Minim Invasive Neurosurg 2005; 48(2): 67-72
DOI: 10.1055/s-2004-830229
Original Article
© Georg Thieme Verlag Stuttgart · New York

Application of a Newly Developed Visual Navigation System in Humans. First Results

M.  Scholz1 , S.  Tombrock2 , W.  Konen2 , B.  Fricke3 , I.  Pechlivanis1 , M.  Engelhardt1 , K.  Schmieder1 , A.  Harders1
  • 1Department of Neurosurgery, Ruhr-University Bochum, Bochum, Germany
  • 2Center of Neuroinformatics, Bochum, Germany
  • 3Department of Neuroanatomy, Institute of Anatomy, Ruhr-University Bochum, Bochum, Germany
Weitere Informationen

Publikationsverlauf

Publikationsdatum:
24. Mai 2005 (online)

Abstract

A visual navigation system (VN) was developed which uses intraoperatively stored endoscopic images together with their specific 3D-address. A special calibration enables one to recalculate the distortion of the endoscopic images. Several modules (e. g., landmark tracking, virtual back-movement) are offered to the neurosurgeon. The system was tested in 12 human subjects during neuroendoscopic interventions and worked without problems in nearly all cases. The possibilities of digital image navigation can be used especially for control of instrument movement in case of red-out situations or blurred vision. Many further developments of the VN system are possible in order to increase the safety of neuroendoscopic interventions.

References

  • 1 Tsai R Y. A versatile camera calibration technique for high accuracy 3D vision metrology using off-the-shelf TV cameras and lenses.  IEEE J Robotics Autom. 1987;  3 323-344
  • 2 Longuet-Higgins H C. A computer algorithm for reconstructing a scene from two projections.  Nature. 1981;  293 61-62
  • 3 Scholz M, Fricke B, Tombrock S, Hardenack M, Schmieder K, Düring M von, Konen W, Harders A. Virtual image navigation: a new method to control intraoperative bleeding in neuroendoscopic surgery.  J Neurosurg. 2000;  93 342-350
  • 4 Konen W, Scholz M, Tombrock S. The VN-project: endoscopic image processing for neurosurgery.  Comput Aided Surg. 1998;  3 144-148
  • 5 Kowdley K V, Silverstein F E. Future developments in endoscopic imaging.  Bailliere's Clinical Gastroenterology. 1995;  9 173-183
  • 6 Bruno D, Vecchio F C Del, Preminger G M. Digital still image recording during video endoscopy.  J Endourol. 1999;  13 353-356
  • 7 Aslan P, Kuo R L, Hazel K, Babayan R K, Preminger G M. Advances in digital imaging during endoscopic surgery.  J Endourol. 1999;  13 251-255
  • 8 Kato T, O'Brien C B, Nishida S, Hoppe H, Gasser M, Berho M, Rodriguez M J, Ruiz P, Tzakis A G. The first case report of the use of a zoom videoendoscope for the evaluation of small bowel graft mucosa in a human after intestinal transplantation.  Gastrointest Endosc. 1999;  50 257-261
  • 9 Gunkel A R, Freisinger W, Thurmfart W F, Truppe M J. Application of the ARTMA image guided navigation system to endonasal sinus surgery. Proc Computer Assisted Radiology (Berlin) 1995: 1146-1151
  • 10 Dorward N L, Alberti O, Zhao J, Dijkstra A, Buurman J, Palmer J D, Hawkes D, Thomas D G. Interactive image-guided neuroendoscopy: development and early clinical experience.  Minim Invasive Neurosurg. 1000;  42 74-78
  • 11 Rhode V, Reinges M H, Krombach G A, Gilsbach J M. The combined use of image-guided frameless stereotaxy and neuroendoscopy for the surgical management of occlusive hydrocephalus and intracranial cysts.  Br J Neurosurg. 1998;  12 531-538
  • 12 Rhoten P, Luciano M, Barnett G. Computer-assisted endoscopy for neurosurgical procedures: technical note.  Neurosurgery. 1997;  40 632-638
  • 13 Lemke A-J, Schurig-Urbaniak A M, Liebig T, Niehues S M, Haberl H, Lehmann T-N, Felix R. Virtual MR endoscopy of the ventricles prior to neurosurgical interventional endoscopy - evaluation of different presentation techniques.  Fortschr Röntgenstr. 2004;  176 1106-1113
  • 14 Kawamata T, Iseki H, Shibasaki T, Hori T. Endoscopic augmented reality navigation for endonasal transsphenoidal surgery to treat pituitary tumors: technical note.  Neurosurgery. 2002;  50 1393-1397

PD Dr. med. Martin Scholz

Neurochirurgische Universitätsklinik · Knappschaftskrankenhaus

In der Schornau 23-25

44892 Bochum

Germany

Telefon: +49/234/299/0

Fax: +49/234/299/3609 ·

eMail: martin.scholz@ruhr-uni-bochum.de

    >