Rofo 2007; 179 - A2
DOI: 10.1055/s-2007-972132

Imaging cerebrovascular reactivity using BOLD MRI at 1.5T and 3.0T: comparison of spiral and EPI combined with parallel imaging

A Kassner 1, J Poublanc 1, A Crawley 1
  • 1Department of Medical Imaging and Neurology, University of Toronto & UHN, Canada

Purpose: Combining inhaled CO2 manipulation with BOLD MRI is a promising method for assessing regional differences in cerebrovascular reactivity (CVR) which is a measurement of the brains autoregulatory capacity. Since the entire grey matter in the brain responds to the CO2 stimulus, CVR measurements provide a way to evaluate different BOLD acquisition schemes with respect to signal drop-out and distortion. Specifically we compared spiral to a single-shot EPI technique combined with parallel imaging in normal healthy subjects.

Methods: Five healthy male (age range 25–42 years) volunteers were imaged on a 1.5T and 3.0T GE Signa MR system on 2 separate days using a rebreathing circuit as described previously [1]. Each subject was placed inside the scanner and the rebreathing device was applied. At each field strength, scanning was performed using a standard single-shot BOLD protocol with a spiral read out (TE=40ms, TR=2100ms, FA=85°, FOV=20mm) for run 1 and an EPI technique combined with parallel imaging (TE=40ms, TR=2100ms, FA=85°, FOV=20mm, ASSET factor=2) for run 2. Scanning duration for each run was 8 minutes for an acquisition of 230 volumes. Each volume contained 28 slices and spatial resolution of the BOLD data was approx 3×3mm with a slice thickness of 4.5mm for both techniques. In addition, high resolution T1 weighted images were acquired for co-registration purposes. Changes in pETCO2 were achieved by controlling the subjects inspired gases with the aid of a nose clip, a mouthpiece, the rebreathing circuit, and a gas sequencer. To ensure that end-tidal gases are representative of lung gas concentrations, subjects were instructed to breathe deeply during the test. The test itself consisted of eight cycles of hypercapnia (45sec at ˜ 50mmHg) interspersed with eight cycles of hypocapnia (45sec at ˜ 30mmHg) all of which was regulated by an automated sequencer. Hypercapnia was induced by administering a gas mixture of 8% CO2/92% O2 at 14L/min for 15s and maintained at plateau for a subsequent 30s by reducing gas flow to 1.5–2 L/min of O2. During the plateau phase, the decreased inflow of fresh gases resulted in rebreathing of previously exhaled gases contained in the expiratory reservoir tube. Intervals of low CO2 were achieved by supplying subjects with 15sec of high flow (16–18 L/min) of 100% O2 and maintained by O2 flow at a rate of 12–14 L/min. Partial pressures of end-tidal CO2 (pETCO2) and O2 (pETO2) were monitored continuously using a commercially available capnograph and recorded digitally at a sampling rate of 60Hz/channel. After completion of the measurements, the collected pETCO2 data was reduced to one measure of pETCO2 per breath and correlation analysis with the BOLD data was performed. Prior to this, the BOLD data was co-registered to compensate for motion artifacts. Signal of the whole brain was used as a reference to determine the shift needed to bring the CO2 and the MR data sets in phase. Once in phase, CVR maps were calculated on a pixel by pixel basis from the slope of the regression of the percentage change of MR signal on the pETCO2. This provides a measure of reactivity expressed in units of % Δ MR signal/mmHg pETCO2. Signal drop-out was measured as the sum of all the pixels within a mask automatically generated by AFNI [2]. Differences in reactivity, residual noise and signal drop-out between the 2 acquisition schemes were assessed using a paired student's t-test.

Results: For 1.5T, there was no significant difference between spiral and EPI for any of the measures (% signal changes, residuals and signal drop-out) (p>0.74). For 3T, differences between the techniques were only significant for signal drop-out (p<0.043). All results are summarized in table 1. Figure 1 shows an example of more signal drop-out using spiral at 3T.

Conclusion: Our results show no significant difference in sensitivity and systematic error between techniques. While signal-drop out was not significantly different at 1.5T, differences were more pronounced at 3T. In this study we used a conventional spiral-out technique. The more recently developed spiral in-out technique [3], however, has promise to markedly reduce signal drop-out. We will compare this method with single-shot parallel EPI in a future study.

References:

1. Vesely et al. MRM 2002,

2. AFNI – http://afni.nimh.nih.gov/afni,

3. Preston et al. NeuroImage 2004

Table 1: Summary of results

Spiral

II EPI

P

1.5T

% signal change/mmHg

0.164

0.160

0.777

Residuals (% signal)

1.592

1.601

0.982

# of pixels in mask

7239

7256

0.736

3.0T

% signal change/mmHg

0.266

0.251

0.264

Residuals

2.529

2.209

0.124

# of pixels in mask

5927

6693

0.043

Figure 1: Signal drop-out at 3.0T – EPI vs. Spiral