Implementation of a novel efficacy score to compare sealing and cutting devices in a porcine model

L Brecht
1   St. Josef's Hospital, Department for internal medicine, Heidelberg
,
M Wallwiener
2   University of Heidelberg, Department for Gynaecology and Obstetrics, Heidelberg
,
S Schott
2   University of Heidelberg, Department for Gynaecology and Obstetrics, Heidelberg
,
C Domschke
2   University of Heidelberg, Department for Gynaecology and Obstetrics, Heidelberg
,
C Dinkic
2   University of Heidelberg, Department for Gynaecology and Obstetrics, Heidelberg
,
M Golatta
2   University of Heidelberg, Department for Gynaecology and Obstetrics, Heidelberg
,
F Schuetz
2   University of Heidelberg, Department for Gynaecology and Obstetrics, Heidelberg
,
H Fluhr
2   University of Heidelberg, Department for Gynaecology and Obstetrics, Heidelberg
,
A Stenzinger
3   University of Heidelberg, Department of Pathology, Heidelberg
,
M Kirchner
4   University of Heidelberg, Institute of Medical Biometry and Informatics, Heidelberg
,
C Sohn
2   University of Heidelberg, Department for Gynaecology and Obstetrics, Heidelberg
,
J Rom
2   University of Heidelberg, Department for Gynaecology and Obstetrics, Heidelberg
› Author Affiliations
Further Information

Publication History

Publication Date:
24 August 2017 (online)

 

Einleitung/Falldarstellung:

In general surgery, minimally invasive laparoscopic procedures have been steadily increasing over the last decade. The application of advanced bipolar and ultrasonic energy devices for sealing and cutting of blood vessels plays a vital role in routine clinical procedures. The advantages of energy-based instruments are enhanced sealing capability combined with both fast sealing-time and minimal thermal injury. The purpose of this study was to compare the safety and efficacy profiles of nine laparoscopic sealing and cutting devices in a porcine model, with a new scoring system.

Methodik/Material:

Comparative studies in a porcine model were performed to assess vessel sealing, burst pressure, thermal spread, maximum heat, sealing/cooling time, and compression strength over the full jaw. Nine different devices from five manufacturers were tested in this study.

The Sealing and Cutting Devices (SCD) score has been developed to enable standardized comparisons of various devices. For this purpose, the most important parameters were identified through a consensus approach.

Resultate/Ergebnisse:

All sealed vessels with different devices could withstand a median pressure of more than 300 mmHg (range: 112–2046mmHg). The time for the sealing procedure was 7.705s (range: 5.305–18.38s) for the ultrasonic and 7.860s (range: 5.08–10.17s) for the bipolar devices. The ultrasonic instruments reached a median temperature of 218.1°C (range: 81.3–349.75°C) and the bipolar devices a temperature of 125.5°C (range: 94.1–133.35°C). The tissue reached a median temperature of 61.9 (range: 47.1–80.6°C) after ultrasonic sealing and 76.7°C (range: 63.1–94.2°C) after bipolar sealing. The median SCD score was 10.47 (range: 7.16–13.72).

Schlussfolgerung/Zusammenfassung:

All the instruments used seemed safe for use on the patient. The SCD score allows an indirect comparability of the instruments.