Rapid Deployment Valve Implantation using the EDWARDS Intuity Valve System: A Word of Caution regarding Sizing in Calcified Sinotubular Junctions

Thorsten C. Wahlers
1   Department of Cardiothoracic Surgery, Heart Center University Hospital Cologne, Cologne, Germany
,
Parwis B. Rahmanian
1   Department of Cardiothoracic Surgery, Heart Center University Hospital Cologne, Cologne, Germany
› Author Affiliations
Further Information

Publication History

14 October 2014

31 December 2014

Publication Date:
10 March 2015 (online)

Abstract

In aortic valve replacement, severe calcification of the sinotubular junction is a limitation in the application of the rapid deployment Edwards Intuity valve system. This is illustrated by a case presentation in which discrepancies of shape and diameter between sizer and valve mounted on the deployment system resulted in discarding the prosthesis and performing a biocomposite root replacement instead. Modification of the valve sizer will allow for safer sizing because the sizing process will mimic the implantation process more accurately.

 
  • References

  • 1 Brown JM, O'Brien SM, Wu C, Sikora JA, Griffith BP, Gammie JS. Isolated aortic valve replacement in North America comprising 108,687 patients in 10 years: changes in risks, valve types, and outcomes in the Society of Thoracic Surgeons National Database. J Thorac Cardiovasc Surg 2009; 137 (1) 82-90
  • 2 Folliguet TA, Laborde F, Zannis K, Ghorayeb G, Haverich A, Shrestha M. Sutureless perceval aortic valve replacement: results of two European centers. Ann Thorac Surg 2012; 93 (5) 1483-1488
  • 3 Kocher AA, Laufer G, Haverich A , et al. One-year outcomes of the Surgical Treatment of Aortic Stenosis with a Next Generation Surgical Aortic Valve (TRITON) trial: a prospective multicenter study of rapid-deployment aortic valve replacement with the EDWARDS INTUITY Valve System. J Thorac Cardiovasc Surg 2013; 145 (1) 110-115 , discussion 115–116
  • 4 Borger MA, Dohmen P, Misfeld M, Mohr FW. Current trends in aortic valve replacement: development of the rapid deployment EDWARDS INTUITY valve system. Expert Rev Med Devices 2013; 10 (4) 461-470
  • 5 Leon MB, Smith CR, Mack M , et al; PARTNER Trial Investigators. Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery. N Engl J Med 2010; 363 (17) 1597-1607
  • 6 Bakir I, Casselman FP, Wellens F , et al. Minimally invasive versus standard approach aortic valve replacement: a study in 506 patients. Ann Thorac Surg 2006; 81 (5) 1599-1604
  • 7 Bartels C, Leyh RG, Matthias Bechtel JF, Joubert-Hübner E, Sievers HH. Discrepancies between sizer and valve dimensions: implications for small aortic root. Ann Thorac Surg 1998; 65 (6) 1631-1633