Determination of Plasma ET-1 by RIA - Risks and Limitations

S. M. Baumgartner-Parzer1 , P. Nowotny1 , O. Wagner2 , W. Waldhäusl1
  • 1Department of Internal Medicine III, Division of Endocrinology and Metabolism, University of Vienna, Austria
  • 2Institute of Clinical Chemistry and Pathobiochemistry, University of Leipzig, Germany
Weitere Informationen

Publikationsverlauf

1998

1998

Publikationsdatum:
20. April 2007 (online)

Plasma concentrations of the vasoconstrictor peptide endothelin-1 (ET-1) were determined in nonobese healthy subjects as well as in unspecified patients using three different commercially available radioimmunoassays (RIAs) and the respective recommended prepurification procedures. Depending on the antibody and extraction procedure applied different ET-1 concentrations were measured in identical plasma samples. Thus, this study clearly documents that ET-1 concentrations measured in plasma samples represent apparent concentrations which have to be interpreted with caution.