Thorac Cardiovasc Surg 2011; 59(8): 493-495
DOI: 10.1055/s-0030-1250505
Case Reports
Cardiovascular
© Georg Thieme Verlag KG Stuttgart · New York

Detection of Valve Malfunction after Mechanical MVR using Digital Phonocardiography

M. Schlömicher1 , A. Brensing2 , A. Laczkovics1
  • 1Department of Cardiothoracic Surgery, Ruhr University Hospital Bergmannsheil Bochum, Bochum, Germany
  • 2Hochschule RheinMain University of Applied Sciences, Wiesbaden, Germany
Weitere Informationen

Publikationsverlauf

received July 20, 2010

Publikationsdatum:
21. März 2011 (online)

Abstract

We present the case of a 63-year-old female patient who developed severe chronic dyspnea two years after mechanical mitral valve replacement (MVR). Conventional cardiac diagnostics including echocardiography, fluoroscopy, right and left heart catheterization could not detect any malfunction. Only digital phonocardiography was able to show alterations of the typical acoustic signal. Based on the symptoms, the clinical history and these abnormal findings the decision was taken to reoperate. Intraoperatively, a ruptured tendinous chord was found, impeding the inferior leaflet. Redo MVR was performed

References

  • 1 Ben Zekry S, Sagie A, Ben-Dor I et al. Initial clinical experience with a hand-held device (Thrombocheck) for the detection of bileaflet prosthetic valve malfunction.  J Heart Valve Dis. 2005;  14 (4) 476-480
  • 2 Fritzsche D, Eitz T, Minami K et al. Digital frequency analysis of valve sound phenomena in patients after prosthetic valve surgery: its capability as a true home monitoring of valve function.  J Heart Valve Dis. 2005;  14 (5) 657-663
  • 3 Fritzsche D, Eitz T, Laczkovics A et al. Early detection of mechanical valve dysfunction using a new home monitoring device.  Ann Thorac Surg. 2007;  83 (2) 542-548
  • 4 Daniel L B, Gigg L E, Weisel R D et al. Comparison of transthoracic and transesophageal assessment of prosthetic valve dysfunction.  J Echocardiography. 1990;  7 83-95
  • 5 Khandheria B K, Seward J B, Oh J K et al. Value and limitations of transesophageal echocardiography in assessment of mitral valve prostheses.  Circulation. 1991;  83 1956-1968
  • 6 Alton M E, Pasierski T J, Orsinelli D A et al. Comparison of transthoracic and transesophageal echocardiography in evaluation of 47 Starr-Edwards prosthetic heart valves.  J Am Coll Cardiol. 1992;  20 1503-1511
  • 7 Barbetseas J, Nagueh S F, Pitsavos C et al. Differentiating thrombus from pannus formation in obstructed mechanical prosthetic valves: an evaluation of clinical, transthoracic and transesophageal echocardiographic parameters.  J Am Coll Cardiol. 1996;  32 1410-1417
  • 8 Montorsi P, Cavoretto D, Repossini A, Bartorelli A L, Guazzi M D. Valve design characteristics and cine-fluoroscopic appearance of five currently available bileaflet prosthetic heart valves.  Am J Card Imaging. 1996;  10 (1) 29-41
  • 9 Wilkins G T, Gillam L D, Kritzer G L, Levine R A, Palacios I F, Weyman A E. Validation of continuous-wave Doppler echocardiographic measurements of mitral and tricuspid prosthetic valve gradients: a simultaneous Doppler-catheter study.  Circulation. 1986;  74 (4) 786-795
  • 10 Straub U, Feindt P, Huwer H et al. Postoperative assessment of chordal preservation and changes in cardiac geometry following mitral valve replacement.  Eur J Cardiothorac Surg. 1996;  10 (9) 734-740
  • 11 David T E, Burns R F, Bacchus C M, Druck M N. Mitral valve replacement for mitral regurgitation with and without preservation of the chordae tendineae.  J Thorac Cardiovasc Surg. 1984;  88 718-725
  • 12 Muthialu N, Varma S K, Ramanathan S, Padmanabhan C, Rao K M, Srinivasan M. Effect of chordal preservation on left ventricular function.  Asian Cardiovasc Thorac Ann. 2005;  13 (3) 233-237