Endoscopy 2003; 35(11): 957-961
DOI: 10.1055/s-2003-43488
Original Article
© Georg Thieme Verlag Stuttgart · New York

Is Esophageal Capsule Endoscopy Feasible? Results of a Pilot

B.  Neu1 , E.  Wettschureck1 , T.  Rösch1
  • 1Dept. of Internal Medicine II, Technical University of Munich, Munich, Germany
Further Information

Publication History

Submitted 2 April 2003

Accepted after Revision 8 August 2003

Publication Date:
07 November 2003 (online)

Background and Study Aims: Capsule endoscopy (CE) has been shown to be accurate in the evaluation of small-bowel bleeding and possibly also other small-bowel disorders. It is commonly believed that other organs are not suitable for CE. We report here on our experience in studying the distal esophagus using CE under various conditions.
Patients and Methods: A prospective evaluation of CE was carried out in three groups: evaluation of the distal esophagus in routine patients (n = 58) mostly examined for suspected small-intestinal bleeding (group 1); in eight patients with signs of grade I - II reflux esophagitis on upper gastrointestinal endoscopy, who were examined in a supine position for esophageal passage (group 2); and in four volunteers who swallowed the capsule attached to a string (group 3).
Results: In 62 routine patients initially included in group 1, the median CE exposure time was 2 s (range 0 - 217 s; median of four pictures, range 0 - 434), excluding four patients with extremely delayed esophageal transit (esophageal times: 45 - 226 min); at least one image of the Z line was obtained in 24.1 % of cases, but adequate assessment of 50 % and 100 % of the circumference of the distal esophagus was possible in only 10.4 % and 0 % of these cases, respectively. In group 2, the values were better (adequate visibility rates of 50 % and 100 % in 12.5 % and 37.5 % of the eight patients, respectively), but the correct diagnosis of reflux lesions was obtained in only three of the eight. In group 3, the visibility of the Z line was good, but all four volunteers experienced the procedure with the attached string as being quite unpleasant.
Conclusions: Distal esophageal assessment by CE with the aim of providing an easy screening method for reflux lesions is not at present feasible. Technical developments will be necessary to achieve this.

References

  • 1 Iddan G, Meron G, Glukhovsky A, Swain P. Wireless capsule endoscopy.  Nature. 2000;  405 417
  • 2 Ell C, Remke S, May A. et al . A prospective controlled trial comparing wireless capsule endoscopy with push enteroscopy in chronic gastrointestinal bleeding.  Endoscopy. 2002;  34 685-689
  • 3 Lewis B S, Swain P. Capsule endoscopy in the evaluation of patients with suspected small intestinal bleeding: results of a pilot study.  Gastrointest Endosc. 2002;  56 349-353
  • 4 Costamagna G, Sha S K, Riccioni M E. et al . A prospective trial comparing small bowel radiographs and video capsule endoscopy for suspected small bowel disease.  Gastroenterology. 2002;  123 999-1005
  • 5 Lewis B S. Capsule endoscopy in clinical practice [abstract].  Gastrointest Endosc. 2002;  55 AB 125
  • 6 Pennazio M, Santucci R, Rondonotti E. et al . Wireless capsule endoscopy in patients with obscure gastrointestinal bleeding: results of the Italian multicenter experience [abstract].  Gastrointest Endosc. 2002;  55 AB 87
  • 7 Saurin J C, Delvaux M, Gaudin J L. et al . Diagnostic value of endoscopic capsule in patients with obscure digestive bleeding: blinded comparison with video push-enteroscopy.  Endoscopy. 2003;  35 576-584
  • 8 Cave D R, Wolff R, Mitty R. et al . Indications, contraindications, and an algorithm for the use of the M2A video capsule in obscure gastrointestinal bleeding [abstract].  Gastrointest Endosc. 2002;  55 AB 136
  • 9 Janowski D, Toth L, Wolff R. et al . Video capsule endoscopy: early observations on its role in the diagnosis and management of obscure gastrointestinal bleeding [abstract].  Gastrointest Endosc. 2002;  55 AB 128
  • 10 Lo S K, Fisher H A, Tabibzadeh S. et al . Evaluation of bleeding of undetermined origin by wireless endoscopy [abstract].  Gastrointest Endosc. 2002;  55 AB 129
  • 11 Mascarenhas-Saraiva M N, Lopes L. Wireless-capsule endoscopy in obscure gastrointestinal bleeding: worthwhile? [abstract].  Gastrointest Endosc. 2002;  55 AB 137
  • 12 Jensen D M, Dulai G, Lousuebsakul V. et al . Diagnostic yield of capsule endoscopy in patients with severe GI bleeding of obscure origin, subsequent recommendations, and outcomes [abstract].  Gastrointest Endosc. 2002;  55 AB 127
  • 13 Van Gossum A, Francois E, Schmit A. et al . A prospective comparative study between push-enteroscopy and wireless video capsule in patients with obscure digestive bleeding [abstract].  Gastrointest Endosc. 2002;  55 AB 88
  • 14 Chutkan R K, Balba N H, Adams T L. et al . Video capsule endoscopy in the evaluation of obscure gastrointestinal bleeding [abstract].  Gastrointest Endosc. 2002;  55 AB 133
  • 15 Yousti M, Sharma V K, Leighton J A. et al . Video capsule endoscopy (VCE) for obscure gastrointestinal bleeding (OGIB) and iron deficiency anemia (IDA) [abstract].  Gastroenterology. 2002;  122 A 18
  • 16 Hahne M, Schilling D, Hartmann D. et al . Wireless capsule endoscopy: a new dimension in small bowel diagnostics [abstract].  Gastrointest Endosc. 2002;  55 AB 147
  • 17 Demedts I, Gevers A M, Hiele M. et al . Wireless capsule enteroscopy is superior to push enteroscopy for identification of small intestinal source of obscure gastrointestinal bleeding [abstract].  Gastrointest Endosc. 2002;  55 AB 146
  • 18 Sant’anna A MG de A, Dubois J, Miron M C. et al . Diagnostic accuracy of wireless capsule videoendoscopy in occult small bowel disorders of childhood [abstract].  Gastroenterology. 2002;  122 A 330
  • 19 Kraus K, Hollerbach S, Schulmann K. et al . Die Diagnostik okkulter gastrointestinaler Blutungen - neue Optionen durch die Videokapselendoskopie [abstract].  Z Gastroenterol. 2002;  40 707
  • 20 Rösch T. DDW report 2002: small-bowel endoscopy.  Endoscopy. 2002;  34 896-900
  • 21 Rossini F P, Pennazio M. Small-bowel endoscopy.  Endoscopy.. 2002;  34 13-20
  • 22 Benz C, Jakobs R, Riemann J F. Do we need the overtube for push-enteroscopy?.  Endoscopy. 2001;  33 658-661
  • 23 Benz C, Jakobs R, Riemann J F. Does the insertion depth in push enteroscopy depend on the working length of the enteroscope?.  Endoscopy. 2002;  34 543-545
  • 24 Lewis B. Enteroscopy: endangered by the capsule?.  Endoscopy. 2002;  34 416-417
  • 25 Fisher H, Lo S K, Deleon V P. Gastrointestinal transit of the wireless endoscopic capsule [abstract].  Gastrointest Endosc. 2002;  55 AB 134
  • 26 Halpern M, Jacob H. Atlas of capsule endoscopy. Yoqneam, Israel; Given Imaging, Inc 2003
  • 27 Savary M, Miller G. Der Ösophagus. Lehrbuch und endoskopischer Atlas. Solothurn; Gassmann 1977
  • 28 Götzberger M, Klose J, Lewerenz B. et al . Bilddokumentation in der gastroenterologischen Endoskopie - Vergleich von Standbild- und Videodokumentation des gastroösophagealen Übergangs [abstract].  Z Gastroenterol. 2002;  40 694
  • 29 Van Gossum A, Devière J. Wireless endoscope: methodological features [abstract].  Gastrointest Endosc. 2002;  55 AB 135
  • 30 Balba N H, Chutkan R K. Measurement of gastric and small intestine transit time using the video capsule endoscope [abstract].  Gastrointest Endosc. 2002;  55 AB 136
  • 31 Leighton J A, Sharma V K, Yousfi M. et al . Video capsule endoscopy (VCE): new information and limitations defined [abstract].  Gastrointest Endosc. 2002;  55 AB 134
  • 32 Voderholzer W A, Ortner M, Hagenmüller F. et al . Transit times of the wireless capsule endoscopy: no effect of bowel preparation together with metoclopramide compared to control patients [abstract].  Endoscopy. 2002;  34 (Suppl 2) A70
  • 33 Swain P C, Mosse A, Burke P. et al . Remote propulsion of wireless capsule endoscopes [abstract].  Gastrointest Endosc. 2002;  55 AB 88

T. Rösch, M.D.

Dept. of Internal Medicine II, Technical University of Munich

Klinikum rechts der Isar, Ismaningerstrasse 22 · 81675 Munich · Germany

Fax: +49-89-4140-4872 ·

Email: Thomas.Roesch@lrz.tu-muenchen.de

    >