Klinische Neurophysiologie 1996; 27(1): 6-8
DOI: 10.1055/s-2008-1060179
© Georg Thieme Verlag Stuttgart · New York

SEP: Anatomische und Elektrophysiologische Grundlagen

SEP: Basic anatomical and electrophysiological principalsM. Stöhr
  • Neurologische Klinik mit klinischer Neurophysiologie des Zentralklinikums Augsburg
Further Information

Publication History

Publication Date:
18 March 2008 (online)

Summary

Somatosensory evoked potentials are a valuable tool in functional testing of the lemniscal part of the somatosensory system. After adequate stimulation, the ascending impulse volley may partly be recorded as near-field activity in the vicinity of different generators. Using far-field recordings from the scalp with non cephalic leads, potentials generated further away from the recording electrodes within different sites of the somatosensory system can be registrated. SEP recording both can help to identify different types of pathological impulse conduction and to localize lesions within the somatosensory system.

Zusammenfassung

SEP-Ableitungen stellen eine Funktionsprüfung des spezifischen (lemniskalen) somatosensiblen Systems dar, welches die Hinterstränge samt zuführenden Afferenzen aus der Peripherie, den Lemniscus medialis, den ventrobasalen Thalamuskern, den Tractus thalamo-corticalis und die primäre sensible Rinde umfaßt. Die in dieser Leitungsbahn aszendierende Impulswelle kann teils als Nahfeld-, teils als Fernfeld-Aktivität über der Körperoberfläche abgeleitet werden, was Rückschlüsse auf den Ort und die Eigenart einer Impulsleitungsstörung erlaubt.

    >