CC BY-NC-ND 4.0 · Endosc Int Open 2024; 12(04): E561-E567
DOI: 10.1055/a-2284-9622
Original article

Impact of a new dedicated sheath device for tissue sampling of biliary stricture on pathological diagnostic yield: Retrospective study

1   Division of Endoscopy, Shizuoka Cancer Center, Nagaizumi, Japan
,
Hirotoshi Ishiwatari
1   Division of Endoscopy, Shizuoka Cancer Center, Nagaizumi, Japan
,
Keiko Sasaki
2   Pathology, Shizuoka Cancer Center, Suntogun, Japan
,
Kazuma Ishikawa
3   Department of Medical Oncology, Sapporo Medical University, Sapporo, Japan (Ringgold ID: RIN13035)
,
Junya Sato
1   Division of Endoscopy, Shizuoka Cancer Center, Nagaizumi, Japan
,
1   Division of Endoscopy, Shizuoka Cancer Center, Nagaizumi, Japan
,
Kenichiro Imai
1   Division of Endoscopy, Shizuoka Cancer Center, Nagaizumi, Japan
,
Kinichi Hotta
1   Division of Endoscopy, Shizuoka Cancer Center, Nagaizumi, Japan
,
Hiroyuki Ono
1   Division of Endoscopy, Shizuoka Cancer Center, Nagaizumi, Japan
› Institutsangaben
 

Abstract

Background and study aims Endoscopic transpapillary biliary forceps biopsy (TBFB) is a common method for obtaining specimens from biliary lesions. Its diagnostic yield is unsatisfactory; to overcome this disadvantage, a dedicated sheath has been developed. This study aimed to evaluate the outcomes of conventional TBFB and TBFB with a novel sheath device.

Patients and methods Consecutive patients who underwent TBFB between January 2020 and December 2021 were retrospectively evaluated. The rate of obtaining adequate samples, failed attempts at forceps insertion into the bile duct, and sensitivity were compared between the two groups.

Results Ninety-two patients who underwent 115 endoscopic retrograde cholangiopancreatographies (76 in the conventional group vs. 39 in the dedicated sheath group) were included. The rates of obtaining adequate samples, failed attempts of the forceps into the bile duct, and sensitivity were 72.4% vs. 89.7% (P = 0.03), 28.3% vs. 0% (P < 0.01), and 66.7% vs. 88.9% (P = 0.02), respectively.

Conclusions TBFB with the novel sheath device contributed to improved sensitivity for diagnosis of biliary stricture without insertion of forceps outside the bile duct.


#

Introduction

Pathological diagnosis plays an important role in determining the appropriate treatment strategy for patients with biliary strictures. However, establishing a definitive diagnosis is challenging. Endoscopic retrograde cholangiopancreatography (ERCP) is the most common technique used to obtain biliary tissue samples for pathological diagnosis [1]. Transpapillary biliary forceps biopsy (TBFB) and brush cytology under fluoroscopy are conventional ERCP methods; however, their diagnostic yield is suboptimal, with a diagnostic sensitivity for malignancy of 33% to 88% and 18% to 80%, respectively [2] [3] [4] [5]. In addition, TBFB requires repeated insertion of biopsy forceps through the major duodenal papilla to achieve higher sensitivity [6] [7]. Careless handling of biliary forceps may cause injury to the major duodenal papilla or perforation of the bile duct because the tip of the biopsy forceps is stiffer than that of the standard ERCP catheter. Moreover, frequent attempts to insert the device into the bile duct require additional time. These factors increase the risk of post-ERCP pancreatitis (PEP). PEP occurs in 3.5% to 9.7% of patients after ERCP; furthermore, in 0.1% to 0.7% of patients who develop PEP, it can be fatal [8] [9] [10]. Thus, an alternative tissue sampling method with a higher sensitivity and safer profile is required.

A TBFB method using an existing biliary stent delivery system or a biliary dilator has been reported [11]. Using these methods, adequate specimens can be obtained for pathological evaluation without leading to PEP [11]. However, these devices require modification by endoscopists for use in TBFB, which is a major disadvantage. Recently, a novel, dedicated sheath device for TBFB (Endosheather; Piolax, Kanagawa, Japan) was launched. It consisted of a tapered inner catheter and an outer sheath. After removing the inner catheter following the insertion of the device into the bile duct, biopsy forceps can be inserted into the targeted bile duct lesion through the outer sheath. A retrospective single-center study evaluated TBFBs using this novel delivery device in 14 patients and revealed high diagnostic yields of 90% and 92.3% in sensitivity and accuracy, respectively, making it a promising method for TBFB [12]. However, there have been no comparative studies of TBFB using this novel sheath device and conventional TBFB. Therefore, in this retrospective study, we compared these two techniques and evaluated the efficacy and safety of the delivery device.


#

Patients and methods

Study design

We retrospectively analyzed consecutive patients with suspected bile duct cancer who underwent TBFB using Radial Jaw 4 Pediatric Biopsy Forceps (Boston Scientific Japan, Tokyo, Japan) and FB-39Q-1 forceps (Olympus, Tokyo, Japan) at the Shizuoka Cancer Center Hospital between January 2020 and December 2021.More than 500 ERCP procedures are performed annually. During the study period, conventional TBFB or TBFB using a novel sheath device (Endosheather) was performed to obtain histological evidence of biliary stricture. Regarding the selection of TBFB methods, conventional TBFB was used until March 2021; thereafter, the novel sheath device was utilized. We excluded patients with surgically altered anatomy other than Billroth I reconstruction and with insufficient data. This study was approved by the Institutional Review Board (J2022–14) and performed in accordance with the principles of the Declaration of Helsinki.


#

A dedicated delivery system

The novel sheath device consists of a tapered inner catheter with a 1.13-mm tip diameter and a 2.44-mm outer sheath (7.2F) with a radiopaque marker ([Fig. 1]). After inserting the device into the bile duct over the guidewire, biopsy forceps up to a diameter of 1.9 mm were inserted through the outer sheath following withdrawal of the inner catheter. In addition, this device is compatible with a 0.035-inch guidewire.

Zoom Image
Fig. 1 Image of the novel delivery device. a The device has an inner catheter with a tapered tip (red arrowhead) and an outer sheath. There is almost no gap between the outer sheath and the inner catheter (blue arrow). b The inner catheter can be removed from the outer sheath. c The biopsy forceps (Radial Jaw 4) are inserted through the outer sheath.

#

ERCP and TBFB

ERCP and TBFB were performed by three endoscopists: one with more than 10 years of experience and two with more than 5 years but less than 10 years of experience with ERCP. ERCP was performed using a side-viewing duodenoscope (TJF-Q290V, TJF-260V, or JF-260V; Olympus Medical Systems, Tokyo, Japan). After selective biliary cannulation using wire-guided cannulation, a 0.025-inch guidewire (VisiGlide2; Olympus Medical Systems, Tokyo, Japan) was inserted into the bile duct, and the biliary stricture was detected by cholangiography. Endoscopic sphincterotomy (EST) was performed when there were no contraindications or a history of EST. Regarding conventional TBFB, we generally used either of the two biopsy forceps (Radial Jaw 4 Pediatric Biopsy Forceps; 2.0 mm outer diameter, 5.4-mm opening width and effective length 160 cm, Boston Scientific Japan, or FB-39Q-1 forceps; 1.95-mm opening width, Olympus Medical Systems, Tokyo, Japan). When there was difficulty in inserting Radial Jaw 4 into the bile duct, FB-39Q-1 forceps were used instead because the latter has more flexibility. Biopsy forceps were inserted into the bile duct alongside the guidewire, and bile duct tissue was obtained from the biliary stricture. In principle, biopsies were performed at least three times. When using the novel sheath device, the delivery device was inserted into the biliary stricture over the guidewire, and the inner catheter and guidewire were removed, leaving the outer sheath in place ([Fig. 2] a). Thereafter, biopsy forceps (Radial Jaw 4 Pediatric Biopsy Forceps; 2.0-mm outer diameter, 5.4-mm opening width, and effective length 240 cm, Boston Scientific Japan, Tokyo, Japan) were inserted into the outer sheath, and bile duct tissue was obtained ([Fig. 2] b). Five biopsies were performed because the insertion of biopsy forceps into the bile duct was easier compared with conventional TBFB. The selection of biliary drainage methods, such as endoscopic nasobiliary drainage and endoscopic biliary stenting, was performed depending on patient condition. A prophylactic pancreatic stent was inserted at endoscopist discretion. Diclofenac suppositories and ulinastatin were administered to prevent PEP when there were no contraindications. Fluoroscopic and endoscopic videos during ERCP were recorded for all patients.

Zoom Image
Fig. 2 A cholangiogram image obtained using the novel delivery device. a The outer sheath tip is located above the bile duct stricture (yellow ahead) after removing the inner catheter. b Biopsy forceps (Radial Jaw 4) (white ahead) are inserted into the bile duct through the outer sheath to obtain the bile duct specimen.

#

Pathological assessment of the sample obtained by TBFB and final diagnosis

In the pathological reports, histological classifications were divided into five categories: inadequate, benign, atypical, suspicious for adenocarcinoma, and adenocarcinoma, which were performed by an experienced pathologist (KS). In this study, an adequate sample was defined as a specimen that included tumor cells or epithelial cells with stroma and diagnosed as a benign, atypical, suspicious, or typical adenocarcinoma. When they were classified as suspicious for adenocarcinoma or adenocarcinoma, the samples were categorized as positive for malignancy; however, samples that were considered benign or atypical were categorized as negative for malignancy. The final diagnosis was based on the surgical diagnosis of a resected specimen, positive for malignancy at TBFB diagnosis with a compatible clinical course of more than 6 months, or negative for malignancy at TBFB diagnosis with spontaneous resolution or a lack of deterioration on imaging findings for a clinical follow-up time of at least 6 months. An accurate diagnosis using TBFB was defined as positive for malignancy with a final diagnosis of malignant disease and negative for malignancy with a final diagnosis of benign disease.


#

Outcome measurements

We defined the conventional and Endosheather groups as patients who underwent TBFB using the conventional method and Endosheather, respectively. Patient data were collected from electronic medical records and endoscopy databases. We evaluated the sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and accuracy of TBFB for pathological diagnosis and compared the conventional and Endosheather methods. Furthermore, the rates of adequate sample collection, failed insertion of forceps into the bile duct per attempt, biopsy time per biopsy, biopsy time per ERCP, procedure time, and adverse events (AEs) were compared between the two groups. Regarding failed attempts of the forceps into the bile duct, defined as when the biopsy forceps were not inserted into the bile duct, two gastroenterologists (H.I. and F.N.; Japanese Gastroenterological Endoscopy Society board-certified members) reevaluated it by examining the video of ERCP for the study. When there was a discrepancy between opinions, a decision was made after a discussion between them. Biopsy time per ERCP was measured from the first attempt to insert the biopsy forceps to the end of the last biopsy. Biopsy time per biopsy was defined as biopsy time per ERCP divided by the total number of biopsies. Procedure time was measured from endoscope insertion to withdrawal. AEs were defined and graded according to the American Society for Gastrointestinal Endoscopy Severity Grading system [13].


#

Statistical analyses

Continuous variables were presented as medians and interquartile ranges and were compared using the Mann-Whitney U-test. Categorical variables were shown as proportions and were compared using Fisher’s exact test. Sensitivity, specificity, PPV, NPV, and accuracy were compared using the χ2 test and Fisher’s exact test, where appropriate. P < 0.05 was considered to be statistically significant. All analyses were performed using R version 3.4.1 (The R Foundation for Statistical Computing, Vienna, Austria).


#
#

Results

Patient characteristics

During the study period, 122 ERCP cases with suspected bile duct cancer underwent TBFB. After excluding seven ERCP cases (surgically altered anatomy, n = 3; insufficient data, n = 4), 115 ERCP procedures were analyzed in this study (Conventional group, n = 76; Endosheather group, n = 39). Patient characteristics are shown in [Table 1]. No significant differences were observed between the two groups.

Table 1 Patient characteristics.

Measure

Conventional

n = 76

Endosheather

n = 39

P value

ERCP, endoscopic retrograde cholangiopancreatography; EST, endoscopic sphincterotomy; PGW, pancreatic guidewire; ENBD, endoscopic nasobiliary drainage; PEP, post-ERCP pancreatitis; IQR, interquartile range.

Age, years, median (IQR)

71 (65–77)

72 (65.8–77)

0.25

Location of biliary stricture, n (%)

  • Hilar

54 (71.1)

26 (66.7)

0.67

  • Distal

22 (28.9)

13 (33.3)

Length of biliary stricture, mm, median (IQR)

16 (12–22)

16.5 (10–22.8)

0.43

Cannulation time, min, (median [IQR])

1.25 (0.3–4)

1 (0.5–2.5)

0.08

Naïve major duodenal papilla, n (%)

45 (59.2)

20 (51.3)

0.43

EST history

28 (36.8)

10 (25.6)

0.30

ERCP-related procedure, n (%)

  • PGW-assisted biliary cannulation

13 (17.1)

3 (7.7)

0.28

  • Precutting

2 (2.6)

0 (0)

0.55

  • EST

48 (63.2)

29 (74.4)

0.30

  • Biliary drainage tube or stent

0.24

    • Plastic stent

21 (27.6)

16 (41)

    • Metallic stent

4 (5.3)

0 (0)

    • ENBD

45 (59.2)

22 (56.4)

    • None

6 (7.9)

1 (2.6)

Pancreatic stent placement for the prevention of PEP

4 (5.3)

1 (2.6)

0.66

Final diagnosis, n (%)

  • Malignant

63 (82.9)

36 (92.3)

0.26

    • Bile duct cancer

55 (72.4)

33 (84.6)

    • Gallbladder cancer

8 (10.5)

3 (7.7)

  • Benign disease

13 (17.1)

3 (7.7)

Surgery

23 (30.3)

16 (41)

0.34


#

Procedure outcomes of bile duct biopsy

The median number of biopsies was three and five in the conventional and Endosheather groups, respectively (P < 0.001) ([Table 2]). Rates of failed insertion of the forceps into the bile duct per attempt were 28.3% (95/336) and 0% in the conventional and Endosheather groups, respectively (P < 0.01). Biopsy times per ERCP were 220 and 296.5 sec in the conventional and Endosheather groups (P = 0.56), respectively, whereas biopsy time per biopsy of the conventional group was significantly longer than that of the Endosheather group (68.3 vs. 54 sec, P = 0.03).

Table 2 Study outcomes of TBFB.

Measure

Conventional

n = 76

Endosheather

n = 39

P value

TBFB, transpapillary biliary forceps biopsy; ERCP, endoscopic retrograde cholangiopancreatography; IQR, interquartile range.

Type of forceps for TBFB, n

  • FB-39Q-1

25 (32.9)

0 (0)

< 0.01

  • Radial Jaw 4

51 (67.1)

39 (100)

Number of biopsies, n

  • Median

3

5

< 0.01

  • IQR

3–3

3–5

  • Range

1–6

2–10

Total number of biopsies, n

241

225

Total number of attempts for insertion of forceps into the bile duct, n

336

225

Total failed attempts of the forceps into the bile duct, n

95

0

Rate of failed attempts of the forceps into the bile duct, % (n/N)

28.3 (95/336)

0 (0/225)

< 0.01

Biopsy time per one biopsy, sec median (IQR)

68.3 (53.4–103)

54 (42–69.6)

0.03

Biopsy time per ERCP, sec, median (IQR)

220 (160–330)

296.5 (218.5–375)

0.56

Procedure time, min, median (IQR)

25 (19–37.5)

24 (19–33)

0.7


#

Diagnostic yields of biliary forceps biopsy

Diagnostic yields of biliary forceps biopsies are summarized in [Table 3]. The rate of obtaining adequate samples was significantly lower in the conventional group than in the Endosheather group (72.4% [55/76] vs. 89.7% [35/39], P = 0.03). Sensitivity of the conventional group was significantly lower than that of the Endosheather group (66.7% vs. 88.9%, P = 0.02). There were no differences observed in specificity, PPV, NPV, or accuracy between the two groups.

Table 3 Diagnostic yield of TBFB.

Conventional n = 76

Endosheather n = 39

TBFB, transpapillary biliary forceps biopsy; CI, confidence interval; PPV, positive predictive value; NPV, negative predictive value; NS, not significant.

Final diagnosis

Final diagnosis

Malignant

Benign

Malignant

Benign

TBFB diagnosis

Malignant

42

0

TBFB diagnosis

Malignant

32

0

Benign

1

12

Benign

1

2

Inadequate

20

1

Inadequate

3

1

Measure

P value

Sensitivity %, (n/N)

66.7 (42/63)

88.9 (32/36)

0.02

  • (95% CI)

53.7–78

73.9–96.9

Hilar

70 (31/47)

87 (20/23)

0.08

Distal

68.8 (11/16)

92.3 (12/13)

0.18

Specificity %, (n/N)

92.3 (12/13)

66.7 (2/3)

0.35

  • (95% CI)

64–99.8

9.4–99.2

Hilar

100 (5/5)

66.7 (2/3)

0.38

Distal

87.5 (7/8)

0

NS

PPV %, (n/N)

100% (42/42)

100 (32/32)

NS

  • (95% CI)

91.6–100

89.1–100

Hilar

100 (31/31)

100 (20/20)

NS

Distal

100 (11/11)

100 (12/12)

NS

NPV %, (n/N)

92.3 (12/13)

66.7 (2/3)

0.35

  • (95% CI)

64–99.8

9.4–99.2

Hilar

83.3 (5/6)

66.7 (2/3)

1

Distal

100 (7/7)

0

NS

Accuracy %, (n/N)

71.1 (54/76)

87.2 (34/39)

0.06

  • (95% CI)

59.5–80.9

72.6–95.7

Hilar

69.2 (36/52)

84.6 (22/26)

0.18

Distal

75 (18/24)

92.3 (12/13)

0.39


#

Adverse events

Procedure-related AEs are summarized in [Table 4]. Pancreatitis and cholangitis occurred in 3.9% and 1.3% of patients in the conventional group, respectively; however, these events did not occur in the Endosheather group.

Table 4 Adverse events.

Measure

Conventional

n = 76

Endosheather

n = 39

P value

NS, not significant.

Overall, n (%)

4 (5.2)

0 (0)

0.30

Pancreatitis, n (%)

3 (3.9)

0 (0)

0.55

Cholangitis, n (%)

1 (1.3)

0 (0)

1.00

Bleeding, n (%)

0 (0)

0 (0)

NS

Perforation, n (%)

0 (0)

0 (0)

NS


#
#

Discussion

In this study, we compared outcomes of conventional TBFB with those of TBFB using a novel sheath device. TBFB with a novel sheath device exhibited higher sensitivity than conventional TBFB.

A retrospective study evaluating TBFB with a pusher tube of an existing biliary stent delivery system revealed that the rate of obtaining samples, including submucosal tissue, was 91.4%, whereas another retrospective study using a biliary dilation catheter demonstrated that sensitivity, specificity, and accuracy were 87.5%, 100%, and 93.7%, respectively [14] [15]. Furthermore, a retrospective, single-arm study evaluating the same sheath device used in our study revealed high diagnostic yields (sensitivity, 90%; specificity, 100%; and accuracy, 92.3%). Our study also demonstrated satisfactory diagnostic ability of the novel sheath device, in line with previous studies. In addition, we compared results with those of the conventional TBFB, and no study has compared these two methods. A significantly higher rate of adequate sampling and higher sensitivity were observed when using the novel sheath device.

Insertion of biopsy forceps into the bile duct can be challenging. However, using the novel sheath device, biopsy forceps can be inserted easily into the bile duct, and once inserted, it can be repeated, in line with our study. This resulted in a shorter biopsy time per biopsy and a lower rate of failed insertion of the forceps into the bile duct in the Endosheather group. We presumed that reducing the number of failed insertions of forceps into the bile duct shortened the time required for biopsy.

Although there was no significant difference in the rate of AEs, pancreatitis was observed only in the conventional group. We speculated that the small sample size was the reason for the lack of statistical difference. However, considering the reduction in failed insertion of forceps into the bile duct, it is intuitively understandable that risk of pancreatitis was reduced. We presumed that the favorable outcome resulted from the ability to obtain tissue from the same lesion repeatedly and effortlessly through the outer sheath. In recent years, precision medicine has become increasingly important in bile duct cancer, and it is necessary to obtain a larger sample size for comprehensive genome profiling using next-generation sequencing [16]. Because increasing the number of biopsies is desirable to increase the amount of tissue obtained, TBFB using the novel sheath device can also be a promising technique to safely perform TBFB in a shorter time for comprehensive genome profiling.

Nonetheless, this study has some limitations. First, some bias was inevitable owing to its retrospective nature. However, enrolling consecutive patients and the choice of two methods (conventional TBFB and TBFB using Endosheather), which were performed at different times, probably reduced selection bias. Second, the number of biopsies performed differed between the two groups. For conventional TBFB, to consider the balance of diagnostic yield and safety, we decided that the number of biopsies was set to at least three, although we knew that five or more biopsies were reported to be an independent predictive factor for a positive cancer diagnosis of malignant biliary stricture in a retrospective study [17]. Therefore, if more biopsies had been performed, the conventional method might have had higher sensitivity. However, this may also lead to a higher incidence of pancreatitis. Whether our approach can be adopted as a conventional method in all facilities remains unclear, but we believe that there are not many institutions where biopsy forceps are consistently inserted into the bile duct for tissue acquisition, five times in every procedure. In that regard, this conventional method might be more realistic. Third, this was a single-center study with a small sample size; thus, a multicenter study with a larger sample size is required to confirm our results.


#

Conclusions

In conclusion, TBFB using a novel dedicated sheath device had a higher diagnostic yield without failed insertion into the bile duct than conventional TBFB. This novel sheath device can also facilitate TBFB and produce higher sensitivity.


#
#

Conflict of Interest

The authors declare that they have no conflict of interest.


Correspondence

Hirotoshi Ishiwatari
Division of Endoscopy, Shizuoka Cancer Center
Nagaizumi
Japan   

Publikationsverlauf

Eingereicht: 05. November 2023

Angenommen nach Revision: 23. Februar 2024

Accepted Manuscript online:
11. März 2024

Artikel online veröffentlicht:
15. April 2024

© 2024. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial-License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/).

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany


Zoom Image
Fig. 1 Image of the novel delivery device. a The device has an inner catheter with a tapered tip (red arrowhead) and an outer sheath. There is almost no gap between the outer sheath and the inner catheter (blue arrow). b The inner catheter can be removed from the outer sheath. c The biopsy forceps (Radial Jaw 4) are inserted through the outer sheath.
Zoom Image
Fig. 2 A cholangiogram image obtained using the novel delivery device. a The outer sheath tip is located above the bile duct stricture (yellow ahead) after removing the inner catheter. b Biopsy forceps (Radial Jaw 4) (white ahead) are inserted into the bile duct through the outer sheath to obtain the bile duct specimen.