Abstract
A hip luxation 3 years following a cementless total hip replacement was diagnosed in a 4.7-year-old neutered male Lagotto Romagnolo that presented with a non-weight-bearing left pelvic limb lameness. Orthogonal radiographs revealed a left hip luxation and an osteolytic fracture of the greater trochanter. A mass consistent with an inflammatory granuloma was identified on preoperative computed tomography scan. Surgical revision consisted of granuloma excision, replacement of the loose acetabular cup, and repair of the femoral fracture with a locking plate. Fracture biology was augmented with the use of recombinant human bone morphogenetic protein 2 (rhBMP2) impregnated collagen sponge applied around the acetabulum and femoral fracture site. Osteointegration of the prosthetics and fracture healing were documented within 3 months of revision. Long-term follow-up at 3 years postrevision showed normal limb usage and no further evidence of total hip replacement complication.
Keywords
total hip replacement - revision surgery - polyethylene wear - cementless - trochanter fractureAuthors' Contribution
Both the authors contributed to acquisition of data, data analysis and interpretation, and drafted, revised, and approved the submitted manuscript.
Publikationsverlauf
Eingereicht: 11. März 2024
Angenommen: 04. Oktober 2024
Artikel online veröffentlicht:
04. November 2024
© 2024. Thieme. All rights reserved.
Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany
-
References
- 1 Schiller TD. BioMedtrix total hip replacement systems: an overview. Vet Clin North Am Small Anim Pract 2017; 47 (04) 899-916
- 2 Warburton KJ, Everingham JB, Helms JL. et al. Wear testing of a canine hip resurfacing implant that uses highly cross-linked polyethylene. J Orthop Res 2018; 36 (04) 1196-1205
- 3 Vezzoni L, Montinaro V, Vezzoni A. Use of a revision cup for treatment of Zurich cementless acetabular cup loosening. Surgical technique and clinical application in 31 cases. Vet Comp Orthop Traumatol 2013; 26 (05) 408-415
- 4 Vezzoni L, Bazzo S, Vezzoni A. Revision of a BFX total hip replacement stem using a Kyon stem and a head adaptor in two dogs. Vet Comp Orthop Traumatol 2017; 30 (01) 81-87
- 5 Hummel D. Zurich cementless total hip replacement. Vet Clin North Am Small Anim Pract 2017; 47 (04) 917-934
- 6 Franklin SP, Miller NA, Riecks T. Complications with the Zurich canine total hip replacement system in an initial series of cases performed by a single surgeon. Vet Comp Orthop Traumatol 2021; 34 (05) 346-351
- 7 DiVincenzo MJ, Frydman GH, Kowaleski MP. et al. Metallosis in a dog as a long-term complication following total hip arthroplasty. Vet Pathol 2017; 54 (05) 828-831
- 8 Gifford AB, Lotsikas PJ, Liska WD. et al. Total hip replacement in dogs with contralateral pelvic limb amputation: a retrospective evaluation of 13 cases. Vet Surg 2020; 49 (08) 1487-1496
- 9 Nesser VE, Kowaleski MP, Boudrieau RJ. Severe polyethylene wear requiring revision total hip arthroplasty in three dogs. Vet Surg 2016; 45 (05) 664-671
- 10 Glyn-Jones S, Thomas GER, Garfjeld-Roberts P. et al. The John Charnley Award: highly crosslinked polyethylene in total hip arthroplasty decreases long-term wear—a double-blind randomized trial. Clin Orthop Relat Res 2015; 473 (02) 432-438
- 11 Schmalzried TP, Callaghan JJ. Wear in total hip and knee replacements. J Bone Joint Surg Am 1999; 81 (01) 115-136
- 12 Dammerer D, Schneider F, Renkawitz T, Putzer D, Bogensperger M, Biedermann R. High risk of hip dislocation following polyethylene liner exchange in total hip arthroplasty-is cup revision necessary?. Arch Orthop Trauma Surg 2020; 140 (11) 1837-1845
- 13 Dyce J, Tompkins D, Bhushan B, Olmstead ML. Polyethylene wear in retrieved canine acetabular components. Vet Comp Orthop Traumatol 2000; 13: 78-86
- 14 Learmonth ID, Cunningham JL. Factors contributing to the wear of polyethylene in clinical practice. Proc Inst Mech Eng H 1997; 211 (01) 49-57
- 15 Kowaleski MP. Biomedical considerations in total hip replacement. In: Peck JN, Marcellin-Little DJ. eds. Advances in Small Animal Total Joint Replacement. West Sussex, UK:: Wiley;; 2012: 53-68
- 16 Peck JN. Weight-bearing surfaces. In: Peck JN, Marcellin-Little DJ. eds. Advances in Small Animal Total Joint Replacement. West Sussex, UK:: Wiley;; 2012: 31-38
- 17 Panjeton GD, Kim SE, Chang K, Palm LS, Ifju PG. Deformation of the Zurich cementless acetabular cup caused by implantation in a canine cadaver model. Vet Surg 2017; 46 (03) 441-447
- 18 Pulido L, Restrepo C, Parvizi J. Late instability following total hip arthroplasty. Clin Med Res 2007; 5 (02) 139-142
- 19 Broomfield JAJ, Malak TT, Thomas GER, Palmer AJ, Taylor A, Glyn-Jones S. The relationship between polyethylene wear and periprosthetic osteolysis in total hip arthroplasty at 12 years in a randomized controlled trial cohort. J Arthroplasty 2017; 32 (04) 1186-1191
- 20 Gallo J, Havranek V, Zapletalova J. Risk factors for accelerated polyethylene wear and osteolysis in ABG I total hip arthroplasty. Int Orthop 2010; 34 (01) 19-26
- 21 Wood I, Powell H, Kohm A, Lannutti JJ, Dyce J. Evaluation of femoral head damage during canine total hip replacement: a comparison of four reduction techniques. Vet Comp Orthop Traumatol 2003; 16: 184-190
- 22 Hamilton WG, Hopper Jr RH, Ginn SD, Hammell NP, Engh Jr CA, Engh CA. The effect of total hip arthroplasty cup design on polyethylene wear rate. J Arthroplasty 2005; 20 (7, Suppl 3): 63-72
- 23 Jameson SS, Lees D, James P. et al. Lower rates of dislocation with increased femoral head size after primary total hip replacement: a five-year analysis of NHS patients in England. J Bone Joint Surg Br 2011; 93 (07) 876-880
- 24 Kirker-Head CA, Boudrieau RJ, Kraus KH. Use of bone morphogenetic proteins for augmentation of bone regeneration. J Am Vet Med Assoc 2007; 231 (07) 1039-1055
- 25 Bono JV, Sanford L, Toussaint JT. Severe polyethylene wear in total hip arthroplasty. Observations from retrieved AML PLUS hip implants with an ACS polyethylene liner. J Arthroplasty 1994; 9 (02) 119-125