Open Access
CC BY-NC 4.0 · Arch Plast Surg 2021; 48(04): 457-461
DOI: 10.5999/aps.2019.00311
Research/Experimental
Original Article

Intraoperative near-infrared spectroscopy for pedicled perforator flaps: a possible tool for the early detection of vascular issues

Authors

  • Andrea Marchesi

    Department of Plastic Surgery, Hand Surgery and Reconstructive Microsurgery, Ospedale San Gerardo, Monza, Italy
  • Pietro Garieri

    Department of Plastic Surgery, Hand Surgery and Reconstructive Microsurgery, Ospedale San Gerardo, Monza, Italy
  • Francesco Amendola

    Department of Plastic Surgery, Hand Surgery and Reconstructive Microsurgery, Ospedale San Gerardo, Monza, Italy
  • Stefano Marcelli

    Department of Plastic Surgery, Hand Surgery and Reconstructive Microsurgery, Ospedale San Gerardo, Monza, Italy
  • Luca Vaienti

    Department of Plastic Surgery, Hand Surgery and Reconstructive Microsurgery, Ospedale San Gerardo, Monza, Italy

Background Pedicled perforator flaps can present postoperative complications similar to those encountered in free flap surgery. Beyond a clinical evaluation, there is still no reliable technical aid for the early prediction of vascular issues. The aim of this study was to assess the support of near-infrared spectroscopy technology as an intraoperative tool to anticipate postsurgical flap ischemia.

Methods We prospectively enrolled 13 consecutive patients who were referred to our hospital from March 2017 to July 2018 and required a reconstructive procedure with a pedicled fasciocutaneous perforator flap. We measured flap peripheral capillary oxygen saturation (SpO2) in each patient with a Somanetics INVOS 5100C Cerebral/Somatic Oximeter (Medtronic), both before and after transposition. Patient demographics, operative data, and complications were then recorded during the following 6 months. We analyzed the data using the Wilcoxon signed-rank test and linear regression.

Results The mean flap SpO2 before and after transposition was 92%±3% and 78%±19%, respectively. The mean change in SpO2 was 14%±17%, with a range of 0% to 55%. The change in saturation and mean saturation ratio were significantly different between patients with and without postoperative flap necrosis.

Conclusions An immediate quantitative analysis of flap peripheral capillary SpO2 after transposition has never before been described. In our experience, an intraoperative drop in SpO2 equal to or greater than 15%–20% predicted vascular complications in pedicled perforator flaps. Conversely, flap size and rotation angle were not correlated with the risk of flap necrosis.



Publication History

Received: 09 March 2019

Accepted: 21 May 2021

Article published online:
21 March 2022

© 2021. The Korean Society of Plastic and Reconstructive Surgeons. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonCommercial License, permitting unrestricted noncommercial use, distribution, and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes. (https://creativecommons.org/licenses/by-nc/4.0/)

Thieme Medical Publishers, Inc.
333 Seventh Avenue, 18th Floor, New York, NY 10001, USA