
It is recommended to use a separate rinsing sink of appropriate size in addition to
the cleaning sink.
Rinsing of external surfaces and all channels removes residual debris and detergent
to a level that avoids any critical interactions in the subsequent reprocessing phases.
Depending on the detergent used, this rinsing step may also be performed in the EWD
as a first rinse before starting the automated cleaning and disinfection cycles.
6.5 Cleaning and disinfection
6.5.1 Automated versus manual reprocessing of flexible endoscopes
EWDs compliant with EN ISO 15883 standard series should be the first choice for endoscope
cleaning and disinfection, in order to:
Provide a standardized and validated reprocessing cycle in a closed environment;
Document the process steps automatically (via a printer or electronically);
Provide reliable and reproducible reprocessing;
Minimize staff contact with chemicals and contaminated equipment;
Minimize contamination of the environment;
Facilitate the work involved for personnel;
Lower the risk of damage to endoscopes.
The process set-up in an EWD is standardized and allows automated documentation of
all critical process parameters. (See [Table 2 ] for the advantages and disadvantages of EWDs). Documentation and traceability are
important for verification of reprocessing quality and to achieve the highest possible
EWD level of safety for patients.
Table 2
Advantages and disadvantages of endoscope washer-disinfectors (EWDs).
Advantages
Disadvantages
High level of standardization in reprocessing
Low infection risk for patients and staff
Complete documentation
Full compatibility with latest European norms
Economical use of chemicals and other resources
User-friendly
Reliable
Less workload compared to full manual reprocessing
Validation of full process for increased reliability
Potentially high costs
Requires dedicated user skills/knowledge; more complex and more training required
Additional validation costs to be covered by users
Risk of infection if not regularly maintained
If EWD breaks down, endoscopy procedure may have to be cancelled
Manual reprocessing may also give reliable results, if staff perform the reprocessing
conscientiously, according to defined standard operating procedures. These procedures
should be controlled and documented in order to verify the process.
Manual reprocessing is more difficult to standardize and prone to human error and
the risk of recontamination. Moreover, staff may have increased exposure to chemicals
and infectious material.
6.5.2 Cleaning and disinfection in EWDs
EWDs compliant with the EN ISO 15883 standard series should be used for endoscope
reprocessing.
After completion of bedside and manual cleaning, endoscopes and their components should
be placed correctly in the EWD.
All endoscope channels should be connected to the EWD according to the manufacturer’s
IFU, even if they have not been used during the patient procedure.
The EN ISO 15883 standard series provides specifications and requirements for EWDs
[43 ]
[67 ]
[68 ]. This standard has enabled ESGE and ESGENA as well as European countries (Netherlands,
Germany, Austria, UK) to prepare guidelines on validation [7 ].
However, if EWDs are not maintained appropriately, they may themselves become an infection
risk by contamination of endoscopes during reprocessing [29 ]. Regular maintenance and validation of reprocessing cycles is mandatory in order
to ensure safe performance under the specifications of the EWD [7 ].
In addition to cleaning, disinfection, rinsing steps, and self-disinfection, the following
features of an EWD may be helpful:
Leak testing;
Means for providing water of the required microbiological quality;
Automatic air purging;
Drying function;
Detection of channel obstruction;
Channel non-connection testing;
Elements for providing and maintaining required temperature throughout the cleaning
and disinfection steps;
Means for documentation of cycle parameters, and identification of the endoscope and
the operator.
All users of EWDs should be trained prior to first use. Regular training updates should
be considered, and all training should be documented by the clinical service provider.
The distributor or company installing the EWD should carry out detailed training of
every user. At a minimum, the training should cover:
The EWD settings;
Correct loading and unloading of endoscopes;
Correct adaptation/use of connectors;
User troubleshooting activities required in case of errors;
EWD maintenance (relevant for daily, weekly, or monthly checks).
Manual reprocessing procedures should be in place in case of malfunctioning or defects.
Staff must be trained in manual reprocessing procedures. Additional access to EWDs
in neighboring units may also be an option, provided that access and compatibility
has been proven.
6.5.3 Disinfection in automated disinfection devices (ADDs)
Wherever possible, EWDs complying with EN ISO 15883 standard series, should be used.
If ADDs are used, they should at least comply with the relevant parts of the EN ISO
15883-4 standard.
The automated disinfection process does not usually have an integrated automated cleaning
stage. ADDs are intended to disinfect flexible endoscopes in a closed system after
complete and careful manual cleaning.
Some ADDs offer:
See [Table 3 ] for the advantages and disadvantages of ADDs.
Table 3
Advantages and disadvantages of automated disinfection devices (ADDs).
Advantages
Disadvantages
Greater workload compared with EWDs
No European standard available for design, type testing, performance requirements,
and validation
In the case of reuse of disinfectant, effective concentration must be confirmed, if
applicable
Increased workload of routine testing (i. e., disinfectant efficacy testing)
Traceability and documentation activities are more time-consuming
More complex; more training required
Risk of infection if not regularly maintained
EWDs, endoscope washer-disinfectors.
6.5.4 Manual endoscope disinfection
Wherever possible, EWDs that comply with EN ISO 15883 standard series should be used.
If this is not possible manual reprocessing should be performed based on standard
operating procedures.
ESGE and ESGENA are aware of the varying economic situations in different countries.
Nevertheless, hygiene standards for patient and staff safety should have the highest
priority. National guidelines emphasize the preference for automated reprocessing
after manual cleaning [9 ]
[10 ]
[11 ]
[12 ]
[13 ]
[14 ]
[15 ]
[16 ]
[17 ]
[18 ]
[19 ]
[20 ]
[21 ]. However, the British and Dutch guidelines clearly state that manual disinfection
is no longer acceptable, except in the case of technical problems with the EWD [12 ]
[13 ]
[14 ]
[15 ]
[16 ].
Ultimately, it is the responsibility of the clinical service provider to choose an
effective reprocessing method in line with national laws and regulations.
See [Table 4 ] for the advantages and disadvantages of manual disinfection.
Table 4
Advantages and disadvantages of manual endoscope disinfection.
Advantages
Disadvantages
Easy to establish
Moderate investments
No standards/guideline available for validation of manual disinfection
Validation difficult, however, standardization for all reprocessing steps is possible
Increased risk of human error (inconsistencies, mistakes, etc)
Staff exposure to process chemicals and potentially infectious material; additional
precautionary measures necessary
Increased workload, because staff involved in each reprocessing step
In the case of reuse of disinfectant, efficacy problems to be considered
Traceability and documentation activities are more time-consuming and more difficult
Increased risk of recontamination, followed by increased risk of infections for patients
Increased risk of health problems for staff (infection, injuries, allergies, etc.)
In the case of manual disinfection:
A sufficient number of sinks of suitable size for endoscope reprocessing should be
available.
All cleaning steps should be performed prior to disinfection.
An intermediate rinsing step is necessary between cleaning and disinfection.
For disinfection, the endoscope should be immersed completely, and all channels should
be filled completely with disinfectant.
The manufacturer’s recommendations regarding correct concentration, temperature, contact
time, and number of reuse cycles (if applicable) should be followed, and this compliance
should be documented, in order to ensure adequate disinfection.
If the disinfectant is a concentrated product, it must be diluted, in the correct
dilution ratio, with filtered water or drinking water of defined quality. Freshly
prepared disinfecting solution provides the largest safety margin. Use of the disinfecting
solution for a longer period risks lowering the concentration by, for example:
Decomposition of the active substance:
Adsorption of active substance onto surfaces;
Inactivation of the active substance by reaction with protein;
Dilution of the disinfecting solution by rinse water remaining in the endoscope from
the previous reprocessing step.
6.6 Final rinsing
Disinfectant solution should be rinsed from the internal and external surfaces of
the endoscope with sterile filtered water. National requirements regarding water quality
should be followed.
Rinsing water should not be reused at any time.
Rinse water quality is an important issue. It must be at least of drinking water quality
and should be free of pathogens such as Pseudomonas aeruginosa . Preferably, sterile filtered water may be used for rinsing.
Insufficient rinsing can cause severe damage to patients. Disinfectant residues on
endoscope surfaces can cause severe complications such as colitis, abdominal cramps,
and bloody diarrhea. This occurs mainly after manual reprocessing procedures [69 ]
[70 ]
[71 ]
[72 ]
[73 ].
Up to 50 mL of solution can remain in an endoscope (depending on endoscope type) if
not removed by compressed air.
6.7 Drying of endoscopes
The endoscope and its components should be dried after completion of the cleaning
and disinfection process. The required intensity of drying depends heavily on the
intended further use of the endoscope:
If the endoscope is to be used for the next patient examination within a short period
of time, removal only of major water residues from the endoscope channels and outer
surfaces will be sufficient.
If the endoscope is not to be reused immediately and is to be stored, the endoscope
channels and outer surfaces should be dried thoroughly, in order to avoid any microorganism
growth leading to recontamination.
If the endoscope is used directly after reprocessing, it must be placed in a clean
and covered transport tray.
Thorough drying of endoscope surfaces and channels is necessary to prevent any growth
of waterborne microorganisms [27 ]
[74 ]
[75 ]
[76 ]. Several outbreaks of P. aeruginosa , K. pneumoniae , Acinetobacter spp. and other pathogens have been caused by insufficient drying [25 ]
[26 ]
[27 ]
[74 ]
[75 ]
[76 ]. Furthermore, biofilms and embedded microorganisms need moisture for survival [27 ]
[45 ]
[46 ].
Endoscope valves can also show contamination after reprocessing and may be the source
of infections if cleaning, drying, storage, and hand hygiene are inadequate [56 ]. There is an increasing trend for using detachable endoscope components as single-use
products to enable full traceability and to prevent cross-infection caused by inadequately
reprocessed detachable components such as valves and distal caps [14 ]
[16 ].
All external parts and all endoscope channels must be dried carefully with compressed
air specially provided for drying [11 ]
[16 ]
[21 ]
[76 ].
Manual drying processes can be avoided by using EWDs with a dedicated endoscope drying
function or by use of specialized endoscope storage and/or drying cabinets that comply
with EN 16442 standard.
Flushing of endoscope channels with alcohol for drying purposes is not recommended.
In various countries the use of alcohol is banned, because of potential protein fixation
risks [10 ]
[12 ]
[14 ]
[15 ]
[42 ]. There is no clear evidence that flushing with alcohol is effective in either drying
of endoscopes or in preventing the proliferation of waterborne bacteria [11 ]
[18 ].
However, attitudes to the use of alcohol for drying endoscope channels are quite diverse
[75 ]. Some guidelines still recommended flushing with 70 % – 90 % ethanol or isopropyl
alcohol to facilitate the drying of endoscope channels [20 ]. Updated national guidelines consequently recommend the use of drying cabinets [10 ]
[12 ]
[15 ]
[16 ].
6.8. Storage of endoscopes
Endoscopes should be stored:
Vertically in well-ventilated, closed cupboards; or
In purpose-designed storage cabinets with/without a drying function.
During storage, endoscope components such as valves and distal caps should be disconnected
from the endoscope. Whenever possible, endoscope components should stay with the named
endoscope as a set, to enable full traceability and to prevent cross-infection.
Endoscopes should never be stored wet or before decontamination has been completed
as such storage supports the growth of microorganisms and biofilms.
Outbreaks connected to insufficient drying and storage were mainly reported when instructions
for drying had not been followed [27 ]
[38 ]
[74 ]. Storage in a controlled environment is aimed at preventing any secondary contamination.
For storage of endoscopes, suitable and well-ventilated locations, generally for vertical
placement, should be selected [9 ]
[11 ]
[17 ]
[18 ]
[19 ]
[20 ]
[21 ]. If nonvertical storage is chosen, special attention will be needed to ensure that
no residual moisture will cause recontamination of the endoscope. Updated national
guidelines consequently recommend the use of drying cabinets [10 ]
[12 ]
[15 ]
[16 ].
For reasons of traceability and prevention of cross-infection, endoscope components
such valves or detachable distal caps stay with the endoscope, but are disconnected
in order to avoid any air blockage/any moist chamber in the endoscope channels. There
is a clear trend toward single use of these components [12 ]
[14 ]
[16 ].
6.8.1 Storage cabinets with/without a drying function
In the storage cabinets:
Only fully cleaned and disinfected endoscopes should be stored.
All endoscope channels should be connected using purpose-designed adapters for air
ventilation purposes.
Endoscope components (such as valves) should also be stored and dried with the endoscope
that they have been used with.
If storage in cabinets with/without a drying function is used:
Maximum storage duration should be consistent with the manufacturer’s IFU of the cabinet
and in line with local regulations.
Regular maintenance should be performed.
Routine microbiological surveillance should be done when the maximum storage time
specified by the manufacturer has elapsed.
The European standard for endoscope storage cabinets (EN 16442) sets minimum product
specifications and deals with all aspects of product type testing and performance
qualification. EN 16442 specifies how storage cabinets must be designed in order to
achieve a controlled environment, and to prevent recontamination risks [77 ].
A number of national guidelines recommend the use of storage cabinets [10 ]
[12 ]
[14 ]
[16 ]
[64 ].
The main performance requirements are [77 ]:
Cabinets must be able to at least keep the microbiological quality of cleaned and
disinfected endoscopes unchanged during storage.
The quality of air inside the cabinet must be specified.
The maximum storage period for endoscopes must be determined.
Cabinets without an endoscope drying function must have instructions for the user
on how to dry endoscopes prior to placement in the cabinets.
If drying is part of the cabinet function, maximum drying times must be specified.
The cabinets must be provided with suitable connectors for all compatible endoscopes.
The connectors must assure sufficient airflow though all channels of compatible endoscopes.
6.8.2 Shelf-life of reprocessed endoscopes/Expiration of storage
Local policies should be in place regarding the shelf-life of endoscopes, as the recommended
shelf-life of endoscopes depends on the storage conditions, national guidelines, and
the manufacturer’s IFU for storage cabinets that comply with EN 16442.
The storage time of reprocessed endoscopes (shelf-life) has been the subject of debate
and differing interpretations in many countries. If endoscopes are stored vertically
in closed cabinets, British, Dutch and French guidelines define a time limit up till
when the endoscope may be used. This time limit differs between 3 to 12 hours [10 ]
[12 ]
[15 ]
[16 ]. If this time limit is exceeded, the whole reprocessing cycle must be repeated.
Studies with small numbers have shown contamination after 5 – 7 days, and up to 14
days, identifying mainly common skin organisms rather than significant pathogens [78 ]
[79 ]
[80 ]
[81 ]. The American multisociety guidelines and the German guidelines rated the data as
not significant enough to define any maximum shelf-life [11 ]
[20 ]
[82 ]. They emphasize that the shelf-life depends on the microbiological quality of the
final rinse inside the EWD, the effectiveness of drying, and possibly the risk of
recontamination.
In a systematic review Schmelzer et al. concluded that appropriately disinfected endoscopes
can be stored for up to 7 days, if regular microbiological surveillance confirms the
effectiveness of reprocessing [83 ].
Manufacturers of storage cabinets compliant with EN 16442 specify, based on type test
results [77 ]:
6.9. Routine inspection
Visual inspections of reprocessed endoscopes should be performed after each reprocessing
cycle and/or before each patient use in order to identify small cracks and wear and
tear and to detect any remaining debris.
Routine maintenance programs offered by manufacturers must be followed.
Recent outbreaks related to ERCP suggest that it may be difficult to be detect microlesions
by routine leak tests [48 ]
[49 ]
[50 ]
[51 ]
[52 ]. Therefore, an additional inspection, for example with magnifying glasses, may be
helpful to identify cracks and wear and tear.
This is recommended especially for complex and fragile components such as the elevator
mechanism or glass lenses.
In order to prevent the consolidation of microlesions, manufacturers offer routine
maintenance and exchange of components that are exposed to increased mechanical stress
and wear and tear.
6.10 Sterilization of endoscopes
Only if medical indications show that sterilization of flexible endoscopes may be
appropriate, a low temperature sterilization process can be applied.
Because of their material and design restrictions, most flexible endoscopes are not
temperature-resistant. Therefore, steam sterilization processes at elevated temperatures
cannot be applied for sterilization of flexible endoscopes. The following alternative
low temperature processes are available:
Ethylene oxide gas sterilization;
Hydrogen peroxide gas sterilization with and without plasma;
Low temperature steam and formaldehyde sterilization.
It must be recognized that low temperature sterilization processes are only effective
if thorough cleaning has already been done. Manual reprocessing and use of an EWD
before sending the endoscope to a CSSD for sterilization will be important in order
to protect reprocessing staff.
Most European countries do not accept immersion of endoscopes into liquid chemical
sterilants, because the devices are not wrapped in sterile packages until the next
use. A critical point is also the quality of the final rinse water which might impair
the sterilization effect.
At present the hydrogen peroxide gas sterilization used on some GI endoscopes has
technical limitations. Gastroscopes, colonoscopies, and duodenoscopes have from three
to seven long separate channels and therefore exceed the lumen capacity of existing
sterilizers.
Further development and research will be needed.
6.11 Transport of ready-to-use reprocessed endoscopes
Hand disinfection should be done before reprocessed endoscopes are handled.
Reprocessed endoscopes should be transported in a disinfected closed container, clearly
marked as “clean equipment ready for use.”
Endoscope components should also be transported in this closed container.
Transport in closed containers reduces the risk of recontamination and prevents any
damage to the endoscope during the transportation phase [12 ]
[15 ].
Hand hygiene compliance in endoscopy is a crucial point [34 ]. Reprocessed endoscopes can be recontaminated if hand hygiene is insufficient.
If several endoscopes are used during one procedure, each endoscope should be transported
in a separate container to avoid any damage.
7. Documentation and traceability
7. Documentation and traceability
7.1 Documentation
The complete reprocessing cycle should be documented:
Each reprocessing step (including bedside cleaning, manual cleaning, and automated
reprocessing in an EWD or ADD) should be recorded manually or electronically, including
the names of the persons undertaking each step.
The process parameters of the EWD and storage cabinets should be documented by printouts
or electronically.
All endoscopes should have a record of their reprocessing showing that they are ready
for use on patients.
The reprocessing record should be documented in the patient’s files.
The documentation of the reprocessing procedure should include [12 ]
[15 ]
[16 ]
[19 ]:
The patient on whom the endoscope was last used;
The endoscope identification;
The whole reprocessing cycle including all manual cleaning steps, and identification
of the EWD/ADD and storage cabinet used (if applicable);
The time-frame for reprocessing and storage (see section 6.1.)
Identification of the staff member involved in reprocessing of that endoscope;
Identification of the staff who check the correct performance of the reprocessing
cycle and release the endoscope for use on the next patient.
Quality assurance entails that the evidence of correct reprocessing is included in
the file of the next patient. Therefore an interface between electronic documentation
of medical endoscopy reports and reprocessing is essential to allow data transfer.
In cases of suspicious infection this data exchange is a necessary tool for investigating
nosocomial infections.
7.2 Maintenance
Regular maintenance of all technical equipment, including endoscopes, EWDs, and storage
cabinets, should be defined according to the manufacturer’s IFU.
It is the responsibility of the clinical service provider to contact the relevant
manufacturer as well as regulatory bodies if:
The manufacturer’s recommendations are unclear;
Any problems arise while using or reprocessing their equipment;
Suspicious infections occur in conjunction with a specific device (e. g. endoscope,
EWD, ADD, storage cabinet, sterilization device).
In the case of technical problems, endoscopes, EWD, ADDs, storage cabinets, or sterilization
devices may pose a potential infection risk. Therefore they must:
Be cleaned/disinfected and maintained according to manufacturer’s IFU on a daily basis;
Have regular engineering maintenance;
Undergo regular microbiological surveillance according to EN ISO 15883, and for storage
cabinets according to EN 16442.
7.3. Loan endoscopes and prototypes
Prior to first use on patients, loan endoscopes and prototypes should be reprocessed,
following the whole reprocessing cycle including manual brushing, and should be checked
for correct functioning.
If a loan endoscope and prototypes differ from endoscopes usually used in the department,
endoscopy and reprocessing staff should receive instructions from the supplier about
the equipment, including the channel configuration and reprocessing information.
The clinical service provider must check whether this type of endoscope can be reprocessed
in the local EWD.
The specifications of each loan endoscope and prototypes should be included in the
database of the endoscopy department as well as in the database for the local EWDs
and storage cabinets (if applicable) in order to enable appropriate documentation.
Staff must be familiar with the channel configuration of loan endoscopes and prototypes
in order to ensure safe reprocessing. Because of legal issues, it is necessary to
document the use of loan endoscopes and prototypes in all relevant patient- and hygiene-focused
documentation systems.
The clinical service provider must check whether the type of the loan endoscopes and
prototypes can be reprocessed in the local EWD [8 ]. If necessary, the clinical service provider should contact the EWD/ADD supplier
to receive information about compatibility with the process chemicals, and about necessary
connectors for the EWD, ADD, and storage cabinets in order to ensure safe reprocessing.
8. Outbreak management
The clinical service provider should establish procedures detailing the management
of any suspicious infection as well as suspected or identified breaches in reprocessing.
The procedure should indicate the management of the potentially affected patients,
staff, and equipment.
If any contamination is found, it is the responsibility of the clinical service provider
to take the suspected piece of equipment out of service (e. g. endoscopes, EWD, ADD,
storage cabinet, accessories, etc), until corrective actions have been taken and satisfactory
results have been achieved.
Outbreaks should be managed within the multidisciplinary team of endoscopy departments,
hospital hygiene experts, microbiologists, manufacturers, and regulatory bodies, if
applicable.
Staff training, adherence to guidelines and manufacturers’ IFUs, regular quality assessment
with audits, regular microbiological surveillance, and validation of reprocessing
cycles are important tools in the prevention of infections. European and national
guidelines already provide helpful flowcharts concerning outbreak management [6 ]
[9 ]
[12 ].
9. Reprocessing of endoscopic accessories
9. Reprocessing of endoscopic accessories
9.1 General recommendations
The employment of single-use endoscopic accessories whenever possible is strongly
recommended.
Endoscopic accessories defined as single-use devices should be discarded directly
after use.
Endoscopic accessories defined as reusable critical devices [28 ] should be reprocessed immediately after use by standardized and validated reprocessing
procedures based on the manufacturer’s IFU (EN ISO 17664).
Heat-stable endoscopic accessories should be reprocessed in washer-disinfectors employing
thermal disinfection.
Reusable endoscopic accessories defined as critical devices should undergo sterilization
processes prior to reuse.
Endoscopic accessories are available as reusable and single-use devices. Disposable
medical products are available in many cases. The trend to single-use devices is increasing
in many western European countries, as their use:
Prevents cross-infection both to patients and staff;
Prevents potential staff injuries during cleaning steps;
Ensures a fully functioning accessory each time.
As most reusable accessories are thermostable devices, this equipment should be reprocessed
at a CSSD having appropriate facilities for automated reprocessing of such devices.
Reusable endoscopic accessories require the same level of safety as surgical equipment;
and automated reprocessing has a number of advantages.
Some EWDs offer separate programs for reprocessing thermostable equipment. But reprocessing
of heat-labile endoscopes and other medical devices that are heat-stable should not
be mixed as the devices require different reprocessing conditions.
9.2 Water bottles and their connectors
Water bottles and their connectors should be changed and filled exclusively with sterile
water for each endoscopy session.
Reusable water bottles should be cleaned and sterilized according to the manufacturer’s
IFU at least on a daily basis.
Water bottles should be included in regular microbiological surveillance.
Water bottles can be a source of endoscope contamination. This can be caused by using
tap water instead of sterile water, by inadequate cleaning, and by lack of sterilization
[84 ]. Therefore, water bottles and connecting tubes must be cleaned and sterilized on
a daily basis. Water bottles should be filled with sterile water. It is not recommended
to add any other solutions to water bottles, such as simethicone, as this might leave
residues in small lumina [85 ]
[86 ]. If simethicone is used, if should be applied directly via the instrument channel
[86 ]. Additionally, testing of water bottles should be part of regular quality control
[6 ]
[84 ].
9.2 Reprocessing cycle for endoscopic accessories
9.2.1 Manual cleaning
After manual cleaning with dismantling and brushing, endoscopic accessories should
be cleaned in an ultrasonic cleaner. Further reprocessing can be performed:
Manual cleaning of reusable endoscope accessories is most important. They should be
cleaned manually immediately after use to prevent any body fluids or debris drying
on the instruments. Prolonged delay before cleaning might lead to ineffective reprocessing
or malfunction of the accessory.
Manual cleaning should consist of:
Dismantling of accessories as far as possible (follow manufacturers’ recommendations);
Cleaning of external surfaces using a soft, disposable cloth/sponge and brushes;
Thorough brushing of complex devices;
Flushing all available channel lumens;
Ultrasonic cleaning;
Rinsing.
Ultrasonic cleaning is essential for the removal of debris from inaccessible spaces
of complex accessories. The tray of the ultrasonic cleaner should not be overloaded,
in order to avoid ultrasound “shadows”/dead space. Ultrasonic cleaning must be performed
before any disinfection and/or sterilization. The use of an ultrasonic cleaning device,
dedicated for medical applications, offering a frequency range over 30 kHz (38 to
47 kHz) and a maximum operating temperature of 40 °C, is recommended.
If the accessories will be forwarded to sterilization immediately, without any automated
cleaning and disinfection under thermal process conditions, accurate and thorough
manual cleaning is even more important.
The water quality available in the endoscopy unit should be specified.
9.2.2 Optional reprocessing in washer-disinfectors
Thermal disinfection programs are recommended for the reprocessing of accessories
(EN ISO 15883).
Some washer-disinfectors for flexible endoscopes offer special programs for heat-stable
accessories. Washer-disinfectors for surgical instruments that comply with EN ISO
15883-2 also offer loading systems and programs for heat-stable accessories.
9.2.3 Sterilization
After thorough rinsing and drying, endoscopic accessories should be packed according
to the EN 868 standard and sterilized according to European sterilization standards
(e. g. EN 285) and local regulations.
9.2.4 Storage
Endoscopic accessories should be stored in a closed cupboard. Before use the sterile
package must be checked for any damage and for expiry date (EN 868).
Appendix 1: Endoscopy-related infections
Appendix 1: Endoscopy-related infections
Microorganisms may be spread by inadequately reprocessed equipment from one patient
to another, or from patients to staff members [25 ]
[27 ]. There are a number of weaknesses and potential deficiencies in periendoscopy patient
care and endoscope reprocessing. These include human error and technical features
that can be sources of microbial contamination and transmission of infectious material
([Table 5 ]).
Table 5
Potential weaknesses and deficiencies in periendoscopic patient care and endoscope
reprocessing that may facilitate endoscopy-related infections.
Area
Details
Personnel factors
Lack of knowledge, experience, training, and awareness concerning endoscope reprocessing
and infection control
Insufficient hygiene in periendoscopic patient care and in reprocessing
Inappropriate management of intravenous medication (e. g., contaminated and time-expired
syringes, tubes, or medication)
Design limitations and damage regarding endoscopes and their components
Inadequate reprocessing of endoscopes and accessories
Contaminated or defective endoscope, EWD, or ADD
Contaminated water used in the endoscopy unit
Inappropriate cleaning (e. g., insufficient brushing of endoscope channels, distal
ends, elevator systems, valve ports)
Contaminated pipes, containers, final rinsing water, filters, dosing system, etc.
Inadequate drying, transport and storage of endoscopes
EWD, endoscope washer-disinfector; ADD, automated disinfection device.
Bacterial infections have been acquired during endoscopy, caused for example by Salmonella spp., [87 ], Helicobacter pylori [88 ]
[89 ] and Pseudomonas spp. [76 ].
Sources of Gram-negative bacteria such as P. aeruginosa and K. pneumoniae are not restricted to patients’ colonized bowels. The bacterium may originate from
the environment. Examples include water bottles, EWDs, and duodenoscopes. Moreover,
inadequate cleaning, disinfection, and drying of the elevator wire channel of duodenoscopes
may result in ERCP-related infections [90 ],
Infections by multidrug-resistant organisms have become increasingly problematic for health care systems worldwide. Since 2010
severe nosocomial infections due to multidrug-resistant organisms have also been linked
to ERCP [38 ]
[48 ]
[49 ]
[50 ]
[51 ]
[52 ]. Multidrug-resistant Enterobacteriaceae including K. pneumoniae , E. coli and Enterobacter spp. as well as P. aeruginosa were found in duodenoscopes, especially at the distal end and around the forceps
elevator mechanism. Small cracks as well as wear and tear, which required maintenance
and repairs despite lack of obvious malfunction, were observed in a number of endoscopes,
again especially at the distal end and around the forceps elevator mechanisms [48 ]
[49 ]
[50 ]
[51 ]
[52 ]. These small defects became the reservoir for debris and microorganisms. In some
cases the outbreaks happened despite the use of apparently appropriate reprocessing
protocols [48 ]
[49 ]
[50 ]
[51 ]
[52 ]. In other cases insufficient cleaning and drying permitted the outbreak [38 ]. Insufficient hand hygiene was also identified as a factor which facilitated transmission
from one patient to another [51 ]. European, American, and Australian official bodies as well as professional societies
have published statements to raise awareness among health care professionals that
the complex design, especially of duodenoscopes, may impede effective reprocessing.
They have initiated appropriate actions and published recommendations to improve endoscope
reprocessing [8 ]
[61 ]
[62 ]
[63 ]
[64 ]. Reviews and editorials have discussed these outbreaks, but have not reached in
any consensus regarding causation [57 ]
[58 ]
[59 ]
[60 ].
Virus infections. Only three cases of hepatitis B virus transmission from inadequately disinfected
endoscopes have been reported. Cases of hepatitis C virus transmission have been related
to inadequate cleaning and disinfection of endoscopes and accessories and to the use
of contaminated anesthetic vials or syringes [90 ]
[91 ]
[92 ], Neither the inadequate reprocessing nor the reuse of anesthetic vials or syringes
could definitely be identified as the actual cause of the infection. Hepatitis B and
C transmission have not been associated with endoscopy when appropriate disinfection
procedures have been performed [93 ].
No cases of human immunodeficiency virus (HIV) transmission attributed to endoscopy
have been reported so far [25 ]
[27 ].
Patients with immune deficiency syndrome or severe neutropenia, those undergoing immunosuppressive chemotherapy, and those
having artificial cardiac valves have an increased risk of infection. Therefore, therapeutic
procedures carry a higher risk of infection. Patients harboring clinically latent
infections (e.g. hepatitis, tuberculosis, salmonellosis, infections caused by H. pylori or HIV) may not be aware of their carrier status, and therefore, all patients should
be considered potentially infective.
Additionally, fungi can be transmitted via endoscopic procedures [94 ]
[95 ]
[96 ],
Mycobacterial infection is becoming more common. The emergence of multidrug-resistant strains of Mycobacterium tuberculosis and the high incidence of infections with M. avium intracellulare among HIV-infected patients has led to a greater awareness of the risk of transmission
of mycobacteria during bronchoscopy. Most reports on mycobacterial outbreaks describe
colonization from the endoscope in the absence of infection in immunocompromised patients
with a history of lung cancer, HIV, AIDS, or hematological malignancies [27 ]. Mycobacteria in general, and especially some waterborne mycobacteria (such as M. chelonae ), show resistance to glutaraldehyde and may contaminate EWDs [15 ]
[97 ].
Clostridium difficile
infection is a growing problem in health care facilities. To date endoscopy has not
been considered to be a risk factor for C. difficile transmission [98 ]
[99 ].
Creutzfeldt–Jakob disease (CJD) and variant-CJD are transmitted by infectious agents called prions (protein particles without nucleic
acid) and are extremely resistant to standard reprocessing procedures. In classical
CJD prion proteins are concentrated in the central nervous system, but in variant-CJD
prion proteins accumulate in lymphoid tissue, including in the GI tract [16 ]. Endoscopic transmission of variant-CJD remains theoretically possible, but no reports
of such transmission have been published [16 ]
[27 ].
Appendix 2: Process chemicals
Appendix 2: Process chemicals
A1. Detergents
Detergents can be divided in two main groups:
Detergents containing antimicrobial active substances are used only for the bedside
cleaning and the manual cleaning step.
A1.1 pH-Neutral detergents with or without enzymatic boosters
pH-Neutral detergents are widely used because of their excellent compatibility with
materials. They are available with or without enzymatic boosters. Detergents with
enzymatic boosters contain one or more different types of enzymes, for example protease,
amylase, or lipase. Enzymes are proteins with biological activity. Protease breaks
protein debris into smaller subunits that are more soluble. Amylase catalyzes the
breakdown of starch and lipase breaks up fat-containing debris. These types of detergent
require a specific contact time as recommended by the manufacturer. When enzymatic
detergents are used for ultrasonic cleaning on endoscope accessories the container
must be covered tightly to prevent the inhalation of enzyme-containing aerosols.
A1.2 Detergents with an alkaline booster
Detergents with alkaline boosters contain alkaline chemical substances forming a mild
alkaline cleaner. Alkaline substances lift off soil and help to dissolve it in the
cleaning solution. Strong alkaline cleaners in the pH range > 11 are not recommended
for flexible endoscope cleaning because of possible incompatibility with the materials
of some endoscope parts.
A1.3 Detergents with alkaline and enzymatic booster
Detergents with alkaline and enzymatic boosters combine the properties of enzymatic
and alkaline detergents.
A1.4 Detergents containing antimicrobial active substances
In some European countries detergents containing antimicrobial active substances are
commonly used and recommended by health authorities for the bedside cleaning and the
manual cleaning steps. The application of this product type may reduce the infection
risk to reprocessing personnel. The efficacy of these antimicrobial active detergents
should be assessed according to the European standard EN 14885. Tests demonstrating
disinfection efficacy should be performed under dirty conditions. For minimum efficacy
there should be bactericidal and yeasticidal activity and activity against enveloped
viruses.
Commonly used active substances in this type of detergent are, for example, amine
compounds, peracetic acid and its salts, and quaternary ammonium compounds.
The use of pH-optimized peracetic acid in detergents is currently under discussion
because of the potential fixation of proteins on surfaces. A laboratory study has
shown fixation of fibrin (a polymer protein molecule) to stainless steel surfaces
[100 ]. On the other hand, another laboratory study showed no findings related to fixation
of proteins on polymer surfaces [101 ]. Furthermore, no residue formation by fixation of proteins on endoscope surfaces
was observed in a field study investigating used endoscopes disinfected by pH-optimized
peracetic acid under practical endoscope reprocessing conditions [102 ].
Detergents based on pH-optimized peracetic acid have the advantage of being effective
against bacterial spores, including C. difficile , under clinical use conditions [103 ].
The application of detergents with antimicrobial active substances does not replace
the disinfection step.
A2. Disinfectants
Active substances, such as oxidizing substances and aldehydes, act against microorganisms
by means of chemical reactions. These groups of disinfecting substances show the required
broad efficacy against microorganisms. Examples of aldehydes include glutaraldehyde
and orthophthalaldehyde and examples of oxidizing substances include hypochlorous
acid, chlorine dioxide, and peracetic acid and its salts.
In the United Kingdom and France national guidelines recommend against the use of
aldehyde- and alcohol-based disinfectants in endoscope reprocessing because of their
fixative properties [10 ]
[15 ]
[42 ].
Non-active substances such as alcohols, phenols, and quaternary ammonium compounds
are not recommended for endoscope disinfection as they do not show the required efficacy
against microorganisms.
A2.1 Glutaraldehyde
Disinfectants based on glutaraldehyde are available as concentrated or as ready-to-use
products. They can be used manually, in ADDs, and in EWDs.
Ready-to-use glutaraldehyde solutions range in concentration from 2.4 % to 2.6 % and
have variable maximum use lives. Accurate monitoring of the glutaraldehyde concentration
is required, as lower concentrations do not guarantee efficacy. The required duration
of immersion to cover the full range of microorganisms is variable depending on the
product and should be determined according to EN 14885 or local standards.
The dilution ratio of concentrated glutaraldehyde-based disinfectants depends on their
composition and on the detected concentration/duration relationship as tested according
to EN 14885 or local standards. Concentrated products based on glutaraldehyde are
often used in combination with other aldehydes such as glyoxal and succinic aldehyde
or with other active substances such as quaternary ammonium compounds. Equivalent
microbiological efficacy is achieved with a reduced concentration of glutaraldehyde
in the application solution.
Glutaraldehyde has the advantages that it is effective, relatively inexpensive, and
does not damage endoscopes, accessories, or processing equipment.
However, there are a number of disadvantages, both for clinical staff and patients,
in using glutaraldehyde. It is an irritant and has sensitizing properties. It can
cause allergic reactions such as dermatitis, conjunctivitis, nasal and throat irritation,
and occupational asthma [16 ]
[34 ]
[104 ]. Glutaraldehyde has been found to exhibit cytotoxic properties in cultured human
cells [105 ]. The hazards of glutaraldehyde use for staff are considerable, and toxicity has
been suspected in 35 % of endoscopic units and detrimental effects established in
up to 80 % [16 ]
[106 ]. Use in a well-ventilated area and storage in closed containers with tight-fitting
lids is recommended.
Residues of glutaraldehyde after insufficient rinsing of devices can cause colitis,
abdominal cramps, and bloody diarrhea in patients [71 ]
[72 ]
[73 ].
Another disadvantage of glutaraldehyde is the coagulation and fixation of proteins
in combination with adsorption effects on endoscope surfaces. Glutaraldehyde is adsorbed
by the plastic surfaces of endoscopes, and remains even after thorough rinsing. The
adsorbed glutaraldehyde is not a toxicological risk for patients [107 ]
[108 ], but it can react with proteins during examination of the patient, forming large
molecules and increasing the fixation risk. Deposits on outer surfaces can be visually
detected by yellow/brown discoloration of marking rings up to the point where the
endoscope has been inserted into the patient [109 ].
Furthermore, the isolation of atypical mycobacteria, with less susceptibility to glutaraldehyde,
from ADDs/EWDs has been reported [97 ]. This may create diagnostic problems in bronchoscopy and the risk of cross-infection
in immunocompromised patients with, for instance, organisms of the M. avium complex.
Advantages and disadvantages of glutaraldehyde are listed in [Table 6 ].
Table 6
Advantages and disadvantages of glutaraldehyde.
Advantages
Disadvantages
Extended in-use solution stability
Excellent material compatibility, does not damage endoscopes
Can be used in automated and manual disinfection
Slow action against bacterial spores at 25 °C
Irritant to eyes and mucus membranes including respiratory tract. Sensitizing (vapor
and contact), requires appropriate ventilation
Adverse effects for patients after insufficient rinsing
Adsorbed by endoscope surfaces after disinfection and even after thorough rinsing
Stains insertion tube surfaces and human skin (if inappropriate gloves are worn)
Fixes proteins, promotes residue film creation, requiring thorough rinsing
Environmental controls are expensive
A2.2 Orthophthalaldehyde
Disinfectants based on orthophthalaldehyde (OPA) are offered as ready-to-use solutions
containing 0.55 % active substance. Commercially available products can be utilized
manually, in ADDs, and in EWDs. Studies have shown improved microbiological efficacy
in comparison to glutaraldehyde. OPA does not produce noxious fumes, requires no activation,
and is stable at a wider pH range of 3 to 9.
Exposure to OPA vapors may be irritating to the respiratory tract and eyes [104 ]. Use in a well-ventilated area and in closed containers with tight-fitting lids
is recommended.
In non-GI endoscopy areas OPA has caused ‘anaphylaxis-like’ reactions after repeated
use [110 ].
An advantage of OPA is its higher efficacy compared to glutaraldehyde. Accurate monitoring
of the OPA concentration is always required.
There are some disadvantages of OPA, and its efficacy and properties need to be evaluated
further. Little data are available on safe exposure levels and the hazards of long-term
exposure. OPA causes coagulation and fixation of proteins. Exposure to the agent can
lead to staining of linen, clothing, skin, instruments etc. because of reactions with
amino and thiol groups. Specific detailed instructions are necessary to ensure adequate
rinsing of equipment.
Advantages and disadvantages of OPA are listed in [Table 7 ].
Table 7
Advantages and disadvantages of orthophthalaldehyde (OPA).
Advantages
Disadvantages
Extended in-use solution stability
Excellent material compatibility, does not damage endoscopes
Can be used in automated and manual disinfection
Slow action against bacterial spores
Irritant to eyes and mucus membranes including respiratory tract; requires appropriate
ventilation
Stains human skin if inappropriate gloves are worn
Stains textiles and some equipment
“Anaphylaxis-like” reactions after repeated use have been reported in non-GI endoscopy
application areas [110 ]
A2.3 Peracetic acid
Disinfectants based on peracetic acid (PAA) and its salts are commercially available
as liquids or powder for reprocessing of endoscopes. They are also available as two-component
systems including two liquids or liquid and powder. They are used at room or at elevated
temperatures (typically ≤ 40 °C). Concentrated products must be diluted with water
in a ratio determined by microbiological testing according to European or local standards.
Powdered products should be dissolved completely according to the manufacturer’s IFU,
to avoid interaction of solid particles with endoscopes. The efficacy of PAA is influenced
by the pH value of the disinfecting solution.
With respect to staff safety, pH-optimized PAA is claimed to cause less irritation
than glutaraldehyde and to be safer for the environment. However, skin, eye, and respiratory
irritation and asthma have been linked to PAA [106 ]. Adverse effects are strongly linked to the pH value of the disinfectant solution
with minimal effects observed in a pH range between 7.5 and 10.0. It would, however,
seem unwise to state that PAA can be used safely without adequate ventilation or personal
protective measures, especially during manual reprocessing. In a closed automated
reprocessing system the pH value of the used solution is less relevant with respect
to staff safety and to the environment.
pH-Optimized PAA has the ability to remove hardened material from biopsy channels
that has resulted from the prior use of glutaraldehyde [111 ]
[112 ]. In its long history of use in the food industry and medicine, development of microorganism
resistance has not been reported; its broad spectrum of chemical reactivity suggests
that microorganisms are unlikely to develop resistance to it.
One disadvantage of liquid PAA is that it is less stable than glutaraldehyde. Multiply
used solutions require replacement more often, depending on the PAA concentration
in the solution. Very accurate monitoring of the PAA concentration is required (e. g.
with test strips). The shelf-life of liquid products containing PAA is between 12
and 18 months depending on storage conditions. The shelf-life of powder products is
3 years.
Further disadvantages of PAA are its vinegary odor and corrosive action, depending
on the formulation. Both properties are strongly linked to the pH value, temperature,
PAA concentration, and the composition of the disinfectant (i. e., inclusion of an
anticorrosive agent, etc.). The oxidizing ability of PAA may expose the leaks in internal
channels of the endoscope, especially if the endoscope has previously been disinfected
with glutaraldehyde, where organic layers might have covered minor perforations. PAA
may also cause cosmetic changes to endoscope surfaces, but without any functional
impairment.
It should be noted that various brands of disinfectants based on PAA are available,
with differences in effectiveness and side-effects. There are also PAA-based disinfectants
on the market, that have various label claims depending on composition and the test
procedure applied to check microbiological efficacy.
Recent studies with non pH-adjusted PAA show that a minimum of 1500 ppm in the working
solution (35 °C, 5 minutes) is necessary to guarantee full virucidal activity, including
against poliovirus, complying with EN 14476 [113 ].
In patients, residues of PAA in devices can cause colitis, which appears to be less
severe than that which occurs with glutaraldehyde [73 ].
Advantages and disadvantages of PAA are listed in [Table 8 ].
Table 8
Advantages and disadvantages of peracetic acid (PAA).
Advantages
Disadvantages
Fast disinfection including sporicidal activity
Environmentally friendly substance
No chemical cross-linking of protein residues
Can be used in automated and manual disinfection
Depending on pH value: irritant to eyes and mucus membranes including respiratory
tract, requires appropriate ventilation
Material compatibility depends on the pH value and temperature; endorsement of compatibility
with endoscopes and processor is required
Acid-related coagulation of proteins is possible, depending on pH value
May damage endoscopes depending on pH value
A2.4 Chlorine dioxide
Disinfectants based on chlorine dioxide are commercially available as two-component
systems, applicable manually or in ADDs. Chlorine dioxide is more effective than glutaraldehyde.
Depending on their composition, they can be more damaging to equipment, than glutaraldehyde.
If chlorine dioxide is used in ADDs, contact times are likely to be much longer and,
therefore, damage is more likely. Experience with chlorine dioxide has demonstrated
discoloration of the black plastic casing of flexible endoscopes, but this change
may be only cosmetic. Chlorine dioxide is another possible alternative to glutaraldehyde,
provided it has been approved by the instrument and processor manufacturers.
Advantages and disadvantages of chlorine dioxide are listed in [Table 9 ].
Table 9
Advantages and disadvantages of chlorine dioxide.
Advatages
Disadvantages
Irritant to eyes and mucus membranes including respiratory tract, requires appropriate
ventilation
Endoscope damage has been reported; endorsement of compatibility with endoscopes is
required (an additional coating might be required with some types of endoscopes –
manufacturer-specific)
Waste water restriction for chlorine compounds in some countries
A2.5 Electrolytically generated disinfectants
Electrolytically generated disinfectants are produced on site by electrolysis of sodium
chloride solutions. The efficacy of the disinfectant is influenced by the concentration
and ratio of oxidant constituents governed by the pH value. An advantage of these
disinfectants is that commercially available systems at different pH levels are much
more effective than glutaraldehyde. Additionally electrolytically generated disinfectants
have excellent user and patient safety profiles. A disadvantage of these disinfectants
is that the biocidal effect is decreased in the presence of soil load. To ensure a
full microbicidal effect, it is essential to perform thorough cleaning. Antimicrobial
efficacy and material compatibility are strongly influenced by the pH value and the
oxidant concentration. Similarly to some PAA-based products, electrolytically generated
disinfectants are able to remove organic layers and biofilm from surfaces. Development
of microorganism resistance has not been reported and the broad spectrum of chemical
reactivity suggests that microorganisms are unlikely to develop resistance to it.
Electrolyzed acid water (EAW) systems operate with pH 2.7, oxidation – reduction potential (ORP) > 1000 mV, and
free released chlorine concentration (FRCC) 10 ± 2 ppm. The generation and use of
EAW must take place at the same time in the same device. Since the pH and the oxidation – reduction
potential are constantly monitored, this method minimizes the major disadvantage of
electrolyzed acid water, namely, its instability. In spite of its strong acidity,
EAW rarely shows adverse effects on human skin and mucosa, unlike hydrochloric acid
and other solutions with the same acidity.
Electrolytically generated hypochlorous acid systems operate with pH 5.75 – 6.75 and ≥ 180 ppm of available free chlorine. Typically
the disinfectant is produced and supplied on site via an external generator that directly
supplies the EWD. The generator controls disinfectant production utilizing validated
system monitoring that ensures that only ‘in specification’ product is delivered to
the EWD. The generator controls pH, conductivity, power, and cell flow rate with each
parameter having a specific tolerance range that is continually checked by the monitoring
system. The disinfectant is safe to handle, requiring minimal personal protective
equipment. It is non-toxic, non-sensitizing, non-irritating, and non-mutagenic.
Advantages and disadvantages of electrolytically generated disinfectants are listed
in [Table 10 ].
Table 10
Advantages and disadvantages of electrolytically generated disinfectants.
Advantages
Disadvantages
Fast deactivation of in-use solution in the case of presence of residual organic load
Prior accurate cleaning and rinsing required
May damage endoscopes
Endorsement of material compatibility with endoscopes is required (an additional coating
might be required with some types of endoscopes – manufacturer-specific)
Acid-related coagulation of proteins is possible depending on pH value
Waste water restriction for chlorine compounds in some countries