Thromb Haemost 2007; 97(02): 310-314
DOI: 10.1160/TH06-03-0181
New Technologies, Diagnostic Tools and Drugs
Schattauer GmbH

Performance evaluation of the new CoaguChek XS system compared with the established CoaguChek system by patients experienced in INR-self management

Siegmund Braun
1   Institut für Laboratoriumsmedizin, Deutsches Herzzentrum München, Germany
,
Herbert Watzke
3   Klinik für Innere Medizin I, Abt. für Hämatologie und Hämostaseologie, Medizinische Universität, Wien, Austria
,
Michael J. Hasenkam
4   Department of CardioThoracic and Vascular Surgery, Skejby Sygehus, Aarhus University Hospital, Aarhus, Denmark
,
Michael Schwab
5   Roche Diagnostics GmbH, Mannheim, Germany
,
Tanja Wolf
5   Roche Diagnostics GmbH, Mannheim, Germany
,
Clemens Dovifat
2   Klinik am See, Rüdersdorf, Germany
,
Heinz Völler
2   Klinik am See, Rüdersdorf, Germany
› Author Affiliations

Financial support: The study was initiated and financed by Roche Diagnostics GmbH as part of the evaluation of the new system.
Further Information

Publication History

Received 29 March 2006

Accepted after revision 06 January 2006

Publication Date:
25 November 2017 (online)

Preview

Summary

The aim of the study was to analyze the concordance of INR values obtained by educated lay users with those obtained by professionals and to determine the imprecision of the new system. The new CoaguChek® XS system was tested in a user study over six weeks at four study centres in Austria, Denmark and Germany. Seventy-five patients receiving oral anticoagulant therapy were enrolled in the study. The INR results in capillary whole blood taken by professionals and by patients using the Coagu- Chek XS system were similar, and the mean relative bias was < 1%. The imprecision of the CoaguChek XS system calculated from duplicate testing is low (<6%) and slightly better than for the established CoaguChek S system. The INR results measured during the home testing phase correlated quite well between the established CoaguChek S system and the new CoaguChek XS system with a mean bias of 0.14 INR. This is a remarkably low bias taking into consideration that more than 30 different test strip lots were applied. A questionnaire was filled out by all patients to assess their personal impression. It revealed that patients were very satisfied with the new system and found it easy to operate. The results demonstrate that the agreement between professional and patient INR results for the new Coagu- Chek XS system was excellent and that INR values can be determined by lay users as well as by professionals. The instrument is very well accepted by the patients and their satisfaction even increased after four weeks practice at home.