RSS-Feed abonnieren
DOI: 10.1055/s-0030-1257110
© Georg Thieme Verlag KG Stuttgart · New York
Probleme bei Implantation und Revision von Schrittmacher-/Defibrillatorsonden
Publikationsverlauf
Publikationsdatum:
02. Februar 2012 (online)

Abstract
Pacemaker and ICD leads have complex compound designs and are therefore the weakest link in cardiac rhythm device therapy. Technical problems are mostly insulation defects. Their number increases during follow-up and makes up more than 50 percent of the yearly failure rate of up to 20% after ten years of service life. Design flaws of specific lead models need special management. The implanting physician is first confronted with anatomic barriers to lead introduction, but his technique also impacts the hardware's failure rate, and his training and procedure volume are predictive of the dislodgment risk, the need for revision and, eventually, infection. Every potential hardware contamination urges complete removal of the system. Early explantation improves prognosis. First choice is transvenous lead extraction which achieves high success rates even with left heart leads but also has specific risks of complications.
Literatur
- 1 Benditt D G. Leadless pacing: Is it realistic?. ESC Stockholm; 2010
Reference Ris Wihthout Link
- 2
Bardy G H, Smith W M, Hood M A et al.
An entirely subcutaneous implantable cardioverter-defibrillator.
N Engl J Med.
2010;
363
36-44
Reference Ris Wihthout Link
- 3 Frey B. Transvenöse Pace-Sense-Sonden bei Patienten mit implantierbarem Kardioverter-Defibrillator:
Häufigkeit von Sondendefekten im Langzeitverlauf. [Dissertation] München; 2006
Reference Ris Wihthout Link
- 4
Phillips R, Frey M, Martin R O.
Long-term performance of polyurethane pacing leads: mechanisms of design-related failures.
Pacing Clin Electrophysiol.
1986;
9
166-1172
Reference Ris Wihthout Link
- 5
Stokes K, Urbanski P, Upton J.
The in vivo auto-oxidation of polyether polyurethane by metal ions.
J Biomater Sci Polym Ed.
1990;
1
207-230
Reference Ris Wihthout Link
- 6
Jenney C, Tan J, Karicherla A.
A new insulation material for cardiac leads with potential for improved performance.
Heart Rhythm.
2006;
2
318-319
Reference Ris Wihthout Link
- 7 Aqua Institut GmbH .Bundesauswertung zum Verfahrensjahr 2010 : 09/4 Implantierbare Defibrillatoren – Implantation. Göttingen; 2011
Reference Ris Wihthout Link
- 8
Neuzner J, Carlsson J.
Dual- versus single-coil implantable defibrillator leads: Review of the literature.
Clin Res Cardiol.
2012;
[in press]
Reference Ris Wihthout Link
- 9
Di CA , Bongiorni M G, Zucchelli G et al.
Transvenous extraction performance of expanded polytetrafluoroethylene covered ICD
leads in comparison to traditional ICD leads in humans.
Pacing Clin Electrophysiol.
2010;
33
1376-1381
Reference Ris Wihthout Link
- 10
Hackler J W, Sun Z, Lindsay B D et al.
Effectiveness of implantable cardioverter-defibrillator lead coil treatments in facilitating
ease of extraction.
Heart Rhythm.
2010;
7
890-897
Reference Ris Wihthout Link
- 11
Ratliff H L, Yousufuddin M, Lieving W R et al.
Persistent left superior vena cava: case reports and clinical implications.
Int J Cardiol.
2006;
113
242-246
Reference Ris Wihthout Link
- 12
Kropil P, Lanzman R S, Miese F R et al.
Minimally invasive catheter procedures to assist complicated pacemaker lead extraction
and implantation in the operating room.
Cardiovasc Intervent Radiol.
2011;
34
345-351
Reference Ris Wihthout Link
- 13
Kirkfeldt R E, Johansen J B, Nohr E A et al.
Risk factors for lead complications in cardiac pacing: a population-based cohort study
of 28,860 Danish patients.
Heart Rhythm.
2011;
8
1622-1628
Reference Ris Wihthout Link
- 14 Aqua Institut GmbH .Bundesauswertung zum Verfahrensjahr 2010 : 09/1 – Herzschrittmacher-Erstimplantation. Göttingen; 2011
Reference Ris Wihthout Link
- 15 Aqua Institut GmbH .Bundesauswertung zum Verfahrensjahr 2010 : 09/3 – Herzschrittmacher-Revision/-Systemwechsel/-Explantation. Göttingen; 2011
Reference Ris Wihthout Link
- 16
Lee D S, Krahn A D, Healey J S et al.
Evaluation of early complications related to De Novo cardioverter defibrillator implantation
insights from the Ontario ICD database.
J Am Coll Cardiol.
2010;
55
774-782
Reference Ris Wihthout Link
- 17
vanRees J B, de Bie M K, Thijssen J et al.
Implantation-related complications of implantable cardioverter-defibrillators and
cardiac resynchronization therapy devices: a systematic review of randomized clinical
trials.
J Am Coll Cardiol.
2011;
58
995-1000
Reference Ris Wihthout Link
- 18
Cheng A, Wang Y, Curtis J P, Varosy P D.
Acute lead dislodgements and in-hospital mortality in patients enrolled in the national
cardiovascular data registry implantable cardioverter defibrillator registry.
J Am Coll Cardiol.
2010;
56
1651-1656
Reference Ris Wihthout Link
- 19 Aqua Institut GmbH .Bundesauswertung zum Verfahrensjahr 2010 : 09/6 – Implantierbare Defibrillatoren-Revision/-Systemwechsel/-Explantation. Göttingen; 2011
Reference Ris Wihthout Link
- 20
Rordorf R, Canevese F, Vicentini A et al.
Delayed ICD lead cardiac perforation: comparison of small versus standard-diameter
leads implanted in a single center.
Pacing Clin Electrophysiol.
2011;
34
475-483
Reference Ris Wihthout Link
- 21
Hauser R G, Hayes D L, Kallinen L M et al.
Clinical experience with pacemaker pulse generators and transvenous leads: an 8-year
prospective multicenter study.
Heart Rhythm.
2007;
4
154-160
Reference Ris Wihthout Link
- 22
Kleemann T, Becker T, Doenges K et al.
Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators
over a period of > 10 years.
Circulation.
2007;
115
2474-2480
Reference Ris Wihthout Link
- 23
Hauser R G, Maisel W H, Friedman P A et al.
Longevity of Sprint Fidelis implantable cardioverter-defibrillator leads and risk
factors for failure: implications for patient management.
Circulation.
2011;
123
358-363
Reference Ris Wihthout Link
- 24
Hauser R G, Hayes D L.
Increasing hazard of Sprint Fidelis implantable cardioverter-defibrillator lead failure.
Heart Rhythm.
2009;
6
605-610
Reference Ris Wihthout Link
- 25
Kallinen L M, Hauser R G, Tang C et al.
Lead integrity alert algorithm decreases inappropriate shocks in patients who have
Sprint Fidelis pace-sense conductor fractures.
Heart Rhythm.
2010;
7
1048-1055
Reference Ris Wihthout Link
- 26
Priori S G, Auricchio A, Nisam S, Yong P.
To replace or not to replace: a systematic approach to respond to device advisories.
J Cardiovasc Electrophysiol.
2009;
20
164-170
Reference Ris Wihthout Link
- 27
Kay G N, Brinker J A, Kawanishi D T et al.
Risks of spontaneous injury and extraction of an active fixation pacemaker lead: report
of the Accufix Multicenter Clinical Study and Worldwide Registry.
Circulation.
1999;
100
2344-2352
Reference Ris Wihthout Link
- 28
Schulte B, Carlsson J, Neuzner J.
Electrical interference between two transvenous implantable defibrillator leads.
Heart.
2001;
86
14
Reference Ris Wihthout Link
- 29
Pfitzner P, Trappe H J.
Oversensing in a cardioverter defibrillator system caused by interaction between two
endocardial defibrillation leads in the right ventricle.
Pacing Clin Electrophysiol.
1998;
21
764-768
Reference Ris Wihthout Link
- 30
Lee J C, Epstein L M, Huffer L L et al.
ICD lead proarrhythmia cured by lead extraction.
Heart Rhythm.
2009;
6
613-618
Reference Ris Wihthout Link
- 31
Bracke F A, Meijer A, van Gelder L M.
Malfunction of endocardial defibrillator leads and lead extraction: where do they
meet?.
Europace.
2002;
4
19-24
Reference Ris Wihthout Link
- 32
Amelot M, Foucault A, Scanu P et al.
Comparison of outcomes in patients with abandoned versus extracted implantable cardioverter
defibrillator leads.
Arch Cardiovasc Dis.
2011;
104
572-577
Reference Ris Wihthout Link
- 33
Suga C, Hayes D L, Hyberger L K et al.
Is there an adverse outcome from abandoned pacing leads?.
J Interv Card Electrophysiol.
2000;
4
493-499
Reference Ris Wihthout Link
- 34
Wollmann C G, Bocker D, Loher A et al.
Two different therapeutic strategies in ICD lead defects: additional combined lead
versus replacement of the lead.
J Cardiovasc Electrophysiol.
2007;
18
1172-1177
Reference Ris Wihthout Link
- 35
de Cock C C, Vinkers M, van Campe L C et al.
Long-term outcome of patients with multiple (> or = 3) noninfected transvenous leads:
a clinical and echocardiographic study.
Pacing Clin Electrophysiol.
2000;
23
423-426
Reference Ris Wihthout Link
- 36
Lin G, Nishimura R A, Connolly H M et al.
Severe symptomatic tricuspid valve regurgitation due to permanent pacemaker or implantable
cardioverter-defibrillator leads.
J Am Coll Cardiol.
2005;
45
1672-1675
Reference Ris Wihthout Link
- 37
Wollmann C G, Bocker D, Loher A et al.
Incidence of complications in patients with implantable cardioverter/defibrillator
who receive additional transvenous pace/sense leads.
Pacing Clin Electrophysiol.
2005;
28
795-800
Reference Ris Wihthout Link
- 38
Jacobs D M, Fink A S, Miller R P et al.
Anatomical and morphological evaluation of pacemaker lead compression.
Pacing Clin Electrophysiol.
1993;
16
434-444
Reference Ris Wihthout Link
- 39
Al-Khatib S M, Greiner M A, Peterson E D et al.
Patient and implanting physician factors associated with mortality and complications
after implantable cardioverter-defibrillator implantation, 2002–2005.
Circ Arrhythm Electrophysiol.
2008;
1
240-249
Reference Ris Wihthout Link
- 40
Harcombe A A, Newell S A, Ludman P F et al.
Late complications following permanent pacemaker implantation or elective unit replacement.
Heart.
1998;
80
240-244
Reference Ris Wihthout Link
- 41
Krahn A D, Lee D S, Birnie D et al.
Predictors of short-term complications after implantable cardioverter-defibrillator
replacement: results from the Ontario ICD Database.
Circ Arrhythm Electrophysiol.
2011;
4
136-142
Reference Ris Wihthout Link
- 42
Al-Khatib S M, Lucas F L, Jollis J G et al.
The relation between patients' outcomes and the volume of cardioverter-defibrillator
implantation procedures performed by physicians treating Medicare beneficiaries.
J Am Coll Cardiol.
2005;
46
1536-1540
Reference Ris Wihthout Link
- 43
Johansen J B, Jorgensen O D, Moller M et al.
Infection after pacemaker implantation: infection rates and risk factors associated
with infection in a population-based cohort study of 46.299 consecutive patients.
Eur Heart J.
2011;
32
991-998
Reference Ris Wihthout Link
- 44
da Costa A, Kirkorian G, Cucherat M et al.
Antibiotic prophylaxis for permanent pacemaker implantation: a meta-analysis.
Circulation.
1998;
97
1796-1801
Reference Ris Wihthout Link
- 45
de Oliveira J C, Martinelli M, Nishioka S A et al.
Efficacy of antibiotic prophylaxis before the implantation of pacemakers and cardioverter-defibrillators:
results of a large, prospective, randomized, double-blinded, placebo-controlled trial.
Circ Arrhythm Electrophysiol.
2009;
2
29-34
Reference Ris Wihthout Link
- 46
Klug D, Balde M, Pavin D et al.
Risk factors related to infections of implanted pacemakers and cardioverter-defibrillators:
results of a large prospective study.
Circulation.
2007;
116
1349-1355
Reference Ris Wihthout Link
- 47
Le K Y, Sohail M R, Friedman P A et al.
Clinical predictors of cardiovascular implantable electronic device-related infective
endocarditis.
Pacing Clin Electrophysiol.
2011;
34
450-459
Reference Ris Wihthout Link
- 48
de Bie M K, van Rees J B, Thijssen J et al.
Cardiac Device Infections are associated with a significant mortality risk.
Heart Rhythm.
2011;
[in press]
Reference Ris Wihthout Link
- 49
Hemmer W, Fröhlig G, Markewitz A.
Kommentar zu den NASPE-Empfehlungen zur Entfernung von permanent implantierten, transvenösen
Herzschrittmacher- und Defibrillatorsonden.
Z Kardiol.
2002;
91
956-968
Reference Ris Wihthout Link
- 50
Wilkoff B L, Love C J, Byrd C L et al.
Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities,
training, indications, and patient management: this document was endorsed by the American
Heart Association (AHA).
Heart Rhythm.
2009;
6
085-1104
Reference Ris Wihthout Link
- 51
Le K Y, Sohail M R, Friedman P A et al.
Impact of timing of device removal on mortality in patients with cardiovascular implantable
electronic device infections.
Heart Rhythm.
2011;
8
1678-1685
Reference Ris Wihthout Link
- 52
Williams S E, Arujuna A, Whitaker J et al.
Percutaneous Lead and System Extraction in Patients with Cardiac Resynchronization
Therapy (CRT) Devices and Coronary Sinus Leads.
Pacing Clin Electrophysiol.
2011;
34
1209-1216
Reference Ris Wihthout Link
- 53
Calton R, Cameron D, Cusimano R J et al.
Successful laser-assisted removal of an infected ICD lead with a large vegetation.
Pacing Clin Electrophysiol.
2006;
29
910-913
Reference Ris Wihthout Link
- 54
Sohail M R, Uslan D Z, Khan A H et al.
Infective endocarditis complicating permanent pacemaker and implantable cardioverter-defibrillator
infection.
Mayo Clin Proc.
2008;
83
46-53
Reference Ris Wihthout Link
- 55
Mugge A, Gulba D C, Jost S et al.
Dissolution of a right atrial thrombus attached to pacemaker electrodes: usefulness
of recombinant tissue-type plasminogen activator.
Am Heart J.
1990;
119
1437-1439
Reference Ris Wihthout Link
- 56
Cooper C J, Dweik R, Gabbay S.
Treatment of pacemaker-associated right atrial thrombus with 2-hour rTPA infusion.
Am Heart J.
1993;
126
228-229
Reference Ris Wihthout Link
- 57
Chuang S, Mehta R H, Fay W P.
Prolonged low-dose thrombolytic therapy: a novel adjunctive strategy in the management
of an infected right atrial thrombus.
Clin Cardiol.
2002;
25
346-349
Reference Ris Wihthout Link
- 58
Bongiorni M G, Soldati E, Zucchelli G et al.
Transvenous removal of pacing and implantable cardiac defibrillating leads using single
sheath mechanical dilatation and multiple venous approaches: high success rate and
safety in more than 2000 leads.
Eur Heart J.
2008;
29
2886-2893
Reference Ris Wihthout Link
- 59
Kennergren C, Bjurman C, Wiklund R et al.
A single-centre experience of over one thousand lead extractions.
Europace.
2009;
11
612-617
Reference Ris Wihthout Link
- 60
Wazni O, Epstein L M, Carrillo R G et al.
Lead extraction in the contemporary setting: the LExICon study: an observational retrospective
study of consecutive laser lead extractions.
J Am Coll Cardiol.
2010;
55
579-586
Reference Ris Wihthout Link
- 61
Klug D, Wallet F, Lacroix D et al.
Local symptoms at the site of pacemaker implantation indicate latent systemic infection.
Heart.
2004;
90
882-886
Reference Ris Wihthout Link
Prof. Dr. med. G. Fröhlig
Kraftwerkstraße 24
66424 Homburg
eMail: ingfro@kabelmail.de