Summary
A Bakelite Thorax phantom was used to test the performance of an X-ray image amplifier.
The visual acuity during fluoroscopy in a room with fair illumination after 1 min.
of adaptation and with only 1/5 of the dose rate is almost twice as good as during ordinary fluoroscopy.
Photographs of the amplified image can be made with 1/10 of the energy used for ordinary contactphotographs ans 1/50 of the energy used for ordinary camera photographs. These photographs show a resolution
of detail which lies midway between contactphotography and ordinary fluoroscopy.
The results are in good agreement with calculated data based upon the photon-fluctuation
theory of Morgan and Sturm.
Résumé
Les auteurs ont fait des investigations avec un fantôme de bakelite sur les qualités
d'un tube amplificateur à Rayons X. L'acuité de vision dans une chambre illuminée
modérément en utilisant 1/5 de l'intensité des Rayons X est à peu près le double en comparaison de la radioscopie
normale.
On peut photographier l'image amplifiéc en utilisant seulement 1/50 de l'énergie nécessaire pour une Photographie à contact et 1/50 de l'énergie nécessaire pour une photographie indirecte ordinaire. La résolution
de détail de ces photographies est au milieu de la Photographie à contact et de la
radioscopie.
Les résultats sont en bon accord avec la théorie basée sur les fluctuations des photons
de Morgan et Sturm.