Summary
Our study was made on 49 patients and we used capsules of lithium fluoride. The average
fluoroscopy time was 32 min. and the aver age cinefilm length was 76.5 metres. The
relationship between the entry dose, the fluoroscopy time and the length of cinefilm
allowed us to calculate a fluoroscopy equivalent time (F.E.T) for each patient which
was of 51 min. (32 min. + 76/4) as radiation dose of one minute fluoroscopy equal
radiation dose of 4 metres of cine-film. The average heart dose was 0.70 × 10-3 × FET (GY) or 70 mr/min. There was an absence of relation of the exit dosis to the
diameter of the patient and a proportionality of the heart dose to the exit dose.
Furthermore the maximum eye dose recorded was 200 mrads and the maximum gonadal dose
was 350 mrads.
Zusammenfassung
Unsere Untersuchung wurde bei 49 Patienten ausgeführt, wobei wir Lithium-Fluorid-Kapseln
benutzten. Die durchschnittliche Durchleuchtungszeit betrug 32 Minuten, die durchschnittliche
Kino-Film-Länge 76,5 m. Die Beziehung zwischen Eintrittsdosis, Durchleuchtungszeit
und Kino-Film-Länge gestattete uns, eine Durchleuchtungs-Äquivalenzzeit (DAZ = fluoroscopy
equivalent time) für jeden Patienten zu kalkulieren. Diese Durch-leuchtungs-Äquivalenzzeit
(DAZ) betrug 51 Minuten, 32 Minuten von Durchleuchtungszeit und 76/4 von Kino-Film-Strahlenzeit
wie die Strahlendosis für 1 Minute Durchleuchtungszeit entsprach 4 m Kino-Film. Die
durchschnittliche Herzdosis war 0,70 × 10-3 × D.A.Z. oder 70 mr/min. Eine Relation zwischen Körperaustrittsdosis und Körperdurchmesser
des Patienten bestand nicht. Die Herzdosis war der Austrittsdosis proportional. Ferner
betrug die höchste gemessene Augendosis 200 mrd/350 mrd.