CC BY-NC-ND 4.0 · Indian J Radiol Imaging 2022; 32(03): 314-323
DOI: 10.1055/s-0042-1753467
Original Article

Diagnostic Accuracy of Standalone T2 Dixon Sequence Compared with Conventional MRI in Sacroiliitis

1   Department of Radiology, Kovai Medical Center and Hospital, Coimbatore, Tamil Nadu, India
,
Seetharaman Cannane
1   Department of Radiology, Kovai Medical Center and Hospital, Coimbatore, Tamil Nadu, India
,
R. Thushara
1   Department of Radiology, Kovai Medical Center and Hospital, Coimbatore, Tamil Nadu, India
,
Santhosh Poyyamoli
1   Department of Radiology, Kovai Medical Center and Hospital, Coimbatore, Tamil Nadu, India
,
Meena Nedunchelian
1   Department of Radiology, Kovai Medical Center and Hospital, Coimbatore, Tamil Nadu, India
› Author Affiliations

Abstract

Aim The aim of this article was to assess the profile of T2-weighted (T2W) multipoint Dixon sequence and conventional sequences in magnetic resonance imaging (MRI) of sacroiliac joints for the diagnosis of active and chronic sacroiliitis.

Settings and Design Prospective observational study.

Materials and Methods Thirty-seven patients with sacroiliitis underwent MRI with conventional coronal oblique short tau inversion recovery, T1W sequences, and T2W multipoint Dixon sequences. T1 fat-saturated postcontrast sequences were added in active cases. Comparisons were made between conventional and T2 Dixon sequences both quantitatively and qualitatively.

Statistical Analysis Paired t-test was used to study the difference in contrast–noise ratio (CNR) between two groups. Chi-squared analysis with p-value of ≤ 0.05 was used to test the significant association of different sequences.

Results Water only images had highest mean CNR (296.35 ± 208.28) for the detection of bone marrow edema/osteitis. T1W (186.09 ± 96.96) and opposed-phase (OP) images (279.22 ± 188.40) had highest mean CNR for the detection of subchondral sclerosis and periarticular fat deposition, respectively. OP images (p-value <0.001) followed by fat-only (FO) images (p-value = 0.001) were superior to T1W sequences in detecting periarticular fat deposition. In-phase (IP) images in detecting subchondral sclerosis and IP and FO images in detecting cortical erosions were comparable to conventional T1W sequences (p-value < 0.001).

Conclusions T2 Dixon sequences are superior or comparable to conventional MR sequences in detection of sacroiliitis, except ankylosis. Hence, Dixon can be used as a single sequence to replace the multiple sequences used in conventional imaging protocol of acute sacroiliac joints due to higher image quality. It can be used as an additional sequence in case of chronic sacroiliitis to increase the confidence and accuracy of diagnosis.

Sources of Support

Nil.




Publication History

Article published online:
23 August 2022

© 2022. Indian Radiological Association. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)

Thieme Medical and Scientific Publishers Pvt. Ltd.
A-12, 2nd Floor, Sector 2, Noida-201301 UP, India

 
  • References

  • 1 Alqatari S, Visevic R, Marshall N, Ryan J, Murphy G. An unexpected cause of sacroiliitis in a patient with gout and chronic psoriasis with inflammatory arthritis: a case report. BMC Musculoskelet Disord 2018; 19 (01) 126
  • 2 Özgen A. The value of the T2-weighted multipoint Dixon sequence in MRI of sacroiliac joints for the diagnosis of active and chronic sacroiliitis. AJR Am J Roentgenol 2017; 208 (03) 603-608
  • 3 Murphey MD, Wetzel LH, Bramble JM, Levine E, Simpson KM, Lindsley HB. Sacroiliitis: MR imaging findings. Radiology 1991; 180 (01) 239-244
  • 4 Slobodin G, Hussein H, Rosner I, Eshed I. Sacroiliitis - early diagnosis is key. J Inflamm Res 2018; 11: 339-344
  • 5 Tilvawala K, Kothari K, Patel R. Sacroiliac joint: a review. Indian J Pain. 2018; 32 (01) 4
  • 6 Puhakka KB, Melsen F, Jurik AG, Boel LW, Vesterby A, Egund N. MR imaging of the normal sacroiliac joint with correlation to histology. Skeletal Radiol 2004; 33 (01) 15-28
  • 7 Montandon C, Costa MAB, Carvalho TN, Montandon Júnior ME, Teixeira K-I-SS. Sacroiliíte: avaliação por imagem. Radiol Bras 2007; 40 (01) 53-60
  • 8 Kang Y, Hong SH, Kim JY. et al. Unilateral sacroiliitis: differential diagnosis between infectious sacroiliitis and spondyloarthritis based on MRI findings. AJR Am J Roentgenol 2015; 205 (05) 1048-1055
  • 9 Prakash D, Prabhu SM, Irodi A. Seronegative spondyloarthropathy-related sacroiliitis: CT, MRI features and differentials. Indian J Radiol Imaging 2014; 24 (03) 271-278
  • 10 Rudwaleit M, Jurik A-G, Hermann KG. et al. Defining active sacroiliitis on magnetic resonance imaging (MRI) for classification of axial spondyloarthritis: a consensual approach by the ASAS/OMERACT MRI group. Ann Rheum Dis 2009; 68 (10) 1520-1527
  • 11 Hermann K-G, Bollow M. Magnetic resonance imaging of sacroiliitis in patients with spondyloarthritis: correlation with anatomy and histology. Röfo Fortschr Geb Röntgenstr Nuklearmed 2014; 186 (03) 230-237
  • 12 Pezeshk P, Alian A, Chhabra A. Role of chemical shift and Dixon based techniques in musculoskeletal MR imaging. Eur J Radiol 2017; 94: 93-100
  • 13 Guerini H, Omoumi P, Guichoux F. et al. Fat suppression with Dixon techniques in musculoskeletal magnetic resonance imaging: a pictorial review. Semin Musculoskelet Radiol 2015; 19 (04) 335-347
  • 14 Ma J. Dixon techniques for water and fat imaging. J Magn Reson Imaging 2008; 28 (03) 543-558
  • 15 Tsoi C, Griffith JF, Lee RKL, Wong PCH, Tam LS. Imaging of sacroiliitis: current status, limitations and pitfalls. Quant Imaging Med Surg 2019; 9 (02) 318-335
  • 16 Orr KE, Andronikou S, Bramham MJ, Holjar-Erlic I, Menegotto F, Ramanan AV. Magnetic resonance imaging of sacroiliitis in children: frequency of findings and interobserver reliability. Pediatr Radiol 2018; 48 (11) 1621-1628
  • 17 Takatsu Y, Akasaka T, Miyati T. The Dixon technique and the frequency-selective fat suppression technique in three-dimensional T1 weighted MRI of the liver: a comparison of contrast-to-noise ratios of hepatocellular carcinomas-to-liver. Br J Radiol 2015; 88 (1050): 20150117
  • 18 Kalovidouri A, Firmenich N, Delattre BMA. et al. Fat suppression techniques for breast MRI: Dixon versus spectral fat saturation for 3D T1-weighted at 3 T. Radiol Med (Torino) 2017; 122 (10) 731-742
  • 19 Gaddikeri S, Mossa-Basha M, Andre JB, Hippe DS, Anzai Y. Optimal fat suppression in head and neck MRI: comparison of multipoint Dixon with 2 different fat-suppression techniques, spectral presaturation and inversion recovery, and STIR. AJNR Am J Neuroradiol 2018; 39 (02) 362-368
  • 20 Ding Y, Rao SX, Chen CZ, Li RC, Zeng MS. Usefulness of two-point Dixon fat-water separation technique in gadoxetic acid-enhanced liver magnetic resonance imaging. World J Gastroenterol 2015; 21 (16) 5017-5022
  • 21 Beddy P, Rangarajan RD, Kataoka M, Moyle P, Graves MJ, Sala E. T1-weighted fat-suppressed imaging of the pelvis with a dual-echo Dixon technique: initial clinical experience. Radiology 2011; 258 (02) 583-589
  • 22 Farrelly C, Shah S, Davarpanah A, Keeling AN, Carr JC. ECG-gated multiecho Dixon fat-water separation in cardiac MRI: advantages over conventional fat-saturated imaging. AJR Am J Roentgenol 2012; 199 (01) W74-83
  • 23 de Boer A, Leiner T, Vink EE, Blankestijn PJ, van den Berg CAT. Modified Dixon-based renal dynamic contrast-enhanced MRI facilitates automated registration and perfusion analysis. Magn Reson Med 2018; 80 (01) 66-76
  • 24 Maeder Y, Dunet V, Richard R, Becce F, Omoumi P. Bone marrow metastases: T2-weighted Dixon spin-echo fat images can replace T1-weighted spin-echo images. Radiology 2018; 286 (03) 948-959
  • 25 Yoo HJ, Hong SH, Kim DH. et al. Measurement of fat content in vertebral marrow using a modified Dixon sequence to differentiate benign from malignant processes. J Magn Reson Imaging 2017; 45 (05) 1534-1544
  • 26 Kim YP, Kannengiesser S, Paek M-Y. et al. Differentiation between focal malignant marrow-replacing lesions and benign red marrow deposition of the spine with T2*-corrected fat-signal fraction map using a three-echo volume interpolated breath-hold gradient echo Dixon sequence. Korean J Radiol 2014; 15 (06) 781-791