Subscribe to RSS
DOI: 10.1055/s-0044-1795156
The Use of Shape Memory Alloys in Cages for Cervical Spinal Surgery
Funding This study was funded by the Federal Ministry of Education and Science Germany (project Zwanzig20).
Abstract
Background Degenerative changes in the cervical spine can include the gradual loss of functionality of the intervertebral disks, development of osteophytes and ligament hypertrophy. Removal of the intervertebral disk and replacement with a cage (anterior discectomy and fusion [ACDF]) is a standardized operative procedure in these patients. The implant should provide structural support, should restore the physiologic lordosis, and enable a solid fusion. In this context, shape memory materials have great potential in the development of implants in spinal surgery.
Methods We designed and developed a cage that automatically adapts to the cross-section of the intervertebral disk space and simultaneously ensures mechanical support for load transfer between the adjacent vertebral bodies. A special mechanism (shape memory alloy [SMA]) should allow the implant to adapt to the geometric configuration of the intervertebral disk space. The cage developed was tested in an artificial cervical spine.
Results The base body of the cage consists of polyether ether ketone (PEEK) with a width of 14 mm, length of 16 mm, and height of 4 mm. A shape memory actuator, made of nickel–titanium alloy, is used to realize the geometry adaptation. Utilizing this, the transformation from martensite to austenite is completed at 35°C. Biomechanical testing with lateral bending and compression was performed. Subsequent cyclic loading results in a constant hysteresis curve, indicating stable implant positioning.
Conclusions We feel confident about having developed an alternative cage for ACDF that can potentially reduce peri- and postoperative morbidity and provide long-term stability by reducing bone removal during cage implantation. Therefore, we are encouraged to proceed with further biomechanical testing in cadaver specimens to eventually reach the goal of in vivo application.
Publication History
Received: 03 January 2022
Accepted: 19 April 2022
Article published online:
11 March 2025
© 2025. Thieme. All rights reserved.
Georg Thieme Verlag KG
Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany
-
References
- 1 Malo JL, Cartier A, L'Archevêque J. et al. Prevalence of occupational asthma and immunologic sensitization to guar gum among employees at a carpet-manufacturing plant. J Allergy Clin Immunol 1990; 86 (4, Pt 1): 562-569
- 2 Patel PD, Arutyunyan G, Plusch K, Vaccaro Jr A, Vaccaro AR. A review of cervical spine alignment in the normal and degenerative spine. J Spine Surg 2020; 6 (01) 106-123
- 3 Yong-Hing K, Kirkaldy-Willis WH. The pathophysiology of degenerative disease of the lumbar spine. Orthop Clin North Am 1983; 14 (03) 491-504
- 4 Kim LH, D'Souza M, Ho AL. et al. Anterior techniques in managing cervical disc disease. Cureus 2018; 10 (08) e3146
- 5 Smith GW, Robinson RA. The treatment of certain cervical-spine disorders by anterior removal of the intervertebral disc and interbody fusion. J Bone Joint Surg Am 1958; 40-A (03) 607-624
- 6 Frey HH. Use of anticonvulsants in small animals. Vet Rec 1986; 118 (17) 484-486
- 7 DiSclafani M, Lederman IR. Intraocular malignant melanoma simulating uveal prolapse. Ann Ophthalmol 1988; 20 (02) 68-70
- 8 Govaerts PJ, Jacob WA, Marquet J. Histological study of the thin replacement membrane of human tympanic membrane perforations. Acta Otolaryngol 1988; 105 (3–4): 297-302
- 9 Rewerski W, Kubikowski P. Wpływ amin biogennych na centralna regulacje temperatury ciała u szczurów. Acta Physiol Pol 1969; 20 (06) 925-928
- 10 Weir WR, Hodges JM, Wright JF, Higgins AF, Corringham RE. Atypical falciparum malaria: case report. Br Med J (Clin Res Ed) 1984; 289 (6438) 178
- 11 Wertz JM, Seward CR, Hove EL, Adkins JS. Interrelationship of dietary glycine, methionine and vitamin E in the rat. J Nutr 1968; 94 (02) 129-137
- 12 Bogduk N, Mercer S. Biomechanics of the cervical spine. I: normal kinematics. Clin Biomech (Bristol, Avon) 2000; 15 (09) 633-648
- 13 Llopis E, Belloch E, León JP, Higueras V, Piquer J. La columna cervical degenerativa. Radiologia (Madr) 2016; 58 (Suppl. 01) 13-25
- 14 Scheer JK, Tang JA, Smith JS. et al; International Spine Study Group. Cervical spine alignment, sagittal deformity, and clinical implications: a review. J Neurosurg Spine 2013; 19 (02) 141-159
- 15 Tan LA, Riew KD, Traynelis VC. Cervical spine deformity: part 1—biomechanics, radiographic parameters, and classification. Neurosurgery 2017; 81 (02) 197-203
- 16 Costanza G, Tata ME. Shape memory alloys for aerospace, recent developments, and new applications: a short review. Materials (Basel) 2020; 13 (08) 1856
- 17 Hattori Y, Taguchi T, Kim HY, Miyazaki S. Effect of stoichiometry on shape memory properties and functional stability of Ti−Ni−Pd alloys. Materials (Basel) 2019; 12 (05) 798
- 18 Pushin V, Kuranova N, Marchenkova E, Pushin A. design and development of Ti-Ni, Ni-Mn-Ga and Cu-Al-Ni-based alloys with high and low temperature shape memory effects. Materials (Basel) 2019; 12 (16) 2616
- 19 Yuan B, Zhu M, Chung CY. Biomedical porous shape memory alloys for hard-tissue replacement materials. Materials (Basel) 2018; 11 (09) 1716
- 20 Gil FJ, Planell JA. Shape memory alloys for medical applications. Proc Inst Mech Eng H 1998; 212 (06) 473-488