Thromb Haemost 2006; 95(03): 567-572
DOI: 10.1160/TH05-01-0042
New Technologies, Diagnostic Tools and Drugs
Schattauer GmbH

A model for the harmonisation of test results of different quantitative D-dimer methods

Piet Meijer
1   ECAT Foundation, Leiden, The Netherlands
2   TNO Prevention and Health, Leiden, The Netherlands
,
Frits Haverkate
1   ECAT Foundation, Leiden, The Netherlands
,
Cornelis Kluft
2   TNO Prevention and Health, Leiden, The Netherlands
,
Philippe de Moerloose
3   University Hospital, Geneva, Switzerland
,
Bert Verbruggen
4   Academic Medical Center St. Radboud, Nijmegen, The Netherlands
,
Michael Spannagl
5   University Hospital, Munich, Germany
6   Instand, Düsseldorf, Germany
› Author Affiliations
Further Information

Publication History

Received 20 January 2005

Accepted after revision 04 January 2006

Publication Date:
29 November 2017 (online)

Preview

Summary

The numerical test results of different D-dimer assays vary widely. Because of the complexity of the analyte of target as well as the variability in specificity of different D-dimer assays, only harmonisation of the test results seems to be feasible. The use of a single conversion factor does not take into account for several methods the lack of commutability between test results and consensus values at different D-dimer levels. This is probably related to the mutually different response of methods to high and low levels. We therefore designed a harmonisation model based on the transformation of a method-specific regression line to a reference regression line. We used the data for the measurement of a set of plasma samples with different D-dimer levels by 353 different laboratories using 7 of the most frequently used quantitative D-dimer methods. For each method we calculated the method-specific consensus value for each sample. The overall median value was also estimated. Per method linear regression was applied throughout the method-specific consensus values using the amount of patient pooled plasma added to the different plasma samples as the independent variable. The line through the overall median values of all 7 methods was used as the reference line. Harmonisation between the methods was obtained by transformation of the method-specific regression line to the reference line. This harmonisation resulted in a reduction of the variability between the method-specific consensus values from about 75% to about 5.5%. Clinical validation of this concept had shown significant improvement of the test result comparability. We conclude that this model isa feasible approach in the harmonisation of D-dimer methods. If the harmonisation procedure is included in the calibration procedure by the manufacturers, customers will automatically obtain harmonised test results.