Endoscopy 2021; 53(05): E179-E180
DOI: 10.1055/a-1216-0897
E-Videos

A rare skin burn during endoscopic submucosal dissection

De-feng Li
Shenzhen People’s Hospital, Shenzhen, Guangdong, China
,
Zhi-yuan Zou
Shenzhen People’s Hospital, Shenzhen, Guangdong, China
,
Sha Nie
Shenzhen People’s Hospital, Shenzhen, Guangdong, China
,
Zheng-lei Xu
Shenzhen People’s Hospital, Shenzhen, Guangdong, China
,
Li-sheng Wang
Shenzhen People’s Hospital, Shenzhen, Guangdong, China
› Author Affiliations

A 54-year-old woman was diagnosed with early gastric cancer (EGC) in the gastric antrum during screening examination ([Fig. 1 a]). The EGC was successfully removed using an insulated-tip knife (Endocut I, Effect 3, Duration 2, Interval 4; ERBE Elektromedizin, Tübingen, Germany) ([Fig. 2]). Unfortunately, we noticed after the procedure that the skin on the patient’s back, where the electrode pad had been placed, was severely burnt ([Fig. 3], [Video 1]). The patient recovered in a week after applying recombinant human epidermal growth factor gel to the burn.

Zoom Image
Fig. 1 The early gastric cancer.
Zoom Image
Fig. 2 Endoscopic submucosal resection was performed for early gastric cancer. a The defect after removal of the tumor. b The resection specimen.
Zoom Image
Fig. 3 The skin burn at the site of electrode pad placement.

Video 1 The skin burn.


Quality:

Endoscopic submucosal dissection (ESD) is extensively used to treat lesions of the gastrointestinal tract, with bleeding and perforation being the most common complications [1]. However, rare cases of skin burn have also been recorded. Miyaga et al. reported that five patients experienced precordial skin burns after ESD when removing lesions from the upper or middle portion of a reconstructed gastric tube through the presternal route, owing to the gastric tube being positioned too close to the precordial skin [2]. However, no case of skin burn from the electrode pad has been reported. To the best of our knowledge, this was the first case of skin burn out of 1560 ESD-treated gastric lesions over 12 years in our clinical center. A possible explanation could be a wet electrode pad during ESD as a result of humid weather and a broken air conditioning unit. We would therefore like to remind endoscopists of the possibility of skin burns from the electrode pad and the need to keep the electrode pad dry to prevent such complication.

Endoscopy_UCTN_Code_CPL_1AN_2AG

Endoscopy E-Videos
https://eref.thieme.de/e-videos

Endoscopy E-Videos is a free access online section, reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high
quality video and all contributions are
freely accessible online.

This section has its own submission
website at
https://mc.manuscriptcentral.com/e-videos



Publication History

Article published online:
20 August 2020

© 2020. Thieme. All rights reserved.

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany

 
  • References

  • 1 Pimentel-Nunes P, Dinis-Ribeiro M, Ponchon T. et al. Endoscopic submucosal dissection: European Society of Gastrointestinal Endoscopy (ESGE) Guideline. Endoscopy 2015; 47: 829-854
  • 2 Miyagi M, Yoshio T, Hirasawa T. et al. Precordial skin burns after endoscopic submucosal dissection for gastric tube cancer. Dig Endosc 2015; 27: 743-747