Subscribe to RSS

DOI: 10.1055/s-0034-1383827
A Rare Case of Dacron Graft Rupture due to Friction against a Rib
Publication History
01 April 2014
06 May 2014
Publication Date:
16 July 2014 (online)

Abstract
A 54-year-old man underwent aortic repair for the infected thoracoabdominal aneurysm with a woven Dacron graft (Vascutek, Renfrewshire, Scotland) treated with gentian violet. Four months later, he complained of sudden back pain, resulting in preshock status. Computed tomographic scans showed massive hematoma around the Dacron graft, suggesting graft rupture. Initially, emergency thoracic endovascular aortic repair was performed, which was subsequently followed by open repair. The Dacron graft had a small hole, which was completely compatible with the site contacting with the rib. The graft rupture was considered due to its friction against the rib. We report on a rare event of mechanical Dacron graft rupture after the thoracoabdominal aortic replacement.
-
References
- 1 Nunn DB, Pourdeyhimi B. Intrinsic Dacron graft failure 19 years post-implantation. Cardiovasc Surg 1997; 5 (3) 333-338
- 2 Tanaka H, Okada K, Yamashita T, Kawanishi Y, Matsumori M, Okita Y. Disruption of the vascular prosthesis caused by aortic calcification after replacement of the thoracoabdominal aortic aneurysm. Ann Thorac Surg 2006; 82 (3) 1097-1099
- 3 Yamamoto H, Yamamoto F, Ishibashi K, Noishiki Y. Vascular prosthesis rupture caused by contact with rib stump after thoracic aorta replacement. J Vasc Surg 2009; 50 (1) 195-197
- 4 Ozaki N, Wakita N, Tanaka Y, Yuji D. The life-threatening complication of a vascular graft rupture early after a thoracic aortic replacement. Eur J Cardiothorac Surg 2012; 42 (4) 737-738