Abstract
Objective The purpose of the present study was to evaluate retrospectively the clinical and
radiographic results of total hip arthroplasty (THA) performed with the Exeter technique
and using the Exeter prosthesis.
Methods Between March 2000 and December 2006, 504 THAs were performed in 477 patients, with
several etiological diagnoses. A total of 260 surgeries were performed on the right
side, 244 on the left side, and 27 were performed bilaterally. The mean age of the
patients was 58.9 (17.7–86.8) years old, with a median of 69.0 years old. The preoperative
planning was performed with appropriate templates. All of the surgeries were performed
through the posterolateral approach with the patient placed on lateral decubitus.
The clinical evaluation was performed according to the Harris hip score (HHS). In
the radiographic evaluation, the bone cement interface in the three zones of DeLee
and Charnley on the acetabular side and in the seven zones of Gruen on the femoral
side were studied. Subsidence of the femoral component, presence of diaphyseal hyperthrophy,
and heterotopic ossification were also observed.
Results The mean follow-up of 441 surgeries (87.5%) was of 7.2 (1.0-16.6) years, with a median
of 7.1 years. The incidence of complications was: dislocation, 3.2%; infection, 2.2%;
peripheral nerve disorders 2.0%; thromboembolism 2.7%, acetabular cup loosing 2.0%;
diaphyseal hypertrophy, 1.26%; distal migration of the femoral component between 2,0
mm and 2.9 mm, 0.5%; 1 case of aseptic loosing of the femoral component and 9 (1.8%)
of the acetabular component. The mean HHS was mean 92.3 (50–100) points.
Conclusion Cemented THA, with the methodology applied, proved to be an effective treatment for
this group of patients. The results were satisfactory with acceptable complication
rates.
Keywords
arthroplasty, replacement, hip - hip prosthesis - bone cement - hip osteoarthritis
- hip joint