Purpose: Several studies have shown that the detection of multiple sclerosis plaques is improved
at higher magnetic fields such as 3.0T if compared with 1.5T. Triggered by the positive
experiences that have been obtained at 3.0T, efforts are made to allow clinical MR
imaging at even higher magnetic fields. MR systems operating at a field strength of
7.0T are meanwhile available and are ready for clinical use. Our aim was to do the
first steps for establishing an MRI protocol that is suitable for imaging multiple
sclerosis patients at 7.0T. Prerequisite was to include all pulse sequences that are
required for diagnosing MS by MR imaging and for rating disease activity in a clinical
setting.
Methods: First experiences were obtained in 6 healthy volunteers (mean age, 35 years, all
male); thereafter, 2 patients (one male, one female, aged 56 and 49 years) with confirmed
multiple sclerosis underwent MR imaging at 7.0T with the „optimized“ protocol. All
imaging studies were performed on a 7.0T system (Philips Medical Systems, Cleveland,
OH, USA). To get a (very preliminary) idea of the diagnostic yield, both patients
and the 6 volunteers also underwent MR imaging at 3.0T, with the same or equivalent
imaging protocol. The final protocol consisted of the pulse sequences: Axial FLAIR,
axial and sagittal, high-resolution thin-section T2-weighted TSE, axial T1-weighted
TSE, T1-weighted 3D Turbo Gradient Echo (TFE), T1-weighted Modified Driven Equilibrium
Fourier Transform (MDEFT). T1-weighted pulse sequences were obtained prior to and
after injection of contrast agent (0.05 mmol/kg BW Gd-DTPA, i.e. half standard dose).
Results: Since no parallel imaging was available for 7.0T at the time when these studies were
performed, RF power deposition, and image homogeneity across the imaging field of
view were the main concerns. RF power management was achieved by compromising on the
acquisition speed (prolongation of TR) and on anatomic coverage (reduction of number
of sections in T1-weighted images). In addition, refocusing RF pulse modulation („Flip
Angle Sweep“) was used to reduce RF deposition in all pulse sequences. High order
shimming was successfully used to improve image homogeneity. We were able to keep
pulse sequence parameters regarding geometric image resolution (slice thickness, imaging
matrix) equivalent to the respective 3.0T and 1.5T imaging protocols. Due to B1 inhomogeneities
and dielectric effects, there was only limited visibility of the posterior fossa structures
(arrow in Fig.1a). Some „shadowing“ was also observed in the lowermost parts of the
brain (basal temporal lobes), in particular in gradient echo images (T1-weighted TFE);
this latter effect did, however, not interfere with the diagnostic utility of the
images in these areas. Otherwise, image homogeneity was considered excellent (see
Figure 1) and comparable to the 3.0T images. Image contrast was comparable, with the
notable exception of T1-weighted TSE (not: TFE or MDEFT) images. T1-w TSE images exhibited a reduced grey/white matter
contrast because, in order to accommodate 20 sections within SAR limits, we had to
increase the TR to 1000ms. In addition, the signal intensity loss in iron-containing
cerebral structures like the basal ganglia, red nucleus or substantia nigra was more
pronounced at 7.0T (comparable to the difference between 1.5T and 3.0T). Conspicuity
of enhancing structures (choroid plexus, MS plaques) was equivalent to 3.0T, with
the notable inclusion of also the T1w TSE images. Overall, 7.0T image quality in the
volunteers as well as in the two patients was rated excellent. Visual SNR was higher
compared with the same patients' examinations obtained at 3.0T. In the two patients,
conspicuity of MS plaques and the overall number of demyelinating lesions („black
holes“ as well as fresh ED plaques) were comparable.
Conclusion: These very initial results suggest that 7.0T imaging of multiple sclerosis patients
is feasible, even without parallel imaging techniques. Coverage of the posterior fossa
is still insufficient. With the advent of multichannel RF transmit coils and tailored
pulses there is reason to assume that this will improve in the foreseeable future.
This study has to be considered as a „proof of principle“ of clinical MR imaging at
7.0T. Next step is to investigate the stability of the protocol in an entire cohort
of patients, and to develop pulse sequences that should yield higher anatomic or contrast
resolution compared with 3.0T.