Correlation of first-trimester thymus size with chromosomal anomalies

M Möllers
1   Universitätsklinikum Münster, Gynäkologie und Geburtshilfe, Münster, Deutschland
,
S Kleemann
1   Universitätsklinikum Münster, Gynäkologie und Geburtshilfe, Münster, Deutschland
,
J Braun
1   Universitätsklinikum Münster, Gynäkologie und Geburtshilfe, Münster, Deutschland
,
K Oelmeier
1   Universitätsklinikum Münster, Gynäkologie und Geburtshilfe, Münster, Deutschland
,
J Steinhard
1   Universitätsklinikum Münster, Gynäkologie und Geburtshilfe, Münster, Deutschland
,
HA Köster
1   Universitätsklinikum Münster, Gynäkologie und Geburtshilfe, Münster, Deutschland
,
R Schmitz
1   Universitätsklinikum Münster, Gynäkologie und Geburtshilfe, Münster, Deutschland
› Author Affiliations
 

Introduction The aim of this study was to investigate the correlation between fetal thymus size measured during first-trimester screening and chromosomal anomalies.

Methods This study is a retrospective evaluation, in which the anterior-posterior diameter of the thymus in a midsagittal plane was measured in first-trimester ultrasound between 11+0 and 13+6 weeks of gestation in 168 fetuses with chromosomal anomalies (study group) and 593 healthy fetuses (control group). The included cases were subdivided into six groups: (1) trisomy 21, (2) trisomy 18, (3) trisomy 13, (4) Turner syndrome, (5) triploidy and (6) normal controls. Thymus size measurements were adjusted to the week of gestation, which was determined by ultrasound using crown-rump-length (CRL), by calculating a ratio between CRL and thymus size (CRL-thymus-ratio). Each study group was compared with the control group separately.

Results Thymus size in fetuses affected by trisomy 18 or trisomy 13 was noticeably smaller compared to the control group (1.4 mm [1.3, 1.5] and 1.3 mm [1.2, 1.4] vs. 1.8 mm [1.6, 2.1]; all p<0,001; respectively). The thymus size of fetuses with trisomy 21 and Turner syndrome did not differ from healthy fetuses. Between the CRL-thymus-ratios of the separate study groups no statistically noticeable differences could be found.

Discussion Fetal thymus size appeared to be smaller in pregnancies affected by trisomy 18 and trisomy 13. The predictive value of fetal thymus size in first-trimester screening should be evaluated prospectively.