Minim Invasive Neurosurg 2011; 54(04): 187-190
DOI: 10.1055/s-0031-1280832
Case Report
Georg Thieme Verlag KG Stuttgart · New York

Visualization and Monitoring of Acute Epistaxis during Endovascular Treatment using a Flat Detector CT

S. Ott
1   Department of Neuroradiology, University of Erlangen-Nuremberg, Erlangen, Germany
,
T. Struffert
1   Department of Neuroradiology, University of Erlangen-Nuremberg, Erlangen, Germany
,
V. Hertel
2   Department of Otolaryngology, Head- and Neck Surgery, University of Erlangen-Nuremberg, Erlangen, Germany
,
T. Engelhorn
1   Department of Neuroradiology, University of Erlangen-Nuremberg, Erlangen, Germany
,
P Gölitz
1   Department of Neuroradiology, University of Erlangen-Nuremberg, Erlangen, Germany
,
V. Arlt
3   Department of Neurosurgery, University of Erlangen-Nuremberg, Erlangen, Germany
,
A. Dörfler
1   Department of Neuroradiology, University of Erlangen-Nuremberg, Erlangen, Germany
› Author Affiliations
Further Information

Publication History

Publication Date:
15 September 2011 (online)

Abstract

Background:

Epistaxis is one of the most common emergencies of ENT surgery and can be managed conservatively in most cases. However, transarterial embolization is an accepted treatment option for intractable epistaxis, if conservative management fails. But often, direct detection of the bleeding point by obvious contrast extravasation is not possible in conventional subtracted angiographic series (DSA). Then the suspected bleeding point is treated by endovascular embolization based on the clinical suspicion.

Case Report:

We here present the case of a young woman with intractable epistaxis where hemorrhage with contrast extravasation was only faintly visible. We used the new imaging modality of flat detector computed tomography (FD-CT) to visualize acute hemorrhage and treatment effect accurately.

Conclusion:

FD-CT was helpful to visualize both irregular branches of the bleeding mucosa and active hemorrhage and also to monitor an effective embolization within the angio suite. Acquisition of FD-CT imaging in addition to conventional 2-D imaging with the same system enhances the possibility to a better understanding of the individual patient’s anatomy and could be beneficial in the accurate depiction of active bleeding, and it might also improve surgical management, if endovascular attempts should fail.

 
  • References

  • 1 Willems PW, Farb RI, Agid R. Endovascular treatment of epistaxis. AJNR Am J Neuroradiol 2009; 30: 1637-1645
  • 2 Pope LE, Hobbs CG. Epistaxis: an update on current management. Postgrad Med J 2005; 81 (955) 309-314
  • 3 Klein GE, Köle W, Karaic R et al. Endovascular embolization treatment for intractable epistaxis. Laryngo Rhino Otol 1997; 76: 83-87
  • 4 Fahrig R, Fox AJ, Lownie S et al. Use of a C-arm system to generate true 3-D computed rotational angiograms: Preliminary in vitro and in vivo results. AJNR Am J Neuroradiol 1997; 8: 1507-1514
  • 5 Benndorf G, Klucznik RP, Strother CM. Images in cardiovascular medicine. Angiographic computed tomography for imaging of underdeployed intracranial stent. Circulation 2006; 114: e499-e500
  • 6 Struffert T, Hertel V, Kyriakou Y et al. Imaging of cochlear implant electrode array with flat-detector CT and conventional multislice CT: comparison of image quality and radiation dose. Acta Otolaryngol 2010; 130: 443-452
  • 7 Struffert T, Richter G, Engelhorn T et al. Visualisation of intracerebral haemorrhage with flat-detector CT compared to multislice CT: results in 44 cases. Eur Radiol 2009; 19: 619-625
  • 8 Struffert T, Buhk JH, Buchfelder M et al. Coil migration after endovascular coil occlusion of internal carotid artery pseudoaneurysms within the sphenoid sinus. Minim Invas Neurosurg 2009; 52: 89-92
  • 9 Engelhorn T, Rennert J, Richter G et al. Myelography using flat panel volumetric computed tomography: a comparative study in patients with lumbar spinal stenosis. Spine 2007; 32: E523-E527
  • 10 Engelhorn T, Struffert T, Richter G et al. Flat panel detector angiographic CT in the management of aneurysmal rupture during coil embolization. AJNR Am J Neuroradiol 2008; 29: 1581-1584