CC BY-NC-ND 4.0 · Ultraschall Med 2023; 44(02): 169-178
DOI: 10.1055/a-1724-7289
Original Article

Point Shear Wave Elastography by ElastPQ for Fibrosis Screening in Patients with NAFLD: A Prospective, Multicenter Comparison to Vibration-Controlled Elastography

Punkt-Schwerwellen-Elastographie mittels ElastPQ für das Fibrosescreening in NAFLD Patienten: eine prospektive, multizentrische Studie
1   Vienna Hepatic Hemodynamic Lab, Medical University of Vienna, Vienna, Austria
2   Division of Gastroenterology and Hepatology, Medical University of Vienna, Wien, Austria
,
Vladimir Matic
3   School of Medicine, University of Zagreb, Zagreb, Croatia
,
4   Division of Gastroenterology & Hepatology, Center for Advanced Research in Gastroenterology and Hepatology, University of Medicine and Pharmacy Victor Babes Timisoara, Timisoara, Romania
,
5   Department of Clinical, Surgical, Diagnostic and Pediatric Sciences, University of Pavia, Pavia, Italy
,
1   Vienna Hepatic Hemodynamic Lab, Medical University of Vienna, Vienna, Austria
2   Division of Gastroenterology and Hepatology, Medical University of Vienna, Wien, Austria
,
1   Vienna Hepatic Hemodynamic Lab, Medical University of Vienna, Vienna, Austria
2   Division of Gastroenterology and Hepatology, Medical University of Vienna, Wien, Austria
,
1   Vienna Hepatic Hemodynamic Lab, Medical University of Vienna, Vienna, Austria
2   Division of Gastroenterology and Hepatology, Medical University of Vienna, Wien, Austria
,
Sandra Mustapic
3   School of Medicine, University of Zagreb, Zagreb, Croatia
,
6   Department of Internal Medicine II,Division of Gastroenterology & Hepatology, Center for Advanced Research in Gastroenterology and Hepatology, University of Medicine and Pharmacy Victor Babes Timisoara, Timisoara, Romania
,
5   Department of Clinical, Surgical, Diagnostic and Pediatric Sciences, University of Pavia, Pavia, Italy
,
Ivica Grgurevic
3   School of Medicine, University of Zagreb, Zagreb, Croatia
,
1   Vienna Hepatic Hemodynamic Lab, Medical University of Vienna, Vienna, Austria
2   Division of Gastroenterology and Hepatology, Medical University of Vienna, Wien, Austria
7   Christian-Doppler Laboratory for Portal Hypertension and Liver Fibrosis, Medical University of Vienna, Wien, Austria
› Author Affiliations
Supported by: Philips Healthcare

Abstract

Background Since nonalcoholic fatty liver disease (NAFLD) has become the leading cause of liver disease in the Western world, clinicians need reliable noninvasive tools for the identification of NAFLD-associated fibrosis. Limited evidence on the performance of the novel shear wave elastography technique Elast-PQ (EPQ) in NAFLD is available. Method In this prospective, European multinational study we assessed the diagnostic accuracy of EPQ using vibration-controlled transient elastography (VCTE) as a reference standard. Results Among 353 NAFLD patients, 332 (94.1%) fulfilled reliability criteria of VCTE and EPQ (defined by IQR/median ≤0.3; 41.3% female, mean age: 59 [IQR: 16.5], mean BMI: 29.0 (7.1)). 4/353 (1.1%) and 17/353 (4.8%) had unreliable VCTE and EPQ measurements, respectively. VCTE-based NAFLD fibrosis stages were F0/F1: 222(66.9%), F2: 41 (12.3%), F3: 30 (9.1%), F4: 39 (11.7%). We found a strong correlation (Pearson R=0.87; p<0.0001) and concordance (Lin’s concordance correlation coefficient =0.792) of EPQ with VCTE. EPQ was able to identify NAFLD-fibrosis risk with the following EPQ cutoffs: ≥6.5 kPa for significant fibrosis (≥F2) (≥1.47 m/s; sensitivity: 78%; specificity: 95%; AUROC: 0.94), ≥6.9 kPa for advanced fibrosis (≥F3) (≥1.52 m/s; sens.: 88%, spec.: 89%; AUROC: 0.949), and ≥10.4 kPa for cirrhosis (F4) (≥1.86 m/s; sens.: 87%; spec.: 94%; AUROC: 0.949). Conclusion The point shear wave elastography technique EPQ shows excellent correlation to and concordance with VCTE. EPQ can reliably exclude NAFLD fibrosis <6.0 kPa (<1.41 m/s) and indicate a high risk of advanced fibrosis ≥10.4 kPa (≥1.86 m/s).

Zusammenfassung

Hintergrund Die nichtalkoholische Fettleber (NAFLD) ist mittlerweile die häufigste Lebererkrankung in der westlichen Welt. Daher besteht ein Bedarf an nichtinvasiven Methoden zur Einschätzung des Fibrose-Risikos bei NAFLD. Über die Aussagekraft der neue Scherwellen-Technologie Elast-PQ (EPQ) liegen nur wenige Daten vor. Methoden Diese prospektive, europäische multinationale Studie untersuchte die diagnostische Genauigkeit von EPQ bei NAFLD-PatientInnen im Vergleich zur Vibrations-kontrollierten transienten Elastografie (VCTE). Resultate Von 353 NAFLD-Patient*innen erfüllten 332 (94,1%) die Zuverlässigkeitskriterien für VCTE und EPQ (definiert als IQR/median ≤0,3). Die Studienpopulation war zu 41,3% weiblich, mit einem medianen Alter von 59 [IQR: 16,5] Jahren und einem medianen BMI von 29 [IQR: 7] kg/m2. Es bestand eine starke Korrelation (Pearson R=0,87; p<0.0001) und Konkordanz (Lin-Konkordanz-Korrelationskoeffizient=0,792) von EPQ mit VCTE. EPQ konnte das NAFLD-Fibrose-Risiko anhand folgender Grenzwerte abschätzen: Signifikante Fibrose (≥F2) ab ≥6,5 kPa (≥1,47 m/s; Sensitivität: 80%; Spezifität: 95%; AUROC: 0,940), fortgeschrittene Fibrose (≥F3) ab ≥6,9 kPa (≥1,52 m/s; Sensitivität: 88%, Spezifität: 89%; AUROC: 0,949), und Zirrhose (F4) ab ≥10,4 kPa (≥1,86 m/s; Sensitivität: 87%; Spezifität: 94%; AUROC: 0,949). Schlussfolgerungen Die Punkt-Scherwellen-Elastografie-Methode EPQ zeigt eine exzellente Korrektion und Konkordanz mit der VCTE. Bei NAFLD-Patient*innen wird durch EPQ <6,0 kPa (<1,41 m/s) eine signifikante Fibrose ausgeschlossen und durch EPQ ≥10,4 kPa (≥1,86 m/s) ein hohes Risiko für eine fortgeschrittene Fibrose angezeigt.

Supporting information



Publication History

Received: 07 April 2021

Accepted after revision: 08 December 2021

Article published online:
28 February 2022

© 2022. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial-License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/).