The Efficacy of the Cook–Swartz Implantable Doppler in the Detection of Free-Flap Compromise: A Systematic Review and Meta-Analysis

Riaz A. Agha
1   Department of Plastic Surgery, Guy's and St. Thomas' NHS Foundation Trust, London, United Kingdom
,
Buket Gundogan
2   University College London Medical School, London, United Kingdom
,
Alexander J. Fowler
3   Department of Vascular Surgery, Guy's and St. Thomas' NHS Foundation Trust, London, United Kingdom
,
Thomas W. H. Bragg
4   Department of Plastic Surgery, Stoke Mandeville Hospital, Aylesbury, Bucks, United Kingdom
,
Dennis P. Orgill
5   Division of Plastic Surgery, Brigham and Women's Hospital, Boston, Massachusetts and Harvard Medical School, Boston, Massachusetts
› Institutsangaben
Weitere Informationen

Publikationsverlauf

22. Februar 2016

26. Mai 2016

Publikationsdatum:
02. September 2016 (online)

Abstract

Background Reducing free-flap failure rates is a key goal of any microsurgical unit. The Cook–Swartz implantable Doppler (CSD) can be used to monitor flap vascularity. We conducted a systematic review and meta-analysis to compare the efficacy of the CSD with clinical monitoring to prevent flap failure.

Methods A comprehensive literature search was performed using MEDLINE, EMBASE, PsycINFO, EBSCO, the Cochrane Library, CINAHL, SCOPUS, SciELO, NHS evidence, and online clinical trial registers from 1966 until December 31, 2015. Studies comparing flap failure rates in the CSD and clinically monitored groups were considered. Screening and data extraction was performed by two independent researchers.

Results Overall, eight articles met the inclusion criteria, involving 3,756 patients and 3,801 flaps. The average failure rate in the clinical group was 3.5% and in the Doppler group was 2.0%. A fixed effects meta-analysis was performed and found a reduced failure rate with the use of the CSD (odds ratio= 0.37 [0.21–0.64], p = 0.0005).

Conclusion Deployment of the CSD can lower flap failure rates and has the potential to be a useful adjunct to clinical monitoring of free flaps. Further research is needed to confirm its benefits and refine its indications to optimize cost-effectiveness.

 
  • References

  • 1 Jallali N, Ridha H, Butler PE. Postoperative monitoring of free flaps in UK plastic surgery units. Microsurgery 2005; 25 (6) 469-472
  • 2 Schmulder A, Gur E, Zaretski A. Eight-year experience of the Cook-Swartz Doppler in free-flap operations: microsurgical and reexploration results with regard to a wide spectrum of surgeries. Microsurgery 2011; 31 (1) 1-6
  • 3 Liu Y, Zhao YF, Huang JT , et al. Analysis of 13 cases of venous compromise in 178 radial forearm free flaps for intraoral reconstruction. Int J Oral Maxillofac Surg 2012; 41 (4) 448-452
  • 4 Hidalgo DA, Jones CS. The role of emergent exploration in free-tissue transfer: a review of 150 consecutive cases. Plast Reconstr Surg 1990; 86 (3) 492-498 , discussion 499–501
  • 5 Ali R, Bernier C, Lin YT , et al. Surgical strategies to salvage the venous compromised deep inferior epigastric perforator flap. Ann Plast Surg 2010; 65 (4) 398-406
  • 6 Yu P, Chang DW, Miller MJ, Reece G, Robb GL. Analysis of 49 cases of flap compromise in 1310 free flaps for head and neck reconstruction. Head Neck 2009; 31 (1) 45-51
  • 7 Chen KT, Mardini S, Wei FC. Intraoperative vascular findings predict salvage outcome in free flaps with circulatory compromise. J Reconstr Microsurg 2006; 22 (07) A020
  • 8 Utley DS, Koch RJ, Goode RL. The failing flap in facial plastic and reconstructive surgery: role of the medicinal leech. Laryngoscope 1998; 108 (8 Pt 1) 1129-1135
  • 9 Bibbo C, Fritsche T, Stemper M, Hall M. Flap infection associated with medicinal leeches in reconstructive surgery: two new drug-resistant organisms. J Reconstr Microsurg 2013; 29 (7) 457-460
  • 10 Smit JM, Zeebregts CJ, Acosta R, Werker PM. Advancements in free flap monitoring in the last decade: a critical review. Plast Reconstr Surg 2010; 125 (1) 177-185
  • 11 Rozen WM, Enajat M, Whitaker IS, Lindkvist U, Audolfsson T, Acosta R. Postoperative monitoring of lower limb free flaps with the Cook-Swartz implantable Doppler probe: A clinical trial. Microsurgery 2010; 30 (5) 354-360
  • 12 Dragu A, Linke R, Kuwert T , et al. Interesting image. Tc-99m sestamibi SPECT/CT as a new tool for monitoring perfusion and viability of buried perforator based free flaps in breast reconstruction after breast cancer. Clin Nucl Med 2010; 35 (1) 36-37
  • 13 Lin SJ, Nguyen MD, Chen C , et al. Tissue oximetry monitoring in microsurgical breast reconstruction decreases flap loss and improves rate of flap salvage. Plast Reconstr Surg 2011; 127 (3) 1080-1085
  • 14 Smit JM, Whitaker IS, Liss AG, Audolfsson T, Kildal M, Acosta R. Post operative monitoring of microvascular breast reconstructions using the implantable Cook-Swartz doppler system: a study of 145 probes & technical discussion. J Plast Reconstr Aesthet Surg 2009; 62 (10) 1286-1292
  • 15 Swartz WM, Jones NF, Cherup L, Klein A. Direct monitoring of microvascular anastomoses with the 20-MHz ultrasonic Doppler probe: an experimental and clinical study. Plast Reconstr Surg 1988; 81 (2) 149-161
  • 16 Rozen WM, Chubb D, Whitaker IS, Acosta R. The efficacy of postoperative monitoring: a single surgeon comparison of clinical monitoring and the implantable Doppler probe in 547 consecutive free flaps. Microsurgery 2010; 30 (2) 105-110
  • 17 Paydar KZ, Hansen SL, Chang DS, Hoffman WY, Leon P. Implantable venous Doppler monitoring in head and neck free flap reconstruction increases the salvage rate. Plast Reconstr Surg 2010; 125 (4) 1129-1134
  • 18 Kind GM, Buntic RF, Buncke GM, Cooper TM, Siko PP, Buncke Jr HJ. The effect of an implantable Doppler probe on the salvage of microvascular tissue transplants. Plast Reconstr Surg 1998; 101 (5) 1268-1273 , discussion 1274–1275
  • 19 Poder TG, Fortier PH. Implantable Doppler in monitoring free flaps: a cost-effectiveness analysis based on a systematic review of the literature. Eur Ann Otorhinolaryngol Head Neck Dis 2013; 130 (2) 79-85
  • 20 Moher D, Liberati A, Tetzlaff J, Altman DG ; PRISMA Group. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. J Clin Epidemiol 2009; 62 (10) 1006-1012
  • 21 Agha RA, Gundogan B, Fowler AJ, Bragg TW, Orgill DP. The efficacy of the Cook-Swartz implantable Doppler in the detection of free-flap compromise: a systematic review protocol. BMJ Open 2014; 4 (3) e004253
  • 22 Balshem H, Helfand M, Schünemann HJ , et al. GRADE guidelines: 3. Rating the quality of evidence. J Clin Epidemiol 2011; 64 (4) 401-406
  • 23 Higgins JPT, Green S. Cochrane Handbook for Systematic Reviews of Interventions Version 5.1.0 [updated March 2011]. The Cochrane Collaboration, 2011. Available at: www.cochrane-handbook.org . Accessed January 26, 2014
  • 24 Sterne JA, Sutton AJ, Ioannidis JP , et al. Recommendations for examining and interpreting funnel plot asymmetry in meta-analyses of randomised controlled trials. BMJ 2011; 343: d4002
  • 25 Ferguson Jr RE, Yu P. Techniques of monitoring buried fasciocutaneous free flaps. Plast Reconstr Surg 2009; 123 (2) 525-532
  • 26 Frost MW, Niumsawatt V, Rozen WM, Eschen GE, Damsgaard TE, Kiil BJ. Direct comparison of postoperative monitoring of free flaps with microdialysis, implantable cook-swartz Doppler probe, and clinical monitoring in 20 consecutive patients. Microsurgery 2015; 35 (4) 262-271
  • 27 Ho MW, Cassidy C, Brown JS, Shaw RJ, Bekiroglu F, Rogers SN. Rationale for the use of the implantable Doppler probe based on 7 years' experience. Br J Oral Maxillofac Surg 2014; 52 (6) 530-534
  • 28 Han ZF, Guo LL, Liu LB , et al. A comparison of the Cook-Swartz Doppler with conventional clinical methods for free flap monitoring: A systematic review and a meta-analysis. 21;32:109–115