Open Access
CC BY-NC-ND 4.0 · Endosc Int Open 2022; 10(02): E209-E214
DOI: 10.1055/a-1729-0048
Original article

Partially covered self-expandable metal stent with antimigratory single flange plays important role during EUS-guided hepaticogastrostomy

Masahiro Yamamura
2nd Department of Internal Medicine, Osaka Medical College, Osaka, Japan
,
Takeshi Ogura
2nd Department of Internal Medicine, Osaka Medical College, Osaka, Japan
,
Saori Ueno
2nd Department of Internal Medicine, Osaka Medical College, Osaka, Japan
,
Atsushi Okuda
2nd Department of Internal Medicine, Osaka Medical College, Osaka, Japan
,
Nobu Nishioka
2nd Department of Internal Medicine, Osaka Medical College, Osaka, Japan
,
Masanori Yamada
2nd Department of Internal Medicine, Osaka Medical College, Osaka, Japan
,
Kazuya Ueshima
2nd Department of Internal Medicine, Osaka Medical College, Osaka, Japan
,
Jun Matsuno
2nd Department of Internal Medicine, Osaka Medical College, Osaka, Japan
,
Yoshitaro Yamamoto
2nd Department of Internal Medicine, Osaka Medical College, Osaka, Japan
,
Kazuhide Higuchi
2nd Department of Internal Medicine, Osaka Medical College, Osaka, Japan
› Author Affiliations
Preview

Abstract

Background and study aims Stent migration into the abdominal cavity, which can occur due to stent shortening or stomach mobility, is a critical adverse event (AE) in EUS-HGS. To prevent this AE due to stent shortening, a novel, partially covered self-expandable metal stent with an antimigratory single flange has recently become available in Japan. The present study evaluated the clinical feasibility and safety of EUS-HGS using this novel stent.

Patients and methods We measured stent length in the abdominal cavity and the luminal portion after EUS-HGS using computed tomography (CT) performed 1 day after EUS-HGS (early phase). To evaluate stent shortening and the influence of stomach mobility, we also measured stent length at the same sites on CT performed at least 7 days after EUS-HGS (late phase).

Results Thirty-one patients successfully underwent EUS-HGS using this stent. According to CT in the early phase, stent length in the abdominal cavity was 7.13 ± 2.11 mm and the length of the luminal portion was 53.3 ± 6.27 mm. Conversely, according to CT in the late phase, stent length in the abdominal cavity was 8.55 ± 2.36 mm and the length of the luminal portion was 50.0 ± 8.36 mm. Stent shortening in the luminal portion was significantly greater in the late phase than in the early phase (P = 0.04).

Conclusions CT showed that stent migration can occur even with successful stent deployment, due to various factors such as stent shortening. The antimigratory single flange may be helpful to prevent stent migration, but further prospective comparative studies are needed to confirm our results.



Publication History

Received: 10 September 2021

Accepted after revision: 19 October 2021

Article published online:
15 February 2022

© 2022. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)