Evaluation of Low-kVp Low-Volume Iodinated Contrast Protocol for Coronary CT Angiography Using Retrospective ECG Gating

Atul Kapoor
1   Advanced Diagnostics and Institute of Imaging, Amritsar, Punjab, India
,
Goldaa Mahajan
1   Advanced Diagnostics and Institute of Imaging, Amritsar, Punjab, India
,
Aprajita Kapoor
1   Advanced Diagnostics and Institute of Imaging, Amritsar, Punjab, India
› Institutsangaben
Preview

Abstract

Objective The aim of this study was to evaluate the use of low peak kilovoltage (kVp) low-volume iodinated contrast protocol for performing coronary computed tomography (CT) angiography (CCTA) in patients using retrospective electrocardiogram (ECG) gating.

Materials and Methods Hundred prospective patients undergoing CCTA were studied in two groups, A and B, using 70 kilovoltage (kV) and 120 kV protocols with half and standard intravenous volumes of injected iodinated contrast, respectively. All patients had heart rates less than 100 beats/min and body mass index (BMI) less than 31 kg/m2. Both the groups were evaluated for signal-to-noise (S/N) and contrast-to-noise (C/N) ratios along with radiation dose delivered in millisievert (mSv), and for image quality (IQ), on per patient and per segment basis.

Results Patients with group A showed statistically reduced radiation dose of 1.86 mSv compared with 6.86 mSv in group B patients. Marked reduction in image noise with statistically improved S/N and C/N ratios in all coronary vessels was seen in group A. S/N ratios in group A were 20.25, 18.68, 19.04, 17.41, and 18.69 for aorta, left main, left anterior descending, right coronary, and left circumflex arteries while they were 13.34, 11.12, 10.96, 9.74, and 8.67 in group B patients. C/N ratios were also higher in all vessels in group A patients, that is, 19.48, 19.48, 19.04, 19.48, and 17.68, compared with group B patients, who had 12.43, 10.03, 9.23, 9.57, and 8.23 ratios (p < 0.0001). No significant difference in IQ per patient and per vessel was seen between both the groups.

Discussion Retrospective ECG-gated low-kVp low-volume iodinated contrast protocol provides good diagnostic quality angiograms in patients with BMI up to 31 kg/m2 and with heart rates of less than 100 beats/min with three times reduced radiation dose. The reduced volume of contrast reduces the cost as well as the chance of contrast-induced nephropathy.

Declaration of Patient Consent

The authors certify that they have obtained all appropriate patient consent forms. In the consent forms, the patients have given their consent for their images and other clinical information to be reported in the journal. The patients understand that their names and initials will not be published and due efforts will be made to conceal their identity, but anonymity cannot be guaranteed.




Publikationsverlauf

Artikel online veröffentlicht:
10. Januar 2022

© 2022. Indian Radiological Association. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)