Case Report

Detection of ABO Discrepancy in a Case of Coronary Artery Disease by Conventional Tube Technique: A Miss by Column Agglutination Technology

Suman Sudha Routray
1   Department of Transfusion Medicine, All India Institute of Medical Sciences, Bhubaneswar, Odisha, India
,
Satya Prakash
1   Department of Transfusion Medicine, All India Institute of Medical Sciences, Bhubaneswar, Odisha, India
,
Gopal Krushna Ray
1   Department of Transfusion Medicine, All India Institute of Medical Sciences, Bhubaneswar, Odisha, India
,
1   Department of Transfusion Medicine, All India Institute of Medical Sciences, Bhubaneswar, Odisha, India
,
1   Department of Transfusion Medicine, All India Institute of Medical Sciences, Bhubaneswar, Odisha, India
› Institutsangaben
Funding None.

Abstract

ABO and Rh blood grouping of donors and recipients is the first and foremost step in pretransfusion compatibility testing. Conventional tube technique (CTT) is used to test for blood grouping and Rh D typing. But the procedure is cumbersome, and there may be subjective variation during the interpretation of the test results. The other disadvantage is that it is not adaptable to automation. Many newer techniques, such as the column agglutination technique (CAT) used for pretransfusion testing, are amenable to automation. It is being preferred to shift from CTT to semiautomated or fully automated CAT platforms or other newer technologies in many blood centers. The CAT has the added advantage of increased sensitivity and stable end-point results. The results in automated platforms using CAT are equally efficient and reliable as CTT. However, sometimes it is noted that CAT misses subgroups detection. Here, we report a case with a subgroup of A that was failed to be detected by the CAT using dextran acrylamide gel, signifying the use of CTT in evaluating blood group discrepancy.



Publikationsverlauf

Artikel online veröffentlicht:
18. Januar 2022

© 2022. The Indian Association of Laboratory Physicians. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)

Thieme Medical and Scientific Publishers Pvt. Ltd.
A-12, 2nd Floor, Sector 2, Noida-201301 UP, India

 
  • References

  • 1 Rumsey DH, Ciesielski DJ. New protocols in serologic testing: a review of techniques to meet today's challenges. Immunohematology 2000; 16 (04) 131-137
  • 2 Dara RC, Tiwari AK, Mitra S. et al. Comparison of a column agglutination technology-based automated immunohematology analyzer and a semiautomated system in pretransfusion testing. Asian J Transfus Sci 2019; 13 (02) 115-119
  • 3 Shin SY, Kwon KC, Koo SH. et al. Evaluation of two automated instruments for pre-transfusion testing: AutoVue Innova and Techno TwinStation. Korean J Lab Med 2008; 28 (03) 214-220
  • 4 Tiwari AK, Setya D, Arora D, Mehta SP, Aggarwal G, Mitra S. An algorithmic approach to serological work-up of ABO sub-groups which present as ABO discrepancies in resource constraint settings. J Immunol Methods 2020; 487: 112895
  • 5 Thakral B, Saluja K, Bajpai M, Sharma RR, Marwaha N. Importance of weak ABO subgroups. Lab Med 2005; 36 (01) 32-34 DOI: 10.1309/X59TAAYPEPCNBLUJ.
  • 6 Garretta M, Muller A, Gener J, Matte C, Moullec J. Reliability in automatic determination of the ABO group by the groupamatic system. Vox Sang 1974; 27 (02) 141-155
  • 7 Weisbach V, Ziener A, Zimmermann R, Glaser A, Zingsem J, Eckstein R. Comparison of the performance of four microtube column agglutination systems in the detection of red cell alloantibodies. Transfusion 1999; 39 (10) 1045-1050
  • 8 Sawierucha J, Posset M, Hähnel V, Johnson CL, Hutchinson JA, Ahrens N. Comparison of two column agglutination tests for red blood cell antibody testing. PLoS One 2018; 13 (12) e0210099
  • 9 Langston MM, Procter JL, Cipolone KM, Stroncek DF. Evaluation of the gel system for ABO grouping and D typing. Transfusion 1999; 39 (03) 300-305
  • 10 Zhang H, Mooney CJ, Reilly MP. ABO blood groups and cardiovascular diseases. Int J Vasc Med 2012; 2012: 641917