Open Access
CC BY-NC-ND 4.0 · Endosc Int Open 2017; 05(07): E670-E674
DOI: 10.1055/s-0043-111720
Original article
Eigentümer und Copyright ©Georg Thieme Verlag KG 2017

Evaluation of endoscopic submucosal dissection using a new endosurgical knife DN-D2718B: a first clinical feasibility study

Yoshiki Sakaguchi
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Yosuke Tsuji
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Mitsuhiro Fujishiro
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
2   Department of Endoscopy and Endoscopic Surgery, Graduate School of Medicine, The University of Tokyo,Tokyo, Japan
,
Yosuke Kataoka
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Itaru Saito
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Satoki Shichijo
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Chihiro Minatsuki
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Itsuko Asada-Hirayama
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Daisuke Yamaguchi
2   Department of Endoscopy and Endoscopic Surgery, Graduate School of Medicine, The University of Tokyo,Tokyo, Japan
,
Keiko Niimi
3   Center for Epidemiology and Preventive Medicine, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Satoshi Ono
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Shinya Kodashima
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Nobutake Yamamichi
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Kazuhiko Koike
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
› Institutsangaben
Weitere Informationen

Publikationsverlauf

submitted 21. September 2016

accepted after revision 29. März 2017

Publikationsdatum:
06. Juli 2017 (online)

Preview

Abstract

Background and study aims Endosurgical devices with injection function have been reported to decrease endoscopic submucosal dissection (ESD) operation times for experts, but the efficacy of these devices for inexperienced endoscopists is unclear. The aim of this study was to evaluate the feasibility of ESD using a novel ESD knife (DN-D2718B).

Patients and methods This is a single-center prospective pilot clinical feasibility study. Patients diagnosed with superficial gastrointestinal neoplasms were enrolled. A pre-specified group of ESD trainees with ESD experience on a porcine gastric model and fewer than 30 cases of ESD in their selected fields performed ESD under expert supervision, using the DN-D2718B. En bloc resection rates, R0 resection rates, procedure times, and incidence of intra-operational/post-operational adverse events were assessed.

Results Between June 2015 and January 2016, 13 esophageal, 27 gastric, and 14 colorectal ESD cases were performed per-protocol with mean resection speeds of 10.2, 12.0, and 15.5 mm2/min, respectively. There were no intra-operational complications.

Conclusion ESD with this novel knife is feasible even when performed by non-experts.